510(k) Data Aggregation

    K Number
    K240334

    Validate with FDA (Live)

    Date Cleared
    2024-10-30

    (268 days)

    Product Code
    Regulation Number
    876.1500
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The LAGIS Endoscopic Instruments-Grasper has applications in a variety of minimally invasive procedures to facilitate grasping and mobilization of tissue.

    Device Description

    The LAGIS Endoscopic Instruments - Grasper is a single use device sterilized by ethylene oxide. It is composed of an insulating shaft, a rotation knob, grasping forceps, and a ratchet handle. The 5 mm diameter insulating shaft is designed for use through appropriate size trocars. The rotation knob located on the handle rotates the insulating shaft 360 degrees in either direction for better maneuverability. With the ratchet handle, the grasping forceps are allowed to be locked in place. The grasping forceps are activated by compression and release of the ring of the handle.

    AI/ML Overview

    The provided text is a 510(k) Summary for the LAGIS Endoscopic Instruments - Grasper. This type of FDA submission is for demonstrating substantial equivalence to a legally marketed predicate device, rather than proving performance against specific acceptance criteria through a clinical study that would involve assessing diagnostic or treatment effectiveness.

    Therefore, the information typically requested in your prompt (such as acceptance criteria for diagnostic performance, sample sizes for test sets, expert qualifications, and ground truth establishment) is not applicable to this document. This document focuses on demonstrating that the new device is as safe and effective as a previously cleared predicate device based on design characteristics, material similarities, and bench testing of functional performance.

    Here's an breakdown of the relevant information provided in the document based on your prompt, with an explanation of why certain criteria are not met by the available text:

    1. A table of acceptance criteria and the reported device performance

    There isn't a table of specific acceptance criteria in the sense of diagnostic performance or clinical outcomes. Instead, the "performance" section describes the types of bench tests conducted to ensure the device functions as intended and is comparable to the predicate device. The general acceptance criterion is that the device "performed acceptably compared to the predicate device."

    Acceptance Criteria (Implied from testing)Reported Device Performance
    Biocompatibility standards per ISO 10993 and 21 CFR Part 58Materials used were tested for cytotoxicity, sensitization, irritation, and systemic toxicity, demonstrating compliance.
    Sterilization validationSterilization by ethylene oxide was validated.
    Material characteristics match predicate or perform acceptablyMaterials are "similar to those in the predicate device." Tested for various characteristics.
    Functional performance (e.g., grasping force, tip hardness, corrosion resistance, bending performance, mobilization)"Test results showed that the device performed acceptably compared to the predicate device in terms of grasping force, tip hardness, corrosion resistance, mobilization, and bending performance, among other parameters."
    Mechanical integrity and functionality (e.g., rotation, locking)The device's design features (rotation knob, ratchet handle) imply these functions were tested and validated to perform as intended. "These studies included assessments of (...) functional performance, along with bending and ex vivo tissue testing."
    No new safety or effectiveness concerns compared to predicateMinor differences (additional tip types, wider tip opening, longer shaft, greater grip force) are stated not to "raise any safety or performance issues" or "introduce new questions regarding safety or effectiveness," as confirmed by performance testing.

    2. Sample size used for the test set and the data provenance

    • Sample Size for Test Set: Not specified. Bench testing typically involves a set number of devices to demonstrate consistency and compliance, but specific numbers are not mentioned here.
    • Data Provenance: The tests are "Performance bench tests" conducted by Lagis Enterprise Co., Ltd. The document does not specify a country of origin for data in the sense of patient data, as this is a device for grasping tissue and not a diagnostic or AI-driven tool. The data is from laboratory testing. It is a pre-market submission, not a post-market study or clinical trial.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    Not applicable. This device is an endoscopic instrument for grasping tissue, not a diagnostic tool that requires human expert interpretation for ground truth establishment. The "ground truth" for this device's performance would be engineering specifications and validated test methods.

    4. Adjudication method for the test set

    Not applicable for the reasons mentioned above.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This is not an AI-assisted diagnostic device.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. This is a mechanical surgical instrument.

