LIAC S is mobile electron linear accelerator for intra-operative radiotherapy (IORT). Known as intraoperative radiation therapy (IORT), this technique allows delivery of high doses of radiation directly aimed at tumours or other sites. The LIAC S is meant to be used for radiotherapy in the operating theatre on a patient to which the surgeon has just removed a neoplasm. The device must be sold by or on the order of a physician. The device is not for use by the general public or over-the-counter. LIAC S is an electron accelerator that can work in an operating theatre respecting its specific characteristics in terms of cleaning possibility, noise and heat dissipation; in particular, they do not require to move the patient.
LIAC S is an electron linear Accelerator used to perform radiation therapy during surgical procedures in an operating room on a patient from whom the Surgeon has just removed a tumor. This technique is known as Intra-Operative Radiation Therapy (IORT) and it is used for the treatment of malignant and benign conditions. IORT allows delivering of high doses of radiation directly to turnour site, while normal tissues are protected against dosage impact. The available version of LIAC S is the 12 MeV one, with selectable energies of 6, 8, 10, 12 MeV and with electrical power requirement of 230 V, 50 Hz and 3kVA. LIAC S model employs the same technology as its predicate, LIAC HWL, with the following differences: 1. The Control Unit frame; 2. The total length of the applicator; 3. The design of the remote controller. All the differences have no impact on the product performances. The Equipment, LIAC S, is a system for Intraoperative Radiotherapy comprised of two main parts, Control and Mobile Unit, and the applicators. The main parts, connected by a ten meters-long cable which transfers the signals and the power are: 1) MOBILE UNIT (STAND). The Mobile Unit is a mobile component which emits the radiation. It is placed in the operating room near the patient with a remote-control device; 2) CONTROL UNIT. The Control Unit is a mobile component too. It is connected to the mobile unit through a group of cables. It is the part of the device that controls the equipment. Both the above listed components, the mobile unit (1) and the control unit (2), are joined by a cable of about 10 meters, to allow positioning the mobile unit in the operating theatre close to the patient, while the console control unit remains outside of the operating theatre allowing the operator to control the radiation. The equipment can perform its action only if the Mobile Unit and the Control Unit are connected adequately and the applicator is installed.
Here's a breakdown of the acceptance criteria and study information for the LIAC S device, based on the provided text:
1. Table of Acceptance Criteria and Reported Device Performance
The document does not explicitly state formal "acceptance criteria" in the sense of specific thresholds for clinical performance metrics (e.g., sensitivity, specificity, accuracy) for a new algorithm. Instead, it focuses on demonstrating substantial equivalence to a predicate device (LIAC HWL) through technical and performance comparisons.
The "acceptance criteria" are implicitly met if the LIAC S device's performance parameters are comparable or superior to those of the predicate device, or if any differences are deemed not to impact safety and effectiveness.
Here's a table summarizing the comparison, which serves as the basis for demonstrating equivalence:
| Characteristic | Proposed Device (LIAC S) Performance | Predicate Device (LIAC HWL) Performance | Assessment / Equivalence |
|---|---|---|---|
| Device Name | LIAC S (12 MeV) | LIAC HWL (Model: 12 MeV) | Substantially Equivalent |
| Manufacturer | S.I.T. - SORDINA IORT TECHNOLOGIES SPA | S.I.T. - SORDINA IORT TECHNOLOGIES SPA | Same |
| Classification | Class II | Class II | Same |
| Intended Use | Mobile electron linear accelerator for intra-operative radiotherapy (IORT) for malignant/benign conditions. | Mobile electron linear accelerator for intra-operative radiotherapy (IORT) for malignant/benign conditions. | Same |
| Emission | Electron beam | Electron beam | Same |
| Type of structure | LINAC | LINAC | Same |
| Docking | Hard docking | Hard docking | Same |
