Found 5 results

510(k) Data Aggregation

    K Number
    K201052

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2020-08-31

    (132 days)

    Product Code
    Regulation Number
    882.4310
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The KLS Martin Individual Patient Solutions (IPS) Planning System is intended for use as a software system and image segmentation system for the transfer of imaging information from a computerized tomography (CT) medical scan. The input data file is processed by the IPS Planning System and the result is an output data file that may then be provided as digital models or used as input to a rapid prototyping portion of the system that produces physical outputs including anatomical models, guides, and case reports for use in the marking and cutting of cranial bone in cranial surgery. The IPS Planning System is also intended as a pre-operative software tool for simulating / evaluating surgical treatment options. Information provided by the software and device output is not intended to eliminate, replace, or substitute, in whole or in part, the healthcare provider's judgment and analysis of the patient's condition.

    Device Description

    The KLS Martin Individual Patient Solutions (IPS) Planning System is a collection of software and associated additive manufacturing (rapid prototyping) equipment intended to provide a variety of outputs to support reconstructive cranial surgeries. The system uses electronic medical images of the patients' anatomy (CT data) with input from the physician, to manipulate original patient images for planning and executing surgery. The system processes the medical images and produces a variety of patient specific physical and/or digital output devices which include anatomical models, guides, and case reports for use in the marking and cutting of cranial bone in cranial surgery.

    AI/ML Overview

    The provided text is a 510(k) summary for the KLS Martin Individual Patient Solutions (IPS) Planning System. It details the device, its intended use, and comparisons to predicate and reference devices. However, it does not describe specific acceptance criteria and a study dedicated to proving the device meets those criteria in the typical format of a diagnostic AI/ML device submission.

    Instead, the document primarily focuses on demonstrating substantial equivalence to a predicate device (K182889) and leveraging existing data from that predicate, as well as two reference devices (K182789 and K190229). The "performance data" sections describe traditional medical device testing (tensile, biocompatibility, sterilization, software V&V) and a simulated design validation testing and human factors and usability testing rather than a clinical study evaluating the accuracy of an AI/ML algorithm's output against a ground truth.

    Specifically, there is no mention of:

    • Acceptance criteria for an AI/ML model's performance (e.g., sensitivity, specificity, AUC).
    • A test set with sample size, data provenance, or ground truth establishment details for AI/ML performance evaluation.
    • Expert adjudication methods, MRMC studies, or standalone algorithm performance.

    The "Simulated Design Validation Testing" and "Human Factors and Usability Testing" are the closest sections to a performance study for the IPS Planning System, but they are not framed as an AI/ML performance study as requested in the prompt.

    Given this, I will extract and synthesize the information available regarding the described testing and attempt to structure it to address your questions, while explicitly noting where the requested information is not present in the provided document.


    Acceptance Criteria and Device Performance (as inferred from the document)

    The document primarily states that the device passes "all acceptance criteria" for various tests, but the specific numerical acceptance criteria (e.g., minimum tensile strength, maximum endotoxin levels) and reported performance values are generally not explicitly quantified in a table format. The closest to "performance" is the statement that "additively manufactured titanium devices are equivalent or better than titanium devices manufactured using traditional (subtractive) methods."

    Since the document doesn't provide a table of acceptance criteria and reported numerical performance for an AI/ML model's accuracy, I will present the acceptance criteria and performance as described for the tests performed:

