Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K172483

    Validate with FDA (Live)

    Date Cleared
    2018-06-21

    (309 days)

    Product Code
    Regulation Number
    880.5570
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    This device is intended for use to inject fluids into or withdraw fluids from parts of the surface of the skin

    Device Description

    The device consists of a metal tube that is beveled at one end and at the other end join to a female connector (hub). The needle cap covers intended to provide physical protection to the needle tube. This product is packed by blister paper and sterilized by E.O. gas.

    AI/ML Overview

    The provided text describes the acceptance criteria and performance data for the Kopac Sterile Needle as part of its 510(k) submission (K172483) to the FDA.

    Here's the breakdown of the requested information:

    1. Table of Acceptance Criteria and Reported Device Performance

    Requirement – Test (ISO 7864)Acceptance Criteria (Implied)Reported Device Performance
    Inner and Outer SurfaceNo bumps and flaws on the outer surface. Smooth, electrolytically polished or equivalent. No hazardous oxides, dusts, and powders on the inner surface. Sharply polished blade without visible bending, uniform cross-section and thickness.Pass
    When injecting glycerin into the needle tube or needle hub, coloring of glycerin should not take place.Glycerin should not cause coloring.Pass
    When a lubricant is applied on the needle tube, the lubricant must be non-toxic and satisfy ISO 7864. When observed by unaided eyes, no lubricant drops should be found on the inner and outer surfaces of the needle tube.Lubricant must be non-toxic, satisfy ISO 7864, and no lubricant drops visible on inner/outer surfaces.Pass
    Dimensions (Measured using digital vernier calipers and digital thickness gauge)Outer diameter of the needle tube must be satisfied. Length of the needle tube must be satisfied.Pass
    The needle hub and hub holeWhen the test gauge indicated on the figure below is inserted into the hub hole at low pressure, the taper must fit with the taper gauge of the hub hole. Also, the nozzle of the needle hub must be within the gauge limit.Pass
    ElasticityReturn to original position when bent to 12° using a protractor.Pass
    Flexural strengthNo snapping when bent to 90° using a protractor.Pass
    PulloutMeasurement of maximum value using a tension and compression tester (specific criteria not detailed, but implied to meet an acceptable threshold).Pass
    Stiffness (ISO 9626 Annex B)When tested in accordance with the method given in ISO 9626 Annex B, apply a downward force given in table at a speed 1 mm/min. Measure the deflection of the tubing at the point of application of the force (specific deflection limits not detailed, but implied to meet an acceptable threshold within the standard).Pass
    Resistance to breakage (ISO 9626 Annex C)The cannula edge shall be sharply ground to a regular cross-section and thickness, with no visible bending. No visible breakage when examined by normal or corrected vision.Pass
    Resistance to corrosion (ISO 9626 Annex D)The tubing is partially immersed in sodium chloride solution ($\text{c(NaCl)} = 0.5mol/L$) at ($\text{23±2}$)°C, for 7h±5min. The immersed half of the tubing shall show no evidence of corrosion resulting from the test compared visually with the unimmersed portion.Pass
    Biocompatibility (ISO 10993 Series)Cytotoxicity: Pass (ISO 10993-5 criteria). Skin Sensitization: Pass (ISO 10993-10 criteria). Intracutaneous Reactivity: Pass (ISO 10993-10, maximization test for delayed hypersensitivity criteria). Acute Systemic Toxicity: Pass (ISO 10993-11 criteria). Pyrogen Test: Pass (ISO 10993-11, Annex F criteria). Hemolysis Test: Pass (ISO 10993-4 criteria).Pass
    Sterility and LAL testLAL test: Pass (USP39 <85>, Bacterial Endotoxins Test, Unit: EU/Device - specific limit not provided but implied to be below the threshold). E.O sterilization validation: Pass (ISO 11135:2014, EO 30%, CO2 70%, Temp: 55°C, Exposure time: 5 hours - implying successful sterilization). Sterility test: Pass (ISO 11737-2 criteria - implying no microbial growth). E.O Residual test: Pass (ISO 10993-7:2008 criteria).Pass

    2. Sample size used for the test set and the data provenance

    The document does not specify the exact sample sizes for each of the bench tests. It broadly states “Bench test were performed.” and lists various types of tests. Data provenance is not explicitly mentioned (e.g., country of origin, retrospective/prospective), but these are standard bench tests typically performed in a laboratory setting by the manufacturer (Poonglim Pharmatech Inc., Korea).

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    This information is not applicable to this submission. The "ground truth" for a device like a hypodermic needle is established through objective physical, chemical, and biological measurements against international standards (ISO, USP) and predefined specifications, not through expert consensus on interpretation.

    4. Adjudication method for the test set

    Not applicable. The tests are quantitative or qualitative assessments against defined standards, not subjective interpretations requiring adjudication.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This is a submission for a medical device (hypodermic needle), not an AI/software as a medical device (SaMD) that involves human readers or AI assistance.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

    Not applicable. This is a physical medical device.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    The "ground truth" for the performance tests relies on:

    • International Standards: Primarily ISO 7864 (Hypodermic needles for single use), ISO 9626 (Stainless steel needle tubing for the manufacture of medical devices — Requirements and test methods), ISO 10993 series (Biological evaluation of medical devices), ISO 11135 (Sterilization of health-care products — Ethylene oxide), ISO 11737-2 (Sterilization of medical devices — Microbiological methods), and USP39 <85> (Bacterial Endotoxins Test).
    • Defined Specifications: Manufacturer's internal specifications for dimensions, material properties, and visual characteristics.
    • Objective Measurements: Using instruments like digital vernier calipers, digital thickness gauges, protractors, tension and compression testers, and laboratory analyses for chemical and biological properties.

    8. The sample size for the training set

    Not applicable. This is a physical medical device, not an AI model requiring a training set.

    9. How the ground truth for the training set was established

    Not applicable. No training set for an AI model was used.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1