Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K100109

    Validate with FDA (Live)

    Date Cleared
    2010-07-15

    (182 days)

    Product Code
    Regulation Number
    876.1500
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Use in laparoscopic procedures for blunt dissection between tissue layers, to pull suture threads, to manipulate tubing and to generally assist with the placement of a gastric band.

    Device Description

    The LogiFlex Laparoscopic Band Positioning Device is a sterile, single-use, laparoscopic, blunt dissection instrument. It comprises a shaft with a distal articulating tip incorporating a slot that forms a hook for capturing suture threads and tubing. The LogiFlex Laparoscopic Band Positioning Device is designed to be connected to and used with the reusable, ratchet handle of the Logic Monopolar Laparoscopic Scissors in order to effect articulation of the distal tip. The LogiFlex Laparoscopic Band Positioning Device is not designed to be used with an electrocautery connection.

    AI/ML Overview

    The provided text describes nonclinical bench testing for the LogiFlex Laparoscopic Band Positioning Device. There is no information about clinical studies with human participants, thus questions related to expert ratings, MRMC studies, or multi-reader performance cannot be answered.

    Here's the information extracted from the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    TestAcceptance CriteriaReported Device Performance
    Distal Tip Pull TestingA tensile axial load > 2kg when pulling suture.Achieved a load of 2.96kg (pre-production), 2.6kg (production standard)
    Product straightens to -5° to 0° for cannula introduction/removal.Passed (articulating from 0° to 90°)
    Product articulates to ≥ 90° to allow positioning around the esophagus.Passed (articulating from 0° to 90°)
    Sideway Force TestingA sideway force ≥ 0.54kg when pulling suture.Achieved a sideway force of 0.54kg
    Product straightens to -5° to 0° for cannula introduction/removal.Passed (articulating from 0° to 90°)
    Product articulates to ≥ 90° to allow positioning around the esophagus.Passed (articulating from 0° to 90°)
    Crimped Nipple Pull Out TestingCrimped nipple to withstand 5kg amount of force.Achieved an average force of 5.23kg
    Tip Deflection TestingProduct straightens to -5° to 0° for cannula introduction/removal.Passed (articulating from 0° to 90°)
    Product articulates to ≥ 90° to allow positioning around the esophagus.Passed (articulating from 0° to 90°)
    Lifecycle TestingAll 10 specimens to survive 2 irradiation cycles and 50 actuations (articulating from 0° to 90°) without any mechanical defects.Passed (survived 2 irradiation cycles and 50 actuations without mechanical defects)
    Drop TestingPass in accordance with pass criteria ASTM D4169-1996. After 100% inspection of product after dropping there should be no visible defects found in the product and the package seal should be intact.Passed (packaging capable of maintaining device sterility and free from damage during transit)
    Shelf Life (Accelerated Ageing)Packaging seals meet ASTM F1886-98(2004), burst strength meets ASTM F1140-07, dye penetration meets ASTM F1929-98(2004) after accelerated aging equivalent to 5 years.Not explicitly stated as "passed" for all specific tests but the report TN 100312 is provided, implying compliance.
    Sterilization ValidationAchieve a SAL of 10⁻⁶ at a sterilization dose of 25 kilogray (validated in accordance with ANSI / AAMI / ISO 11137-2:2006, using the VDmax2 method).Validation confirmed; 25 kilogray applied.
    Biocompatibility - CytotoxicityNon-cytotoxic.Indicated to be non-cytotoxic (negative control non-cytotoxic, positive control cytotoxic).
    Biocompatibility - SensitizationNon-sensitiser (stimulation of less than 3).Considered to be a non-sensitiser (stimulation less than 3 for polar and non-polar extracts).
    Biocompatibility - Intracutaneous ReactivityNon-irritant (average reaction not greater than corresponding control).Considered to be non-irritant (extracts meet ISO 10993-10:2002 requirements).

    2. Sample Size for the Test Set and Data Provenance

    • Distal Tip Pull Testing: "13 samples" for production standard product. The provenance of the data is not specified (e.g., country of origin, retrospective/prospective), but it is implied to be laboratory bench testing.
    • Lifecycle Testing: "All 10 specimens". The provenance of the data is not specified.
    • Shelf Life (Accelerated Ageing): "Thirty test samples at each time point". The provenance of the data is not specified.
    • Other tests do not explicitly state sample sizes but generally refer to "the LogiFlex laparoscopic band positioning device," implying testing on representative samples.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications of Experts

    • Not applicable. The reported studies are nonclinical bench tests, not involving expert review for diagnostic purposes.

    4. Adjudication method for the test set

    • Not applicable. The reported studies are nonclinical bench tests.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done

    • No. The provided text describes nonclinical bench tests.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

    • Not applicable. The device is a physical laparoscopic instrument, not an AI algorithm.

    7. The type of ground truth used

    • Ground truth was based on pre-defined engineering and safety specifications, as documented in technical notes and industry standards (e.g., ASTM, ISO). For example:
      • Mechanical strength (tensile load, sideway force, pull out force).
      • Articulation angles (-5° to 0°, ≥ 90°).
      • Absence of mechanical defects.
      • Maintenance of sterility and package integrity.
      • Biocompatibility based on established biological response criteria (cytotoxicity, sensitization, irritation).

    8. The sample size for the training set

    • Not applicable. This is a physical medical device, not a machine learning model.

    9. How the ground truth for the training set was established

    • Not applicable. This is a physical medical device, not a machine learning model.
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1