    Test CategoryAcceptance Criteria (as described)Reported Device Performance (as described)
    Tensile & Bending TestingPolyamide guides can withstand multiple sterilization cycles without degradation and can maintain 85% of initial tensile strength. Titanium devices must be equivalent or better than those manufactured using traditional methods.Polyamide guides meet criteria. Additively manufactured titanium devices are equivalent or better than traditionally manufactured ones.
    Biocompatibility TestingAll biocompatibility endpoints (cytotoxicity, sensitization, irritation, chemical/material characterization, acute systemic, material-mediated pyrogenicity, indirect hemolysis) must be within pre-defined acceptance criteria.All conducted tests were within pre-defined acceptance criteria, adequately addressing biocompatibility.
    Sterilization TestingSterility Assurance Level (SAL) of 10^-6 for dynamic-air-removal cycle. All test method acceptance criteria must be met.All test method acceptance criteria were met.
    Pyrogenicity TestingEndotoxin levels must be below the USP allowed limit for medical devices that have contact with cerebrospinal fluid (< 2.15 EU/device) and meet pyrogen limit specifications.Devices contain endotoxin levels below the USP allowed limit (< 2.15 EU/device) and meet pyrogen limit specifications.
    Software Verification and ValidationAll software requirements and specifications are implemented correctly and completely, traceable to system requirements. Conformity with pre-defined specifications and acceptance criteria. Mitigation of potential risks. Performs as intended based on user requirements and specifications.All appropriate steps have been taken to ensure mitigation of any potential risks and performs as intended based on the user requirements and specifications.
    Simulated Design Validation Testing"Passed all acceptance criteria regardless of age or size" for representative cranial case extrapolated to six age ranges. Manufacturable at a high and acceptable level of fidelity, independent of feature size, age of patient, and device size.Demonstrated that the subject devices passed all acceptance criteria regardless of age or size. Confirms manufacturability at a high and acceptable level of fidelity, independent of feature size, age of patient, and device size.
    Human Factors and Usability TestingNo potential risks or concerns, outside of those previously raised and mitigated in the IFU, are found. Clinical experts confirm testing and outputs are applicable to real life situations and can be used to effectively execute a planned cranial procedure.No potential risks or concerns were found (outside of those mitigated in IFU). All clinical experts confirmed the testing and outputs were applicable to real life situations and could be used to effectively execute a planned cranial procedure (pediatric or adult patients).

    Detailed Study Information (Based on available text):

    1. Sample size used for the test set and the data provenance:

      • Test Set for Simulated Design Validation Testing: A "representative cranial case" was "extrapolated to six (6) distinct age ranges for input data (CT scan) equals output data validation." This implies 6 simulated cases were tested, but no further details on the number of actual CT scans or patients are provided.
      • Test Set for Human Factors and Usability Testing: "Eighteen (18) cases were analyzed" (6 distinct age ranges, with outputs sent to 3 clinical experts, meaning 6 (age ranges) x 3 (experts) = 18 cases analyzed in total by the experts).
      • Data Provenance: Not specified for the "representative cranial case" in simulated design validation. For human factors, it implicitly used outputs derived from the "six (6) distinct age ranges" based on the system's processing. The document does not specify if the data was retrospective or prospective, or the country of origin.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

      • Simulated Design Validation Testing: Not explicitly stated that experts established ground truth for this test. It seems to be a technical validation against the design specifications.
      • Human Factors and Usability Testing: "Three separate clinical experts" were used to review the outputs. Their qualifications are not specified beyond being "clinical experts." Their role was to analyze for potential use problems and make recommendations, and confirm applicability to real-life situations. This is not the establishment of ground truth in the sense of a diagnostic classification.
    3. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

      • No adjudication method is described for either the simulated design validation or human factors/usability testing. The human factors testing involved reviews by multiple experts, but no process for reconciling disagreements or establishing a consensus "ground truth" among them is mentioned; they each provided independent feedback.
    4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

      • An MRMC comparative effectiveness study was not conducted according or described in this document. The device is not presented as an AI-assisted diagnostic tool that improves human reader performance in the traditional sense. It's a pre-operative planning system that processes CT data to create physical/digital outputs. The "Human Factors and Usability Testing" involved multiple readers (clinical experts) and multiple cases, but it was for usability assessment rather than a comparative effectiveness study of AI assistance.
    5. If a standalone (i.e. algorithm only without human-in-the loop performance) was done:

      • The document describes "Software Verification and Validation" which is a form of standalone testing for the software applications used. It states that "all software requirements and specifications were implemented correctly and completely." However, this is a validation of the software's functionality and adherence to specifications, not a performance study of an AI/ML algorithm's accuracy in a diagnostic context. The system is explicitly described as requiring "trained employees/engineers who utilize the software applications to manipulate data and work with the physician to create the virtual planning session," indicating a human-in-the-loop process for generating the final outputs.
    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc):

      • Simulated Design Validation Testing: The "ground truth" appears to be the "initial input data (.STL)" and the design specifications; the test verifies that "output data (CT scan) equals output data validation" (likely intended to mean input equals output, or input from CT leads to correct output).
      • Human Factors and Usability Testing: The "ground truth" is effectively the "expert opinion" of the three clinical experts regarding the usability and applicability of the outputs, rather than a definitive medical diagnosis.
    7. The sample size for the training set:

      • The document describes the KLS Martin IPS Planning System as using "commercially off-the-shelf (COTS) software applications" (Materialise Mimics and Geomagic® Freeform PlusTM) for image segmentation and manipulation. This implies that the core algorithms were pre-existing and not developed by KLS Martin as a novel AI/ML model that would require a distinct training set outlined in this submission. Therefore, no information on a training set size is provided for the device.
    8. How the ground truth for the training set was established:

      • Not applicable, as no training set for a novel AI/ML model by KLS Martin is described. The COTS software validation would have been performed by their respective developers.
    Ask a Question

    Ask a specific question about this device

    K Number
    K192979

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2020-03-11

    (139 days)

    Product Code
    Regulation Number
    888.3030
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The KLS Martin Individual Patient Solutions (IPS) Planning System is intended for use as a software system and image segmentation system for the transfer of imaging information from a medical scanner such as a CT based system. The input data file is processed by the IPS Planning System and the result is an output data file that may then be provided as digital models or used in an additive manufacturing portion of the system that produces physical outputs including anatomical models, guides, and case reports for use in thoracic (excluding spine) and reconstructive surgeries. The IPS Planning System is also intended as a pre-operative software tool for simulating surgical treatment options.

    Device Description

    The KLS Martin Individual Patient Solutions (IPS) Planning System is a collection of software and associated additive manufacturing equipment intended to provide a variety of outputs to support thoracic (excluding spine) and reconstructive surgeries. The system uses electronic medical images of the patients' anatomy (CT data) with input from the physician. to manipulate original patient images for planning and executing surgery. The system processes the medical images and produces a variety of patient specific physical and/or digital output devices which include anatomical models, guides, and case reports.

    AI/ML Overview

    The KLS Martin Individual Patient Solutions (IPS) Planning System is a medical device for surgical planning. The provided text contains information about its acceptance criteria and the studies performed to demonstrate its performance.

    1. Table of Acceptance Criteria and Reported Device Performance

    The provided document describes specific performance tests related to the materials and software used in the KLS Martin Individual Patient Solutions (IPS) Planning System. However, it does not explicitly provide a table of acceptance criteria with numerical targets and corresponding reported device performance values for the device's primary function of surgical planning accuracy or effectiveness. Instead, it relies on demonstrating that materials withstand sterilization, meet biocompatibility standards, and that software verification and validation were completed.

    Here's a summary of the performance claims based on the provided text:

    Acceptance Criteria CategoryReported Device Performance (Summary from text)
    Material PerformancePolyamide (PA) Devices: Demonstrated ability to withstand multiple sterilization cycles and maintain ≥85% of initial tensile strength, leveraging data from K181241. Testing provides evidence of shelf life. Titanium Devices (additively manufactured): Demonstrated substantial equivalence to titanium devices manufactured using traditional (subtractive) methods, leveraging testing from K163579. These devices are identical in formulation, manufacturing processes, and post-processing.
    BiocompatibilityAll testing (cytotoxicity, sensitization, irritation, and chemical/material characterization) was within pre-defined acceptance criteria, in accordance with ISO 10993-1. Adequately addresses biocompatibility for output devices and intended use.
    SterilizationSteam sterilization validations performed for each output device for the dynamic-air-removal cycle in accordance with ISO 17665-1:2006 to a sterility assurance level (SAL) of 10-6 using the biological indicator (BI) overkill method. All test method acceptance criteria were met.
    PyrogenicityLAL endotoxin testing conducted according to AAMI ANSI ST72. Results demonstrate endotoxin levels below USP allowed limit for medical devices and meet pyrogen limit specifications.
    Software Verification & ValidationPerformed on each individual software application (Materialise Mimics, Geomagic® Freeform Plus™) used in planning and design. Quality and on-site user acceptance testing provided objective evidence that all software requirements and specifications were correctly and completely implemented and traceable to system requirements. Testing showed conformity with pre-defined specifications and acceptance criteria. Software documentation ensures mitigation of potential risks and performs as intended based on user requirements and specifications.
    Guide SpecificationsThickness (Cutting/Marking Guide): Min: 1.0 mm, Max: 20 mm. Width (Cutting/Marking Guide): Min: 7 mm, Max: 300 mm. Length (Cutting/Marking Guide): Min: 7 mm, Max: 300 mm. Degree of curvature (in-plane): N/A Degree of curvature (out-of-plane): N/A Screw hole spacing (Cutting/Marking Guide): Min: ≥4.5 mm, Max: No Max. No. of holes (Cutting/Marking Guide): N/A
    Screw SpecificationsDiameter (Temporary): 2.3 mm - 3.2 mm. Length (Temporary): 7 mm - 17 mm. Style: maxDrive (Drill-Free, non-locking, locking).

