Search Results
Found 1 results
510(k) Data Aggregation
(49 days)
Kindguard Powder Free Vinyl Patient Examination Gloves, Light Yellow Color is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.
PVC films form a barrier to prevent contamination between patient and examiner. The PVC rubber is water tight under normal conditions of use. It's tensile properties cause it to conform to the hand, allowing movements necessary for a medical procedure. The Poly (vinyl chloride) glove acts as a barrier to prevent contamination between patient and examiner. The glove is manufactured in accordance with the requirements of ASTM D5250 and ASTM D5151.
The provided document describes a 510(k) premarket notification for Kindguard Powder Free Vinyl Patient Examination Gloves. This is a Class I medical device, and the evaluation for substantial equivalence primarily relies on non-clinical performance data and comparison to a predicate device, rather than a study involving an AI algorithm or human readers.
Therefore, many of the requested elements for describing the acceptance criteria and the study that proves the device meets those criteria (especially those related to AI, human readers, ground truth establishment, sample sizes for training/test sets, and expert adjudication) are not applicable to this type of device and submission.
However, I can extract the relevant information regarding the acceptance criteria and the performance of this specific device based on the provided text.
Here's a breakdown of the requested information, with NA (Not Applicable) for elements not relevant to this submission type:
Acceptance Criteria and Device Performance for Kindguard Powder Free Vinyl Patient Examination Gloves
This submission is for a Class I medical device (patient examination gloves), which does not involve an AI algorithm or complex human interpretation tasks. Thus, the "study" proving acceptance is primarily a set of non-clinical performance tests and a comparison to a predicate device.
1. Table of Acceptance Criteria and Reported Device Performance:
| Characteristics | Standard/Acceptance Criteria | Reported Device Performance (Kindguard Gloves) | Comparison Result |
|---|---|---|---|
| Form Factor / Dimensions | |||
| Length | ≥230mm (ASTM D 5250-06) | 231-241mm (Small, Medium, Large, X-Large) | Similar |
| Width (Small) | 80-90 mm (ASTM D 5250-06) | 81-89mm | Similar |
| Width (Medium) | 90-100 mm (ASTM D 5250-06) | 93-99mm | Similar |
| Width (Large) | 100-110 mm (ASTM D 5250-06) | 102-110mm | Similar |
| Width (X-Large) | 110-120 mm (ASTM D 5250-06) | 111-119mm | Similar |
| Thickness (Fingertip) | ≥0.05mm (ASTM D 5250-06) | 0.05-0.10mm | Similar |
| Thickness (Palm) | ≥0.08mm (ASTM D 5250-06) | 0.09-0.13mm | Similar |
| Physical Properties | |||
| Tensile Strength (Before & After aging) | ≥11MPa (ASTM D 5250-06) | 15-25 MPa (Before & After aging) | Same |
| Elongation Rate (Before & After aging) | ≥300% (ASTM D 5250-06) | 340-410% (Before & After aging) | Same |
| Freedom from Pinholes | Passed Standard Acceptance Criteria (21 CFR 800.20, ASTM D5250-06, ASTM D5151-06) | Holes at Inspection Level I, AQL2.5 | Passed / Same |
| Powder Residual | <2mg/glove (ASTM D 5250-06, D6124-06) | 0.3mg/glove | Meets |
| Biocompatibility | |||
| Primary Skin Irritation | Passes (ISO 10993-10: 2010-08-01 - "Not a primary skin irritant") | Passes (Under the conditions of the study, the device is not a primary skin irritant) | Passed |
| Dermal Sensitization | Passes (ISO 10993-10: 2010-08-01 - "Not a skin sensitizer") | Passes (Under the conditions of the study, the device is not a skin sensitizer) | Passed |
2. Sample size used for the test set and the data provenance:
- Sample Size for Testing: The specific number of gloves tested for each characteristic is not explicitly stated in this summary. However, tests were conducted according to the relevant ASTM and ISO standards, which define the required sample sizes for quality control and batch testing (e.g., AQL levels for pinholes).
- Data Provenance: The device manufacturer, Anhui Kindguard Medical Supplies Technology Co.,Ltd, is based in China. The testing would have been performed by or for the manufacturer. The data is based on prospective testing of the manufactured product to demonstrate compliance with standards.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
- NA. For this type of device (patient examination gloves), "ground truth" is established through standardized physical, chemical, and biological tests (e.g., measuring tensile strength, water leak tests, biocompatibility assays) according to international standards (ASTM, ISO). There is no "expert consensus" on image interpretation or diagnostic performance in the traditional sense. The "experts" would be the qualified personnel performing these laboratory tests.
4. Adjudication method (e.g., 2+1, 3+1, none) for the test set:
- NA. Adjudication methods like 2+1 or 3+1 are used for establishing ground truth in subjective assessments, particularly in medical imaging where multiple readers interpret cases. This is not applicable to the objective physical and chemical testing of medical gloves.
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
- NA. This device does not involve AI or human readers for diagnostic purposes.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:
- NA. This device does not use an algorithm.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):
- The "ground truth" for this device is based on objective, standardized measurements and laboratory test results performed according to specified ASTM and ISO standards (e.g., ASTM D5250 for dimensions and physical properties, ASTM D5151 for freedom from pinholes, D6124 for powder residual, ISO 10993-10 for biocompatibility). These standards define the acceptable range or pass/fail criteria.
8. The sample size for the training set:
- NA. This device does not involve a "training set" in the context of machine learning or AI algorithms.
9. How the ground truth for the training set was established:
- NA. See point 8.
Ask a specific question about this device
Page 1 of 1