Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K110455

    Validate with FDA (Live)

    Date Cleared
    2011-04-29

    (73 days)

    Product Code
    Regulation Number
    878.4040
    Reference & Predicate Devices
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Kimberly-Clark KC100 Face Mask(s) is intended to be worn to protect both the patient and healthcare personnel from transfer of microorganisms, body fluids, and particulate material. These face masks are intended for use in infection control practices to reduce the potential exposure of the wearer to blood and body fluids. The Kimberly-Clark KC100 face mask(s) is a single use, disposable device(s), provided non-sterile.

    Device Description

    The Kimberly-Clark KC100 Procedure Mask(s) is a three layer mask, constructed of nonwoven polyester blends and polypropylene materials. Bindings are nonwoven polyester. The mask is provided with earloops or ties in either knifted polyester/lycra or nonwoven polyester. A malleable nosepiece is placed within the bindings for comfort and individualized fit around the wearer's nose and may or may not contain a fog free strip. Kimberly-Clark KC100 Procedure Mask(s) will be provided in a variety of colors, with and without a protective visor. Kimberly-Clark KC100 Procedure Mask(s) is a single use, disposable device, provided non-sterile.

    AI/ML Overview

    The acceptance criteria and study proving K110455 device meets those criteria are detailed below, based on the provided text.

    1. Table of Acceptance Criteria and Reported Device Performance

    StandardTitleAcceptance CriteriaReported Device Performance
    Mil-M369454CMilitary Specifications: Surgical Mask, disposable 1992Not explicitly stated as an acceptance criterion in the provided text, but implies compliance with military specifications for surgical masks."All testing conducted on the Kimberly-Clark KC100 Procedure Mask(s) met acceptance criteria." This statement broadly covers compliance with this standard.
    ASTM F2101-07Standard Test Method for Evaluating the Bacterial Filtration Efficiency (BFE) of Medical Face Mask Materials, Using a Biological Aerosol, Staphylococcus AureusNot explicitly stated in the provided text, but generally, acceptance criteria for BFE are a minimum percentage (e.g., ≥ 95% or ≥ 98%) depending on the mask type."All testing conducted on the Kimberly-Clark KC100 Procedure Mask(s) met acceptance criteria."
    PSC CS-191-53Flammability Test Method (16 CFR 1610) for Flammability of Clothing TextilesNot explicitly stated in the provided text, but relates to flammability resistance (e.g., Class 1 for fabrics)."All testing conducted on the Kimberly-Clark KC100 Procedure Mask(s) met acceptance criteria."
    ASTM F2299Standard Test Method for Evaluating the Initial Efficiency of Materials Used in Medical Masks to Penetration of Particulates Using Latex SpheresNot explicitly stated in the provided text, but generally involves a minimum particle filtration efficiency (e.g., ≥ 95% for 0.1 µm particles)."All testing conducted on the Kimberly-Clark KC100 Procedure Mask(s) met acceptance criteria."
    ISO 10993Standards for evaluating the biocompatibility of a medical deviceNot explicitly stated in the provided text, but implies the device passed various biocompatibility tests (e.g., cytotoxicity, irritation, sensitization)."All testing conducted on the Kimberly-Clark KC100 Procedure Mask(s) met acceptance criteria."
    F 1862Test Method for Resistance of Medical Face Masks to Penetration by Synthetic BloodNot explicitly stated in the provided text, but would typically involve a specific level of resistance to synthetic blood penetration (e.g., passing at a certain pressure)."All testing conducted on the Kimberly-Clark KC100 Procedure Mask(s) met acceptance criteria."

    Note: While the document states "All testing conducted on the Kimberly-Clark KC100 Procedure Mask(s) met acceptance criteria," the specific quantitative acceptance criteria for each standard are not explicitly detailed in the provided text. These would typically be defined within the test protocols for each standard.

    2. Sample Size Used for the Test Set and Data Provenance

    The provided text does not specify the sample sizes used for the test set for each of the listed standards. It also does not mention the data provenance (e.g., country of origin, retrospective or prospective). Safety and performance testing of medical devices like masks is typically performed prospectively, but this is not explicitly stated.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

    This information is not applicable to this type of device and testing. The "ground truth" for mask performance testing is established by objective measurements against standardized test methods (e.g., measuring BFE with a particle counter, resistance to blood penetration under specific pressure). It does not involve expert consensus on visual interpretations or diagnoses.

    4. Adjudication Method for the Test Set

    This information is not applicable as the evaluation of mask performance is based on objective, standardized tests, not subjective interpretations requiring adjudication.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    A multi-reader multi-case (MRMC) comparative effectiveness study was not performed as this type of study is relevant for diagnostic imaging AI devices where human readers interpret medical images. This document describes a physical medical device (surgical mask) and its performance based on laboratory testing, not human-AI interaction for interpretation.

    6. Standalone Performance Study

    This document describes standalone performance testing for the physical device itself (Kimberly-Clark KC100 Face Masks). The tests performed (BFE, particle filtration, flammability, synthetic blood penetration, biocompatibility) are all designed to evaluate the mask's inherent protective properties and material characteristics in a laboratory setting, without human interaction. Therefore, a standalone performance study was done for the device.

    7. Type of Ground Truth Used

    The ground truth used for these studies is based on objective, quantitative measurements obtained through standardized laboratory test methods (e.g., BFE measured by specific equipment, particle penetration quantified by particle counters, flammability determined by burn rates, synthetic blood penetration by visual inspection at defined pressures, biocompatibility by cellular responses). It is not based on expert consensus, pathology, or outcomes data in the typical sense.

    8. Sample Size for the Training Set

    This information is not applicable as the device is not an AI algorithm requiring a training set. The performance is assessed through laboratory testing of physical samples.

    9. How the Ground Truth for the Training Set Was Established

    This information is not applicable for the same reason as above; there is no training set for a physical device.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1