    2. Sample size used for the test set and the data provenance

    The document specifies "simulated use of guides intended for use in the thoracic region was validated by means of virtual planning sessions with the end-user." However, it does not provide any specific sample size for a test set (e.g., number of cases or patients) or details about the provenance of data (e.g., retrospective or prospective, country of origin). The studies appear to be primarily focused on material and software validation, not a clinical test set on patient data.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    The document does not specify the number of experts used to establish ground truth for a test set, nor their qualifications. The "virtual planning sessions with the end-user" implies input from clinical professionals, but no details are provided.

    4. Adjudication method for the test set

    The document does not describe any adjudication method (e.g., 2+1, 3+1, none) for a test set.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    The document states that "Clinical testing was not necessary for the determination of substantial equivalence." Therefore, an MRMC comparative effectiveness study involving AI assistance for human readers was not performed or reported for this submission. The device is a planning system for producing physical outputs, not an AI-assisted diagnostic tool for human readers.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    The device is described as "a software system and image segmentation system for the transfer of imaging information... The system processes the medical images and produces a variety of patient specific physical and/or digital output devices." It also involves "input from the physician" and "trained employees/engineers who utilize the software applications to manipulate data and work with the physician to create the virtual planning session." This description indicates a human-in-the-loop system, not a standalone algorithm. Performance testing primarily focuses on the software's ability to implement requirements and specifications and material properties, rather than an independent algorithmic assessment.

    Software verification and validation were performed on "each individual software application used in the planning and design," demonstrating conformity with specifications. This can be considered the standalone performance evaluation for the software components, ensuring they function as intended without human error, but it's within the context of supporting a human-driven planning process.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    For the materials and sterilization parts of the study, the "ground truth" or reference is established by international standards (ISO 10993-1, ISO 17665-1:2006) and national standards (AAMI ANSI ST72, USP allowed limit for medical devices).

    For the "virtual planning sessions with the end-user," the ground truth is implicitly the expert judgment/agreement of the end-user (physician) on the simulated surgical treatment options and guide designs. No further specifics are given.

    8. The sample size for the training set

    The document does not mention any training set or its sample size. This is expected as the device is not described as a machine learning or AI device that requires a training set for model development in the typical sense. The software components are commercially off-the-shelf (COTS) applications (Materialise Mimics, Geomagic® Freeform Plus™) which would have their own internal validation and verification from their developers.

    9. How the ground truth for the training set was established

    Since no training set is mentioned for the device itself, the establishment of ground truth for a training set is not applicable in this document.

    Ask a Question

    Ask a specific question about this device

    K Number
    K182889

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2019-07-26

    (284 days)

    Product Code
    Regulation Number
    882.4310
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The KLS Martin Individual Patient Solutions (IPS) Planning System is intended for use as a software system and image segmentation system for the transfer of imaging information from a computerized tomography (CT) medical scan. The input data file is processed by the IPS Planning System and the result is an output data file that may then be provided as digital models or used as input to a rapid prototyping portion of the system that produces physical outputs including anatomical models, guides and case reports for use in the marking of cranial surgery. The IPS Planning System is also intended as a pre-operative software tool for simulating surgical treatment options.

    Device Description

    The KLS Martin Individual Patient Solutions (IPS) Planning System is a collection of software and associated additive manufacturing (rapid prototyping) equipment intended to provide a variety of outputs to support reconstructive cranial surgeries. The system uses electronic medical images of the patients' anatomy (CT data) with input from the physician, to manipulate original patient images for planning and executing surgery. The system processes the medical images and produces a variety of patient specific physical and/or digital output devices which include anatomical models, guides, and case reports for use in the marking of cranial bone in cranial surgery.

    AI/ML Overview

    The KLS Martin Individual Patient Solutions (IPS) Planning System is a software system and image segmentation system used for transferring imaging information from a CT scan. The system processes input data to produce output data files, which can be digital models or physical outputs like anatomical models, guides, and case reports for cranial surgery. It is also a pre-operative software tool for simulating surgical treatment options.

