Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K032413

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2003-10-10

    (66 days)

    Product Code
    Regulation Number
    888.3030
    Age Range
    All
    Reference & Predicate Devices
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    KLS-Martin Sternal Plating System is intended for use in stabilization and fixation of anterior chest wall fractures including Sternal Fixation subsequent to Sternotomy and Sternal reconstructive procedures.

    Device Description

    KLS - Martin Sternal Plating System consists of plates having a thickness of 1.0mm to 3.0mm and screws having a diameter of 2.3mm to 3.2mm.

    AI/ML Overview

    The KLS-Martin Sternal Plating System is a Class II device intended for stabilization and fixation of anterior chest wall fractures, including sternal fixation subsequent to sternotomy and sternal reconstructive procedures.

    Here's an analysis of the provided information:

    1. Table of Acceptance Criteria and Reported Device Performance

    The provided document does not explicitly state quantitative acceptance criteria in a dedicated table format. However, it establishes substantial equivalence and superiority based on specific biomechanical performance compared to a predicate device and cerclage wires.

    Acceptance Criteria (Implied)Reported Device Performance
    Substantially equivalent in strength to the Lorenz Sternal Closure System with Modular Screws (K011079).The KLS-Martin Sternal Plating System's plates and screws were asserted to be "substantially equivalent in strength to the Lorenz Sternal Closure System with Modular Screws (K011079)."
    Superiority to sternum wiring (cerclage wires) in terms of fixation strength, stiffness, and lower lateral displacement.Clinical studies supporting superiority of sternal plating over wiring were cited (Sargent et al., Cheng et al., Buchman et al.). In the Ozaki et al. study (Phase II), "sterna fixed with rigid custom titanium 'H' plates [KLS-Martin prototype] showed a greater stiffness and a lower lateral displacement when compared with sterna fixed with stainless steel cerclage wires."
    Biomechanically superior to wire fixation.The Ozaki et al. study concluded: "... this study has led to the development of a rigid Sternal Plating System that is biomechanically superior to wire fixation."

    2. Sample Size Used for the Test Set and Data Provenance

    Only one specific study with biomechanical testing is detailed: "Biomechanical Study of Sternal Closure Using Rigid Fixation Techniques in Human Cadavers. Ozaki et al.".

    • Sample Size: Not explicitly stated for the cadaver tests in the provided text.
    • Data Provenance: Human cadavers were used for the biomechanical study. The location or institution where the study was conducted is not specified beyond listing "Ozaki et al.". The study appears to be retrospective in the sense that existing systems were compared, but the testing itself was prospective on the cadaveric samples.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    This device's acceptance is based on biomechanical testing and comparison to predicate devices, not on a diagnostic accuracy study requiring expert adjudication of "ground truth" in the traditional sense of medical image analysis. Therefore, this question is not directly applicable. The "ground truth" here is the physical measurement of strength, stiffness, and displacement.

    4. Adjudication Method for the Test Set

    Not applicable. The biomechanical study involved quantitative measurements, not subjective evaluations requiring adjudication.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    No. The provided document concerns a Sternal Plating System, which is a surgical implant, not an AI or imaging device that would typically undergo MRMC studies. The "clinical studies" mentioned (Sargent et al., Cheng et al., Buchman et al.) refer to studies on sternal plating in general versus wiring, not a specific MRMC study comparing human performance with and without this particular device.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    Not applicable. This is a physical surgical device, not an algorithm.

    7. The Type of Ground Truth Used

    For the Ozaki et al. biomechanical study, the "ground truth" was established through physical biomechanical measurements of stiffness and lateral displacement on human cadavers. For the overall claim of substantial equivalence, it was based on comparison of performance data against predicate devices and clinical literature assessment.

    8. The Sample Size for the Training Set

    Not applicable. This is a physical device, and there is no "training set" in the context of machine learning or AI.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable for the same reasons as above.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1