Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K161888

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2016-09-12

    (63 days)

    Product Code
    Regulation Number
    888.3080
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    JULIET® LL Lumbar Interbody Device is indicated for intervertebral body fusion procedures in skeletally mature patients with degenerative disc disease (DDD) at one or two contiguous levels from L2-S1. DDD is defined as discogenic back pain with degeneration of the disc confirmed by patient history and radiographic studies. These DDD patients may also have up to Grade 1 spondylolisthesis or retrolisthesis at the involved level(s). These spinal implants are to be used with autogenous bone graft. JULIET® LL Lumbar Interbody Device is to be used with supplemental fixation. Patients should have at least six (6) months of non-operative treatment prior to treatment

    Device Description

    The JULIETLL Titanium Lateral Lumbar cages are rectangle-shaped intervertebral body fusion devices with a central cavity that can be filled with bone graft (autograft) to facilitate fusion. The JULIETLL intervertebral body fusion spacer comes in various sizes and footprints in order to fulfill surgeons' needs and accommodate different patient anatomies.
    The JULIETLL system is made of Titanium Ti 6Al-4V ELI conforming to ASTM F136 and ISO 5832-3.
    The JULIETLL Titanium Lateral Lumbar cages are supplied with dedicated surgical instruments (reusable - provided non sterile).

    AI/ML Overview

    This document describes the FDA 510(k) clearance for the JULIET® LL Lateral Lumbar Cage, an intervertebral body fusion device. The acceptance criteria and supporting studies are primarily related to the mechanical performance and biocompatibility of the device, rather than diagnostic accuracy as would be presented for an AI/ML medical device.

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria CategorySpecific Test/StandardAcceptance CriteriaReported Device PerformanceStudy Type
    Mechanical PerformanceStatic axial compression (ASTM F2077)Not explicitly stated (implied to be equivalent to predicate)Not explicitly stated (implied to be equivalent to predicate)Non-clinical (bench testing)
    Mechanical PerformanceStatic shear compression (ASTM F2077)Not explicitly stated (implied to be equivalent to predicate)Not explicitly stated (implied to be equivalent to predicate)Non-clinical (bench testing)
    Mechanical PerformanceSubsidence testing (ASTM F2267)Not explicitly stated (implied to be equivalent to predicate)Not explicitly stated (implied to be equivalent to predicate)Non-clinical (bench testing)
    BiocompatibilityBacterial Endotoxin Testing (USP <85>)Endotoxin limit of 20 EU / deviceNot explicitly stated (implied to meet 20 EU/device)Non-clinical (laboratory testing)
    Design & Material EquivalenceSimilar material (Titanium Ti 6Al-4V ELI conforming to ASTM F136 and ISO 5832-3)Equivalent to predicateMeets ASTM F136 and ISO 5832-3Design/Material Comparison
    Intended Use EquivalenceSame indicated useEquivalent to predicateSame indications for useClinical/Regulatory Comparison

    2. Sample Size Used for the Test Set and Data Provenance

    This document describes a medical device (spinal implant), not a diagnostic AI/ML system. Therefore, the concepts of "test set" and "data provenance" in the context of clinical images or patient data are not directly applicable.

    • Mechanical Testing: For mechanical tests (static axial compression, static shear compression, subsidence testing), the sample size would refer to the number of physical devices tested. This information is not explicitly stated in the provided document. The tests were performed on the predicate device (JULIET®LL Titanium Lateral Lumbar cages, K141135), and a rationale of "verification activity and validation activity" was used to demonstrate equivalence for the modified devices.
    • Bacterial Endotoxin Testing: The sample size would refer to the number of devices or material samples tested for endotoxins. This information is not explicitly stated.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of those Experts

    Not applicable for a physical medical device clearance based on mechanical testing and biocompatibility. Ground truth in this context is established by adherence to engineering standards and material specifications, not by expert interpretation of data.

    4. Adjudication Method for the Test Set

    Not applicable for a physical medical device clearance. Adjudication methods are typically used for establishing ground truth in clinical studies involving interpretation (e.g., by radiologists).

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done

    No, an MRMC comparative effectiveness study was not conducted. This type of study is relevant for evaluating the impact of AI on human reader performance in diagnostic tasks, which is not the subject of this 510(k) clearance for a spinal implant.

    6. If a Standalone Study (i.e., algorithm only without human-in-the-loop performance) was done

    No, a standalone study in the context of an AI algorithm was not done. This 510(k) is for a physical medical device. The "standalone" performance here refers to the device's mechanical integrity and biocompatibility, which were evaluated through non-clinical bench testing.

    7. The Type of Ground Truth Used

    The ground truth for this device clearance is based on:

    • Engineering Standards: Adherence to established ASTM (American Society for Testing and Materials) and ISO (International Organization for Standardization) standards for mechanical testing (ASTM F2077, ASTM F2267) and material specification (ASTM F136, ISO 5832-3).
    • Biocompatibility Standards: Adherence to USP (United States Pharmacopeia) standard for bacterial endotoxin testing (USP <85>).
    • Predicate Device Equivalence: The primary ground truth for the modified device is its substantial equivalence to the performance and safety of the previously cleared predicate device (JULIET® LL, K141135).

    8. The Sample Size for the Training Set

    Not applicable. There is no AI model or "training set" in the context of this traditional medical device clearance. The "training" for such devices involves adherence to manufacturing processes and quality control systems.

    9. How the Ground Truth for the Training Set was Established

    Not applicable. As there is no "training set," there is no ground truth established in this manner for the submitted device.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1