Search Filters

Search Results

Found 3 results

510(k) Data Aggregation

    K Number
    K153011
    Device Name
    iViewDose R1.0
    Manufacturer
    Date Cleared
    2015-12-11

    (58 days)

    Product Code
    Regulation Number
    892.5050
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    iViewDose is a standalone software tool, available as an optional accessory of the linear accelerator. It is intended to assist in reducing the clinical risk in the delivery of radiotherapy treatments and does not alter the treatment delivery, or impact the clinical workflow.

    The software is to be used for the purposes of detecting gross errors during the delivery of radiation therapy. The software acquires data using the iViewGT Electronic Portal Imaging Device at the time of treatment) and subsequently processes it. The processed data is compared with data calculated by the treatment planning system. The comparison is derived from the application of dose conversion and reconstruction algorithms to the EPID data, which is back-projected to a plane normal to a beam (2D) or multiple planes (3D) inside the patient (in vivo) or phantom (pretreatment).

    iViewDose is not a treatment planning system, and provides an independent means of checking the dose delivered to the patient.

    iViewDose therefore provides an added level of treatment assurance, thus giving clinicians greater confidence especially when complex treatment techniques are employed.

    iViewDose is intended to support decision making in relation to the delivery of radiation to defined target volumes performed with Elekta linear accelerators only.

    Device Description

    This Traditional 510(k) describes a new standalone software to be used with the Elekta range of medical linear accelerators, to further enhance the existing functionality.

    The iViewDose program uses iViewGT™ images to reconstruct the dose in 3D in the volume (patient or phantom). The program then uses a gamma analysis to compare the reconstructed dose with the TPS dose in 3D. iViewDose uses a dose reconstruction algorithm to calculate the dose in the patient from the dose delivered to the MV detector panel. The algorithm corrects for scattered radiation from the patient and the MV detector panel. You can use iViewDose in vivo to make an estimate of the delivered dose to a patient. You can also use iViewDose for pre-treatment checks to make an estimate of the delivered dose to a phantom.

    AI/ML Overview

    The provided text is a 510(k) summary for the Elekta iViewDose R1.0 device. While it describes the device's functionality and its comparison to predicate devices, it does not contain a detailed study report with specific acceptance criteria or an analysis of its performance against those criteria in a formal study.

    The document states: "Validation of the system under clinically representative conditions has been performed by competent and professionally qualified personnel. Results from verification and validation testing demonstrate that conformance to applicable technical design specification have been met and safety & effectiveness have been achieved." However, it does not provide the specifics of these tests.

    It also mentions: "Although no clinical testing has been performed as part of the development, it can be shown that the algorithm at the core of the iViewDose product has been used extensively for a number of years. The data reviewed as part of this evaluation has been obtained from clinical sites. The data is presented in section 11 of this submission." This indicates that while there's no new clinical study specifically for this device, existing clinical data related to the underlying algorithm was reviewed. The specifics of "section 11 of this submission" are not included in the provided text.

    Therefore, I cannot provide the requested information in a table format with specific metrics because the detailed acceptance criteria and performance study results are not present in the given document.

    Here's what I can extract from the text regarding the nature of the evaluation, even without specific numbers:

    Device Overview:

    • Device Name: iViewDose R1.0
    • Intended Use: A standalone software tool to assist in reducing clinical risk in radiotherapy treatments by detecting gross errors during delivery. It acquires data from the iViewGT Electronic Portal Imaging Device (EPID), processes it, and compares it with data calculated by the treatment planning system. It provides an independent means of checking delivered dose (in vivo or pre-treatment) and offers an added level of treatment assurance.
    • Functionality: Reconstructs 2D or 3D dose from EPID images, uses gamma analysis to compare reconstructed dose with TPS dose, generates reports, shows % difference, volume analysis, total dose assessment, dose profiles, and EPID images. It's compatible initially with Monaco and Pinnacle TPS and Elekta linear accelerators.

    Evaluation Information (based on general statements, not specific study data):

    1. A table of acceptance criteria and the reported device performance:

      • Acceptance Criteria: Not explicitly stated in the provided text. The document broadly mentions "conformance to applicable technical design specification have been met and safety & effectiveness have been achieved."
      • Reported Device Performance: Not numerically reported in the provided text. The document states "This testing has shown that iViewDose is safe in use... Also, the use of iViewDose has been shown to have equivalent clinical efficacy when compared to the predicate devices..."
    2. Sample sized used for the test set and the data provenance: Not specified in the provided text. It mentions "data reviewed as part of this evaluation has been obtained from clinical sites," but details about the size or nature (retrospective/prospective, country of origin) of this data are absent.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not specified in the provided text. However, it indicates "Validation of the system under clinically representative conditions has been performed by competent and professionally qualified personnel."

