Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K182771

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2019-03-12

    (165 days)

    Product Code
    Regulation Number
    888.3070
    Reference & Predicate Devices
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The ISS-JAZZ Screw System is intended to provide immobilization of spinal segments in skeletally mature patients as an adjunct to fusion of the thoracic, lumbar and/or sacral spine. The ISS-JAZZ Screw System is intended for posterior, non-cervical pedicle and non-pedicle fixation for the following indications: degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies), spondylolisthesis, trauma (i.e., fracture or dislocation), spinal deformities (i.e., scoliosis, kyphosis and/or lordosis), tumor, pseudoarthrosis, or revision of a failed fusion attempt.

    When used for posterior non-cervical pedicle screw fixation in pediatric patients, the ISS-JAZZ Screw System implants are indicated as an adjunct to fusion to treat adolescent idiopathic scoliosis. These devices are to be used with autograft and/or allograft. Pediatric pedicle screw fixation is limited to a posterior approach.

    The JAZZ systems (including the JAZZ Claw Connector) are temporary implants to be used in orthopedic surgery. The JAZZ systems are intended to provide temporary stabilization as a bone anchor during the development of solid bony fusion and aid in the repair of bone fractures.

    The indications for use include the following applications:

      1. Spinal trauma surgery, used in sublaminar or facet wiring techniques;
      1. Spinal reconstructive surgery, incorporated into constructs for the purpose of correction of spinal deformities such as adolescent idiopathic scoliosis, adult scoliosis, kyphosis and spondylolisthesis:
      1. Spinal degenerative surgery, as an adjunct to spinal fusions.

    The JAZZ Systems may also be used in conjunction with other medical implants made of titanium alloy or Cobalt-chromium-molybdenum alloy whenever "wiring" may help secure the attachment of other implants.

    Device Description

    The JAZZ CAP SP is composed of three components, the Cap SP, the Base SP, and the Insert Lock. The head shape ISS pedicle screws is modified from the current tulip shape to a cylindrical shape to match the Cap SP, resulting in the JAZZ Screw. The ISS-JAZZ Screw System is composed of monoaxial and polyaxial screws and straight and pre bent rods. The JAZZ CAP SP is added to the proximal end of the JAZZ Screw and provides for an alternate method to attach the JAZZ Band and JAZZ Passer Band to skeletal structures as compared to the current method of using a JAZZ Connector (K171881).

    AI/ML Overview

    This document is a 510(k) summary for a medical device called the ISS-JAZZ Screw System and JAZZ CAP SP. It describes the device, its intended use, comparison to predicate devices, and performance data to demonstrate substantial equivalence.

    Here's an analysis of the requested information based on the provided text, indicating when information is not provided:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not explicitly state acceptance criteria in a quantitative table format. Instead, it refers to performance testing conforming to a standard and demonstrating substantial equivalence.

    Performance CharacteristicAcceptance Criteria (Implicit)Reported Device Performance
    Dynamic CompressionEquivalence to predicate device per ASTM F1717Performed on worst-case construct, established equivalence to predicate.
    Static Axial LoadCharacterize maximum load by Band on JAZZ CAP SP to cause displacement on JAZZ ScrewTesting conducted to characterize this load.
    SterilizationNot constituting a new worst-case compared to predicateDetailed rationales provided to demonstrate this.
    CleaningNot constituting a new worst-case compared to predicateDetailed rationales provided to demonstrate this.
    BiocompatibilityNot constituting a new worst-case compared to predicateDetailed rationales provided to demonstrate this.

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size: Not explicitly stated. The document mentions "worst-case construct" for dynamic compression testing, implying a limited number of samples representative of the most challenging configuration. For static axial load, it states "Additional static axial load testing" without specifying the number of samples.
    • Data Provenance: Not provided, but generally, such performance testing for medical devices is conducted in a controlled laboratory environment. The document is for submission to the U.S. FDA, so the data is intended for U.S. regulatory review. It is a premarket notification, so the data would be based on pre-clinical testing, not clinical data from patients.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    • This information is not applicable to this type of submission. This document describes the performance testing of a spinal implant system. The "ground truth" here is based on engineering principles, standardized mechanical tests (ASTM F1717), and material science, not clinical expert consensus on diagnostic images or patient outcomes.

    4. Adjudication Method for the Test Set

    • This information is not applicable. Adjudication methods (like 2+1, 3+1) are typically used in clinical studies or studies involving human readers/interpreters where there might be disagreement in interpretations. This document pertains to mechanical performance testing.

    5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve with AI vs Without AI Assistance

    • This information is not applicable. This is not an AI/software as a medical device (SaMD) submission. It is a traditional medical device (implant) submission.

    6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) Was Done

    • This information is not applicable. This is not an AI/software as a medical device (SaMD) submission.

    7. The Type of Ground Truth Used

    • The "ground truth" in this context is defined by engineering specifications, material properties, and established mechanical testing standards (ASTM F1717). The goal is to demonstrate that the device meets predefined physical and mechanical performance characteristics, and is substantially equivalent to predicate devices based on these objective measures.

    8. The Sample Size for the Training Set

    • This information is not applicable. This is not an AI/machine learning device. There is no "training set" in the context of mechanical performance testing for an implant.

    9. How the Ground Truth for the Training Set Was Established

    • This information is not applicable. As there is no training set for an AI/ML model, there is no ground truth derivation process for such a set.
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1