Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K181408

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2018-08-27

    (89 days)

    Product Code
    Regulation Number
    884.5300
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    JO Classic Hybrid Personal Lubricant is a personal lubricant for penile, anal and/or vaginal application, intended to lubricate and moisturize, to enhance the ease and comfort of intimate sexual activity and supplement the body's natural lubrication. This product is compatible with natural rubber latex and polyisoprene condoms. This product is not compatible with polyurethane condoms.

    Device Description

    JO Classic Hybrid Personal Lubricant is an off-white, semi-viscous, translucent gel type personal lubricant that is compatible with condoms made of natural rubber latex and polyisoprene. This product is not compatible with polyurethane condoms. This device is a non-sterile personal lubricant for penile, anal and/or vaginal application, to lubricate and moisturize, to enhance the ease and comfort of intimate sexual activity and supplement the body's natural lubrication. This product is provided in clear, Polyethylene Terephthalate (PET) cylinder bottles. The 1 oz. size bottles are capped with natural polypropylene disc top dispenser caps. The 2 oz., 4 oz. and 8 oz. size bottles are capped with silver polypropylene disc top dispenser caps. The individual bottles are hermetically sealed during the production process. This product is also available in polyester-faced laminated pouches.

    AI/ML Overview

    This document describes the premarket notification (510(k)) for the JO Classic Hybrid Personal Lubricant. Here's a breakdown of the requested information:

    1. Table of acceptance criteria and the reported device performance

    PropertyAcceptance Criteria (Specification)Reported Device Performance (all time points within shelf-life study)
    AppearanceOpaque thick liquidMet
    ColorOff-whiteMet
    OdorOdorlessMet
    Viscosity (cps)15,000 cps to 34,000 cpsMet
    pH6.20 to 7.20Met
    Specific Gravity0.950 to 1.075Met
    Osmolality (diluted 1:10)350 - 599 mOsm/kgMet
    Osmolality (undiluted)600 - 990 mOsm/kgMet
    Antimicrobial effectiveness per USP <51>Meets US <51> acceptance criteria for Category 2 productsMet
    Total aerobic microbial count (TAMC) per USP <61> and <1111>Less than 10 cfu/gMet
    Total yeast and mold count (TYMC) per USP <61> and <1111>Less than 10 cfu/gMet
    Presence of Pseudomonas aeruginosaAbsentMet
    Presence of Staphylococcus aureusAbsentMet
    Presence of Salmonella/ShigellaAbsentMet
    Presence of Escherichia coliAbsentMet
    Presence of Candida albicansAbsentMet
    Biocompatibility (Cytotoxicity)Pass (ISO 10993-5:2009)Met
    Biocompatibility (Sensitization)Pass (ISO 10993-10:2010)Met
    Biocompatibility (Vaginal Irritation)Pass (ISO 10993-10:2010)Met
    Biocompatibility (Penile Irritation)Pass (ISO 10993-10:2010)Met
    Biocompatibility (Acute Systemic Toxicity)Pass (ISO 10993-11:2006)Met
    Condom Compatibility (Natural Rubber Latex)CompatibleCompatible
    Condom Compatibility (Polyisoprene)CompatibleCompatible
    Condom Compatibility (Polyurethane)Not CompatibleNot Compatible
    Shelf-Life3 years3 years

    2. Sample size used for the test set and the data provenance

    The document does not explicitly state a "test set" in the context of a typical AI/software study. This is a medical device (personal lubricant), and the performance is evaluated through material properties and biocompatibility testing.

    • For Shelf-Life and Device Specifications: The "test set" would be samples of the lubricant tested at different time points (0, 1, 2, and 3 years). The specific number of samples tested at each time point is not provided. The data provenance is internal testing performed by the manufacturer (United Consortium).
    • For Biocompatibility: The "test set" involves biological models (in vitro and in vivo studies). The specific sample sizes for each biocompatibility test (e.g., number of cells in cytotoxicity, number of animals in sensitization/irritation/toxicity studies) are not provided in this summary. The data provenance is from studies conforming to ISO 10993 standards.
    • For Condom Compatibility: The "test set" would be lubricant samples tested with natural rubber latex, polyisoprene, and polyurethane condoms. The specific number of samples / tests is not provided. The data provenance is from studies conforming to ASTM D7661-10.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    This information is not applicable to this type of medical device submission. "Ground truth" in this context is established by standardized material testing and biocompatibility protocols, not expert consensus on interpretations. Laboratories performing these tests would have qualified personnel, but they are not "experts establishing ground truth" in the same way radiologists might for an imaging AI.

    4. Adjudication method for the test set

    Not applicable as it's not an expert-based subjective assessment. The results are from objective laboratory tests.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This is a personal lubricant, not an AI-powered diagnostic or assistive device requiring human interpretation.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. This is a physical, chemical, and biological performance evaluation of a personal lubricant, not an algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    The "ground truth" for this device is based on:

    • Standardized Material/Chemical Specifications: Adherence to defined physical and chemical properties (e.g., pH range, viscosity range, osmolality range, microbial limits).
    • International Standards: Compliance with recognized international standards for biocompatibility (ISO 10993 series) and condom compatibility (ASTM D7661-10). These standards define the test methods and acceptance criteria.
    • Defined Acceptance Criteria: The device must meet the predefined specifications for each tested property.

    8. The sample size for the training set

    Not applicable. This is not an AI/machine learning device. There is no "training set" in the context of software development.

    9. How the ground truth for the training set was established

    Not applicable.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1