| Remote control | Yes | Yes | Same |
| Nominal energies | 6, 8, 10, 12 MeV (12 MeV) | 6, 8, 10, 12 MeV (12 MeV) | Same |
| Maximum operating temperature | 25°C | 25°C | Same |
| Electrical power requirement | 230 V, 50 Hz, 3 kVA | 230 V, 50 Hz, 3 kVA | Same |
| Applied Standards | IEC 60601-1, IEC 60601-1-2, IEC 60601-2-1, IEC 62304, ISO 14971 | IEC 60601-1, IEC 60601-1-2, IEC 60601-2-1, IEC 62304 | Conformance claimed for LIAC S |
| Surface dose | ≥ 85% | ≥ 90% (12 MeV) | Difference noted, but deemed not relevant for safety/effectiveness (implicitly accepted as equivalent) |
| Field symmetry | ≤ 3 % | ≤ 3 % | Same |
| Pulse repetition frequency | 5 - 50 Hz | 5 - 50 Hz (depended on the selected energy) | Equivalent |
| Beam current | 1.5 mA | ≤ 1.5 mA | Equivalent |
| Long term stability | ≤ 3 % | ≤ 3 % | Same |
| Short term stability | ≤ 1 % | ≤ 1 % | Same |
| Dosimetric system linearity | ≤ 1 % | ≤ 1 % | Same |
| Dose rate (applicator Ø 10 cm) | 2 - 20 Gy/min | 10-30 Gy/min | Difference noted, but deemed not relevant for safety/effectiveness (implicitly accepted as equivalent) |
| Stray X-radiation (PDD Bremsstrahlung tail) | ≤ 0.7% | ≤ 0.4% | Difference noted, but deemed not relevant for safety/effectiveness (implicitly accepted as equivalent) |
| Source Surface Distance (SSD) | 60 cm | 64.5 cm | Difference noted, but deemed not relevant for safety/effectiveness (implicitly accepted as equivalent) |
| Uniformity or Field Flatness (value in conformity with 60601-2-1 at max energy bevel angle 0° for applicator Ø 10 cm) | ≤ 5% | ≤ 7% | LIAC S shows better performance, deemed equivalent or superior |
| Mobile Unit movement capability | 2 degrees of movement (Translation), 1 degree (Roll), 1 degree (Pitch), 1 degree (Elevation) | 2 freedom degrees (Translation), 1 freedom degree (Roll), 1 freedom degree (Pitch), 1 freedom degree (Elevation) | Equivalent (different wording, same functionality) |
| Hard docking velocity | Up to 0.01 m/s | 0.1 m/s | Difference noted, but deemed not relevant for safety/effectiveness (implicitly accepted as equivalent) |
| Mobile Unit length | 185 cm | 210 cm | Difference noted, but deemed not relevant for safety/effectiveness (implicitly accepted as equivalent) |
| Mobile Unit Width | 76 cm | 76 cm | Same |
| Mobile Unit height | 180 cm | 180 cm | Same |
| Mobile Unit weight | < 550 Kg | 570+ (< 230, shield) kg | Difference noted, but deemed not relevant for safety/effectiveness (implicitly accepted as equivalent) |
| Control unit dimensions | 30 cm x 80 cm x 170 cm (length, width, height) | 80 cm x 60 cm x 120 cm (width, depth, height) | Difference in dimensions, but deemed not relevant for safety/effectiveness |
| Control unit weight | < 50 Kg | 120 kg | Difference in weight, but deemed not relevant for safety/effectiveness |
| Applicator length | 50 cm (total length) | 40 cm | Difference in length, but deemed not relevant for safety/effectiveness |
| Applicator size | Ø: 3,4,5,6,7,8 and 10 cm each with cutting angles of 0°, 15°, 30°, 45° | Ø: 3, 4, 5, 6, 7, 8, 9, 10 and 12 cm each with bevel angels of 0°, 15°, 30°, 45° | Differences in available diameters, but deemed not relevant for safety/effectiveness |
| Applicator Material | PMMA | PMMA | Same |
2. Sample size used for the test set and the data provenance
The provided document describes a device comparison/equivalence study rather than a study on an AI/algorithm's performance. Therefore, there is no specific "test set" in the context of clinical data for an algorithm. The evaluation relies on comparing the technical specifications and performance characteristics of the LIAC S device against its predicate device (LIAC HWL).
The "data provenance" mentioned is the technical specifications and performance measurements derived from the LIAC S device during its product development and verification, and similar data for the predicate device. These are likely from internal testing and engineering assessments rather than clinical trials with patient data.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
This information is not applicable as there is no clinical "ground truth" derived from expert review in this type of submission. The ground truth for device performance is based on physical measurements and engineering standards.
