    K Number
    K081089

    Validate with FDA (Live)

    Date Cleared
    2008-07-14

    (89 days)

    Product Code
    Regulation Number
    880.6250
    Age Range
    All
    Reference & Predicate Devices
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Based upon 21CFR§880.6250 "Patient examination glove" Use: A patient examination glove is a medical device intended for medical purposes that is worn on the examiner's hand to prevent contamination between patient and examiner.

    Device Description

    Light gray nitrile, chlorinated, powder-free, textured fingertip, ambidextrous patient examination glove that meets all of the requirements of ASTM D 6319-00a, Standard Specification for Nitrile Examination Gloves for Medical Application

    AI/ML Overview

    The provided text describes the acceptance criteria and the study conducted for the Kimberly-Clark* STERLING* Nitrile & STERLING* Nitrile-Xtra* Powder-Free Exam Gloves with a Chemotherapy Drug Use Claim. This information is from a Traditional 510(k) summary filed with the FDA.

    Here's a breakdown of the requested information:

    1. Table of acceptance criteria and the reported device performance

    TestStandardAcceptance Criteria / Reported Performance
    Non-Clinical Tests
    DimensionsASTM D 6319-00aMeets
    Physical PropertiesASTM D 6319-00aMeets
    Freedom from pinholesASTM D 6319-00aASTM D 5151-06Meets
    Powder FreeASTM D 6124-06ASTM D 6319-00aMeets
    ISO Skin Irritation StudyISO 10993, Part 10Meets
    Murine Local Lymph Node AssayISO 10993, Part 10Meets
    ISO Systemic Toxicity StudyISO 10993, Part 11Meets
    Resistance to Permeation (Protective Clothing)ASTM F 739-07See data below (specific breakthrough times reported for various chemotherapy drugs)
    Resistance to Permeation (Medical Gloves)ASTM D 6978-05See data below (specific breakthrough times reported for various chemotherapy drugs)
    Chemotherapy Drug Permeation (Average Breakthrough Detection Time in minutes)
    Cyclophosphamide (20.0 mg/ml)No breakthrough up to 240 minutes
    Doxorubicin HCI (2.0 mg/ml)No breakthrough up to 240 minutes
    Etoposide (20.0 mg/ml)No breakthrough up to 240 minutes
    5-Fluorouracil (50.0 mg/ml)No breakthrough up to 240 minutes
    Paclitaxel (Taxol) (6.0 mg/ml)No breakthrough up to 240 minutes
    ThioTEPA (10.0 mg/ml)Avg. minutes before breakthrough = 54.2
    Cisplatin (1.0 mg/ml)No breakthrough up to 240 minutes
    Dacarbazine (10.0 mg/ml)No breakthrough up to 240 minutes
    Ifosfamide (50.0 mg/ml)No breakthrough up to 240 minutes
    Mitoxantrone (2.0 mg/ml)No breakthrough up to 240 minutes
    Vincristine sulfate (1.0 mg/ml)No breakthrough up to 240 minutes

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    The document does not explicitly state the sample sizes for each of the specific non-clinical tests (e.g., number of gloves tested for pinholes, dimensions, or physical properties). For the chemotherapy drug permeation tests, it lists "Average Breakthrough Detection Time," implying multiple tests were conducted for each drug, but the exact number of samples for each is not specified.

    The data provenance is not mentioned (e.g., country of origin). The testing would be considered prospective in the sense that the tests were conducted specifically to demonstrate compliance for this 510(k) submission.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    This product is a medical glove, not an AI/imaging device requiring expert consensus for ground truth. The "ground truth" for the device's performance is established by the standardized test methods themselves and the quantitative results obtained from those tests. Therefore, experts in establishing "ground truth" in the context of diagnostic interpretation are not applicable here. The experts involved would be those qualified to execute the specific ASTM and ISO standards for glove testing and biocompatibility.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Not applicable. Adjudication methods are typically relevant for studies involving human interpretation (e.g., in diagnostic imaging or clinical trials) where disagreements in judgment need to be resolved to establish ground truth. For this type of product and testing, the results are objectively measured per the specified standards.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This is not an AI-assisted diagnostic or interpretation device that would involve human readers.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. This is a medical glove, not an algorithm. The "standalone performance" is essentially the results of the physical and chemical tests conducted on the glove itself.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    The "ground truth" for the performance of these gloves is based on:

    • Standardized Test Methods: Adherence to established and recognized industry standards like ASTM D 6319-00a, ASTM D 5151-06, ASTM D 6124-06, ASTM F 739-07, ASTM D 6978-05.
    • Biocompatibility Standards: Compliance with ISO 10993 (Parts 10 and 11) for skin irritation, sensitization, and systemic toxicity.
    • Quantitative Measurements: Direct measurements of physical properties (dimensions, tensile strength, elasticity), freedom from defects (pinholes), powder content, and breakthrough times for specific chemotherapy drugs.

