Found 5 results

510(k) Data Aggregation

    K Number
    K240266

    Validate with FDA (Live)

    Date Cleared
    2024-09-09

    (222 days)

    Product Code
    Regulation Number
    880.6250
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    This device is a disposable device intended for medical purposes that is worn on the examiner's hand to prevent contamination between patient and examiner.

    The tested chemotherapy drugs and permeation times are as follows:

    Bleomycin Sulfate 15 mg/mL, >240 minutes
    Busulfan 6 mg/mL, >240 minutes
    Carboplatin (Paraplatin) 10 mg/mL, >240 minutes
    Carmustine (BCNU) 3.3 mg/mL, 1.9 minutes
    Cisplatin 1.0 mg/mL, >240 minutes
    Cyclophosphamide (Cytoxan) 20 mg/mL, >240 minutes
    Cytarabine HC1 100 mg/mL, >240 minutes
    Dacarbazine (DTIC) 10 mg/mL, >240 minutes
    Daunorubicin HC1 5 mg/mL, >240 minutes
    Docetaxel 20 mg/mL, >240 minutes
    Doxorubicin Hydrochloride 2 mg/mL, >240 minutes
    Epirubicin HC1 (Ellence) 2 mg/mL, >240 minutes
    Etoposide (Toposar) 20 mg/mL, >240 minutes
    Fludarabine 25 mg/mL, >240 minutes
    Fluorouracil 50 mg/mL, >240 minutes
    Gemcitabine 38 mg/mL, >240 minutes
    Idarubicin HCl 1 mg/mL, >240 minutes
    lfosfamide 50 mg/mL, >240 minutes
    Irinotecan 20 mg/mL, >240 minutes
    Mechlorethamine HC1 1 mg/mL, >240 minutes
    Melphalan 5 mg/mL, >240 minutes
    Methotrexate 25 mg/mL, >240 minutes
    Mitomycin C 0.5 mg/mL, >240 minutes
    Mitoxantrone 2 mg/mL, >240 minutes
    Paclitaxel (Taxol) 6 mg/mL, >240 minutes
    Rituximab 10 mg/mL, >240 minutes
    Thiotepa 10 mg/mL, 3.7 minutes
    Trisenox 1 mg/mL . >240 minutes
    Vincristine Sulfate 1 mg/mL, >240 minutes

    Fentanyl tested as follows: Fentanyl citrate 100 mcg/2mL, >240 minutes

    WARNING: Not for use with Carmustine or Thiotepa.

    Device Description

    This device is a disposable device intended for medical purposes that is worn on the examiner's hand to prevent contamination between patient and examiner.

    AI/ML Overview

    I cannot fulfill this request. The provided text is an FDA 510(k) clearance letter for examination gloves. It does not contain any information about acceptance criteria or a study proving a device meets acceptance criteria related to a medical imaging AI device or similar complex medical technology.

    The text exclusively details the clearance of Latex Examination Gloves Powder Free (Ocean Blue and Natural White), citing regulation numbers, product codes, and the review process for market clearance. It mentions general controls, quality system regulations, UDI rules, and reporting of adverse events, all pertaining to the manufacturing and marketing of physical medical devices (gloves in this case), not performance studies of an AI algorithm or diagnostic device.

    Therefore, I cannot extract the information required to answer your specific questions about acceptance criteria, study design, ground truth, or expert involvement, as this information is not present in the provided document.

    Ask a Question

    Ask a specific question about this device

    K Number
    K161833

    Validate with FDA (Live)

    Date Cleared
    2017-05-23

    (322 days)

    Product Code
    Regulation Number
    880.6250
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    A patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.

    Device Description

    It is the powder-free variation of the class I latex patient examination gloves made by on-line polymercoating and mild on-line chlorination process. The process modifies the surface characteristics and causes it to remain tack-free without the use of any dusting or donning powder.

    AI/ML Overview

    The document is a 510(k) Premarket Notification for "Latex Examination Gloves Powder Free". It does not describe an AI/ML medical device, therefore, the requested information for acceptance criteria and study details for such a device is not applicable.

    The document focuses on demonstrating substantial equivalence to a predicate device (K152593) for patient examination gloves.

    However, I can extract the acceptance criteria and performance of the device based on the non-clinical tests mentioned, as they serve a similar purpose of demonstrating safety and effectiveness for this medical device.

