Found 1 results

510(k) Data Aggregation

    K Number
    K243866

    Validate with FDA (Live)

    Date Cleared
    2025-05-21

    (155 days)

    Product Code
    Regulation Number
    870.2200
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    InVision Precision Cardiac Amyloid is an automated machine learning-based decision support system, indicated as a screening tool for adult patients aged 65 years and over undergoing cardiovascular assessment using echocardiography.

    When utilized by an interpreting physician, this device provides information alerting the physician for referral to confirmatory investigations.

    InVision Precision Cardiac Amyloid is indicated in adult populations over 65 years of age. Patient management decisions should not be made solely on the results of the InVision Precision Cardiac Amyloid.

    Device Description

    The InVision Precision Cardiac Amyloid (InVision PCA) is a Software as a Medical Device (SaMD) machine-learning screening algorithm to identify high suspicion of cardiac amyloidosis from routinely obtained echocardiogram videos. The device assists clinicians in the diagnosis of cardiac amyloidosis.

    The InVision PCA algorithm uses a machine learning process to identify the presence of cardiac amyloidosis. The device inputs images and videos from echocardiogram studies, and it outputs a report suggestive or not suggestive of cardiac amyloidosis.

    The device has no physical form and is installed as a third-party application to an institution's PACS system.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and the study that proves the device meets them, based on the provided FDA 510(k) clearance letter:


    InVision Precision Cardiac Amyloid: Acceptance Criteria and Performance Study

    InVision Precision Cardiac Amyloid (InVision PCA) is a Software as a Medical Device (SaMD) machine-learning screening algorithm developed to identify a high suspicion of cardiac amyloidosis from routinely obtained echocardiogram videos. It acts as a decision support system, alerting interpreting physicians for referral to confirmatory investigations for adult patients aged 65 years and over undergoing cardiovascular assessment using echocardiography.

    The device's performance was validated through a comprehensive study, demonstrating its substantial equivalence to the predicate device.

    1. Acceptance Criteria and Reported Device Performance

    The primary acceptance criteria for the InVision PCA device were established based on its ability to reliably screen for cardiac amyloidosis. The reported performance metrics from the validation study are as follows:

    Acceptance CriteriaReported Device Performance
    Sensitivity0.607 (60.7%)
    Specificity0.990 (99.0%)

    Note: While specific numerical acceptance thresholds are not explicitly stated as "passing" values (e.g., "must achieve >X% sensitivity"), these reported values are presented as the results that successfully met the predefined endpoints of the validation study, implying they satisfied the implicit acceptance criteria deemed necessary for clearance.

    2. Sample Size and Data Provenance for Test Set

    • Sample Size: 1221 unique echocardiogram studies.
    • Data Provenance: The data were selected from three geographically different U.S. sites. The study was conducted on "previously acquired" images, indicating it was a retrospective study.

    3. Number of Experts and Qualifications for Ground Truth

    The provided document does not explicitly state the number of experts used to establish the ground truth nor their specific qualifications. It mentions "confirmatory reference data," which could imply a consensus of expert opinion but does not detail the process.

    4. Adjudication Method for Test Set

    The document does not explicitly state the adjudication method used (e.g., 2+1, 3+1). It refers to the ground truth being established by "confirmatory reference data, such as diagnostic imaging or pathology," suggesting a definitive diagnostic pathway rather than a multi-reader visual interpretation adjudication.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    No Multi-Reader Multi-Case (MRMC) comparative effectiveness study was explicitly described in the provided text, meaning there is no information on how much human readers improve with AI vs. without AI assistance. The study focused on the standalone performance of the AI model.

    6. Standalone (Algorithm Only) Performance

    Yes, a standalone (algorithm only) performance study was conducted. The reported sensitivity of 0.607 and specificity of 0.990 are directly attributable to the InVision PCA algorithm's performance in analyzing echocardiogram studies against confirmed ground truth.

    7. Type of Ground Truth Used

    The ground truth for the test set was established using confirmatory reference data, such as diagnostic imaging or pathology. This indicates a high-fidelity ground truth derived from definitive diagnostic procedures rather than solely expert consensus on images.

    8. Sample Size for Training Set

    The document does not specify the sample size used for the training set. It only details the sample size for the validation/test set.

    9. How Ground Truth for Training Set Was Established

    The document does not explicitly state how the ground truth for the training set was established. It is assumed that similar rigorous methods involving confirmatory diagnostic imaging or pathology would have been used for the training data, consistent with the approach for the test set, but this is not directly mentioned.