    8. The sample size for the training set

    Not applicable. This is not a machine learning or AI device that requires a training set.

    9. How the ground truth for the training set was established

    Not applicable. As this device does not use a training set, the establishment of ground truth for such a set is irrelevant.

    Ask a Question

    Ask a specific question about this device

    K Number
    K081027

    Validate with FDA (Live)

    Date Cleared
    2008-06-30

    (81 days)

    Product Code
    Regulation Number
    880.6250
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Kimberly-Clark* STERLING Nitrile Powder-Free Exam Glove, Sterile, is a disposable device intended for medical purposes that is worn on the examiner's hand to prevent contamination between patient and examiner.

    Based upon 21CFR§880.6250 "Patient examination glove" Use: A patient examination glove is a medical device intended for medical purposes that is worn on the examiner's hand to prevent contamination between patient and examiner.

    Device Description

    Light gray nitrile, chlorinated, powder-free, sterile, textured fingertip, ambidextrous patient examination glove

    AI/ML Overview

    The Kimberly-Clark* Sterile STERLING* Nitrile Powder-Free Exam Glove (K081027) is a patient examination glove. The following information details its acceptance criteria and the study that proves the device meets these criteria:

    1. A table of acceptance criteria and the reported device performance

    The acceptance criteria for this device are based on compliance with several ASTM and ISO standards for properties typical of patient examination gloves. The device's performance is reported as meeting these requirements based on non-clinical tests.

    Acceptance CriteriaStandardReported Device Performance
    DimensionsASTM D 6319-00aMeets the standard
    Physical PropertiesASTM D 6319-00aMeets the standard
    Freedom from pinholesASTM D 6319-00aMeets the standard AQL (Acceptable Quality Level) of 2.5 for freedom from pinholes, as defined in ASTM D 5151-06.
    Powder FreeASTM D 5151-06Meets the standard. The device is distinctly labeled as "Powder-Free."
    SterilityASTM D 6124-06; ISO 11137-2Meets the sterility requirements. The device is specified as "sterile" and sterilized by gamma irradiation.
    ISO Skin Irritation StudyISO 10993, Part 10Meets the standard. No significant skin irritation observed.
    Murine Local Lymph Node AssayISO 10993, Part 10Meets the standard. No sensitization potential observed.
    ISO Systemic Toxicity StudyISO 10993, Part 11Meets the standard. No systemic toxicity observed.

    2. Sample size used for the test set and the data provenance

    The document does not specify the exact sample sizes for each non-clinical test. The data provenance is derived from internal non-clinical laboratory and animal-based biocompatibility tests conducted by Kimberly-Clark Corporation to confirm compliance with established standards. The data is retrospective in the sense that it pertains to the product being submitted for 510(k) clearance, reflecting the outcome of these tests.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    Not applicable. The ground truth for this device is established through compliance with recognized industry standards (ASTM and ISO) for physical properties, sterility, and biocompatibility. These standards define the objective measures for performance, rather than requiring expert consensus for a "ground truth" in the diagnostic sense.

    4. Adjudication method for the test set

    Not applicable. As the evaluation relies on objective measurements against established standards, an adjudication method in the context of expert review is not required. The results of the non-clinical tests are directly compared to the specified limits of the standards.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This device is a sterile patient examination glove and does not involve AI assistance or human-reader interpretations.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done

    Not applicable. This device is a physical medical device and does not involve an algorithm.

    7. The type of ground truth used

    The ground truth is established through objective measurements compared against published national and international standards (ASTM, ISO) for physical properties, sterility, and biocompatibility. For example, for "Freedom from pinholes," the ground truth is defined by the AQL (Acceptable Quality Level) of 2.5 specified in ASTM D 5151-06.

    8. The sample size for the training set

    Not applicable. This is a physical medical device, not a machine learning model, so there is no "training set."

    9. How the ground truth for the training set was established

    Not applicable. As there is no training set, this question is not relevant.