    4. Adjudication method for the test set: Not specified in the provided text.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not mentioned as performed. The device is described as a "standalone software tool" for checking dose, not primarily as an AI assistant for human reading.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Yes, the device is explicitly described as a "standalone software tool" and its function is to objectively compare reconstructed dose with planned dose. The performance claims ("conformance to applicable technical design specification have been met") likely refer to this standalone performance.

    7. The type of ground truth used:

      • The primary comparison is between the dose calculated by the device (reconstructed from EPID data) and the data calculated by the treatment planning system (TPS dose).
      • The TPS dose serves as the reference or "ground truth" for the comparison within the iViewDose system's operation.
    8. The sample size for the training set: Not specified in the provided text. The document refers to the algorithm having "been used extensively for a number of years" and data being "obtained from clinical sites," but does not provide details on training set size.

    9. How the ground truth for the training set was established: Not specified in the provided text. Given the device's function, it's likely that the "training" (or calibration/validation) of the dose reconstruction algorithm would involve comparisons to established dosimetry measurements (e.g., ionization chambers, film, TPS calculations on phantoms), but this is not detailed.

    Ask a Question

    Ask a specific question about this device

    K Number
    K053306
    Device Name
    IVIEW
    Manufacturer
    Date Cleared
    2006-04-19

    (142 days)

    Product Code
    Regulation Number
    892.1560
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The iView™ is intended to be used by or under the direction of qualified medical personnel during prenatal monitoring, to visualize fetal features in a reconstructed 3D image that they may wish to exam more closely during routine 2D fetal diagnostic ultrasound imaging examinations, and to assist patient's understanding in communicating diagnostic results in a form that may be more easily understood. It does not provide quantitative measurements or diagnostic interpretations.

    The subject device may only be used in an adjunctive capacity with standard diagnostic ultrasound imaging techniques and that no clinical decisions or diagnostic interpretations should be made based solely on the use of this device.

    3D image reconstructed from this software is referring use only and is not responsible for any diagnosis.

    Device Description

    Adds 3D imaging capability to commercial 2D ultrasound imaging systems

    AI/ML Overview

    I am sorry, but the provided text does not contain the detailed information necessary to answer your request about the acceptance criteria and the study proving the device meets those criteria.

    The document is a 510(k) summary for the OnDemandSoft iView, which focuses primarily on establishing substantial equivalence to a predicate device (SonoReal system). It includes a comparison table of features between the two devices, but it does not specify any acceptance criteria, performance metrics, or details of a clinical study designed to demonstrate the device meets acceptance criteria.

    Specifically, the document lacks information for the following points:

    1. A table of acceptance criteria and the reported device performance: This is entirely missing. The comparison table is for features, not performance metrics against acceptance criteria.
    2. Sample size used for the test set and the data provenance: Not mentioned.
    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not mentioned.
    4. Adjudication method: Not mentioned.
    5. Multi reader multi case (MRMC) comparative effectiveness study: Not mentioned.
    6. Standalone (i.e. algorithm only without human-in-the loop performance) performance: Not mentioned.
    7. Type of ground truth used: Not mentioned.
    8. Sample size for the training set: Not mentioned.
    9. How the ground truth for the training set was established: Not mentioned.

    The 510(k) summary generally highlights the device's technical specifications and how they compare to a previously cleared device, focusing on "substantial equivalence" rather than presenting detailed performance study results against predefined acceptance criteria.

    Ask a Question

    Ask a specific question about this device

    K Number
    K981790
    Date Cleared
    1998-08-07

    (79 days)

    Product Code
    Regulation Number
    892.5050
    Reference & Predicate Devices
    N/A
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The FOS iView EPID, as with the predicate SKI 100, is intended to be used with radiation therapy treatments of malignant neoplastic diseases, as determined by a licensed medical practitioner.

    Device Description

    The iView EPID is an enhancement to the existing SRI 100 electronic portal imaging device which has previously been cleared for commercial distribution. The primary reason for the introduction of this device is to address obsolescence issues associated with the image acquisition components and the computer hardware/operating system.

