510(k) Data Aggregation

    K Number
    K191267

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2019-08-01

    (83 days)

    Product Code
    Regulation Number
    892.5050
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The ImagingRing System on rails (IRr) is an X-ray device consisting of imaging components, supports and software, designed to support patient localization and intended to be used as image guidance equipment as part of the radiation therapy (RT) treatment process, in all areas of the body where such image guidance is determined by a licensed physician. IRr facilitates patient localization by means of inaging patient anatomy including the therapeutic target volume, critical structures, bone and soft tissue with or without the use of implanted markers or contrast agents.

    IRr provides 2D planar imaging (including recording sequences of 2D frames for motion analysis) and 3D volumetric imaging before, during and after irradiation, support pationing, monitoring and management of internal target motion, and decision making as a function of target position, size, shape and displacement resulting from patient set-up deviation, organ deformation and anatomical movement.

    With the goal to accurately perform patient alignment with respect to an external treatment beam and to spare critical organs surrounding the target volume, the IRr is intended to be embedded in the radiotherapy workflow, i.e. being able to communicate with third party systems used in particle therapy or LINAC based RT such as Oncology Information Systems (OIS) or Radiotherapy Control Systems (RTCS) - receiving commands and data from and providing information to these party systems (e.g., receiving command from RTCS to start an image acquisition and providing image data. providing input for third party RTCS for repositioning of the patient).

    Device Description

    This Traditional 510(k) describes modifications made to the medPhoton ImagingRing System. The kev differences are:

    • In contrast to the existing ImagingRing System. where the ring (imager gantry) is mounted ● on the patient table, in the current system, the ring is mounted on a ceiling-mounted rail which allows longitudinal movement of the ring.
    • Thus, the ImagingRing System on Rails does not include the patient table. ●
    • The diameter of the gantry (ring structure carrying the imaging components X-ray generator ● and flat-panel detector) has been increased, providing bigger gantry clearance. This is facilitating positioning of the patient relative to the ImagingRing System on Rails using a Patient Positioning System (PPS) that is provided and controlled by a system integrator since not part of the ImagingRing System on Rails.
    • A new control console, being used as human machine interface, has been developed to ● improve usability and to simplify imaging workflows (Control Console 2.0).
    • In addition to the existing basic and slightly extended software (ImagingRing Software ● Suite), front-end software (medPhoton controls) was added to simplify imaging workflows and integration with overall radiation therapy workflow software that is provided and controlled by a svstem integrator.

    The remaining characteristics and the basic technology have remained the same. The same essential components (X-ray generator and -detector, control system, etc.) are used.

    Similar to the ImagingRing System, the ImagingRing System on Rails (Rr) is an X-ray imager that is intended to be integrated into a radiation therapy treatment system (e.g. a proton therapy system with its subsystems and workflows) by a system integrator. A major reason for the further development of the ImagingRing System has been to provide a simpler X-ray imager integration option for system integrators.

    The medPhoton ImagingRing System on Rails (IRr) is an imaging system designed to obtain Two-Dimensional (2D) and Three-Dimensional (3D) X-ray images. The images visualize patient's anatomy whose information can be used to accurately position or re-position the patient undergoing radiotherapy treatment.

    The IRr consists of

    • . hardware called ImagingRing (IR) consisting of rails for longitudinal movement, ring gantry, flat-panel detector, X-ray source, generator, control cabinet, control console;
    • software comprising the ImagingRing Software Suite (ImRiSS) and the medPhoton ● Controls (mPc) to orchestrate image acquisition and data processing;
    • programmable logic controller (PLC) components and related software (ImagingRing ● Control System - ImRiCS), to execute and control motion and image acquisition.
    AI/ML Overview

    The provided text does not contain information about studies proving that the ImagingRing System on Rails meets specific acceptance criteria related to its performance in terms of diagnostic accuracy or clinical effectiveness. The document is a 510(k) summary for a medical device, which focuses on demonstrating substantial equivalence to a predicate device rather than presenting detailed performance studies against specific acceptance criteria for AI-assisted or standalone diagnostic capabilities.

