510(k) Data Aggregation

    K Number
    K052824

    Validate with FDA (Live)

    Date Cleared
    2006-05-02

    (209 days)

    Product Code
    Regulation Number
    878.4040
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Kimberly-Clark Procedure Gown is a non-sterile, disposable, single use item of apparel intended to be worn by healthcare professionals during the preparation and administration of selected chemotherapy drugs. The gown is not intended to be worn during surgical procedures.

    Device Description

    The Kimberly-Clark Comfort Gown is an open back gown manufactured from a non-woven fabric with a film laminate coating. The non-woven fabric is a polypropylene spunbond with a polyethylene coating. The composition of the gown fabric resists tearing and permeation by various chemotherapy drugs for up to four hours of use.

    AI/ML Overview

    The provided text describes a 510(k) premarket notification for the Kimberly-Clark Procedure Gown, a non-sterile, disposable gown intended for use by healthcare professionals during the preparation and administration of selected chemotherapy drugs. The submission demonstrates substantial equivalence to a predicate device, the Kimberly-Clark Impervious Open Back Gown (K880382).

    The core of this submission is demonstrating that the new device meets specific performance criteria through testing.

    1. Table of Acceptance Criteria and Reported Device Performance

    TestAcceptance Criteria (Implied by meeting standard)Reported Device Performance
    Dermal Irritation & SensitizationNo evidence of dermal irritation or allergic contact sensitization (as expected for patient contact devices)No evidence of dermal irritation or allergic contact sensitization
    FlammabilityMeets Class I flammability requirements per NFPA Standard #702-1980Meets Class I flammability requirements per NFPA Standard #702-1980
    Liquid Chemical PermeationNo permeation of specified chemotherapy drugs for up to 240 minutes (4 hours) per ASTM F 739-99aResults showed no permeation of the specified chemotherapy drugs (carmustine, cisplatin, cyclophosphamide, dacarbazine, doxorubicin hydrochloride, etoposide, fluorouracil, paclitaxel, thioTEPA and vincristine sulfate) for up to 240 minutes.
    Fluid Penetration"Pass" determination based on visual detection of synthetic blood penetration per ASTM-F 1670-03Resistance of the gown fabric to penetration by blood under conditions of continuous liquid contact; "pass" determination based on visual detection of synthetic blood penetration.
    Viral Penetration"Pass" determination based on detection of viral penetration per ASTM-F 1671-03Resistance of the gown fabric to penetration by blood-borne pathogens under conditions of continuous liquid contact; "pass" determination based on detection of viral penetration.

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not explicitly state the specific sample sizes used for each test (e.g., how many gown samples were tested for permeation, flammability, etc.). It refers to industry standards (e.g., ASTM, NFPA), which would have their own defined sample size requirements for each test method.

    The data provenance is retrospective, as the tests were conducted on the Kimberly-Clark Procedure Gown to demonstrate its characteristics. The country of origin for the data is implicitly the United States, as the submission is to the U.S. FDA and references U.S. standards (ASTM, NFPA).

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

    This information is not applicable in this context. This is a premarket notification for a physical medical device (a gown), not an AI/software device that requires expert review for ground truth establishment. The performance is assessed through standardized physical and chemical tests rather than expert interpretation.

    4. Adjudication Method for the Test Set

    This information is not applicable. Adjudication methods (like 2+1, 3+1) are typically used in clinical studies or studies involving human readers/interpreters to resolve discrepancies in diagnoses or assessments. For physical device testing, the results are objectively determined by the test method (e.g., a permeation threshold, visual observation for penetration, or a flammability measurement).

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done. This type of study is relevant for AI-powered diagnostic tools where human readers (e.g., radiologists) interpret images with and without AI assistance. The Kimberly-Clark Procedure Gown is a physical barrier device, and its effectiveness is determined by its physical and chemical properties, not by human interpretation of data.

    6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Was Done

    This question is not applicable. This device is a physical medical device. The concept of "standalone algorithm performance" refers to the performance of an AI algorithm without human involvement, which is not relevant to a medical gown.

    7. The Type of Ground Truth Used

    The "ground truth" for the performance of the Kimberly-Clark Procedure Gown is established by objective, standardized laboratory test results and their comparison against the defined acceptance criteria (e.g., "no permeation for up to 240 minutes," "meets Class I flammability"). These tests quantify the physical and chemical properties of the gown, such as its barrier performance against liquids, chemicals, and viruses, and its flammability.

    8. The Sample Size for the Training Set

    This question is not applicable. There is no "training set" in the context of a physical medical device like this gown. Training sets are used to develop and train AI algorithms. The performance of this gown is based on its material properties and manufacturing.

    9. How the Ground Truth for the Training Set Was Established

    This question is not applicable for the same reasons as #8. There is no training set for this device.