Search Filters

Search Results

Found 2 results

510(k) Data Aggregation

    K Number
    K110588

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2011-05-25

    (84 days)

    Product Code
    Regulation Number
    888.3070
    Reference & Predicate Devices
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use
    Device Description
    AI/ML Overview
    Ask a Question

    Ask a specific question about this device

    K Number
    K103102

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2011-02-08

    (111 days)

    Product Code
    Regulation Number
    888.3070
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The LEUCADIA™ Pedicle Screw System is intended to be used in skeletally mature patients as an adjunct to fusion using autograft or allograft in posterior, non-cervical fixation for the following conditions: severe spondylolisthesis (grade 3 and 4) at L5-S1; degenerative spondylolisthesis with objective evidence of neurologic impairment, trauma (i.e. fracture or dislocation); spinal stenosis; curvatures (i.e. scoliosis, kyphosis and/or lordosis); tumor; pseudarthrosis; and/or failed previous fusion.

    Device Description

    The LEUCADIA™ Pedicle Screw System is intended to help provide correction, immobilization and stabilization of spinal segments as an adjunct to fusion of the thoracic, lumbar and/or sacral space. The LEUCADIA™ Pedicle Screw System consists of a variety of rods and screws, which can be rigidly locked into a variety of configurations, with each construct being tailor made for the individual case. Polyaxial screws are supplied in winged and non-winged configurations, in a variety of length, ranging from 30 mm to 100 mm and in 5 mm, 7 mm. 8mm and 9 mm diameter sizes. All sizes are able to receive 5.5mm connecting rods only. Pre-bent rods are available in 10 mm increments, ranging from 30 to 120 mm in length. The LEUCADIA™ Pedicle Screw System implant components are fabricated from medical grade titanium alloy per ASTM F136.

    AI/ML Overview

    The document provided is a 510(k) summary for the LEUCADIA™ Pedicle Screw System. It details the device's indications for use and states that biomechanical performance tests were conducted to demonstrate substantial equivalence to a predicate device.

    However, the provided text does not include an acceptance criteria table, reported device performance metrics against specific criteria, or details regarding studies that would involve human readers, ground truth establishment, or sample sizes related to clinical or image-based performance. The study described is a series of biomechanical tests.

    Therefore, many of the requested items cannot be extracted from this document.

    Here's what can be extracted based on the provided text:

    1. Table of acceptance criteria and the reported device performance

    Acceptance CriteriaReported Device Performance
    (Not explicitly defined in terms of specific pass/fail values in the document)"The test results were equivalent to the predicate device and/or other similar implants and are sufficient for in vivo loading."

    2. Sample size used for the test set and the data provenance

    • Sample Size for test set: Not applicable for biomechanical tests in the context of clinical or image-based studies. The "test set" here refers to the biomechanical testing of the device components.
    • Data Provenance: The tests are biomechanical in nature, conducted on the device components themselves, not on patient data.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    • Not applicable. Ground truth in this context would be related to biomechanical standards and measurements, not expert clinical assessment.

    4. Adjudication method for the test set

    • Not applicable. Biomechanical tests follow standardized protocols (e.g., ASTM F1717, F1798, F543). The "adjudication" is based on meeting the requirements of these standards or demonstrating equivalence to a predicate device.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • No, a multi-reader multi-case (MRMC) study was not done. This device is a pedicle screw system, a physical implant, not an AI-powered diagnostic or assistive tool.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done

    • No, a standalone (algorithm only) performance study was not done. This device is a physical implant.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    • The "ground truth" for the biomechanical tests would be the quantitative measurements dictated by the ASTM standards (e.g., force, displacement, torque values) and comparison to established predicate device performance or industry benchmarks for in vivo loading sufficiency.

    8. The sample size for the training set

    • Not applicable. This is not an AI/ML device that requires a training set.

    9. How the ground truth for the training set was established

    • Not applicable. This is not an AI/ML device that requires a training set.

    Study details (from the document):

    The study to "prove the device meets the acceptance criteria" involved the following biomechanical tests:

    • Static Compression Bending Tests per ASTM F1717
    • Static Tension Bending Tests per ASTM F1717
    • Static Torsion Tests per ASTM F1717
    • Dynamic Axial Compression Bending Tests per ASTM F1717
    • Static Axial Gripping Capacity Tests per ASTM F1798
    • Static Axial Torque Tests per ASTM F1798
    • Static A-P Screw Pull Tests per ASTM F1798
    • Static Screw Pullout Tests per ASTM F543

    The conclusion of these tests was that the results were equivalent to the predicate device (LAGUNA® Pedicle Screw System, K083826 and K050060) and/or other similar implants and were deemed "sufficient for in vivo loading."

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1