Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K241480

    Validate with FDA (Live)

    Device Name
    JBS-LVO
    Manufacturer
    Date Cleared
    2024-09-27

    (126 days)

    Product Code
    Regulation Number
    892.2080
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    JBS-LVO is a notification-only, parallel workflow tool for use by hospital networks and trained clinicians to identify and communicate images of specific patients to a specialist, independent of standard of care workflow.

    JBS-LVO uses an artificial intelligence algorithm to analyze images for findings suggestive of a pre-specified clinical condition and to notify an appropriate medical specialist of these findings in parallel to standard of care image interpretation. Identification of suspected positive findings is not for diagnostic use beyond notification. Specifically, the device analyzes CT angiogram images of the brain acquired in the acute setting and sends notifications to a neurovascular specialist that a suspected large vessel occlusion has been identified and recommends a review of those images. Images can be previewed through a mobile application. JBS-LVO is intended to analyze terminal ICA and MCA-M1 vessels for LVOs.

    Images that are previewed through the mobile application are compressed and for informational purposes only. They are not intended for diagnostic use beyond notification. The JBS-LVO device does not alter the original medical image. Notified clinicians are responsible for viewing non-compressed images on a diagnostic viewer and engaging in appropriate patient evaluation and relevant discussion with a treating physician before making care-related decisions or requests. JBS-LVO is limited to the analysis of imaging data and should not be used in-lieu of full patient evaluation or relied upon to make or confirm a diagnosis.

    Limitations:
    The device does not process scans containing metallic artifacts.

    Device Description

    JBS-LVO is a radiological computer aided triage and notification (CAD) software package compliant with the DICOM standard. IBS-VV is a notification-only, parallel workflow tool for use by hospital networks and trained clinicians to analyze computed tomography angiography (CTA) images for findings suggestive of a suspected large vessel occlusion (LVO) and to notify an appropriate medical specialist of these findings in parallel to standard of care image interpretation. Specifically, JBS-LVO is optimized to evaluate occlusions of the intracranial carotid artery (ICA) and proximal middle cerebral artery (MCA-M1 segment). It is important to clarify that this quantification is solely used within the device's Al module to facilitation process. The output provided to heathcare professionals is stiritly a flag indicating the presence (positive) of an LVO, in accordance with regulatory guidelines.

    JBS-LVO is a combination of software modules that allow for detection and notification of patients with a suspected LVC. JBS-LVO consists of an algorithm and mobile application software module.

    The JBS-LVO Image Analysis Algorithm (LVO Detection Algorithm) is a locked, artificial intelligence (Al) software algorithm utilizing convolutional neural network (CNN) that analyzes CTA images of the brain for a suspected LVO. The LVO Detection Algorithm is hosted on the ILK-server and analyzes applicable CTA images of the brain that are acquired on CT scanners and are automatically transmitted to the ILK-server. Upon detection of a suspected LVO, the LVO notification module sends a notification of the suspected finding.

    The JBS-LVO notification functionality enable medical professionals and clinicians to preview compressed and informational images through via mobile application notification. Image viewing through the mobile application interface is for informational purposes only and is not for diagnostic use.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and the study proving the device meets them, based on the provided text:

    Acceptance Criteria and Device Performance

    1. Table of Acceptance Criteria and Reported Device Performance

    Performance MetricAcceptance Criteria (Target)Reported Device Performance (JBS-LVO)
    Sensitivity> 80%91.8% (95% CI: 85.8% - 95.8%)
    Specificity> 80%92.8% (95% CI: 87.2% - 96.5%)
    Area Under the Curve (AUC)Not explicitly stated as a target, but reported95% CI: 93.0% - 98.1%
    CTA to Notification Time≤ 3.5 minutes (compared to predicate)Ranged from 2.32 to 3.29 minutes (95% CI: 2.89 - 3.02)

    Notes:

    • The document implies that the sensitivity and specificity acceptance criteria were "exceeded," suggesting a >80% threshold was the minimum.
    • The AUC is reported as a performance metric, indicating its importance, even if a specific numerical acceptance criterion wasn't explicitly listed alongside the other two.
    • The CTA to notification time criterion is established by comparison to a reference predicate device (Rapid LVO, K221248).

    Study Details

    2. Sample Size Used for the Test Set and Data Provenance

    • Test Set Sample Size: Not explicitly stated as a numerical count of cases or patients in the provided text. The document mentions "a retrospective study" was conducted and "each case output from the JBS-LVO device was compared with a ground truth."
    • Data Provenance: Retrospective study. The origin of the data (e.g., country of origin) is not explicitly mentioned for the test set, but it states that "the images used to train the algorithm were sourced from datasets that included equipment from various manufacturers, such as Siemens, Philips, Toshiba, and GE", implying a diverse source for the overall dataset.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    • Number of Experts: Three (two initial ground truthers, with a third intervening in cases of disagreement).
    • Qualifications of Experts: All truthers were US board-certified neuroradiologists.

    4. Adjudication Method for the Test Set

    • Adjudication Method: 2+1 adjudication. The ground truth was determined by two ground truthers, with a third ground truther intervening in cases of disagreement.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    • MRMC Study: No, an MRMC comparative effectiveness study was not conducted. The document describes a "standalone performance evaluation" of the algorithm only.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    • Standalone Performance Study: Yes, a standalone performance evaluation of the algorithm without human-in-the-loop performance was conducted. The text states: "The performance of the device's AI algorithms was validated in a standalone performance evaluation, utilizing an independent dataset different from the one used for algorithm training." and "JLK, Inc. performed a standalone performance with the §892.2080 special controls to show acceptance of the clinical performance of the JBS-LVO module."

    7. The Type of Ground Truth Used

    • Ground Truth Type: Expert consensus. Specifically, ground truth was "established by US board-certified neuro-radiologists."

    8. The Sample Size for the Training Set

    • Training Set Sample Size: Not explicitly stated as a numerical count in the provided text. The document refers to it as "datasets."

    9. How the Ground Truth for the Training Set Was Established

    • Training Set Ground Truth: The training data was "labeled by trained radiologists to identify the presence of LVO." The specific number or adjudication method for these "trained radiologists" is not detailed, but it implies expert labeling.
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1