4. Adjudication method for the test set
This is not applicable for the same reasons as #3.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
An MRMC study is not applicable because the LIAC S is a medical radiation therapy device, not an AI or imaging diagnostic tool that relies on human readers interpreting output.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
This is not applicable as the LIAC S is a physical medical device, not a standalone algorithm. Its function is to deliver radiation therapy directly.
7. The type of ground truth used
The "ground truth" here refers to engineering specifications, physical measurements, and conformity to recognized technical standards (e.g., IEC 60601 series, ISO 14971). The evaluation confirms that the LIAC S device meets these established performance benchmarks and standards for safety and effectiveness, and that its differences from the predicate are not significant enough to raise new safety or effectiveness concerns.
8. The sample size for the training set
This is not applicable as the submission describes a physical medical device, not an AI model that requires a training set of data.
9. How the ground truth for the training set was established
This is not applicable for the same reasons as #8.
In summary: This 510(k) submission for the LIAC S device focuses on demonstrating substantial equivalence to an existing predicate device (LIAC HWL) through a detailed comparison of technical specifications, intended use, and performance parameters. The "study" here is essentially a technical validation and verification process, with "acceptance criteria" met by showing that the new device is as safe and effective as the predicate, despite some minor design and numerical differences which were deemed not to impact overall safety or effectiveness. It does not involve AI, clinical efficacy trials, or expert-based ground truth establishment in the way typically seen for diagnostic algorithms.
Ask a specific question about this device
(258 days)
LIAC HWL are mobile electron linear accelerators for intra-operative radiotherapy (IORT). Known as intraoperative radiation therapy (IORT), this technique allows delivery of high doses of radiation directly aimed at tumors or other sites. The LIAC HWL are meant to be used for radiotherapy in the operating theatre on a patient to which the surgeon has just removed a neoplasm. The devices must be sold by or on the order of a physician. They are not for use by the general public or over-the-counter. LIAC HWL are electron accelerators that can work in an operating theater respecting its specific characteristics in terms of cleaning possibility, noise and heat dissipation; in particular, they do not require to move the patient.
LIAC HWL are mobile electron linear accelerators for intra-operative radiotherapy (IORT). The device consists of a STAND (irradiating unit), CONTROL UNIT, and APPLICATORS. It is a LINAC type structure with HARD DOCKING. It emits an ELECTRON BEAM.
Here's an analysis of the provided text to extract information about the acceptance criteria and study proving device performance:
It is important to note that the provided text is a 510(k) summary for a medical device (LIAC HWL), which primarily focuses on demonstrating substantial equivalence to predicate devices. It does not present a detailed clinical study with performance metrics in the typical sense of an AI/human-in-the-loop diagnostic accuracy study. Instead, the performance is evaluated by comparing the technical specifications and radiation performance of the new device (LIAC HWL) against its legally marketed predicate devices.
Therefore, many of the requested categories for acceptance criteria and study details (like multi-reader multi-case studies, expert consensus for ground truth, training set details, adjudication methods) are not applicable in this context, as this is a medical device clearance based on engineering and physics performance, not a diagnostic algorithm.
Acceptance Criteria and Reported Device Performance
The acceptance criteria are implicitly defined by the parameters of the predicate devices. The new device (LIAC HWL) must perform comparably or better than the predicate devices within ranges considered safe and effective for its indicated use. The "reported device performance" are the values measured for the LIAC HWL, which are then compared to the predicates.
