Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K213424

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2022-01-19

    (90 days)

    Product Code
    Regulation Number
    874.3325
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Jabra Enhance Plus self-fitting hearing aid is intended to amplify sound for individuals 18 years of age or older with perceived mild to moderate hearing impairment. It is adjusted by the user to meet the user's hearing needs. No pre-programming or hearing test is necessary. The device is intended for direct-to-consumer sale and use without the assistance of a hearing care professional.

    Device Description

    The Jabra Enhance Plus is a wireless, self-fitting air-conduction hearing aid system. It incorporates microphones on the earbuds for audio input into the ear, and it can be controlled wirelessly via Bluetooth Low Energy using the mobile app, the Jabra Enhance, installed on a compatible iPhone, iOS 14 or later. Further control of the earbuds is possible via an on-device user control button on both the Left and Right earbud. In addition to hearing aid functionality for environmental listening, the Jabra Enhance Plus earbuds can be used for placing and receiving telephone calls and for streaming audio from a compatible, Bluetooth compliant mobile device that has been paired with the earbuds. The controls accessible through the Jabra Enhance mobile app and on the earbuds are used to configure parameters, settings, and listening modes of the earbuds. The earbuds integrate a rechargeable 3.7V/15mAh li-ion battery coin cell inside each earbud, and they are recharged by the on-the-go charging case that also serves as a carrying case. The mobile app is connected to the Internet Services that enable remote upgrades to the earbud firmware in support of continued enhancements.

    AI/ML Overview

    The provided document is a 510(k) Summary for the Jabra Enhance Plus self-fitting air-conduction hearing aid. It outlines the device's characteristics, comparison to a predicate device (Bose Hearing Aid), and the studies conducted to demonstrate its safety and effectiveness for substantial equivalence.

    Here's an analysis of the acceptance criteria and the study that proves the device meets them, based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    The FDA 510(k) summary does not explicitly state acceptance criteria in numerical targets for all aspects, but rather compares the Jabra Enhance Plus (Subject Device) to a predicate device (Bose Hearing Aid) and relevant standards. The "Discussion" column often acts as a statement of meeting implied criteria (e.g., "Same as predicate," "Comparable to predicate and suitable for the intended user"). Performance is reported against technical standards and clinical equivalence.