    The document discusses the device's technical specifications and adherence to safety and performance standards (e.g., IEC 60601 series, IEC 62304, IEC 62366, ISO 14971) and outlines performance characteristics like spatial resolution, low contrast visibility, and geometric accuracy for both 2D and CBCT imaging. However, these are presented as technical specifications, not as results from a study against specific acceptance criteria for diagnostic performance or human reader improvement with AI assistance.

    Specifically, the document states:

    • "Summary of performance testing: The system is subject to compliance testing to voluntary consensus safety standards. Details of the standards employed in the design are specified in chapter 9... The results from verification and validation testing prove the conformance to applicable standards and demonstrate that safety & effectiveness have been achieved."
    • "medPhoton GmbH did not perform any animal testing for the ImagingRing System on Rails."
    • "medPhoton GmbH did not perform any clinical testing for the ImagingRing System on Rails."

    Based on the provided text, it's not possible to populate the requested table or answer the questions related to clinical performance studies, such as sample size, data provenance, expert ground truth establishment, MRMC studies, or standalone algorithm performance, as these types of studies were explicitly stated as not performed.

    Therefore, I cannot provide a detailed answer to your request as the information is not present in the provided text.

    Ask a Question

    Ask a specific question about this device

    K Number
    K161400

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2016-12-08

    (202 days)

    Product Code
    Regulation Number
    892.5050
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The ImagingRing System (IRS) is an X-rav device consisting of (1) imaging components, supports and (2) patient support structure (couch top), designed to support patient localization in the Couch Coordinate System (CCS) and intended to be used as image guidance equipment as part of the radiation therapy (RT) treatment process, in all areas of the body where such image guidance is determined by a licensed physician.

    IRS facilitates patient localization by means of imaging patient anatomy including the therapeutic target volume, critical structures, bone and soft tissue with or without the use of implanted markers or contrast agents.

    IRS provides 2D planar imaging (including streaming and recording sequences of 2D frames for motion and 3D volumetric imaging before, during and after irradiation, and is intended to support pationing, monitoring and management of internal target motion, and decision making as a function, size, shape and displacement resulting from patient set-up deviation. organ deformation and anatomical movement.

    With the goal to accurately perform patient alignment with respect to an external treatment beam and to spare critical organs surrounding the target volume, IRS is intended to be:

    · embedded in the radiotherapy workflow, i.e. being able to communicate with third party systems used in particle therapy or LINAC based RT such as Oncology Information Systems (OIS) or Radiotherapy Control Systems (RTCS) in order to receive commands and data from and provide information to (e.g., receiving command from RTCS to start an image acquisition and providing image data to RTCS for the purpose to calculate and execute the patient (re)-positioning vector externally in the third party system)

    · interfaced with a Patient Positioning Systems (PPS), like - but not limited to - a robotic multi-axis system, to allow the calculated (re)-positioning vector to be applied with respect to Coordinate Systems the treatment delivery unit relates to.

    Device Description

    The medPhoton IRS is a standalone imaging system designed to obtain TwoDimensional (2D) and Three-Dimensional (3D) X-ray images. The images visualize patient's anatomy whose information can be used to accurately position or re-position the patient undergoing radiotherapy treatment.

    The IRS consists of

    • · hardware called ImagingRing (IR) consisting of couch top, ring, detector, X-ray source, generator, etc .;
    • · software called ImagingRing Software Suite (ImRiSS), to orchestrate image acquisition and data processing:
    • . programmable logic controller (PLC) components and related software (e.g., ImagingRing Control System - IRCS), to execute and control image acquisition.