    Here's an analysis of the acceptance criteria and supporting studies based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance:

    Acceptance Criteria CategoryReported Device Performance
    Tensile & Bending TestingPolyamide guides withstand multiple sterilization cycles without degradation and maintain 85% of initial tensile strength after 6 months. Additively manufactured titanium devices are equivalent to or better than traditionally manufactured titanium devices.
    Biocompatibility TestingPolyamide devices meet pre-defined acceptance criteria (cytotoxicity, sensitization, irritation, chemical/material characterization, acute systemic toxicity, material-mediated pyrogenicity, indirect hemolysis). Titanium devices (including acute systemic toxicity, material-mediated pyrogenicity, indirect hemolysis) meet pre-defined acceptance criteria.
    Sterilization TestingAll output devices (polyamide, epoxy/resin/acrylic, titanium) achieve a sterility assurance level (SAL) of $10^{-6}$ using the biological indicator (BI) overkill method for steam sterilization.
    Pyrogenicity TestingDevices contain endotoxin levels below the USP allowed limit for medical devices in contact with cerebrospinal fluid (< 2.15 EU/device) and meet pyrogen limit specifications.
    Software Verification & ValidationAll software requirements and specifications are correctly and completely implemented, traceable to system requirements. Testing (including risk analysis and impact assessments) conforms to pre-defined specifications and acceptance criteria. Potential risks are mitigated.

    2. Sample Size Used for the Test Set and Data Provenance:

    The document does not explicitly state the sample sizes used for the non-clinical performance tests (tensile, bending, biocompatibility, sterilization, pyrogenicity, software V&V). The data provenance for these tests is internal to the manufacturer, as they were "performed on the subject polyamide guides," "conducted on devices manufactured from polyamide," and for titanium, "leveraged from the reference device." This indicates the studies were likely conducted in a controlled lab or manufacturing environment. The studies are not described as retrospective or prospective clinical studies, but rather as non-clinical engineering and material performance tests.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts:

    This information is not applicable. The device is a planning and segmentation system that outputs digital models or physical guides/anatomical models. The acceptance criteria and performance data described relate to the physical and software characteristics of these outputs (e.g., material strength, sterility, software functionality), not to a diagnostic or treatment outcome for which expert ground truth would be established from patient data. The process involves a physician providing input for model manipulation and interactive feedback, but this is part of the planning workflow, not establishing ground truth for a test set in the context of algorithm performance.

    4. Adjudication Method for the Test Set:

    This information is not applicable, as the described studies are non-clinical performance tests focused on material properties, sterility, and software functionality, not on qualitative assessments requiring adjudication from expert readers.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done:

    No, an MRMC comparative effectiveness study was not done. The document explicitly states: "Clinical testing was not necessary for the determination of substantial equivalence."

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Study was Done:

    Yes, the "Software Verification and Validation" section describes testing of individual software applications. This indicates a standalone assessment of the software's performance in implementing requirements and specifications, separate from a human-in-the-loop clinical scenario.

    7. The Type of Ground Truth Used (expert consensus, pathology, outcomes data, etc.):

    For the non-clinical tests:

    • Tensile & Bending Testing: Ground truth is based on established engineering standards and material science principles (e.g., 85% retention of initial tensile strength after sterilization, comparison to traditional manufacturing methods).
    • Biocompatibility Testing: Ground truth refers to pre-defined acceptance criteria in accordance with ISO 10993-1.
    • Sterilization Testing: Ground truth is based on achieving a sterility assurance level (SAL) of $10^{-6}$ as per ISO 17665-1:2006.
    • Pyrogenicity Testing: Ground truth is based on meeting USP allowed limits for endotoxin levels (< 2.15 EU/device) and pyrogen limit specifications.
    • Software Verification & Validation: Ground truth is defined by "all software requirements and specifications" and "pre-defined specifications and acceptance criteria" derived from user requirements and risk analysis.

    8. The Sample Size for the Training Set:

    The document does not mention a training set. This device is described as utilizing a collection of software and additive manufacturing equipment for planning and producing physical outputs based on patient CT data, not a machine learning model that would require a dedicated training set. The software applications used (Materialise Mimics, Geomagic® Freeform Plus™) are "comedically off-the-shelf (COTS) software applications," implying they were developed and validated independently, prior to their integration into this system.

    9. How the Ground Truth for the Training Set Was Established:

    This information is not applicable, as there is no mention of a training set for a machine learning algorithm in the context of this device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K182789

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2019-03-11

    (161 days)

    Product Code
    Regulation Number
    872.4120
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The KLS Martin Individual Patient Solutions (IPS) Planning System is intended for use as a software system and image segmentation system for the transfer of imaging information from a medical scanner such as a CT based system. The input data file is processed by the IPS Planning System and the result is an output data file that may then be provided as digital models or used as input to a rapid prototyping portion of the system that produces physical outputs including anatomical models, guides, splints, and case reports for use in maxillofacial surgery. The IPS Planning System is also intended as a pre-operative software tool for simulating / evaluating surgical treatment options.

