    Intended Use

    The LeadCare® Plus™ Blood Lead Testing System is intended for the quantitative measurement of lead in a whole blood sample. The LeadCare Plus Blood Lead Testing System is intended for in vitro (external) use only. The test kit components are for use with both the LeadCare Plus and LeadCare Ultra® Blood Lead Testing Systems.

    This test system is for prescription use only. This system is not intended for point of care use.

    Device Description

    The LeadCare Plus Blood Lead Testing System is a new instrument to the LeadCare line of instruments. The predicate, LeadCare Ultra, is a 6-channel analyzer which has the same intended use as the LeadCare Plus. The LeadCare Plus instrument is a one-channel version of this system. The LeadCare Plus Analyzer is used in conjunction with the same test kit components, including: sensors, control materials, and single-use packaged reagent tubes as utilized with the predicate device, LeadCare Ultra. The single-channel LeadCare Plus analyzer is an in vitro diagnostic device that relies on electrochemistry (Anodic Stripping Voltammetry or ASV).

    The analyzer is a low voltage potentiostat that runs on AC power or batteries and has dimensions of 9"x6.5"x3.5". It is equipped with a Liquid Crystal Display (LCD) Screen, Sensor Connector and Calibration Button reader. The Screen displays instructions for the blood lead assay, result, lot code and error messages. The Calibration Button reader allows for the download of all calibration information, analytical test parameters, and lot code information for any given Sensor lot.

    For measurement of lead, the whole blood sample is pipetted to a packaged reagent tube. Upon mixing, the red blood cells are ruptured and the lead is released from the proteins and becomes labile and available for electrochemical detection. After the sample mixture is applied to the sensor, the analyzer applies an electrical potential that causes the lead to reduce (collect) on the sensor. After three minutes, the analyzer applies potentials that cause the lead to oxidize back into solution. The current produced is measured and the amount of lead in the sample is calculated. The analytical result is displayed on the screen in micrograms per deciliter (ug/dL).

    AI/ML Overview

    Here's an analysis of the acceptance criteria and study details for the LeadCare® Plus™ Blood Lead Testing System, based on the provided document:


    1. Table of Acceptance Criteria & Reported Device Performance

    The document implicitly defines acceptance criteria through the reported performance metrics that demonstrate substantial equivalence to the predicate device and the reference method. Explicit, pre-defined numerical acceptance criteria are not always stated as "acceptance criteria" but are demonstrated by the device meeting statistical thresholds or showing strong correlation.

    Performance MetricAcceptance Criteria (Implicit/Demonstrated)Reported Device Performance
    PrecisionWithin-run and Total CVs demonstrating acceptable variability for blood lead testing.Reported:
    - Mean 3.1 µg/dL: WR CV 14.1%, Total CV 15.6%
    - Mean 5.1 µg/dL: WR CV 8.5%, Total CV 9.6%
    - Mean 11.7 µg/dL: WR CV 5.3%, Total CV 6.0%
    - Mean 24.7 µg/dL: WR CV 3.2%, Total CV 4.0%
    - Mean 45.4 µg/dL: WR CV 3.5%, Total CV 3.7%
    - Mean 59.1 µg/dL: WR CV 3.2%, Total CV 4.0%
    Linearity (vs. LeadCare Plus)Strong linear relationship, R² close to 1.0.Reported:
    Higher order coefficients not statistically significant (CLSI EP06-A Section 5.32).- Lot 1312B/021214U: Y = 1.11x - 1.49, R² = 0.998
    - Lot 1310A/041014U: Y = 1.08x - 0.803, R² = 0.997
    Limit of Blank (LoB)Stated as 1.5 µg/dL (matching predicate).Calculated Average LoB: 0.98 µg/dL (claimed 1.5 µg/dL)
    Limit of Detection (LoD)Stated as 1.9 µg/dL (matching predicate).Calculated Average LoD: 1.2 µg/dL (claimed 1.9 µg/dL)
    Limit of Quantification (LoQ)Stated as 1.9 µg/dL (matching predicate).Calculated Average LoQ: 1.6 µg/dL (claimed 1.9 µg/dL), Total Error at LoQ 18%
    Method Comparison (vs. LeadCare Ultra)Strong linear correlation, slope near 1.0, intercept near 0, R² close to 1.0.Reported: Slope 0.985, Intercept -0.10, R² 0.994
    Method Comparison (vs. GFAAS)Strong linear correlation, slope near 1.0, intercept near 0, R value close to 1.0.Reported: Slope 1.029, Intercept -0.98, R 0.994
    Matrix Comparison (Micro-capillary tubes vs. GFAAS)Strong linear correlation, slope near 1.0, intercept near 0, R² close to 1.0.Reported: Slope 1.018, Intercept 0.023, R² 0.983

