    Intended Use

    The Jetstream System is intended for use in atherectomy of the peripheral vasculature and to break apart and remove thrombus from upper and lower extremity peripheral arteries. It is not intended for use in coronary, carotid, iliac or renal vasculature.

    Device Description

    The Jetstream G3 System is an atherectomy catheter system designed with an expandable cutting tip for use in debulking and treating vascular disease in the peripheral vasculature. Separate lumens within the Catheter allow for continuous aspiration and infusion during device use. Excised tissue, thrombus, and fluid are aspirated from the peripheral treatment site through a port in the Catheter tip to a collection bag located on the Console. The distal portion of the Catheter also possesses infusion ports that provide continuous infusion of sterile saline during the atherectorny procedure.

    The Jetstream G3 System consists of two primary components: (1) a Catheter with Control Pod and (2) a Console, which are packaged separately. Each of these system components is described generally as follows:

    • . Jetstream G3 Catheter with Control Pod: A sterile, single-use unit consisting of an electrically driven Catheter and Control Pod. The Catheter utilizes a differentially cutting tip and includes both aspiration and infusion capabilities. The Control Pod provides a user interface with keypad controls. The unit, its electrical connectors, tubing, and aspirant collection bag are packaged in a double-pouched tray.
    • PV Console: A reusable compact Console, with two (2) peristaltic pumps for . aspiration and infusion, power supply, system controller, keypad interface, and LED indicators for device operational status. The Console mounts on a standard I.V. stand and remains outside the sterile field during the procedure.

    This 510(k) is for the same device most recently cleared under 510(k) K093456, but with the primary modifications intended to ensure any ingress of aspirated material into the guidewire lumen is minimized. In addition, the PATHWAY JETWIRE™ has been added as a compatible guidewire in the product labeling.

    AI/ML Overview

    The Pathway Medical Technologies Jetstream G3™ System (K101221) is a peripheral atherectomy catheter system. This 510(k) submission describes modifications primarily intended to minimize ingress of aspirated material into the guidewire lumen and the addition of the PATHWAY JETWIRE™ as a compatible guidewire.

    1. Table of Acceptance Criteria and Reported Device Performance

    The provided document does not explicitly present a table of acceptance criteria with corresponding performance values. Instead, it lists the types of in vitro testing performed to demonstrate substantial equivalence to predicate devices. The "reported device performance" is summarized as the tests demonstrating that the "technological and performance characteristics of the modified Jetstream G3 System are comparable to the predicate devices."

    Here's a summary of the types of tests mentioned as "acceptance criteria" through their performance evaluation:

    Acceptance Criteria (Test Type)Reported Device Performance Summary
    System Reliability/LifeDemonstrated comparable performance to predicate devices.
    Speed Drop FlexibilityDemonstrated comparable performance to predicate devices.
    Accessory CompatibilityDemonstrated comparable performance to predicate devices.
    Infusion and Aspiration Flow RatesDemonstrated comparable performance to predicate devices.
    Catheter Trackability and PushabilityDemonstrated comparable performance to predicate devices.
    Aspiration Efficiency and Crossing TimeDemonstrated comparable performance to predicate devices, particularly focusing on minimizing ingress into the guidewire lumen with the modifications.
    Thrombus Aspiration ChallengeDemonstrated comparable performance to predicate devices.
    Guidewire and Device WearDemonstrated comparable performance to predicate devices.
    Sterilant (EO) ResidualsDemonstrated comparable performance to predicate devices.
    Control Pod Logic VerificationDemonstrated comparable performance to predicate devices.
    Intended UseConfirmed device performs for its stated intended use: atherectomy of peripheral vasculature and breaking apart/removing thrombus from upper and lower extremity peripheral arteries, comparable to predicate devices.
    PyrogenicityDemonstrated comparable performance to predicate devices.

    The document states, "The results from these tests: demonstrate that the technological and performance characteristics of the modified Jetstream G3 System are comparable to the predicate devices, support the safety and effectiveness of the modifications... and ensure the modified device can perform in a manner equivalent to the listed predicate devices with the identical intended use."

    2. Sample Size Used for the Test Set and Data Provenance

    The document only states that "in vitro testing" was performed.

    • Sample Size: The specific sample sizes for each in vitro test are not provided in this summary.
    • Data Provenance: The tests were "in vitro," meaning conducted in a controlled laboratory environment, not on human or animal subjects. The country of origin for the data is implied to be the manufacturer's location or its testing facilities, which is Kirkland, Washington, USA. The data is not retrospective or prospective patient data, but rather performance testing data.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications of Those Experts

    This information is not applicable as the study involved in vitro performance testing of a medical device, not a diagnostic or AI-driven system requiring expert assessment for 'ground truth' in the clinical sense. The "ground truth" for these engineering and performance tests would be defined by established engineering and biological standards, measurement protocols, and comparison to the predicate devices.