    Evaluation Criterion / CharacteristicAcceptance Criteria (Implied/Standard)Reported Device Performance (Jabra Enhance Plus)Discussion (Meeting Acceptance)
    Indications for UseSame as predicate (amplify sound for individuals 18+ with perceived mild to moderate hearing impairment, self-adjusted, no pre-programming/hearing test, direct-to-consumer, no professional assistance)Same as predicateMet: "The Indications for Use are the same."
    Intended UseSame as predicate (amplify sound for individuals 18+ with perceived mild to moderate hearing impairment)Same as predicateMet: "The intended uses are the same."
    Technological Characteristics
    HousingBiocompatibility and safety characteristics comparable to predicate.In-ear earbuds (different from predicate's neckband)Met: "difference in housing does not raise different questions of safety or effectiveness. Biological safety characteristics same as predicate device. Nonclinical data from biological safety testing and data from a clinical validation study support substantial equivalence."
    Wireless communicationWireless communication with handheld device via BluetoothWireless communication with handheld device via BluetoothMet: "Same as predicate"
    Wireless coexistenceComplies with AAMI TIR 69 (temporary loss of Bluetooth communication a negligible risk)Uses 2.4GHz Classic Bluetooth and BLE; risk assessment determined temporary loss of Bluetooth communication from interfering RF signals is negligible.Met: "Same as predicate"
    Wireless user control functions via mobile appFunctionally similar to predicate. Fine-tuning options typical for mild-moderate hearing loss.Volume Control (-12dB to +6dB), Listen Mode (Surround, adaptive, focus), Preferred filter (full, normal, clear)Met: "Bose Hearing Aid allows for Left/Right balance control, however Left/Right balance is achieved through the self-fitting process in the subject device and as such, is inherent in the self-fitted device. The additional Bose Left/Right balance does not raise different questions of safety or effectiveness. Nonclinical data from a formative usability study and data a clinical validation study support substantial equivalence."
    Bluetooth pairing, control, streamingVerification with paired mobile devicePairing, control, streaming verification with the paired mobile device.Met: "Same as predicate"
    Battery lifeRechargeable, providing ~10 hours of battery lifeRechargeable 3.7V/15mAh li-ion battery coin cells, 10 hours battery life (implied from discussion)Met: "While battery capacity is different, both devices are rechargeable and allow for 10 hours of battery life on a full charge, and as such the difference in battery capacity does not raise different questions of safety or effectiveness."
    ChargingSafe charging method not raising new questions of safety/effectivenessProprietary On-The-Go charging case via physical connection (Pogo pins), with internal 3.7V/129mAh li-ion battery for on-the-go charging, supports USB power.Met: "Charging the subject device hearing aids in a portable charging case with or without connecting the charging case to a power supply via a USB cable does not raise different questions of safety or effectiveness. Non-clinical data from a formative usability study support substantial equivalence."
    MicrophonesOmnidirectional and directional modes, suitable for intended use.Omnidirectional or directional modes (Surround, adaptive, focus). Supports adaptive directional mode.Met: "The added automatic selection of directionality does not raise different questions of safety or effectiveness... Clinical data from peer-reviewed literature (Wu et al. Ear Hear. 2019) and a clinical validation study support substantial equivalence."
    Device controlComparable to predicate, not raising new safety/effectiveness questions.On-Device user controls for volume, mute, mobile calls. Bluetooth pairing via removal from charging case.Met: "The added call controls and different means of engaging BlueTooth pairing do not raise new questions for safety or effectiveness. Nonclinical data from a summative usability study and data from a clinical validation study support substantial equivalence."
    CompressionChannels allowing for similar spectral tilt as predicate.17 channel wide band dynamic range compressionMet: "The 5 additional channels... offer similar spectral tilt as predicate device. The 5 additional channels do not raise different questions of safety or effectiveness. Data from a clinical validation study support substantial equivalence."
    Noise reductionProven effective, not raising new safety/effectiveness questions.Steady-state noise reduction, impact noise control. No active noise reduction. (Different from predicate).Met: "While differences in the implementation of noise cancellation exist, the methods developed by GN Hearing have proven effective in products already on the market, and therefore do not raise different questions of safety or effectiveness. Clinical data from peer-reviewed literature (Wu et al. Ear Hear. 2019) and a clinical validation study support substantial equivalence."