Here's a table based on the "DEVICE DESIGN – RADIATION PERFORMANCE" section:
| Characteristic | Acceptance Criteria (Predicate LIAC) | Acceptance Criteria (Predicate NOVAC11) | Reported Device Performance (LIAC HWL) |
|---|---|---|---|
| Surface dose (10 MeV) | ≥ 85 % | ≥ 80-85% | ≥ 88 % |
| Surface dose (12 MeV) | ≥ 87 % | N/A (for 12 MeV) | ≥ 90 % |
| Field symmetry | ≤ 3 % | ≤ 2 % | ≤ 3 % |
| Pulse repetition frequency | 5 - 50 Hz | 9 Hz | 5 - 50 Hz |
| Beam current | ≤ 1.5 mA | ≤ 1.5 mA | ≤ 1.5 mA |
| Long term stability | ≤ 3% | ≤ 2 % | ≤ 3% |
| Short term stability | ≤ 1% | ≤ 1 % | ≤ 1% |
| Dosimetric system linearity | ≤ 1 % | ≤ 1 % | ≤ 1 % |
| Dose rate (applicator ∅ 10 cm) | 3 - 20 Gy/min | According to applicator (6-39 Gy/min) | 10 - 30 Gy/min |
| Stray X-radiation | ≤ 0.7 % | ≤ 0.2 % | ≤ 0.4 % |
| Source Surface Distance (SSD) | 71.3 cm | 65-80 cm | 64.5 cm |
| Uniformity or Field Flatness | ≤3% Ø 10, 8, 7, 6 [cm], ≤9% Ø 4,5 [cm], ≤12% Ø 3 [cm] | ≤5% | ≤3% Ø 8,7,6 [cm], ≤7% Ø 10 [cm], ≤9% Ø 4,5 [cm], ≤12% Ø 3, 12 [cm], ≤4% Ø 9[cm] |
Study Details (as applicable to a 510(k) for a linear accelerator)
-
Sample sized used for the test set and the data provenance:
- Test Set: No "test set" in the sense of patient data or images for a diagnostic algorithm. The "test" in this context refers to physical measurements and engineering validations of the device's radiation characteristics and physical specifications. These tests would have been performed on the LIAC HWL device itself in a controlled environment.
- Data Provenance: The data comes from direct measurements and engineering tests of the LIAC HWL prototype/production unit. The country of origin of the manufacturer is Italy. This is a prospective evaluation of the new device's performance.
-
Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
- Not applicable in the context of diagnostic "ground truth." The "ground truth" for these physical parameters is established by standard physics principles, dosimetry protocols, and engineering specifications. While experts (e.g., medical physicists, engineers) would have performed and analyzed these measurements, their role is not to establish "ground truth" in an interpretive sense, but to conduct and verify the measurements against established physical standards.
-
Adjudication method (e.g. 2+1, 3+1, none) for the test set:
- Not applicable. This is not a study requiring human interpretation or adjudication of diagnostic findings. Device performance is determined through reproducible physical measurements.
-
If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
- Not applicable. This is not a diagnostic AI device or a study involving human readers. It's a linear accelerator for radiation therapy application.
-
If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:
- Yes, in spirit. The "standalone" performance here refers to the measured physical and radiation characteristics of the LIAC HWL itself, independent of operator influence on the inherent physics of the beam. The system is designed to produce a specific electron beam with defined characteristics, and these characteristics are measured directly.
-
The type of ground truth used (expert consensus, pathology, outcomes data, etc):
- Physical and Dosimetric Standards: The "ground truth" for this device's performance is established by internationally recognized standards for medical linear accelerator dosimetry and performance (e.g., IAEA TRS-398, AAPM TG-51 protocols, IEC standards like IEC 60601-2-1). Compliance with these physical standards ensures the device performs as expected for its therapeutic application.
-
The sample size for the training set:
- Not applicable. This is not a machine learning model that requires a training set. The device's design and performance are based on physics and engineering principles, not on learned patterns from a dataset.
-
How the ground truth for the training set was established:
- Not applicable. As above, no training set for an algorithm.
Ask a specific question about this device
(389 days)
Liac (Models: 10MeV and 12MeV) is a mobile electron linear accelerator for intra-operative radiotherapy (IORT). Known as intraoperative radiation therapy (IORT), this technique allows delivery of high doses of radiation directly aimed at tumors or other sites. The LIAC is meant to be used for radiotherapy in the operating theatre on a patient to which the surgeon has just removed a neoplasm. The device must be sold by or on the order of a physician. It is not for use by the general public or over-the-counter.
mobile electron linear accelerator for intra-operative radiotherapy (IORT)
This document is an FDA 510(k) clearance letter for a medical device. It does not contain any information about the acceptance criteria or a study proving the device meets those criteria, nor does it contain information about AI/ML performance.
The letter confirms that the device, LIAC (Models: 10MeV and 12MeV), is substantially equivalent to legally marketed predicate devices and can be marketed. It outlines the indications for use of the device, which is a mobile electron linear accelerator for intra-operative radiotherapy (IORT), meaning it delivers high doses of radiation directly aimed at tumors or other sites during surgery after a neoplasm has been removed.
Therefore, I cannot provide the requested information based on the provided text.