    4. Adjudication Method for the Test Set

    This information is not applicable as the study involved in vitro performance testing. Adjudication methods like 2+1 or 3+1 typically apply to clinical studies where multiple human readers or experts assess cases to establish a definitive diagnosis or outcome.

    5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study Was Done, and Effect Size

    No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done. This type of study involves human readers (e.g., radiologists, pathologists) evaluating cases, often with and without AI assistance, to measure diagnostic performance. The Jetstream G3™ System is a medical device for atherectomy, not an AI software diagnostic tool, and therefore, an MRMC study with human readers and AI assistance is not relevant to its evaluation in this 510(k) submission.

    6. If a Standalone (i.e. algorithm only without human-in-the loop performance) Was Done

    No, a standalone (algorithm only) performance study was not done in the context of AI without human-in-the-loop performance. The Jetstream G3™ System is a physical medical device, not a standalone AI algorithm. The performance evaluation focuses on the physical and functional characteristics of the device itself.

    7. The Type of Ground Truth Used

    For the in vitro performance tests, the "ground truth" was established by engineering specifications, standardized test methods, and quantitative measurements for device characteristics (e.g., flow rates, mechanical properties, wear, aspiration efficiency). The primary "ground truth" for demonstrating equivalence was the established performance of the predicate devices. The tests aimed to prove that the modified device met the same performance standards or demonstrated comparable characteristics as previously cleared versions.

    8. The Sample Size for the Training Set

    This information is not applicable. The Jetstream G3™ System is a physical medical device and does not involve AI or machine learning algorithms that require a "training set" in the computational sense. The device's design and manufacturing processes are based on engineering principles and validated through testing, not machine learning training.

    9. How the Ground Truth for the Training Set Was Established

    This question is not applicable for the same reasons above. There is no AI training set for this device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K093456

    Validate with FDA (Live)

    Date Cleared
    2009-12-01

    (26 days)

    Product Code
    Regulation Number
    870.4875
    Age Range
    All
    Reference & Predicate Devices
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Jetstream G3TM System is intended for use in atherectomy of the peripheral vasculature and to break apart and remove thrombus from upper and lower extremity peripheral arteries ≥3.0 mm in diameter. It is not intended for use in coronary, carotid, iliac or renal vasculature.

    Device Description

    The Jetstream G3 System is an atherectomy catheter system designed with an expandable cutting tip for use in debulking and treating vascular disease in the peripheral vasculature. Separate lumens within the Catheter allow for continuous aspiration and infusion during device use. Excised tissue, thrombus, and fluid are aspirated from the peripheral treatment site through a port in the Catheter tip to a collection bag located on the Console. The distal portion of the Catheter also possesses infusion ports that provide continuous infusion of sterile saline during the atherectomy procedure. The Jetstream G3 System consists of two primary components: (1) a Catheter with Control Pod and (2) a Console, which are packaged separately. Each of these system components is described generally as follows: - Jetstream G3 Catheter with Control Pod: A sterile, single-use unit consisting of an electrically driven Catheter and Control Pod. The Catheter utilizes a differentially cutting tip and includes both aspiration and infusion capabilities. The Control Pod provides a user interface with keypad controls. The unit, its electrical connectors, tubing, and aspirant collection bag are packaged in a double-pouched tray. - . PV Console: A reusable compact Console, with two (2) peristaltic pumps for aspiration and infusion, power supply, system controller, keypad interface, and LED indicators for device operational status. The Console mounts on a standard I.V. stand and remains outside the sterile field during the procedure. This 510(k) is for the same device most recently cleared under 510(k) K092332, but changes the motor gear ratio to increase torque and improve cutting efficiency of the device.

    AI/ML Overview

    The provided text is a 510(k) summary for the Jetstream G3 System. It describes the device, its indications for use, and its substantial equivalence to a predicate device. However, it does not contain detailed information regarding acceptance criteria and the specific study that proves the device meets those criteria in the way you've outlined.

    Here's a breakdown of why I cannot fulfill your request based on the provided text, and what information is present:

    Information NOT available in the provided text:

    • A table of acceptance criteria and reported device performance: This document states "Bench testing was completed and provided to support the safety and effectiveness of the modifications," but it does not present the criteria or results in a table format, nor does it quantify performance metrics.
    • Sample size used for the test set and data provenance: No information on the number of cases or subjects studied, or their origin.
    • Number of experts used to establish the ground truth for the test set and their qualifications: No mention of experts or ground truth establishment.
    • Adjudication method: Not mentioned.
    • Multi-reader multi-case (MRMC) comparative effectiveness study: Not mentioned. This device is an atherectomy catheter system, not an AI or imaging diagnostic device where such studies are common.
    • Standalone (algorithm only) performance: Not applicable as this is a physical medical device, not a software algorithm in the AI sense.
    • Type of ground truth used: Not mentioned.
    • Sample size for the training set: Not applicable (not an AI device).
    • How the ground truth for the training set was established: Not applicable.

