    K Number
    K222460

    Validate with FDA (Live)

    Date Cleared
    2023-05-12

    (270 days)

    Product Code
    Regulation Number
    892.1570
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Intraoperative Ultrasound Probe Cover can be used to minimize contamination between the probe during ultrasound scanning procedures for both intact skin and compromised tissue. The cover allows use of the transducer in scanning and needle guided procedures for body surface, Endo cavity, and intra-operative diagnostic ultrasound, while helping to prevent transfer of microorganisms, body fluids, and particulate material to the patient and healthcare worker during reuse of the transducer.

    Device Description

    Intraoperative Ultrasound Probe Cover can be used to minimize contamination between the patient and the probe during ultrasound scanning procedures for both intact skin and compromised tissue. The cover allows use of the transducer in scanning and needle quided procedures for body surface, endocavity, and intra-operative diagnostic ultrasound, while helping to prevent transfer of microorganisms, body fluids, and particulate material to the patient and healthcare worker during reuse of the transducer.

    The probe cover is designed for use in both invasive and non-invasive procedures. It comes in a sterile form only. The sterile cover made used in invasive procedures where it will come into direct contact with compromised tissue. The probe cover may be used with any multiple different probes including, but not limited to, an Endoscope re-usable, any rectal, vaginal, trans-urethral, gastro-entero probes, or pulmonary probe, In-vitro Fertilization instruments, biopsy probes, fiberoptic probes, catheter probes, any probe that is used during a surgical procedure, re-usable vascular (visualization) probes, and re-usable central nervous system (CNS) or cerebral probes.

    AI/ML Overview

    The provided text describes a 510(k) premarket notification for an "Intraoperative Ultrasound Probe Cover." This type of submission focuses on demonstrating substantial equivalence to a legally marketed predicate device, rather than proving novel clinical effectiveness through a comparative effectiveness study. Therefore, the device's acceptance criteria are primarily related to its physical properties, sterility, and performance characteristics consistent with its intended use as a protective barrier, rather than diagnostic accuracy or direct patient outcomes. The study mentioned is a non-clinical performance testing study.

    Here's a breakdown of the information requested based on the provided text:

    1. A table of acceptance criteria and the reported device performance

    Acceptance Criteria CategorySpecific Test/PropertyReported Device Performance
    Barrier PropertiesViral penetrationMet acceptance criteria
    Water resistance/hydrostatic pressureMet acceptance criteria
    Water resistance/impact penetrationMet acceptance criteria
    Synthetic blood penetrationMet acceptance criteria
    Material PropertiesTensile strengthMet acceptance criteria
    ElongationMet acceptance criteria
    LintingMet acceptance criteria
    SterilitySterility (via EO Gas, ISO 11135-1:2014)Met acceptance criteria (SAL 10-6)
    Ethylene oxide residualsMet acceptance criteria
    Bacterial endotoxinsMet acceptance criteria
    BiocompatibilityBiocompatibilityMet acceptance criteria
    AcousticsAcoustics performanceMet acceptance criteria
    Packaging & Shelf LifePackagingTesting completed (results applicable from predicate)
    Shelf lifeTesting completed (results applicable from predicate)

    2. Sample size used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective)

    The document does not explicitly state the specific sample sizes for each non-clinical test conducted. It generally states that "Non-clinical performance testing was conducted." The provenance of the data (country of origin, retrospective/prospective) is also not specified, beyond being part of the 510(k) submission from "Exact Medical Manufacturing" in Lancaster, NY.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    This information is not applicable to this type of non-clinical performance testing for a medical device cover. The "ground truth" for these tests would be established by standardized testing methods and their associated industry-recognized specifications (e.g., ISO standards for sterility, ASTM standards for material properties), not by expert human interpretation of diagnostic images or clinical data.

    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set

    Not applicable. This is not a study involving human interpretation or adjudication. The results are based on objective physical, chemical, and microbiological testing methods.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    No, an MRMC comparative effectiveness study was not done. This device is an "Intraoperative Ultrasound Probe Cover," a physical barrier, not an AI-powered diagnostic tool. Therefore, the concept of human readers improving with AI assistance is not relevant.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done

    No, a standalone (algorithm only) performance study was not done. This device is not an algorithm or software; it is a physical medical device.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    The "ground truth" for the non-clinical tests were the established acceptance criteria for each specific test method, derived from recognized industry standards (e.g., ISO for sterility, unspecified standards for material properties, viral penetration, etc.). These criteria define what constitutes an acceptable performance level for the device's physical and barrier properties.

    8. The sample size for the training set

    Not applicable. This device is not an AI/machine learning model, so there is no "training set."

    9. How the ground truth for the training set was established

    Not applicable, as there is no training set for this type of device.