    The system is designed in a way such that it provides the following key features:

      1. During imaging acquisition, the IR can be moved along the couch top in order to acquire images of any anatomical region of interest.
      1. The X-ray source and detector can move independently in order to enable nonisocentric acquisition trajectories (i.e., to focus on a selectable center of interest within the patient in an axial plane) and variable size of the axial Field Of View (FOV).
      1. The combination of imaging arm and couch top is designed to be directly attached to a range of Patient Positioning Systems (PPS) including multi-axis robotic devices.

    The imaging device is integrated with a couch top. The independently moveable components allow for acquisition of X-ray projective images. These provide the foundation for 2D planar imaging as well as acquisition and reconstruction of 3D ConeBeam Computed Tomography (CBCT) images around customizable centers of interest.

    AI/ML Overview

    The provided text describes the medPhoton ImagingRing System (IRS) and compares its characteristics to predicate devices (Elekta XVI R5.0 and Qfix kVue) to demonstrate substantial equivalence for FDA 510(k) clearance. However, the document does NOT contain details about a specific study proving the device meets acceptance criteria. Instead, it outlines performance characteristics and states that testing was performed to evaluate safety and performance requirements.

    Therefore, the following information will be extracted from the comparison tables and accompanying text. Aspects not mentioned in the document will be explicitly stated as "Not mentioned in the provided text."

    1. A table of acceptance criteria and the reported device performance

    The document presents performance characteristics by comparing the ImagingRing System's specifications against those of predicate devices. The values for the predicate devices essentially serve as the "acceptance criteria" or benchmarks that the new device aims to match or exceed for substantial equivalence.

    CharacteristicAcceptance Criteria (Predicate Device XVI R5.0)Reported Device Performance (ImagingRing System)
    Position Accuracy±0.3°±0.2°
    2D low contrast visibility≥12 disks visible≥12 disks visible
    2D spatial resolution≥1.4 line/mm≥1.6 line/mm
    CBCT spatial resolution≥7 lp/cm≥10 lp/cm (up to 18)
    CBCT low contrast resolution< 1.5%< 1.5%
    CBCT uniformity< 2%< 2%
    Transverse geometric accuracy< 1 mm< 1 mm
    Leakage radiation (loading state)< 1.0 mGy/h0.1979 mGy/h
    Leakage radiation (not loading)< 20 mGy/h0 mGy/h
    X-ray beam must not be larger than boundaries by more than 3% of SID (image receptor plane)YesYes
    Aluminum Equivalence Thickness (AET) @ 100kVpNot specified (kVue context)1.9 mm (compared to kVue)
    Water Equivalence Thickness (WET) @ 6 MVNot specified (kVue context)≤16.5 mm (compared to kVue)
    Meets deflection requirements of IEC 60976< 5 mm (kVue)< 1 mm effect in imaging volume accuracy (compared to kVue)

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    Not mentioned in the provided text. The document refers to "module and integration verification as well as system level validation" but does not detail specific test sets, sample sizes, or data provenance (e.g., country of origin, retrospective/prospective nature).

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience)

    Not mentioned in the provided text.

    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set

    Not mentioned in the provided text.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not mentioned in the provided text. This document describes an imaging hardware system, not an AI-assisted diagnostic tool for human readers.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    The document does not describe a "standalone" algorithmic performance in the context of an AI device. It describes the ImagingRing System as a standalone imaging system (hardware and software) designed to acquire images and support patient localization. Its performance metrics (e.g., spatial resolution, low contrast visibility) characterize the system itself, not an algorithm's diagnostic capabilities.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    Not explicitly mentioned for the reported performance metrics. However, typical "ground truth" for physical performance metrics like spatial resolution, low contrast visibility, and geometric accuracy usually involves physical phantoms (e.g., Catphan 503 mentioned as a phantom for predicate device data) and calibrated measurement systems.

    8. The sample size for the training set

    Not applicable. The document describes a medical imaging device, not a machine learning model that would typically have a "training set."

    9. How the ground truth for the training set was established

    Not applicable. The document describes a medical imaging device, not a machine learning model that would typically have a "training set."