    Device Description

    The KLS Martin Individual Patient Solutions (IPS) Planning System is a collection of software and associated additive manufacturing (rapid prototyping) equipment intended to provide a variety of outputs to support reconstructive and orthognathic surgeries. The system uses electronic medical images of the patients' anatomy (CT data) with input from the physician, to manipulate original patient images for planning and executing surgery. The system processes the medical images and produces a variety of patient specific physical and/or digital output devices which include anatomical models, guides, splints and case reports.

    AI/ML Overview

    The provided text describes the KLS Martin Individual Patient Solutions (IPS) Planning System and its regulatory clearance (K182789) by the FDA. However, it does not contain information about a study proving the device meets specific acceptance criteria in the context of an AI/algorithm's performance.

    Instead, the document is a 510(k) summary demonstrating substantial equivalence to a predicate device (K181241), primarily to expand the patient population to include pediatric subgroups. The core of the device is a planning system involving commercial off-the-shelf (COTS) software and additive manufacturing for physical outputs, with human-in-the-loop interaction from trained employees/engineers and physicians.

    Therefore, most of the requested information regarding AI acceptance criteria, performance metrics, sample sizes, ground truth establishment, expert adjudication, and MRMC studies is not present in the provided document. The device, as described, is not an AI algorithm in the sense that it performs automated diagnostic or treatment recommendations independently based on image analysis with defined metrics. It's a system to assist human planning processes.

    Here's an analysis based on the information available in the document, and a clear indication of what is not available.


    Acceptance Criteria and Device Performance (as far as applicable from the document)

    The document focuses on demonstrating substantial equivalence, not on acceptance criteria for a freestanding AI algorithm's performance. The "performance" described is primarily related to material properties, biocompatibility, and sterilization, and the functioning of the software as a tool for human planning.

    Acceptance Criteria (Inferred from Substantial Equivalence and Safety/Performance)Reported Device Performance (from document)
    Material Degradation (Polyamide Guides)Subject polyamide guides can withstand multiple sterilization cycles without degradation and can maintain 85% of its initial tensile strength. Demonstrates shelf life of 6 months. (p.10)
    Titanium Device EquivalencyAdditively manufactured titanium devices are equivalent to, or better than, titanium devices manufactured using traditional (subtractive) methods, leveraging data from reference device K163579. (p.10)
    BiocompatibilityBiocompatibility endpoints (cytotoxicity, sensitization, irritation, chemical/material characterization) for both polyamide and titanium manufactured devices met pre-defined acceptance criteria (leveraged from predicate K181241 and reference K163579). (p.10)
    Sterility Assurance Level (SAL)Achieved an SAL of 10^-6 for each output device using the biological indicator (BI) overkill method for steam sterilization. Validations for polyamide and titanium leveraged from predicate K181241 and reference K163579. (p.10)
    Pyrogenicity (Endotoxin Levels)Endotoxin levels were below the USP allowed limit for medical devices and met pyrogen limit specifications, leveraging data from reference device K163579 for titanium. (p.10)
    Software Functionality and ValidationQuality and on-site user acceptance testing provided objective evidence that all software requirements and specifications were implemented correctly and completely and are traceable to system requirements. Testing from risk analysis showed conformity with pre-defined specifications and acceptance criteria. Software documentation demonstrates mitigation of potential risks and performance as intended. (p.11, p.14)
    Safety and Effectiveness in Pediatric SubpopulationsA risk assessment based on FDA guidance and supporting peer-reviewed clinical literature was performed. The conclusion is that expanding the patient population to neonates, infants, and children does not identify new issues of safety or effectiveness and the device is substantially equivalent to the predicate. (p.5, p.11, p.14) Note: This is a risk assessment and literature review, not a new clinical performance study.

    Information NOT Available in the Document (as it pertains to an AI/Algorithm performance study):

    1. Sample size used for the test set and data provenance:
      This information is not provided. The study performed was primarily non-clinical (material testing, biocompatibility, sterilization) and a risk assessment for pediatric use, not a clinical trial evaluating algorithm performance on a test set of patient data.

    2. Number of experts used to establish the ground truth for the test set and their qualifications:
      Not applicable, as no dedicated test set for evaluating AI/algorithm performance against a ground truth is described. The system relies on physician input and interaction with trained employees/engineers for planning, not on an autonomous algorithmic output that requires expert adjudication for ground truth.

    3. Adjudication method for the test set:
      Not applicable for the same reasons as above.

    4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done:
      No, an MRMC comparative effectiveness study was not done. The document explicitly states: "Clinical testing was not necessary for the determination of substantial equivalence." (p.11, p.14). The device is a planning system with human-in-the-loop, not an AI intended to improve human readers' performance directly.

