Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K121390

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2012-08-10

    (93 days)

    Product Code
    Regulation Number
    880.6375
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    For Prescription Use: For easing the insertion of medical devices such as scopes and catheters into body orifices.

    Device Description

    The subject device is a terminally gamma sterilized 10cc polypropylene plastic syringe filled with United Guardian lubricating jel and capped with a polypropylene plastic cap. United Guardian has a Master File Reference for its lubrication jel, Master File for Devices MAF-613 pertaining to Lubrajel RR. All components of the device are gamma irradiation stable.

    AI/ML Overview

    The provided document describes a 510(k) premarket notification for a medical device, specifically a "Sterile Lube Jelly Pre-Filled Syringe," asserting its substantial equivalence to a legally marketed predicate device. The information details non-clinical performance data rather than a clinical study involving human patients or a complex AI algorithm. Therefore, many of the requested categories related to clinical trials, AI, ground truth, and expert evaluation are not applicable to this submission.

    Here's a breakdown based on the provided text:

    Acceptance Criteria and Reported Device Performance

    The device, the Nurse Assist Lube Jelly Pre-Filled Syringe, aims to demonstrate substantial equivalence to the predicate device, the Horizon Steri-Lub Lubrication Gel. The acceptance criteria are implicit in showing that the device's characteristics and performance are comparable to the predicate.

    Acceptance Criteria CategoryReported Device Performance (Nurse Assist Lube Jelly Pre-Filled Syringe)Predicate Device Performance (Horizon Steri-Lub Lubrication Gel)
    Classification Product CodeKMJKMJ
    Intended Use: Lubricating Device InsertionYesYes
    Prescription UseYesYes
    SterilityYes (terminally gamma sterilized)Yes
    Shelf Life2 years2 years
    Chemical CompositionWater, Glycerin, Polyacrylic Acid, Propylene Glycol, Sodium Polyacrylate preserved with methyl and propyl paraben and sodium metabisulfiteWater, Glycerin, Polyacrylic Acid, Propylene Glycol, Sodium Polyacrylate preserved with methyl and propyl paraben and sodium metabisulfite
    Mechanism of Dispensing10cc Plastic Syringe, Oral Tip10cc Plastic Syringe, Oral Tip
    Barrel, Plunger, Tip Cap MaterialPolypropylenePolypropylene
    Plunger Grommet Material (Latex-Free)Not made with natural rubber latexNot made with natural rubber latex
    Syringe Seal IntegrityRemained sealed when exposed to a 15 In-Hg vacuum.(Not specified for predicate, but Nurse Assist device met this criterion)
    Stain TestingPost sterile lubricant did not stain gloves.(Not specified for predicate, but Nurse Assist device met this criterion)
    ViscosityPost sterile lubricant remained within the pre-sterile 18,000 to 26,500 cps viscosity range.(Nurse Assist device met this criterion against its own pre-sterile range, implying it's consistent)
    Volume after Accelerated AgingRemained within specification.(Nurse Assist device met this criterion)
    pH after Accelerated AgingRemained within specification.(Nurse Assist device met this criterion)
    Sterility after Accelerated AgingRemained within specification.(Nurse Assist device met this criterion)
    BiocompatibilityComplied with the requirements of ISO 10993.(Nurse Assist device met this criterion)

    Study Details

    This submission relies on non-clinical performance data and a comparison to a predicate device to demonstrate substantial equivalence, rather than a clinical study the way an AI/software device would.

    1. Sample size used for the test set and the data provenance: Not applicable. The "test set" here refers to samples of the device undergoing various performance tests (e.g., syringe seal, stain, viscosity, accelerated aging, biocompatibility). The number of units tested for each specific non-clinical test is not provided in this summary. The data provenance is from internal testing conducted by Nurse Assist Incorporated.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable. Ground truth for non-clinical performance tests is established through standardized testing protocols (e.g., ISO 10993 for biocompatibility, physical property measurements for viscosity, pH, volume, etc.), not expert consensus in the diagnostic sense.
    3. Adjudication method: Not applicable. Standard laboratory testing procedures would be followed.
    4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done: No, this is not an MRMC study. This device is a pre-filled lubricant syringe, not an AI diagnostic tool.
    5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable. This is not an AI algorithm.
    6. The type of ground truth used: For the non-clinical performance tests, the "ground truth" is defined by established scientific and regulatory standards (e.g., specific viscosity ranges, pH levels, sterility tests, ISO 10993 for biocompatibility) that the device must meet.
    7. The sample size for the training set: Not applicable. There is no AI training set.
    8. How the ground truth for the training set was established: Not applicable. There is no AI training set.
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1