    AI/ML Overview

    This document is a 510(k) submission for the Elekta Oncology Systems iView Electronic Portal Imaging Device (EPID). It primarily focuses on demonstrating substantial equivalence to a predicate device (SRI 100 EPID) rather than presenting a detailed study with specific acceptance criteria and performance metrics for the new device.

    Therefore, for many of your requested items, the information is not explicitly provided in the furnished text. The submission relies heavily on the established safety and effectiveness of the predicate device and the fact that the iView EPID addresses obsolescence issues without introducing new safety or effectiveness concerns.

    Here's an analysis based on the provided text:

    1. Table of acceptance criteria and the reported device performance

    This information is not explicitly stated in the provided text. The submission focuses on demonstrating substantial equivalence to an existing device rather than presenting specific performance metrics against pre-defined acceptance criteria for the iView EPID itself. The "reported device performance" is implied to be equivalent to the predicate device.

    Acceptance CriteriaReported Device Performance
    (Not explicitly defined in the document for the new device)(Implied to be equivalent to the predicate SRI 100 EPID)

    2. Sample sized used for the test set and the data provenance

    Not applicable/Not provided. The document does not describe a new study with a test set of data for the iView EPID's performance. The basis for safety and effectiveness is primarily the predicate device's track record and the assertion that the iView EPID does not raise new issues.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    Not applicable/Not provided. No new test set data or ground truth establishment is described for the iView EPID's performance.

    4. Adjudication method for the test set

    Not applicable/Not provided. No new test set data or adjudication method is described.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable/Not provided. This device is an Electronic Portal Imaging Device (EPID), which is a hardware and software system used in radiation therapy for imaging, not an AI-assisted diagnostic tool for human readers. No MRMC study is mentioned.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable/Not provided. This is not an algorithmic diagnostic device in the sense of AI; it's an imaging device. The document states that it's an enhancement to an existing device, implying its performance is comparable to the predicate.

    7. The type of ground truth used

    Not applicable/Not provided. No new ground truth is described for the iView EPID. The "ground truth" for its safety and effectiveness relies on the established use of the predicate device (SRI 100) in treating malignant neoplastic diseases, as determined by a licensed medical practitioner.

    8. The sample size for the training set

    Not applicable/Not provided. There is no mention of a training set as this is not an AI/machine learning device in the context of image analysis or diagnostics that requires a separate training set.

    9. How the ground truth for the training set was established

    Not applicable/Not provided. As there is no mentioned training set, this question is not applicable.


    Summary of what the document does provide regarding acceptance criteria and studies:

    The document demonstrates that the Elekta Oncology Systems iView EPID meets acceptance criteria primarily through:

    • Substantial Equivalence: The core argument is that the iView EPID is an "enhancement" to the existing SRI 100 EPID, which has already been cleared and has a "proven track record for safety." This implies the acceptance criteria for safety and effectiveness are met by virtue of being similar to or better than the predicate.
    • Addressing Obsolescence: The primary reason for the new device is to update components and address obsolescence, not to fundamentally change its function or performance in a way that would require new clinical effectiveness studies.
    • Compliance with Standards:
      • Safety Standards: The iView EPID has been subject to compliance testing as defined in IEC 601-1 and IEC 601-2-1. Proprietary IT equipment complies with IEC 950 and/or UL 1950.
      • Quality System Standards: Elekta Oncology Systems is a registered medical device manufacturer assessed against ISO 9001, EN 46001, and the Medical Device Directive, 93/42/EEC Annex II. Their software quality system meets ISO 9001, EN 46001, the Medical Device Directive, and US 21 CFR 820 GMP.
      • Regulatory Compliance: The device bears the CE mark, affirming compliance with relevant European Directives (Medical Device Directive, Electromagnetic Compatibility Directive).
    • Hazard Analysis: Elekta Oncology Systems conducted a hazard analysis and concluded that the device does not introduce new types of safety or effectiveness considerations.
    • Audits: The quality system is subject to periodic internal quality audits and regular, planned, and documented GMP audits by external auditors (SGS Yarsley and FDA).

    In essence, the "study" proving the device meets acceptance criteria is a documentation of substantial equivalence, compliance with established safety and quality standards, and internal hazard analysis, rather than a clinical performance study with explicit test sets and ground truth.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1