Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K162235

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2017-03-24

    (227 days)

    Product Code
    Regulation Number
    884.5300
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Luvena Vaginal Moisturizer and Lubricant is a personal lubricant, for vaginal application, intended to moisturize and lubricate, to enhance the ease and comfort of intimate sexual activity and supplement the body's natural lubrication. This product is compatible with natural rubber latex and polyisoprene condoms, but is not compatible with polyurethane condoms.

    Device Description

    Luvena Vaginal Moisturizer & Lubricant is non-sterile, translucent, non-staining, non-greasy, aqueous-based, cranberry-flavored gel and is packed in EVA/PE copolymer plastic applicator tube. Each application tube is single-use. It is for personal vaginal use as need for moisturizing and lubrication during intimate sexual activity. This device is not a contraceptive or spermicide, nor does it contain any such component. It is compatible with natural rubber latex and polyisoprene condoms, but is not compatible with polyurethane condoms as demonstrated in condom compatibility testing conducted according to ASTM D7661-10.

    AI/ML Overview

    This document is a 510(k) Summary for the Luvena Vaginal Moisturizer & Lubricant (K162235). The primary purpose of this summary is to demonstrate substantial equivalence to a predicate device, not to provide a detailed study for meeting specific acceptance criteria in the context of an AI/algorithm-based medical device. Therefore, much of the requested information regarding AI device evaluation is not directly available in this text.

    However, I can extract information related to the device's specifications and the testing performed to ensure its safety and effectiveness, which can be interpreted as fulfilling certain acceptance criteria.

    1. Table of Acceptance Criteria and Reported Device Performance

    Since this is a personal lubricant, the "acceptance criteria" are related to its physical properties, microbial safety, biocompatibility, and compatibility with condoms.

    Acceptance Criteria CategorySpecific Criteria / TestReported Device Performance
    Physical PropertiesAppearance, Odor, pH, Viscosity, OsmolalityAppearance: non-sterile, translucent, non-staining, non-greasy Odor: Cranberry-flavored gel pH: 3.8 - 4.8 (Matches predicate) Viscosity, Osmolality: Evaluated (Specific values not provided, but implicitly met specifications)
    Microbiological SafetyTotal Aerobic Microbial Count (TAMC) (USP <61> and <1111>)Total Yeast and Mold Count (TYMC) (USP <61> and <1111>)Absence of pathogenic organisms (Pseudomonas aeruginosa, Staphylococcus aureus, Candida albicans (USP <62>))Met specifications as recommended in USP <1111> for vaginal use products. Also passed Antimicrobial Preservative Effectiveness Validation testing under USP <5> Category 2.
    BiocompatibilityCytotoxicity (ISO 10993-5:2009)Sensitization (ISO 10993-10:2010 - GPMT)Vaginal Irritation (ISO 10993-10:2010)Acute Systemic Toxicity (ISO 10993-11:2006)The results demonstrate that the subject lubricant is biocompatible.
    Condom CompatibilityCompatibility with natural rubber latex and polyisoprene condomsIncompatibility with polyurethane condoms (ASTM D7661-10)Compatible with natural rubber latex and polyisoprene condoms; not compatible with polyurethane condoms. (Matches predicate and tested per ASTM D7661-10)
    Shelf LifeStability testing (ICH-Q1A, ICH-Q1E)Results of accelerated time testing and real-time testing support the proposed shelf life. (Specific shelf life duration not provided but implied as adequate).

    2. Sample Size for Test Set and Data Provenance

    This document does not describe a "test set" in the context of an AI/algorithm evaluation. The testing described falls under laboratory and in vitro testing for a physical product (lubricant). Therefore, information on "country of origin of data" or "retrospective/prospective" does not apply in the typical sense for AI devices. The sample sizes for the various physical, chemical, and biological tests performed on the lubricant formulation are not specified in this summary.

    3. Number of Experts Used to Establish Ground Truth and Qualifications

    This information is not applicable. "Experts" in this context would be laboratory technicians and chemists performing the tests according to established standards (ISO, USP, ASTM). There isn't a "ground truth" to be established by human experts in the way it would be for an AI device.

    4. Adjudication Method

    Not applicable, as there is no "adjudication" in the context of expert review for an AI device. The tests are laboratory-based and yield objective results.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    Not applicable. This is not an AI device, so there is no comparison of human reader performance with or without AI assistance.

    6. Standalone Performance Study (Algorithm Only)

    Not applicable. This is a physical device (lubricant), not an algorithm or AI.

    7. Type of Ground Truth Used

    The "ground truth" in this context is established by:

    • Established standards: ISO, USP, and ASTM test methods define the benchmarks against which the lubricant's properties (biocompatibility, microbial limits, condom compatibility) are measured.
    • Laboratory measurements: Objective physical, chemical, and biological assays directly determine the characteristics of the product.

    8. Sample Size for the Training Set

    Not applicable. There is no "training set" as this is not an AI/algorithm device.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable. There is no training set for this type of device.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1