Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K243363

    Validate with FDA (Live)

    Device Name
    JLK-ICH
    Manufacturer
    Date Cleared
    2025-01-03

    (66 days)

    Product Code
    Regulation Number
    892.2080
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    JLK-ICH is a radiological computer-aided triage and notification software indicated for use in the analysis of non-contrast CT images. JLK-ICH is a notification-only, parallel workflow tool that is intended to assist hospital networks and trained clinicians to identify and communicate images of specific patients to specialists, independent of the standard of care workflow.

    JLK-ICH uses an artificial intelligence algorithm to analyze images for findings suggestive of pre-specified clinical conditions and promptly notifies the appropriate medical specialists of these findings in parallel with the standard of care image interpretation. Identification of suspected findings is not for diagnostic use beyond notification. Specifically, the device analyzes non-contrast CT images of the head to detect intracranial hemorrhage (ICH). The system sends a notification to a clinician that a suspected ICH has been identified and recommends a review of those images. Images can be previewed and compressed through a mobile application.

    Notified clinicians are responsible for viewing non-compressed images on a diagnostic viewer and engaging in appropriate patient evaluation and relevant discussion with a treating clinician before making care-related decisions or requests.

    JLK-ICH is limited to the analysis of imaging data and should not be used in lieu of full patient evaluation or relied upon to make or confirm the diagnosis.

    Device Description

    JLK-ICH is a radiological computer-assisted triage and notification (CADt) software that adheres to the DICOM standard. The device functions as a Non-Contrast Computed Tomography (NCCT) processing module, providing triage and notification for suspected hemispheric intracranial hemorrhage (ICH). This software acts as a notificationonly, parallel workflow tool for hospital networks and trained clinicians, enabling the identification and communication of suspected patient images to relevant specialists, independent of the standard care workflow. JLK-ICH processes non-contrast computed tomography (NCCT) scans, prioritizing triage and notification for suspected hemispheric intracranial hemorrhaqe (ICH). The system utilizes advanced artificial intelligence to automatically analyze NCCT scans for indicators of ICH and promptly notify appropriate medical specialists of potential cases.

    JLK-ICH comprises an image analysis algorithm hosted on JLK servers and a mobile application for notification management. The Al/ML-based algorithm is designed to analyze NCCT of the head scans forwarded from CT scanners to the JLK servers. The mobile software module enables users to receive and toggle notifications for suspected ICH cases identified by the JLK-ICH Image Analysis Algorithm. Users can view a patient list and non-diagnostic CT scans through the mobile application. Image viewing through the mobile application interface is for non-diagnostic purposes only.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and the study that proves the device meets them, based on the provided text:

    Acceptance Criteria and Device Performance

    1. Table of Acceptance Criteria and Reported Device Performance

    The document doesn't explicitly list "acceptance criteria" for the standalone performance in a table, but it states that the primary endpoints, sensitivity and specificity, both exceeded 80%. This implies that 80% or greater for both sensitivity and specificity were the target acceptance criteria. It also sets a time-to-notification target.

    Here's a table summarizing the implicit acceptance criteria and the reported performance:

    Acceptance Criterion (Implicit)Reported Device Performance
    Sensitivity $\ge$ 80%97.3% (95% CI: 94.8% to 99.5%)
    Specificity $\ge$ 80%97.9% (95% CI: 95.5% to 99.5%)
    Time-to-Notification $\le$ 0.49 $\pm$ 0.15 minutes (based on predicate)0.19 $\pm$ 0.04 minutes (95% CI: 0.186 to 0.197)

    2. Sample Size and Data Provenance for the Test Set

    • Sample Size: 376 Non-Contrast CT (NCCT) scans. This included 188 ICH-positive and 188 ICH-negative cases.
    • Data Provenance: Retrospective study. The scans were obtained from various regions in the U.S.

    3. Number of Experts and Qualifications for Ground Truth

    • Number of Experts: Two primary ground truthers, with a third used for adjudication in cases of disagreement.
    • Qualifications of Experts: All truthers were U.S. board-certified neuroradiologists (specifically, American Board of Radiologists (ABR)-certified neuroradiologists).

    4. Adjudication Method for the Test Set

    The adjudication method used was a 2+1 scheme.

    • Ground truth was initially determined by two ABR-certified neuroradiologists.
    • In cases of disagreement between the first two truthers, a third neuroradiologist intervened to reach a consensus (act as a tie-breaker).
    • The document notes that 30 cases were sent to the third truther due to disagreements.

    5. Multi Reader Multi Case (MRMC) Comparative Effectiveness Study

    No, a multi-reader multi-case (MRMC) comparative effectiveness study was not conducted. The study focused on standalone performance. The document explicitly states: "The documentation was provided as recommended by FDA's Guidance for Industry and FDA staff, "Content of Premarket Submissions for Device Software Functions," June 14, 2023. In addition to the software verification and validation testing described in the sections above, JLK, Inc. performed a standalone performance in accordance with the §892.2080 special controls to demonstrate adequate clinical performance of the JLK-ICH module."

    Therefore, there is no effect size reported for human readers improving with AI vs. without AI assistance.

    6. Standalone Performance (Algorithm Only)

    Yes, a standalone performance evaluation (algorithm only, without human-in-the-loop performance) was performed.

    • "The algorithm's performance was validated through a standalone performance evaluation using an independent dataset, distinct from the one for algorithm training data."
    • "JLK, Inc. performed a standalone performance in accordance with the §892.2080 special controls to demonstrate adequate clinical performance of the JLK-ICH module."

    7. Type of Ground Truth Used

    The ground truth used was expert consensus. It was established by U.S. board-certified neuroradiologists following a 2+1 adjudication scheme.

    8. Sample Size for the Training Set

    The training dataset was substantial and diverse:

    • Total ICH cases: 14,462
    • US-based datasets: 14,998 cases (7,499 ICH cases, 7,499 normal cases)
    • Out-of-US datasets: 13,926 cases (6,963 normal cases, 6,963 ICH cases)
    • Total Training Cases (approximate sum): 14,998 (US) + 13,926 (Out-of-US) = 28,924 cases total. (Note: The first line stating 14,462 ICH cases seems to be a subset or summary that might exclude normal cases, given the detailed breakdown follows. The calculated sum provides a better approximation of the total
      images used for training).

    9. How Ground Truth for the Training Set Was Established

    The document states: "The JLK-ICH Al model was trained using a dataset that includes 14,462 ICH cases from various institutions, divided between US-based and Out-of-US sources... All cases are carefully separated from the clinical performance datasets."

    While it mentions the dataset composition, it does not explicitly detail how the ground truth for the training set was established. It only specifies the ground truth establishment method (2+1 neuroradiologist consensus) for the test set. It is common for training data to have ground truth established through similar expert review processes, but this document does not provide those specifics for the training set itself.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1