    What the text does give us about the "study" (bench testing):

    • Study Type: It was "Bench testing."
    • Purpose: "Bench testing was completed and provided to support the safety and effectiveness of the modifications that were the subject of this 510(k)." The 510(k) in question is for a modification to an already cleared device "changes the motor gear ratio to increase torque and improve cutting efficiency of the device." So, the studies were focused on verifying the impact of this specific modification.
    • Conclusion: The FDA reviewed this information and determined the device is "substantially equivalent" to the predicate device. This implies the bench testing successfully demonstrated that the modified device's performance aligned with the predicate and was safe and effective for its intended use, likely through comparison to the predicate's established performance or relevant industry standards.

    In summary: The provided document is a regulatory submission summary, not a detailed technical report of the studies performed. It confirms that bench testing was done to support the device modifications, but it does not provide the specific quantitative acceptance criteria or detailed results of that testing.

    Ask a Question

    Ask a specific question about this device

    K Number
    K092332

    Validate with FDA (Live)

    Date Cleared
    2009-09-03

    (30 days)

    Product Code
    Regulation Number
    870.4875
    Age Range
    All
    Reference & Predicate Devices
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Jetstream G3™ System is intended for use in atherectomy of the peripheral vasculature and to break apart and remove thrombus from upper and lower extremity peripheral arteries ≥3.0 mm in diameter. It is not intended for use in coronary, carotid, iliac or renal vasculature.

    Device Description

    The Jetstream G3 System is an atherectomy catheter system designed with an expandable cutting tip for use in debulking and treating vascular disease in the peripheral vasculature. Separate lumens within the Catheter allow for continuous aspiration and infusion during device use. Excised tissue, thrombus, and fluid are aspirated from the peripheral treatment site through a port in the Catheter tip to a collection bag located on the Console. The distal portion of the Catheter also possesses infusion ports that provide continuous infusion of sterile saline during the atherectomy procedure.

    The Jetstream G3 System consists of two primary components: (1) a Catheter and Control Pod and (2) a Console, which are packaged separately, Each of these system components is described generally as follows:

    • . Jetstream G3 Catheter and Control Pod: A sterile, single-use unit consisting of an electrically driven Catheter and Control Pod. The Catheter utilizes a differentially cutting tip and includes both aspiration and infusion capabilities. The Control Pod provides a user interface with keypad controls. The unit, its electrical connectors, tubing, and aspirant collection bag are packaged in a double-pouched tray.
    • PV Console: A reusable compact Console, with two (2) peristaltic pumps for . aspiration and infusion, power supply, system controller, keypad interface, and LED indicators for device operational status. The Console mounts on a standard I.V. stand and remains outside the sterile field during the procedure.

    This 510(k) is for the same device most recently cleared under 510(k) K091509, but changes the distal cutter design to improve cutting efficiency and manufacturability of the device.

    AI/ML Overview

    The provided text describes a 510(k) summary for the Jetstream G3™ System, a peripheral atherectomy catheter. However, it does not contain information about acceptance criteria, device performance metrics, or a study that specifically proves the device meets acceptance criteria in the context of AI/ML or diagnostic performance.

    The document discusses:

    • General Information: Manufacturer, contact, trade/common name, regulation, product code, predicate devices.
    • Indications for Use: What the device is intended for (atherectomy of peripheral vasculature, thrombus removal from upper and lower extremity peripheral arteries ≥3.0 mm in diameter) and what it is not intended for (coronary, carotid, iliac, or renal vasculature).
    • Device Description: Components (Catheter and Control Pod, Console), how it works (expandable cutting tip, aspiration, infusion).
    • Substantial Equivalence: It states that this 510(k) is for the same device cleared previously (K091509) but with changes to the distal cutter design to improve cutting efficiency and manufacturability. It claims substantial equivalence to predicate devices, identical indications for use, and same technological characteristics.
    • Bench Testing: It mentions "Bench testing was completed and provided to support the safety and effectiveness of the modifications." This implies that the bench testing results supported the claim of improved cutting efficiency and manufacturability, but it does not specify what the acceptance criteria for this bench testing were, nor does it provide any quantitative performance data from this testing.

    Therefore, I cannot fulfill your request for:

    1. A table of acceptance criteria and the reported device performance: This information is not present.
    2. Sample size used for the test set and the data provenance: Not applicable as no specific "test set" for performance evaluation in the requested context is mentioned.
    3. Number of experts used to establish the ground truth... and their qualifications: Not applicable.
    4. Adjudication method: Not applicable.
    5. Multi-reader multi-case (MRMC) comparative effectiveness study information: Not applicable.
    6. Standalone (algorithm only) performance: Not applicable.
    7. Type of ground truth used: Not applicable.
    8. Sample size for the training set: Not applicable.
    9. How the ground truth for the training set was established: Not applicable.

    This document is a regulatory submission for a medical device (atherectomy catheter) and focuses on demonstrating substantial equivalence to previously cleared devices based on design and technical characteristics, supported by bench testing regarding mechanical efficiency and manufacturability. It is not an AI/ML or diagnostic device submission that would typically include the requested performance metrics and study details.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1