    Feedback cancellationFeedback canceller presentFeedback cancellerMet: "Same as Predicate"
    Telephone callsPlacing and receiving telephone calls capabilityPlacing and receiving telephone callsMet: "Same as Predicate"
    Mobile App compatibilityNot raising different safety/effectiveness questions even with reduced compatibility.iOS only (Predicate compatible with iOS & Android).Met: "The lack of compatibility with handheld Android devices does not raise different questions of safety or effectiveness. Nonclinical data from a summative usability study and data from a clinical validation study support substantial equivalence."
    Self-fitting methodValidated algorithm, not raising different safety/effectiveness questions.NAL-NL2 fitting algorithmMet: "NAL-NL2 is widely used by hearing care professionals... As such the difference in fitting does not raise different questions of safety or effectiveness. Data from a clinical validation study support substantial equivalence."
    Remote Firmware updateCybersecurity risks assessed and mitigated, not raising different safety/effectiveness questions.Allows remote firmware update via app/cloud.Met: "This feature does not raise different questions of safety or effectiveness. Nonclinical data documented in the cyber risk register support substantial equivalence."
    Exposure to nonionizing radiation (IEC 62479:2010)Sufficiently safe in terms of human exposure to nonionizing radiation.Passes according to IEC 62479:2010.Met: "Comparable to predicate. Uses same Bluetooth technology. Nonclinical data from verification testing in accordance with IEC 62478:2010 (human exposure to electromagnetic fields) support substantial equivalence."
    Electroacoustic characteristics (ANSI/ASA S3.22-2014 & ANSI/CTA 2051:2017)Comparable to predicate and suitable for intended user. Latency <15ms, THD <5%.Frequency range: 100-8700 Hz. Max output: 110 dB SPL. Eq. noise input: 26dB. HFA FOG: 25 dB. Latency: <15ms (5.6 ms). THD@500Hz: 0.4%, @800Hz: 0.4%, @1600Hz: 1.0%, @3200Hz: 0.3%.Met: "Comparable to predicate and suitable for the intended user... Latency and total harmonic distortion meet the requirements in ANSI/CTA 2051-2017... do not raise different questions for safety or effectiveness."
    Maximum Acoustic Output Sound Pressure Level Input 90dB SPL (Max OSPL90)<= 120 dB SPL110 dB SPLMet: "Same as predicate"
    Harmonic Distortion (Output Distortion)<= 5%500 Hz 0.4%, 800 Hz 0.4%, 1600 Hz 1.0%, 3200 Hz 0.3% (all < 5%)Met: "Same as predicate"
    Equivalent Input Noise (EIN)<= 32 dB SPL24 dB SPLMet: "Same as predicate"
    High Frequency Average Output Sound Pressure Level (HFA OSPL 90)Adequate for fitting mild to moderate hearing loss as prescribed by NAL-NL2.104 dB SPL (Predicate: 112 dB SPL)Met: "Adequate for fitting mild to moderate hearing loss as prescribed by NAL-NL2. Data from a clinical validation study support substantial equivalence."
    High Frequency Average Full On Gain (HFA FOG)Adequate for fitting mild to moderate hearing loss as prescribed by NAL-NL2.26 dB (Predicate: 43 dB)Met: "Adequate for fitting mild to moderate hearing loss as prescribed by NAL-NL2. Data from a clinical validation study support substantial equivalence."
    Reference Test Gain (RTG)Adequate for fitting mild to moderate hearing loss as prescribed by NAL-NL2.25 dB (Predicate: 36 dB)Met: "Adequate for fitting mild to moderate hearing loss as prescribed by NAL-NL2. Data from a clinical validation study supports substantial equivalence."
    Safety (Adverse Events)No AEs or SAEsNo AEs or SAEs recordedMet: "The safety endpoint of the clinical validation study was met. There were no recorded Adverse Events (AEs) or Serious Adverse Events (SAEs) at either study site during the study."
    Reliability of Self-fittingStatistically equivalent REAR and QuickSIN results for repeated self-fit procedure.Statistically equivalent REAR (500, 1000, 4000 Hz) and QuickSIN scores for SELF-FIT A and B.Met: "Repetition of the self-fitting procedure resulted in statistically equivalent REAR and QuickSIN results. These findings... provide evidence that the self-fitting procedure for the Jabra Enhance Plus hearing aid is very reliable."
    Non-inferiority to professional fitting (APHAB-global score)SELF-FIT statistically non-inferior to PRO-FIT.SELF-FIT was statistically non-inferior to PRO-FIT, and scores were nearly identical to predicate.Met: "Analysis of the APHAB-global score revealed that SELF-FIT was statistically non-inferior to PRO-FIT."
    Non-inferiority to professional fitting (QuickSIN scores)Performance with SELF-FIT statistically non-inferior to PRO-FIT.Performance with SELF-FIT was statistically non-inferior to PRO-FIT.Met: "Performance with SELF-FIT was statistically non-inferior to that with PRO-FIT."

