Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K150696

    Validate with FDA (Live)

    Date Cleared
    2015-06-23

    (97 days)

    Product Code
    Regulation Number
    872.3200
    Panel
    Dental
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    A thixotropic 37% phosphoric acid dental etchant gel for etching enamel, dentin, and glass ionomer cements to produce the necessary micro-retentive surface for successful bonding of restorations.

    Device Description

    Thixotropic dental etchant gel with 37% phosphoric acid

    AI/ML Overview

    The provided text is a 510(k) summary for a dental etchant gel, not a study involving AI or complex medical device performance metrics like sensitivity, specificity, or reader studies. Therefore, many of the requested categories related to AI systems, clinical trials, and expert adjudication are not applicable.

    However, I can extract the relevant information regarding the acceptance criteria (comparative performance to a predicate device) and the study that "proves" the device meets these criteria, based on the provided text.

    Device Name: Jade 37% Phosphoric Acid Etchant Gel

    1. A table of acceptance criteria and the reported device performance

    The acceptance criteria for the Jade 37% Phosphoric Acid Etchant Gel are based on demonstrating substantial equivalence to its predicate device, Seity 37% Phosphoric Acid Etchant Gel (K112597). This equivalence is primarily established through comparisons of technological characteristics and bench testing results.

    Acceptance Criteria (Comparison to Predicate Device)Predicate Device Performance (Seity 37% Phosphoric Acid Etchant Gel)Reported Device Performance (Jade 37% Phosphoric Acid Etchant Gel)Outcome
    Phosphoric Acid Concentration37%37%Equivalent
    ConsistencyThick gelThick gelEquivalent
    ColorBlue or GreenBlue or GreenEquivalent
    OdorNo characteristicNo characteristicEquivalent
    Viscosity (cps)40,000 cps60,000 cpsComparable*
    pH1.201.52Comparable*
    Shelf-LifeAppears to be 2-years (based on label example in MISC files)Proposed 2-years (based on existing shelf-life studies)Comparable
    Intended UseA thixotropic 37% phosphoric acid dental etchant gel for etching enamel, dentin, and glass ionomer cements to produce the necessary micro-retentive surface for successful bonding of restorations.A thixotropic 37% phosphoric acid dental etchant gel for etching enamel, dentin, and glass ionomer cements to produce the necessary micro-retentive surface for successful bonding of restorations.Equivalent
    Technological CharacteristicsThixotropic gel with 37% phosphoric acid, thickeners, and emulsifiers.Thixotropic gel with 37% phosphoric acid, thickeners, and emulsifiers. Coloring agents are also similar.Substantially Equivalent

    Note: While viscosity and pH values are not identical, the submission asserts that bench testing demonstrates the equivalence of Jade to the predicate device, implying these differences are not clinically significant for the intended use and do not raise new issues of safety or efficacy. The conclusion specifically states that "Even though the chemical compositions are not identical, bench testing has demonstrated that Jade has comparable material properties and performance."

    2. Sample size used for the test set and the data provenance

    The document describes "bench testing" without specifying a particular sample size for each characteristic tested. The data provenance is derived from these bench tests, which are laboratory-based and not tied to patient data or specific countries of origin in this context. It's prospective in the sense that Dharma Research likely conducted these tests specifically for this 510(k) submission.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    Not applicable. This is not a study involving expert consensus for ground truth but rather laboratory measurements comparing physical and chemical properties.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Not applicable. There was no expert adjudication in the context of clinical outcomes or imaging reviews.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This is not an AI-assisted device or a diagnostic imaging device.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. This device is not an algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    The "ground truth" here is the established properties and performance of the legally marketed predicate device (Seity 37% Phosphoric Acid Etchant Gel, K112597) and the measured chemical/physical properties of both the predicate and the new device through laboratory bench testing.

    8. The sample size for the training set

    Not applicable. This is not an AI/machine learning device.

    9. How the ground truth for the training set was established

    Not applicable. This is not an AI/machine learning device.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1