Found 1 results

510(k) Data Aggregation

    K Number
    K030162

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2003-03-28

    (71 days)

    Product Code
    Regulation Number
    892.5730
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    IsoRay, Inc. Lawrence CSERION seeds are indicated for the treatment of malignant disease (e.g., head and neck, brain, breast, prostate, etc.) and may be used in surface, interstitial, and intracavitary applications for tumors with known radiosensitivity. The seeds may be used as a primary treatment or in conjunction with other treatment modalities, such as external beam radiation therapy, chemotherapy or as treatment for residual disease after excision of primary tumors.

    Device Description

    The IsoRay, Inc. Lawrence CSERION Model CS-1 is a small, cylindrical sealed source which contains the low energy gamma (X-ray) emitting radionuclide, cesium-131, adsorbed onto an internal inorganic substrate. The nominal external seed dimensions (4.5 mm length and 0.8 mm diameter) and patient-contacting material (titanium) are identical to predicate device(s).

    AI/ML Overview

    The provided text is a 510(k) Summary for the IsoRay, Inc. Lawrence CSERION Model CS-1, a brachytherapy source. This document focuses on demonstrating substantial equivalence to existing predicate devices, rather than establishing acceptance criteria and conducting a primary study to prove that the device meets those criteria.

    Therefore, the requested information regarding acceptance criteria, study details, sample sizes, expert qualifications, adjudication methods, MRMC studies, standalone performance, and ground truth for training and test sets is not available in this document.

    The document states that the device is "substantially equivalent" to predicate devices based on a comparison of parameters. The "Comparison Chart" on page 0 outlines these parameters. This comparison is the "study" demonstrating the device meets the criteria for substantial equivalence, implying that if the new device's parameters are similar enough to an already approved device, it is considered safe and effective.

    Here's a breakdown of what is provided, framed in terms of the closest available information to your request:

    1. A table of acceptance criteria and the reported device performance:

    The document doesn't provide "acceptance criteria" in the traditional sense of a performance study with specific thresholds. Instead, it presents a "Comparison Chart" to demonstrate substantial equivalence to predicate devices. The "acceptance criterion" here is essentially that the new device's parameters are comparable to those of already cleared devices.

    Parameters for Evaluating Substantial EquivalenceIsoRay, Inc. Lawrence CSERION (Reported Performance)Predicate Device(s) (Reference)
    Indications for UseMalignant DiseaseSame
    RadionuclideCs-131Same for K924261, I-125 for K914281, Pd-103 for K010283
    Half-Life (days)9.69Same for K924261, 59.4 for K914281, 17.0 for K010283
    Principle Energies (keV)29.5, 29.8, 33.6Same for K924261, 27.4, 31.4, 35.5 for K914281, 20-22 for K010283
    Patient-Contacting Capsule:Welded TitaniumSame
    Nominal External Length (mm)4.5Same
    Nominal External Diameter (mm)0.8Same
    Radiographic MarkerGold WireVarious (Predicate: Silver Rod, Lead Piece)
    Apparent Activity Range (mCi)0.20 to 50.00.1 to 100 (Predicate: 5.0 to 40, 0.1 to 10)
    External Contamination (uCi)< 0.005 uCiSame
    Implantation/Application MethodNeedles, Applicators, Tubing, Catheters, Expanders, etc.Same

    Regarding the other requested points:

    • 2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective): Not applicable. This is not a study assessing performance on a dataset of patient data. It's a comparison of device specifications.
    • 3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience): Not applicable. Ground truth for a clinical dataset is not relevant to a device specification comparison.
    • 4. Adjudication method (e.g. 2+1, 3+1, none) for the test set: Not applicable.
    • 5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable. This device is a brachytherapy seed, not an AI software.
    • 6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable. This device is a brachytherapy seed, not an algorithm.
    • 7. The type of ground truth used (expert concensus, pathology, outcomes data, etc): Not applicable. The "ground truth" here is the established characteristics and performance of the predicate device(s) as cleared by the FDA.
    • 8. The sample size for the training set: Not applicable. There is no AI model or training set involved.
    • 9. How the ground truth for the training set was established: Not applicable.

    In summary, the provided document is a 510(k) submission for a medical device (brachytherapy seeds) that seeks to demonstrate substantial equivalence to already approved devices. It is not a document detailing a performance study with acceptance criteria for an AI or diagnostic device. The "study" here is the comparison of technical specifications against predicate devices.