510(k) Data Aggregation

    K Number
    DEN170010

    Validate with FDA (Live)

    Date Cleared
    2018-04-20

    (423 days)

    Product Code
    Regulation Number
    862.3590
    Type
    Direct
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The LZI Carisoprodol Metabolite (Meprobamate) Enzyme Immunoassay is intended for the qualitative and semi-quantitative determination of carisoprodol metabolite (meprobamate) in human urine at a cutoff value of 100 ng/mL when calibrated against meprobamate. The assay is designed for prescription use with a number of automated clinical chemistry analyzers.

    The semi-quantitative mode is for purposes of enabling laboratories to determine an appropriate dilution of the specimen for verification by a confirmatory method such as GC/MS, LC/MS or permitting laboratories to establish quality control procedures.

    The assay provides only a preliminary analytical result. A more specific alternative chemical confirmatory method (e.g., gas or liquid chromatography and mass spectrometry) must be used to obtain a confirmed analytical result. Clinical consideration and professional judgment must be exercised with any drug of abuse test, particularly when the preliminary test result is positive.

    Device Description

    The LZI Carisoprodol Metabolite (Meprobamate) Enzyme Immunoassay is a homogeneous enzyme immunoassay with ready-to-use liquid reagents. The assay is a kit comprised of two reagents (antibody/substrate reagent R1 and enzyme-drug conjugate reagent R2), which are bottled separately but sold together within the kit. The Ri solution contains mouse monoclonal anti-meprobamate antibody, glucose-6- phosphate (G6P), nicotinamide adenine dinucleotide (NAD), stabilizers, and sodium azide (0.09 %) as a preservative. The R2 solution contains meprobamate-labeled glucose-6-phosphate dehydrogenase (G6PDH) in buffer with sodium azide (0.09 %) as a preservative.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and the study details for the LZI Carisoprodol Metabolite (Meprobamate) Enzyme Immunoassay, based on the provided document:

    Acceptance Criteria and Reported Device Performance

    Acceptance Criteria CategorySpecific Acceptance Criteria (Inferred from study results and regulatory requirements)Reported Device Performance
    Accuracy (Qualitative Mode)High agreement with confirmatory method (LC/MS or GC/MS) for positive and negative samples, particularly around the cutoff.Qualitative Analysis Agreement: - Positive Agreement: 98.5% - Negative Agreement: 96.6% *2 false positives in "Near Cutoff Negative" (92 and 98 ng/mL, both identified as Negative by GC/MS/LC/MS) *1 false negative in "Near Cutoff Positive" (103 ng/mL, identified as Positive by GC/MS/LC/MS)
    Accuracy (Semi-Quantitative Mode)High agreement with confirmatory method (LC/MS or GC/MS) for positive and negative samples, particularly around the cutoff.Semi-Quantitative Analysis Agreement: - Positive Agreement: 98.5% - Negative Agreement: 98.3% *1 false positive in "Near Cutoff Negative" (98 ng/mL, identified as Negative by GC/MS/LC/MS) *1 false negative in "Near Cutoff Positive" (103 ng/mL, identified as Positive by GC/MS/LC/MS)
    Reproducibility/Precision (Qualitative Mode)Consistent results at specific concentrations, especially near the cutoff.At cutoff (100 ng/mL, GC/MS confirmed 94.9 ng/mL): 40 Negative / 48 Positive out of 88 determinations. At -25% Cutoff (75 ng/mL, GC/MS confirmed 76.8 ng/mL): 88 Negative / 0 Positive. At +25% Cutoff (125 ng/mL, GC/MS confirmed 122.3 ng/mL): 0 Negative / 88 Positive.
    Reproducibility/Precision (Semi-Qualitative Mode)Consistent results at specific concentrations, especially near the cutoff.At cutoff (100 ng/mL, GC/MS confirmed 94.9 ng/mL): 60 Negative / 28 Positive out of 88 determinations. At -25% Cutoff (75 ng/mL, GC/MS confirmed 76.8 ng/mL): 88 Negative / 0 Positive. At +25% Cutoff (125 ng/mL, GC/MS confirmed 122.3 ng/mL): 0 Negative / 88 Positive.
    Linearity/Reportable RangeAcceptable recovery across a range of concentrations.Mean Recovery: 66.6% to 109.2% for expected concentrations from 10 ng/mL to 400 ng/mL. (Note: 10 ng/mL showed lower recovery, but higher concentrations were generally good).
    Analytical Specificity (Endogenous Compounds)No interference from common endogenous compounds at specified concentrations.No positive or negative interference observed for the listed endogenous compounds when meprobamate was at -25% or +25% of the cutoff.
    Analytical Specificity (Urine Preservatives)No significant interference from common urine preservatives.Sodium azide and sodium fluoride: No interference. Boric acid: Caused false negative results at +25% cutoff (125 ng/mL) and up to +125% cutoff (225 ng/mL). Labeling mitigation required.
    Analytical Specificity (pH)No interference across a physiological pH range.No positive or negative interference observed at urine pH values ranging from 3 to 11.
    Analytical Specificity (Specific Gravity)No interference across a physiological specific gravity range.No positive or negative interference observed at urine specific gravities ranging from 1.002 to 1.029.
    Cross-Reactivity (Structurally Related Compounds)Acceptable cross-reactivity profile.Carisoprodol (90.9%), Felbamate (25.0%), Meprobamate-N-Glucuronide (0.5%), Hydroxymeprobamate (0.2%). Other listed compounds showed <0.1% cross-reactivity.
    Interference (Structurally Unrelated Compounds)No interference from common structurally unrelated drugs.No positive or negative interference observed from a wide range of common drugs/substances (e.g., Acetaminophen, Codeine, Ibuprofen) at high concentrations (e.g., 100,000 ng/mL) when meprobamate was at -25% or +25% of the cutoff.
    Clinical Validity of CutoffMeprobamate levels measurable for a relevant post-dose period.Pharmacokinetic studies of carisoprodol and meprobamate in plasma, and meprobamate elimination in urine, indicated levels above 100 ng/mL are present for "up to a few days" post-single dose.

