Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K242595

    Validate with FDA (Live)

    Date Cleared
    2024-11-07

    (69 days)

    Product Code
    Regulation Number
    878.4810
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Intense Pulse Light Therapeutic Apparatus is an over-the-counter device intended for removal of unwanted body hair.

    Device Description

    The Intense Pulse Light Therapeutic Apparatus is a personal, light-based, hair reduction device intended to be sold over-the-counter directly to the end user. The device provides hair reduction using Intense Pulsed Light (IPL) technology, and it works below the skin's surface and does not involve any cutting or pulling, reducing hair growth with minimal pain.The Intense Pulse Light Therapeutic Apparatus is only powered by the external power adapter and its IPL emission activation is by finger switch. The device contains a Xenon lamp and its built-in skin sensor to detect appropriate skin contact. If the device is not properly and fully applied to the treated skin, the device will not emit the light pulse.There are IPL-18 BF and IPL-18 FG two models in this application. Their work principle, intended use, structure, appearance, size, and composition are the same, with differences being product function are slight differences, but these parameters are within the predicate device and do not affect or change the intended use of the device.

    AI/ML Overview

    The FDA 510(k) summary for the Intense Pulse Light Therapeutic Apparatus (K242595) does not contain detailed acceptance criteria or a study proving that the device meets specific performance criteria for its intended use (removal of unwanted body hair).

    Instead, the submission primarily focuses on demonstrating substantial equivalence to predicate devices based on:

    • Intended Use: The device shares the same intended use as its primary predicate: "removal of unwanted body hair."
    • Operational Characteristics and Design: The summary compares various physical and technical characteristics (wavelength range, energy density, pulse duration, light source, etc.) to show similarities or non-significant differences with predicate devices.
    • Compliance with Recognized Standards: The substantial equivalence is supported by showing compliance with a suite of international standards related to biocompatibility, electrical safety, electromagnetic compatibility (EMC), photobiological safety, and software verification/validation.

    Here's an breakdown of the available information based on your request:

    1. A table of acceptance criteria and the reported device performance

    The document does not provide specific quantitative acceptance criteria for hair removal effectiveness (e.g., X% hair reduction after Y treatments) or reported performance data against such criteria. The reported performance is implicitly deemed acceptable by virtue of:

    • Substantial Equivalence to Predicate Devices: The key "acceptance criterion" here is that the subject device's characteristics and safety/effectiveness profile are comparable enough to legally marketed predicate devices.
    • Compliance with Safety Standards: Meeting standards for biocompatibility, electrical safety, EMC, and photobiological safety is presented as the primary performance data.
    DomainAcceptance Criterion (Implicit/Standard)Reported Device Performance
    Intended UseSame as predicate: Over-the-counter device intended for removal of unwanted body hair.The subject device states the identical intended use as the primary predicate (K232932).
    Device CharacteristicsSimilar operational characteristics (light source, wavelength, energy density range, pulse duration, spot size) to predicate devices.Differences noted (e.g., exact dimensions, specific energy density/output energy values, exact spot size, pulse duration, materials) were deemed not to raise new safety/effectiveness issues. The "Energy density" of the subject device (IPL-18 BF: 1.284.80J/cm²; IPL-18 FG: 1.525.89J/cm²) is stated to be within acceptable limits and the max is lower than a reference device (3~6 J/cm²). Output energy for the subject device ranges from 6.0±1.2J to 13.0±2.6J across 5 levels. Spot sizes are (3.5±0.25) cm² for IPL-18 BF and (2.9±0.25) cm² for IPL-18 FG. Pulse duration is 10.0±3.0ms.
    BiocompatibilityCompliance with ISO 10993 series (cytotoxicity, sensitization, irritation) for body-contacting components.Performed and passed: ISO 10993-5:2009 (cytotoxicity), ISO 10993-10:2021 (sensitization), ISO 10993-23:2021 (skin irritation).
    Electrical SafetyCompliance with IEC 60601-1, IEC 60601-1-11, IEC 60601-2-83 standards.Performed and passed: IEC 60601-1, IEC 60601-1-11, IEC 60601-2-83.
    EMCCompliance with IEC 60601-1-2.Performed and passed: IEC 60601-1-2.
    Eye SafetyCompliance with IEC 62471 (photobiological safety).Performed and passed: IEC 62471.
    SoftwareSoftware verification and validation demonstrating all requirements met and hazards mitigated to acceptable risk levels (moderate level of concern).System validation testing demonstrated all software requirement specifications are met and all software hazards mitigated.

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    The document does not describe a clinical test set with human subjects to evaluate hair removal efficacy. The "test set" in this submission refers to the components and systems subjected to engineering (biocompatibility, electrical safety, EMC, etc.) and software validation testing. The data provenance for these engineering tests is generally from the manufacturer's testing or accredited labs, not specific countries or patient populations.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    This section is not applicable as there is no clinical test set for efficacy evaluation described in the submitted 510(k) summary. The "ground truth" for the engineering tests would be the specifications and requirements of the relevant IEC and ISO standards. The experts involved would be the engineers and quality personnel conducting the tests and verifying compliance. Their specific qualifications are not detailed in this summary.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    This is not applicable as there is no clinical test set requiring adjudication in this 510(k) summary.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    This is not applicable as the device is an Intense Pulse Light Therapeutic Apparatus for hair removal, which is not an AI-powered diagnostic device requiring human reader interpretation. No MRMC study was performed or is relevant.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This is not applicable as the device is a physical light therapy apparatus, not an algorithm. Performance is demonstrated through compliance with safety and engineering standards and comparison of physical and operational characteristics to predicate devices.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    For the compliance testing:

    • Biocompatibility, Electrical Safety, EMC, Eye Safety: The "ground truth" is compliance with the specified international standards (ISO 10993 series, IEC 60601 series, IEC 62471). This means the device met the predefined thresholds and requirements set out in these widely accepted safety and performance standards.
    • Software Verification and Validation: The "ground truth" is that the software met its specified requirements and mitigated identified hazards, as demonstrated through validation testing against defined specifications.

    There is no "ground truth" for hair removal efficacy (e.g., pre/post treatment hair counts, subject satisfaction scores, expert assessment of hair reduction) presented in this 510(k) summary.

    8. The sample size for the training set

    This is not applicable. The device is an Intense Pulse Light Therapeutic Apparatus. It is not an AI/ML algorithm that requires a training set. The "design" and "development" of the device would involve engineering prototypes and testing, but not in the context of a "training set" as it would apply to AI.

    9. How the ground truth for the training set was established

    This question is not applicable as the device is not an AI/ML system requiring a training set with established ground truth.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1