    Intended Use

    The Intense Pulsed Light System is intended for medical use in the treatment of the following conditions, with treatment filters applied as indicated:

    420-1200nm and 515-1200nm: Moderate inflammatory acne vulgaris.

    515-1200nm and 560-1200nm: Benign pigmented epidermal lesions (including dyschromia, hyperpigmentation, melasma, and ephelides (freckles)).

    515-1200nm and 560-1200nm: Benign cutaneous vascular lesions (including port wine stains, hemangiomas, facial, truncal, and leg telangiectasias, erythema of rosacea, leg veins, spider angiomas, and venous malformations).

    590-1200nm, 640-1200nm and 695-1200nm: Permanent hair reduction (long-term stable reduction in the number of hairs regrowing after a treatment regimen).

    Device Description

    Intense Pulsed light (IPL) System is a type of intensive, broadband, coherent light source which has a wavelength spectrum of 420 nm -1200 nm.

    There are six optical filters that can be using in this system. With special properties, the IPL System has a wide application in non-ablative therapies based on theory of human skin tissue's selective absorption.

    Basic parameters/use conditions/power supply specifications is as follows:

    | light source | Intense Pulsed Light |

    Filter wavelength nm420515560590640695
    Wavelength range(nm)420-1200515-1200560-1200590-1200640-1200695-1200
    Pulse energy J1.9 -87.81.9-236.21.9-2161.9-202.51.9-175.51.9-148.5
    Energy density J/cm²5-135-355-325-305-265-22
    Pulse train width (ms)5-5005-200
    Pulse train interval time s1s-5s
    AI/ML Overview

    N/A

    Ask a Question

    Ask a specific question about this device

    K Number
    K232708

    Validate with FDA (Live)

    Date Cleared
    2023-12-01

    (87 days)

    Product Code
    Regulation Number
    878.4810
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Intense pulsed light therapy device is indicated for use in surgical and aesthetic applications in permanent hair removal. Permanent hair reduction is defined as the long-term, stable reduction in the number of hairs regrowing when measured at 6, 9, and 12 months after the completion of a treatment regimen.

    Device Description

    The Intense pulsed light therapy device is an intense pulsed light system which delivers intense pulsed light at a wavelength ranging from 650-1200nm. Intense Pulsed Light (IPL) systems work on the principles of selective thermolysis. That is, causing thermal damage to target chromophores by using light of appropriate wavelength in pulses that exceeds the chromophores' thermal relaxation time but sparing normal skin by limiting the pulse width below the thermal relaxation time for skin. IPL Systems are different from lasers in that they deliver many wavelengths in each pulse of light instead of just one wavelength. Generally, IPL enhances penetration without using excessive energy levels and enables targeting of specific chromophores. Based on this, the Intense pulsed light therapy device is indicated for use in surgical and aesthetic applications in permanent hair removal.

    AI/ML Overview

    The document is a 510(k) Summary for an Intense Pulsed Light (IPL) therapy device (Model: FI-L06) seeking FDA clearance for permanent hair removal. It compares the proposed device to a predicate device (K192519) to establish substantial equivalence.

    Here's an analysis based on your request:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not explicitly state "acceptance criteria" in a numerical or pass/fail format for clinical performance. Instead, it relies on demonstrating substantial equivalence to a predicate device through non-clinical testing and comparison of technical specifications. The key performance comparison is presented in tables 6-2 and 6-3.

    Feature/ParameterProposed Device PerformancePredicate Device (K192519) PerformanceDifference Analysis
    Product CodeONFONFSAME
    Regulation No.21 CFR 878.481021 CFR 878.4810SAME
    ClassIIIISAME
    Indication for UseSurgical and aesthetic applications in permanent hair removal. Permanent hair reduction defined as long-term, stable reduction measured at 6, 9, and 12 months post-treatment.Surgical and aesthetic applications in permanent hair removal, reduction of benign pigmented lesions and benign vascular lesions. Permanent hair reduction defined as long-term, stable reduction measured at 6, 9, and 12 months post-treatment.SAME (for permanent hair removal indication)
    Light sourceIntense pulsed light (Xenon Lamp)Intense pulsed lightAnalysis 1: Both use Xenon lamps; considered SAME type.
    Wavelength650-1200nm520-650nm, 800-1200nm, 540-800nm, 640-1200nmSAME (for hair removal, proposed device's range is within the broader predicate ranges)
    Deliver systemSapphireSapphireSAME
    Energy density1-50J/cm²1-50 J/cm²SAME
    Pulse Width0.1-25ms1-25msSAME (within the common range)
    Max. Power1500W3500VAAnalysis 2: Different, but non-clinical tests (AAMI/ANSI/ES 60601-1 and IEC 60601-1-2) showed no safety/effectiveness issues.
    Spot size8×40mmSmall: 40×12 mm, Large: 46×16mm, Ex-Large: 60×20mmAnalysis 3: Different, but only affects treatment area, not therapeutic effect.
    Wavelength Range (for Hair removal)650-1200640-1200SAME
    Energy Range (for Hair removal)1-505-40Analysis 4: Slight difference, but within allowable error range of predicate and bench tests support substantial equivalence.
    Pulse Width (for Hair removal)0.1-251-25SAME
    Spot Size (for Hair removal)8×40Small: 40×12 mm, Large: 46×16mm, Ex-Large: 60×20mmDifferent (same as above for spot size analysis).

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size: Not applicable. The submission explicitly states: "No clinical study is included in this submission." The evaluation for substantial equivalence relies on non-clinical (bench) testing and comparison to the predicate device's established performance and specifications.
    • Data Provenance: Not applicable, as no clinical data was submitted. Non-clinical tests were conducted by the manufacturer, Hebei Zhemai Technology Co., Ltd., in China.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts

    • Not applicable. No clinical test set requiring expert ground truth was created for this submission. The submission relies solely on non-clinical testing and comparison to an existing predicate device.

    4. Adjudication Method for the Test Set

    • Not applicable. No clinical test set requiring adjudication was used.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • Not applicable. This device is an Intense Pulsed Light therapy device for hair removal, not an AI-powered diagnostic or assistive tool for human readers. No MRMC study was conducted or is relevant for this device type.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

    • Not applicable. This device is a physical IPL therapy machine; it does not involve an algorithm running in a standalone mode without human interaction, nor an AI component to evaluate. Its operation is managed by a user.

    7. The Type of Ground Truth Used (expert consensus, pathology, outcomes data, etc.)

    • Not applicable. No clinical ground truth was established or used as no clinical studies were performed. The "ground truth" for demonstrating substantial equivalence was based on bench testing and comparison against recognized industry standards (IEC 60601-1-2:2020, AAMI/ANSI ES 60601-1: 2012, IEC 60601-2-57: 2011, ISO 10993-5: 2009, ISO 10993-10: 2010, ISO 14971: 2019) and the established specifications and indications for use of the predicate device.

    8. The Sample Size for the Training Set

    • Not applicable. No AI model or algorithm requiring a training set was part of this submission.

    9. How the Ground Truth for the Training Set Was Established

    • Not applicable. No training set was used.