510(k) Data Aggregation

    K Number
    K183505

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2019-05-17

    (150 days)

    Product Code
    Regulation Number
    884.5300
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    This product is intended for penile, vaginal and/or anal application to moisturize and lubricate, to enhance the ease and comfort of intimate sexual activity and supplement the body's natural lubrication. This product is compatible with natural rubber latex and polyisoprene condoms. This product is not compatible with polyurethane condoms.

    Device Description

    KY Banksy Moisture will be marketed as a non-sterile, non-prescription medical device for over-thecounter (OTC) use. This device is a water-based personal lubricant that contains: Water, Glycerin, Propanediol, Xanthan Gum, Benzoic Acid, Sodium Hyaluronate, Potassium Lactate and Lactic Acid. The tube material is composed of High-Density Polyethylene (HDPE) and Low-Density Polyethylene (LDPE), covered with a 50-micron aluminum/inear low-density polyethylene seal and fitted with the Polypropylene (PP) cap. The tube will be packaged in an outer cardboard carton.

    AI/ML Overview

    The provided document is a 510(k) premarket notification for a medical device called "KY Banksy Moisture," a personal lubricant. It doesn't contain information about AI/ML models, human-in-the-loop performance, or a comparative effectiveness study. Therefore, I cannot provide details on those aspects.

    However, I can extract the acceptance criteria and the studies performed to demonstrate substantial equivalence to a predicate device.

    Here's the information based on the provided text:

    1. Table of acceptance criteria and the reported device performance

    PropertyAcceptance Criteria (Specification)Reported Device Performance
    AppearanceClear colorless and homogeneous gel, free from extraneous matterClear colorless and homogeneous gel, free from extraneous matter
    OdorOdorlessOdorless
    ViscosityRelease: 3,000 – 6,500 cps; Stability: 3,000 - 8,000 cpsMaintained within specified ranges for release and stability.
    pH at 25°C3.5 – 4.5Maintained within the specified range.
    Benzoic AcidRelease: 0.207 – 0.253 %w/w; Stability: 0.150 - 0.253 %w/wMaintained within specified ranges for release and stability.
    Osmolality1050 to 1250 mOsm/kgMaintained within the specified range.
    Total Yeast and Mold Count (TYMC)≤10 cfu/g (per EP 8.0 Section: 2.6.12)<10 cfu/g
    Total Aerobic Microbial Count (TAMC)≤100 cfu/g (per EP 8.0 Section: 2.6.12)<100 cfu/g
    Preservative Effectiveness (Escherichia coli)NLT 3.0 log reduction from the initial count at 7 days, and no increase from the 7 days' count at 28 days (per EP 8.0 Section: 5.1.3)Device met the criteria.
    Preservative Effectiveness (Pseudomonas aeruginosa)NLT 3.0 log reduction from the initial count at 7 days, and no increase from the 7 days' count at 28 days (per EP 8.0 Section: 5.1.3)Device met the criteria.
    Preservative Effectiveness (Staphylococcus aureus)NLT 3.0 log reduction from the initial count at 7 days, and no increase from the 7 days' count at 28 days (per EP 8.0 Section: 5.1.3)Device met the criteria.
    Preservative Effectiveness (Candida albicans)2.0 log reduction from the initiation count at 14 days, and no increase from the 14 days' count at 28 days (per EP 8.0 Section: 5.1.3)Device met the criteria.
    Preservative Effectiveness (Aspergillus niger)2.0 log reduction from the initiation count at 14 days, and no increase from the 14 days' count at 28 days (per EP 8.0 Section: 5.1.3)Device met the criteria.
    Total Specified Organisms (e.g., Pseudomonas aeruginosa)Absent (per EP 8.0 Section: 2.6.13)Absent
    Total Specified Organisms (e.g., Staphylococcus aureus)Absent (per EP 8.0 Section: 2.6.13)Absent
    Total Specified Organisms (e.g., Candida albicans)Absent (per EP 8.0 Section: 2.6.13) -----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------Absent -------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
    Biocompatibility (Cytotoxicity)Showed no evidence of causing cell lysis or cytotoxicity to L-929 cells, indicating non-cytotoxic.Not cytotoxic.
    Biocompatibility (Vaginal Irritation)An Irritation Index of 0, with macroscopically normal vaginal tissue.Non-irritant to vaginal tissue.
    Biocompatibility (Sensitization)No evidence of causing delayed dermal contact sensitization.Not a sensitizer.
    Biocompatibility (Acute Systemic Toxicity)No mortality or evidence of systemic toxicity.Met the requirements of ISO10993-11.
    Shelf-LifeMaintain specifications over a 2-year duration.Maintained specifications for 2 years. -----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
    Condom Compatibility (Natural Rubber Latex)Compatible. -----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------Compatible.
    Condom Compatibility (Polyisoprene)Compatible. -----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------Compatible.
    Condom Compatibility (Polyurethane)Not compatible. -----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------Not compatible.

    2. Sample size used for the test set and the data provenance

    • Sample Size: Not explicitly stated for each test, but studies were conducted.
    • Data Provenance: The document does not specify the country of origin of the data or whether the studies were retrospective or prospective.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    • This information is not applicable as the studies are for physical/chemical properties and biological compatibility, not for expert interpretation of medical images or data. No "ground truth" in the clinical imaging sense is established by experts here.

    4. Adjudication method for the test set

    • Not applicable for these types of laboratory and physical property tests.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, if so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • No, an MRMC comparative effectiveness study was not done. This device is a personal lubricant, not an AI-assisted diagnostic tool.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • Not applicable as this is not an algorithm-based device. The performance tests are for the physical, chemical, and biological properties of the lubricant itself.

    7. The type of ground truth used

    • The "ground truth" for demonstrating the device meets acceptance criteria is based on:
      • Scientific Standards: Conformance to specified physical, chemical, and microbiological parameters (e.g., viscosity, pH, osmolality, microbial counts).
      • Biological Standards: Results from established biocompatibility tests (cytotoxicity, irritation, sensitization, acute systemic toxicity) according to ISO 10993 standards.
      • Performance Standards: Compatibility testing with condoms according to ASTM D7661-10.
      • Stability Studies: Accelerated shelf-life testing.

    8. The sample size for the training set

    • Not applicable as this is not an AI/ML device that requires a training set.

    9. How the ground truth for the training set was established

    • Not applicable as this is not an AI/ML device that requires a training set or its associated ground truth establishment.