    7. The type of ground truth used

    The "ground truth" for this device would be established engineering specifications, material standards, and functional performance benchmarks (e.g., a certain force must be achieved, a certain bending radius must be tolerated). The document states that "performance bench tests were conducted to verify the design characteristics and confirm that the device functions as intended." This implies adherence to pre-defined engineering and safety standards rather than a clinical ground truth like pathology or outcomes data.

    8. The sample size for the training set

    Not applicable. This is not a machine learning device.

    9. How the ground truth for the training set was established

    Not applicable. This is not a machine learning device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K141902

    Validate with FDA (Live)

    Date Cleared
    2015-05-26

    (316 days)

    Product Code
    Regulation Number
    876.1500
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Lagis Endoscopic Instruments Disposable Grasper has application in a variety of general, urologic, gyneoologic and endoscopic procedure for grasping and clamping of tissue and small tubular structures.

    Device Description

    The Lagis Endoscopic Instruments Disposable Grasper with 5mm diameter insulated shaft, 33cm in length, is designed for introduction and use through all appropriately sized trocar sleeves or larger sized trocar sleeves with the use of a converter. Same as predicate devices, the Lagis Endoscopic Instruments Disposable Grasper has a pair of plastic handles, which are compressed and released to close or open the instrument forceps. The rotation knob located on the handle rotates the shaft 360 degrees in either direction. Due to different functional requirements, the instrument forceps can be of various designs such as fenestrated/atraumatic/flat/ Endo-Clinch/Babcock/CROCE-OLMI grasping forceps. The materials of construction are similar to the predicate devices and are satisfactorily tested for cytotoxicity, sensitization, intracutaneous reactivity and systematic toxicity per ISO 10993 and 21 CFR, Part 58 requirements.

    AI/ML Overview

    The provided text is a 510(k) Summary for the Lagis Endoscopic Instruments Disposable Grasper. It describes the device, its intended use, and states that performance bench tests were conducted. However, it does not contain the detailed information necessary to complete most of the requested fields regarding acceptance criteria and the specifics of a study proving device performance in the context of AI/medical imaging.

    Here's an attempt to answer based only on the provided text, with clear indications of what information is missing:

    1. A table of acceptance criteria and the reported device performance

    Acceptance Criteria (Implied)Reported Device Performance
    BiocompatibilityAcceptable results
    Sterilization validationAcceptable results
    Material characteristicsAcceptable results
    Functional characteristicsAcceptable results
    Performance characteristicsAcceptable results
    Corrosion performanceAcceptable results (with respect to predicate devices)
    Bending performanceAcceptable results (with respect to predicate devices)
    SafetyDevice is safe (conclusion)
    EffectivenessDevice is effective (conclusion)
    Performs as intendedDevice performs well as intended (conclusion)
    Substantial Equivalence to Predicate DeviceSubstantially equivalent to the predicate device (conclusion)

    Note: The document generally states "acceptable results" or "safe, effective and performs well as intended" without providing specific thresholds or metrics for the acceptance criteria.

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    • Sample Size for Test Set: Not specified. The document mentions "performance bench tests" but does not detail the number of devices or trials.
    • Data Provenance: Not specified. The document describes bench tests which are typically laboratory-based, not patient data.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    • This information is not applicable and not provided. The study described involved "performance bench tests" of a physical medical instrument, not an AI or imaging device requiring expert ground truth establishment from medical data.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    • This information is not applicable and not provided, as the study was bench testing of a physical device.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • No, a multi-reader multi-case (MRMC) comparative effectiveness study was not done. This study pertains to a physical surgical instrument, not an AI system or imaging device requiring human reader analysis.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • This information is not applicable. The device is a physical endoscopic grasper, not an algorithm. The "performance bench tests" can be considered "standalone" in that they evaluate the device itself.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    • The concept of "ground truth" as typically applied to medical imaging/AI is not directly relevant here. For the bench tests, the "ground truth" would be established engineering specifications, material properties, and functional requirements against which the device's performance was measured. The document states tests were conducted to "verify design characteristics and to ensure that the device can be used as intended."

    8. The sample size for the training set

    • This information is not applicable and not provided. The device is a physical medical instrument, not a machine learning model that requires a training set.