    5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:
      No, a standalone algorithm performance study was not described. The device is presented as a "software system and image segmentation system" where "trained employees/engineers... manipulate data and work with the physician to create the virtual planning session." (p.4, p.13) The core function described is human-assisted planning and production of physical models, not an autonomous algorithmic output.

    6. The type of ground truth used:
      Ground truth in the context of an AI algorithm's diagnostic or predictive performance is not relevant here, as no such AI is described. The "ground truth" for the device's function would relate to the accuracy of the generated physical models and plans relative to the patient's anatomy and surgical intent, which is managed through human interaction and validation within the system's intended use.

    7. The sample size for the training set:
      Not applicable. The document refers to "commercially off-the-shelf (COTS) software applications" (p.4, p.13) which implies existing, validated software tools are being used, not a newly developed AI model requiring a separate training set.

    8. How the ground truth for the training set was established:
      Not applicable, as no new training set for an AI model is described.

    Ask a Question

    Ask a specific question about this device

    K Number
    K181241

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2018-09-13

    (126 days)

    Product Code
    Regulation Number
    872.4120
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The KLS Martin Individual Patient Solutions (IPS) Planning System is intended for use as a software system and image segmentation system for the transfer of imaging information from a medical scanner such as a CT based system. The input data file is processed by the IPS Planning System and the result is an output data file that may then be provided as digital models or used as input to a rapid prototyping portion of the system that produces physical outputs including anatomical models, guides, splints, and case reports for use in maxillofacial surgery. The IPS Planning System is also intended as a pre-operative software tool for simulating / evaluating surgical treatment options.

    Device Description

    The KLS Martin Individual Patient Solutions (IPS) Planning System is a collection of software and associated additive manufacturing (rapid prototyping) equipment intended to provide a variety of outputs to support reconstructive and orthognathic surgeries. The system uses electronic medical images of the patients' anatomy (CT data) with input from the physician, to manipulate original patient images for planning and executing surgery. The system processes the medical images and produces a variety of patient specific physical and/or digital output devices which include anatomical models, guides, splints and case reports.

    AI/ML Overview

    Here's an analysis of the acceptance criteria and study information based on the provided text, focusing on what is present and what is not:

    The document (K181241 510(k) Summary) describes the KLS Martin Individual Patient Solutions (IPS) Planning System, which is a software system for image segmentation and pre-operative planning, and also provides physical outputs like anatomical models, guides, and splints for maxillofacial surgery. The submission focuses on demonstrating substantial equivalence to a predicate device (VSP System K120956) and uses several reference devices.

    Acceptance Criteria and Study Information:

    Based on the provided text, the device itself is a planning system that ultimately produces physical outputs. The "performance" being evaluated relates to the characteristics of these physical outputs (tensile strength, biocompatibility, sterility, pyrogenicity) and the software's functionality. There isn't a direct "device performance" metric in the traditional sense of an AI diagnostic device's sensitivity, specificity, or accuracy against a clinical outcome.

    1. Table of Acceptance Criteria and Reported Device Performance

    Given the nature of the device (planning system with physical outputs, not a diagnostic AI), the performance metrics are primarily related to safety and manufacturing quality.

    Acceptance Criteria CategorySpecific Acceptance Criteria (as implied)Reported Device Performance (as summarized)
    Tensile & BendingPolyamide guides maintain 85% of initial tensile strength after multiple sterilization cycles. Demonstrate a 6-month shelf life. Titanium devices are equivalent or better than traditional methods.Polyamide guides met the criteria, demonstrating resistance to degradation after sterilization and supporting a 6-month shelf life. Titanium test results were leveraged from a reference device (K163579) and confirmed equivalence or superiority to traditional manufacturing.
    BiocompatibilityMeet pre-defined acceptance criteria for cytotoxicity, sensitization, irritation, and chemical/material characterization (according to ISO 10993-1).All conducted tests (cytotoxicity, sensitization, irritation, chemical/material characterization) for subject devices (polyamide and titanium) were within pre-defined acceptance criteria. Titanium results also leveraged from K163579.
    SterilizationAchieve a Sterility Assurance Level (SAL) of 10^-6 for each output device using the BI overkill method for steam sterilization (according to ISO 17665-1:2006 for dynamic-air-removal cycle).All test method acceptance criteria were met, achieving the specified SAL of 10^-6. Validations for titanium were leveraged from K163579.
    PyrogenicityMeet pyrogen limit specifications, with endotoxin levels below USP allowed limit for medical devices (according to AAMI ANSI ST72 for LAL endotoxin testing).The devices contain endotoxin levels below the USP allowed limit for medical devices and meet pyrogen limit specifications. Testing for titanium was leveraged from K163579.
    Software Verification & Validation (V&V)All software requirements and specifications are implemented correctly and completely, traceable to system requirements. Conformity with pre-defined specifications and acceptance criteria based on risk analysis and impact assessments. Mitigation of potential risks and performance as intended based on user requirements.Quality and on-site user acceptance testing provided objective evidence of correct and complete implementation of software requirements, traceability, and conformity with specifications and acceptance criteria. Software documentation demonstrated risk mitigation and intended performance. (Note: Specific quantitative metrics for software performance are not provided in this summary).

