510(k) Data Aggregation

    K Number
    K103513

    Validate with FDA (Live)

    Date Cleared
    2011-10-24

    (328 days)

    Product Code
    Regulation Number
    888.3030
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The AOS Clavicle Plate System provides fixation for fractures, fusions, or osteotomies for the clavicle.

    Device Description

    The AOS Clavicle Plate system consists of bone plates and screws for fractures, fusions, and osteotomies of the clavicle bone. The bone plates are pre-shaped to fit the curves and angles of the clavicle and are provided in small, medium and large sizes. The plate accepts both locking and nonlocking screws.

    AI/ML Overview

    The provided text describes the AOS Clavicle Plate System and its substantial equivalence to predicate devices, focusing on mechanical testing rather than AI/software performance. Therefore, many of the requested criteria (e.g., sample size for test/training sets, expert qualifications, multi-reader multi-case studies) are not applicable or cannot be extracted from this document.

    Here's an analysis based on the information available:

    1. A table of acceptance criteria and the reported device performance

    Acceptance Criteria (What was tested)Reported Device Performance (Results)
    Static 4-point bend testing (in accordance with ASTM F382)Results demonstrate that the subject device is substantially equivalent to the predicates.
    Dynamic 4-point bend testing (in accordance with ASTM F382)Results demonstrate that the subject device is substantially equivalent to the predicates.

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    The document does not specify the sample size for the mechanical tests. The provenance of the data is not explicitly mentioned beyond being "PRECLINICAL TESTING" conducted to compare with predicate devices. This is a technical/mechanical study, not a clinical study involving patient data.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    Not applicable. This is a physical device subject to mechanical testing, not a diagnostic or AI device requiring expert interpretation of medical images to establish ground truth. The "ground truth" here is the performance metrics derived from the ASTM F382 standard.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Not applicable. Adjudication methods are relevant for studies involving human interpretation or subjective assessments, typically in clinical or diagnostic contexts. This is a mechanical testing study.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This device is a bone plate system, not an AI or diagnostic tool. No MRMC study was conducted.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. This is a physical medical device (bone plate), not an algorithm or software. Its performance is inherent in its physical and mechanical properties.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    The "ground truth" for this device's performance is established by validated mechanical testing standards (ASTM F382). The device's performance is compared against the performance of predicate devices as measured by these same standards to demonstrate substantial equivalence.

    8. The sample size for the training set

    Not applicable. This is not a machine learning or AI device that requires a training set.

    9. How the ground truth for the training set was established

    Not applicable. As above, there is no training set for this type of device.