    9. How the ground truth for the training set was established

    • This information is not applicable and not provided (see point 8).
    Ask a Question

    Ask a specific question about this device

    K Number
    K090309

    Validate with FDA (Live)

    Date Cleared
    2009-08-28

    (203 days)

    Product Code
    Regulation Number
    878.4400
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Lagis Endoscopic Instruments are designed for use in a variety of surgical procedures to facilitate grasping, dissecting, and transecting of tissues.

    Device Description

    The Lagis Endoscopic Instruments is a sterile, single use device having an insulated shaft with a diameter of 4.5-5.0 mm for use in 6mm and larger size trocar. The shaft can be rotated through the actions of a roller knob. A pair of plastic handles manipulates closing or opening of the tip scissors or graspers. An electrical connector allows for connection of the instrument to a generator for electrosurgical applications.

    AI/ML Overview

    The provided 510(k) summary describes a new medical device, Lagis Endoscopic Instruments, and its substantial equivalence to predicate devices. However, it does not contain the level of detail regarding acceptance criteria and study particulars that is typically found in a clinical study for AI/software-based medical devices.

    The document describes a physical medical instrument (electrosurgical instruments), not an AI or software-based device. Therefore, many of the typical acceptance criteria and study details relevant to AI/ML devices (like sample size for test sets, data provenance, number of experts for ground truth, MRMC studies, standalone performance, and training set details) are not applicable or not provided in the context of this device.

    Here's an attempt to answer the questions based on the provided text, while acknowledging the inherent differences for a hardware device:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not provide a formal table of acceptance criteria with specific numerical targets. Instead, it states that "Test results demonstrated acceptable results with respect to predicate devices." The performance evaluation focused on mechanical and electrical characteristics.

    Acceptance Criterion (Implied)Reported Device Performance
    Unit Packaging IntegrityAcceptable results
    Functional CharacteristicsAcceptable results (e.g., jaw/scissor movement)
    Performance CharacteristicsAcceptable results (e.g., tip rotation, shaft sturdiness)
    Electrical SafetyAcceptable results (e.g., shaft dielectric strength, monopolar connection capability)
    Grasping ForceAcceptable results (in respect to predicate devices)
    Cutting CapabilityAcceptable results (in respect to predicate devices)
    BiocompatibilitySatisfactorily tested for cytotoxicity, sensitization, intracutaneous reactivity and systematic toxicity per ISO 10993 and US FDA G-95 requirements

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not specify the exact sample sizes used for the bench tests. It broadly mentions "Performance bench tests were carried out." The "data provenance" in this context refers to the origin of the test results, which are from internal bench tests conducted by the manufacturer, Lagis Enterprise Co. Ltd. This is a retrospective evaluation against established engineering and safety standards, rather than prospective clinical data.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications

    This question is not applicable to this type of device and study. Bench tests for electrosurgical instruments typically involve engineering and quality control personnel following defined test protocols and specifications, not clinical experts establishing "ground truth" in the way it's done for diagnostic AI. The "ground truth" here is adherence to engineering specifications and safety standards.

    4. Adjudication Method for the Test Set

    This question is not applicable. The performance tests are objective measurements against specifications (e.g., torque, force, electrical resistance), not subjective assessments requiring adjudication by multiple readers.

    5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    This question is not applicable. This is a hardware medical device, not an AI or software-based diagnostic or assistive system that would involve human readers.

    6. If a Standalone (i.e. algorithm only without human-in-the loop performance) was done

    This question is not applicable. This is a physical instrument, not an algorithm.

    7. The Type of Ground Truth Used

    The "ground truth" for this device's performance evaluation is based on engineering specifications, established safety standards (e.g., ISO 10993 for biocompatibility), and comparison to the performance characteristics of predicate devices. For example, a "grasping force" test would have an engineering specification for the minimum/maximum force the jaws should exert, which serves as the ground truth.

    8. The Sample Size for the Training Set

    This question is not applicable. This is a physical device, and the concept of a "training set" is not relevant to its development or pre-market notification process.

    9. How the Ground Truth for the Training Set Was Established

    This question is not applicable, as there is no "training set" for this type of device.