    2. Sample Size for the Test Set and Data Provenance

    • Clinical Validation Study (Test Set):

      • Sample Size: 38 subjects (19 at each clinical site) for the initial in-lab session. 37 subjects for the wear-time field trial.
      • Data Provenance: Two geographically disparate U.S. clinical audiology sites. The study was prospective.
    • Usability Testing (Summative Usability Study):

      • Sample Size:
        • 21 participants who were not intended device users (for external packaging and labeling).
        • 15 intended users (i.e., those with mild-moderate hearing loss) for functional usability (setting up earbuds).
      • Data Provenance: Not explicitly stated, but implies U.S.-based given the FDA submission context. Prospective.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    • Clinical Validation Study (for PRO-FIT comparison): Ground truth for the "PRO-FIT" (professional fitting) condition was established by an audiologist at each clinical site. The description refers to "an audiologist fit the same device using in situ NAL-NL2 real ear aided response (REAR) targets." This implies at least two audiologists (one per site). Their specific qualifications (e.g., years of experience) are not detailed.
    • Self-fitting method: The "ground truth" for the self-fitting aspect of the device is based on the NAL-NL2 prescription algorithm, which is a validated and widely used algorithm in audiology.

    4. Adjudication Method for the Test Set

    • Clinical Validation Study: The clinical study was a cross-over, within-subject (repeated measures) wear-time field trial. Subjects were blinded as to which fitting (SELF-FIT or PRO-FIT) they were wearing during the field trial. This blinding serves as a form of "adjudication" against bias in subjective reporting. For objective measures like REAR and QuickSIN scores, statistical analysis was used to compare the conditions. No explicit "expert panel" style adjudication is mentioned.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    • No, this was not an MRMC study in the traditional sense. MRMC studies are typically for diagnostic imaging where multiple readers interpret cases.
    • This was a clinical validation study comparing two fitting methods (self-fit vs. professional-fit) of the same hearing aid device in human subjects. The study assessed the effectiveness of one method (self-fitting) compared to another (professional fitting) and showed non-inferiority for the self-fitting method, not an improvement of human readers with AI assistance.
    • Effect Size: The study reported that the SELF-FIT was statistically non-inferior to PRO-FIT for APHAB-global scores and QuickSIN scores. For APHAB-global scores, the mean scores for SELF-FIT and PRO-FIT were "nearly identical," indicating very similar performance. For REAR, while SELF-FIT showed slightly less high-frequency gain than PRO-FIT, it was determined to be "ultimately inconsequential to performance" as evidenced by functional outcomes. The equivalence region for QuickSIN was ±1.5 dB, and equivalence was found within this range.

    6. Standalone (Algorithm Only Without Human-in-the-Loop Performance)

    • Yes, in essence, the "SELF-FIT" condition represents a standalone algorithm performance (without professional human intervention). The device self-fits based on user input and the NAL-NL2 algorithm. The clinical study's primary goal was to validate the reliability and effectiveness of this standalone self-fitting method by comparing it to a professionally-fitted version of the same device. The entire field trial evaluated the device being used "without the assistance of a hearing care professional."

    7. The Type of Ground Truth Used

    • Clinical Validation Study:

      • Expert Consensus/Standard: For the "PRO-FIT" condition, the ground truth was based on a best-practices professional fitting using in situ NAL-NL2 real ear aided response (REAR) targets by an audiologist.
      • Clinical Outcomes Data: For effectiveness endpoints, QuickSIN scores (speech-in-noise test) and APHAB questionnaires (patient-reported outcome of hearing aid benefit) were compared. These are standardized and validated clinical outcome measures reflecting functional performance and patient perception.
      • Objective Measures: REAR measurements were used to objectively assess the acoustic output of the hearing aid in the ear.
    • Biocompatibility: In-vitro and in-vivo testing according to ISO 10993 standards (e.g., cytotoxicity, sensitization, intracutaneous reactivity).

    • Electroacoustic Performance: Ground truth derived from established industry standards like ANSI/ASA S3.22-2014 and ANSI/CTA 2051:2017.

    8. The Sample Size for the Training Set

    • The document does not provide information on the sample size for the training set used to develop or train the Jabra Enhance Plus's self-fitting algorithm. The "self-fitting method" utilizes the validated NAL-NL2 fitting algorithm, which is an established prescription method in audiology, not typically "trained" on a specific dataset in the way a deep learning model would be. The document states NAL-NL2 is "widely used by hearing care professionals."

    9. How the Ground Truth for the Training Set Was Established

    • As mentioned above, the device uses the NAL-NL2 fitting algorithm. This algorithm itself is the "ground truth" for how the device determines personalized gain settings based on user input. The document mentions "NAL-NL2 is widely used by hearing care professionals to fit persons with mild to moderate hearing loss." The development and validation of NAL-NL2 as a prescriptive fitting formula involved extensive research and clinical studies by its developers (National Acoustic Laboratories, Australia) over many years, establishing its effectiveness and empirical basis in audiology rather than being a "training set" in the machine learning sense for this specific device. The device's self-fitting procedure then applies this established algorithm.
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1