    2. Sample size used for the test set and the data provenance

    • Test Set Sample Size: The document does not specify a "test set" in the context of a dataset for evaluating AI performance (e.g., medical images for segmentation accuracy). Instead, "testing" refers to non-clinical performance evaluations of the physical outputs and software.
      • For Tensile & Bending, Biocompatibility, Sterilization, and Pyrogenicity, the sample sizes are not explicitly mentioned, but the tests were performed on "the subject polyamide guides" and "titanium" components. The provenance is internal testing performed by the manufacturer, or results leveraged from previous KLS Martin device submissions.
      • For Software V&V, no specific numerical "test set" of software inputs is given, but testing was performed on "each individual software application."
    • Data Provenance: The data provenance for non-clinical testing is internal to the manufacturer or relied upon previous regulatory submissions for similar materials/processes. It is not patient or country-specific data as would be for clinical studies. The data used by the IPS Planning System itself (CT data) would be patient data, but the evaluation here is of the system's outputs, not its interpretation of patient data in a diagnostic manner. The document states the system "transfers imaging information from a medical scanner such as a CT based system."

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    • This question is not applicable in the context of this 510(k) submission. The "ground truth" for the non-clinical tests (tensile strength, biocompatibility, etc.) is established by standard scientific and engineering methodologies, not by expert medical review of images.
    • For the "Software Verification and Validation," the "ground truth" for software functionality is defined by the established software requirements and specifications, validated by internal quality and user acceptance testing, not by external experts in the medical domain. The document mentions "input from the physician" for manipulation of original patient images, suggesting physicians set the clinical goals for the plan, but the validation of the system's performance is not described as involving experts establishing a "ground truth" concerning image interpretation.

    4. Adjudication method for the test set

    • Not applicable. There is no adjudication method described as would be used for clinical interpretation or diagnostic performance evaluation by multiple experts. The non-clinical tests follow established standards and protocols.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • No, a multi-reader multi-case (MRMC) comparative effectiveness study was not done.
    • This device is a planning system for maxillofacial surgery, not a diagnostic AI that assists human readers in image interpretation or diagnosis. It aids in surgical planning and creates physical outputs. The submission explicitly states: "Clinical testing was not necessary for the determination of substantial equivalence."

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • Given the device's function, it is inherently a "human-in-the-loop" system. The description states: "The system uses electronic medical images of the patients' anatomy (CT data) with input from the physician, to manipulate original patient images for planning and executing surgery." And, "The physician provides input for model manipulation and interactive feedback through viewing of digital models...that are modified by the trained employee/engineer during the planning session."
    • Therefore, performance of the algorithm without human intervention is not the intended use or focus of this submission. The "software verification and validation" (Section 11) is the closest thing to an "algorithm-only" evaluation, but it's about software functionality, not standalone image interpretation performance.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    • For the non-clinical performance tests of the physical outputs (Tensile & Bending, Biocompatibility, Sterilization, Pyrogenicity), the "ground truth" is defined by established engineering and scientific standards (e.g., ISO 10993-1, ISO 17665-1:2006, AAMI ANSI ST72, and internal specifications).
    • For Software Verification & Validation, the "ground truth" is adherence to predefined software requirements and specifications (functional and non-functional, related to image transfer, manipulation, and output file generation). It is not based on medical "ground truth" like pathology or clinical outcomes.

    8. The sample size for the training set

    • This question is not applicable. The KLS Martin IPS Planning System is described as using "validated commercially off-the-shelf (COTS) software applications" for image manipulation. There is no mention of a "training set" in the context of machine learning or AI model development within this summary. It appears to be a rule-based or conventional algorithmic system rather than a deep learning/machine learning model that would require a distinct training set.

    9. How the ground truth for the training set was established

    • This question is not applicable, as no training set for machine learning was mentioned or identified in the document (see point 8).
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1