    Study Details

    2. Sample size used for the test set and the data provenance

    • Sample Size for Test Set:

      • Precision Study: 88 determinations per concentration level for
        each of 8 concentrations (a total of 704 determinations for qualitative and 704 for semi-quantitative).
      • Method Comparison Study: 127 clinical urine samples.
      • Linearity Study: Each sample run in 10 replicates across 11 concentrations.
      • Analytical Specificity (Endogenous, Preservatives, pH, Specific Gravity, Unrelated Compounds): Samples tested in duplicate.
      • Cross-reactivity (Structurally Related Compounds): Not explicitly stated, but typically involves multiple runs.
    • Data Provenance: Retrospective clinical urine samples "from subjects that were (b) (4) taking prescribed or nonprescribed Carisoprodol." The origin country is not explicitly stated, but given this is an FDA submission, it's likely primarily US-based or from regions with comparable medical practices.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    • The ground truth for the test set was established by Gas Chromatography/Mass Spectrometry (GC/MS) or Liquid Chromatography/Mass Spectrometry (LC/MS). These are established analytical methods for drug confirmation. There is no mention of "experts" in the sense of medical professionals (e.g., radiologists) establishing ground truth, as this is an in vitro diagnostic device. The "experts" in this context would be the technicians/scientists operating and interpreting the GC/MS/LC/MS results, who are implied to be qualified in their field based on the use of these validated methods.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    • There was no human adjudication process described. The reference method (GC/MS or LC/MS) provided the definitive "ground truth" for each sample.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • No, this was not a multi-reader multi-case (MRMC) comparative effectiveness study. This device is an automated enzyme immunoassay for chemical analysis, not an imaging-based AI system that would typically involve human readers. Therefore, the concept of "human readers improve with AI vs without AI assistance" does not apply here.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • Yes, this was a standalone (algorithm only) performance study. The LZI Carisoprodol Metabolite (Meprobamate) Enzyme Immunoassay is an automated system designed to provide a preliminary analytical result without human interpretation of the assay itself. Human involvement comes in confirming positive results with alternative chemical methods and clinical judgment, as stated in the intended use.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    • The ground truth used was confirmatory analytical methods, specifically Gas Chromatography/Mass Spectrometry (GC/MS) or Liquid Chromatography/Mass Spectrometry (LC/MS). These methods are considered the gold standard for drug quantification and identification in urine.

    8. The sample size for the training set

    • The document does not explicitly describe a separate "training set" for the device. As an enzyme immunoassay, the "training" (calibration) involves using known calibrator concentrations (0, 50, 100, 200, and 400 ng/mL meprobamate, as mentioned in the linearity section) to establish the standard curve against which unknown samples are measured. There isn't an "algorithm" in the conventional AI sense that undergoes training with a large dataset. The performance characteristics described are related to the analytical validation of the assay itself.

    9. How the ground truth for the training set was established

    • As there isn't a traditional "training set" in the AI sense, the concept of ground truth for it doesn't directly apply. The "ground truth" for the calibrators used in the assay system would be their known, precisely manufactured concentrations of meprobamate, traceable to a commercially available meprobamate source material with 99% analytical purity (as mentioned in section L.1.c).