    2. Sample Size Used for the Test Set and Data Provenance

    • Precision Study: 80 data points per concentration level (6 concentrations), for a total of 480 individual measurements. The samples included bovine blood standards and an additional human blood sample. Blood matrix information suggests samples were from human donors (for linearity, LoD/LoQ, and method comparison) or bovine (for LoB and some precision).
    • Linearity Study: Nine donor blood samples per Sensor/Reagent Lot Pair (concentrations ranging from 0-70 ug/dL). One whole blood, unadulterated in K2EDTA vacutainers, was also included.
    • Limit of Blank (LoB) Study: 60 replicates of near blank bovine blood samples.
    • Limit of Detection (LoD) & Limit of Quantification (LoQ) Study: 60 replicates of 10 different whole blood samples (human, collected in K2EDTA vacutainers).
    • Method Comparison Study: 284 clinical samples were generated, and 169 were within the reportable range of 1.9 - 65 µg/dL for comparison against LeadCare Ultra and GFAAS. These were whole blood samples collected in K2EDTA vacutainers.
    • Matrix Comparison Study (Micro-capillary tubes): 23 native patient samples and 27 contrived (spiked) blood samples, totaling 50 samples.

    Data Provenance: The document does not explicitly state the country of origin for the human or bovine blood samples. The studies appear to be prospective as they were specifically designed and executed to evaluate the performance of the LeadCare Plus device.


    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    The document describes the Graphite Furnace Atomic Absorption Spectroscopy (GFAAS) as the reference method (ground truth) for the Method Comparison and Matrix Comparison studies. For the other studies (Precision, Linearity, LoB/LoD/LoQ), the "ground truth" is typically established by the spiked concentrations or the inherent characteristic of the controls/samples used.

    • Number of Experts: Not applicable in the context of GFAAS being the reference method. GFAAS is an analytical instrument-based method, and its results are considered the objective ground truth for lead concentration.
    • Qualifications of Experts: Not relevant as the ground truth is an analytical measurement, not an expert visual/cognitive assessment. The technicians operating the GFAAS system would be qualified laboratory personnel.

    4. Adjudication Method for the Test Set

    Not applicable. The ground truth (GFAAS) is an objective analytical measurement, not a subjective interpretation requiring adjudication among human readers.


    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This is an in vitro diagnostic device for quantitative measurement of lead, not an imaging or diagnostic AI device that involves human readers interpreting cases.


    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

    Yes, this entire submission focuses on the standalone performance of the LeadCare® Plus™ Blood Lead Testing System. It is an automated analytical device that provides a quantitative result. The performance metrics presented (precision, linearity, LoB/LoD/LoQ, and method comparisons) all reflect the device's algorithmic and instrument performance without direct human interpretation being part of the measurement output itself. The comparison is made against existing analytical methods (LeadCare Ultra and GFAAS).


    7. The Type of Ground Truth Used

    The primary ground truth used for performance evaluation, especially for accuracy claims, is objective analytical measurement by Graphite Furnace Atomic Absorption Spectroscopy (GFAAS). GFAAS is a highly accurate and widely accepted reference method for determining elemental concentrations, including lead in blood.


    8. The Sample Size for the Training Set

    This document does not specify a separate "training set" in the context of machine learning. The LeadCare Plus system is based on electrochemistry (Anodic Stripping Voltammetry or ASV) and operates on a defined algorithm, not a machine learning model that requires a distinct training and test set with ground truth labels. The development and internal validation of the ASV algorithm would have used various samples, but these are not explicitly termed a "training set" in the sense of AI/ML.


    9. How the Ground Truth for the Training Set Was Established

    As this is not an AI/ML device relying on a "training set" in the typical sense, this question is not directly applicable. The device's underlying principles and algorithm are based on established electrochemical theories and analytical chemistry, with development and optimization based on known concentrations of lead and robust calibration protocols.