    1. Table of acceptance criteria and the reported device performance

    ParameterAcceptance Criteria (ASTM Specifications)Reported Device Performance (K161833)Status
    Dimension (Length)
    Size XSmall220 mm min.224mm-236mmMeet Specification
    Size Small220 mm min.225mm-240mmMeet Specification
    Size Medium230 mm min.235mm-246mmMeet Specification
    Size Large230 mm min.235mm-245mmMeet Specification
    Size XLargeN/AN/AN/A
    Thickness
    Palm0.08 mm min.0.12mm-0.14mmMeet Specification
    Finger0.08 mm min.0.15mm-0.17mmMeet Specification
    Width
    Size XSmall70 ± 10 mm73mm-75mmMeet Specification
    Size Small80 ± 10 mm82mm-87mmMeet Specification
    Size Medium95 ± 10 mm94mm-97mmMeet Specification
    Size Large111 ± 10 mm102mm-107mmMeet Specification
    Size XLargeN/AN/AN/A
    Physical Properties
    Tensile Strength, Before Aging18 MPa min.21.18MPa-26.17MPaMeet Specification
    Stress at 500% Elongation5.5 MPa max.3.0MPa-4.2MPaMeet Specification
    Ultimate Elongation, Before Aging650% min.750.20%-820.20%Meet Specification
    Tensile Strength, After Aging14 MPa min.18.28MPa-23.88MPaMeet Specification
    Ultimate Elongation, After Aging500% min.550.40%-700.50%Meet Specification
    Water Leak Test, 1000 ml (AQL)
    Before Aging2.51.5 and belowMeet Specification
    After Aging2.52.5 and belowMeet Specification
    Extractable Protein
    Water Extractable Protein50 µg/dm²50 µg/dm² and belowMeet Specification
    Residual Powder Content
    Residual Powder Content2 mg/glove max.2 mg/glove and belowMeet Specification

    2. Sample size used for the test set and the data provenance
    The document does not explicitly state the sample sizes for each specific test (e.g., how many gloves were tested for length, thickness, or tensile strength). However, the tests were performed "as per ASTM D3578-05, Standard Specification for Rubber Examination Gloves and 21 CFR 800.20." These standards typically define the sampling plans. The data provenance is Malaysia, as the applicant, Careglove Global SDN BHD, is located in Seremban, Negeri Sembilan, Malaysia. The tests are non-clinical and would be considered prospective in the sense that they were conducted specifically for this submission.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
    Not applicable. This is not an AI/ML device that requires expert ground truth for image or data interpretation. The ground truth for the glove's performance is established by objective measurements against established industry standards (ASTM D3578-05).

    4. Adjudication method for the test set
    Not applicable, as it's not an AI/ML device requiring human adjudication of performance. The performance is determined by direct physical and chemical measurements.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
    Not applicable. This is not an AI/ML device.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
    Not applicable. This is not an AI/ML device.

    7. The type of ground truth used
    The ground truth is based on established industry standards and regulatory requirements, specifically:

    • ASTM D3578-05 (Standard Specification for Rubber Examination Gloves)
    • 21 CFR 800.20 (dealing with devices that prevent contamination)
    • ISO 10993-10 (Biological evaluation of medical devices - Tests for irritation and skin sensitization)

    These standards define the acceptable performance limits for the characteristics tested.

    8. The sample size for the training set
    Not applicable. This is not an AI/ML device, so there is no "training set."

    9. How the ground truth for the training set was established
    Not applicable. This is not an AI/ML device, so there is no "training set" or ground truth for it. The ground truth for the device's performance is established by the aforementioned ASTM and ISO standards for physical characteristics and biocompatibility.

    Ask a Question

    Ask a specific question about this device

    K Number
    K152593

    Validate with FDA (Live)

    Date Cleared
    2016-02-25

    (168 days)

    Product Code
    Regulation Number
    880.6250
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    A patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.

    Device Description

    LATEX EXAMINATION GLOVE POWDER FREE

    AI/ML Overview

    This document is a 510(k) premarket notification letter from the FDA regarding a Latex Examination Glove Powder Free. It does NOT contain information about acceptance criteria, device performance, a study, or any of the specific details requested in your prompt (such as sample size, data provenance, ground truth, or MRMC studies).

    This document is solely an approval letter for a medical device that falls under general controls.

    Ask a Question

    Ask a specific question about this device

    K Number
    K992031

    Validate with FDA (Live)

    Date Cleared
    1999-07-12

    (26 days)

    Product Code
    Regulation Number
    880.6250
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    A patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.

    Device Description

    Latex Examination Glove Powder Free, With Protein Labeling Claim (50 Micrograms or Less Total Water Soluble Protein)

    AI/ML Overview

    N/A

    Ask a Question

    Ask a specific question about this device

    K Number
    K974870

    Validate with FDA (Live)

    Date Cleared
    1998-03-13

    (74 days)

    Product Code
    Regulation Number
    880.6250
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    This gloves disposible and intended for medical purpose that is worn on the examiners hand to prevent contamination between patient and examiner.

    Device Description

    SMARTTEX LATEX EXAMINATION GLOVE POWDER FREE (Green, blue and natural color)

    AI/ML Overview

    This document is a 510(k) premarket notification decision letter from the FDA for a medical device, specifically SMARTTEX Latex Examination Gloves. It does not contain information about acceptance criteria or a study proving that the device meets acceptance criteria. The letter confirms that the device is substantially equivalent to legally marketed predicate devices, which means it has met the regulatory requirements for general controls, but it does not detail any specific performance studies or their results.

    Therefore, I cannot provide the requested information based on the provided text. The document is a regulatory approval, not a performance study report.