510(k) Data Aggregation

    K Number
    K231664

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2023-10-19

    (134 days)

    Product Code
    Regulation Number
    882.5550
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The use of IRRAflow Active Fluid Exchange System is intracranial pressure monitoring is required, and for externally draining intracranial fluid, as a means of reducing intracranial pressure in patients where an external drainage and monitoring system is needed.

    Device Description

    The IRRAflow® Active Fluid Exchange System (AFES) is an intracranial pressure (ICP) monitoring and drainage system intended for use by professional medical hospital personnel, trained and experienced in neurosurgical medical care. The drainage flow of cerebrospinal fluid (CSF) into the IRRAflow Catheter is uni-directional and gravity-driven; there is no recirculation of the CSF. A parallel line from the saline infusion bag is used in case clearance at the tip of the catheter is required. The IRRAflow Tube Set has a cassette that clicks on to the IRRAflow Control Unit and aligns the tubing against a peristaltic pump and pinch valve. The IRRAflow Drainage Collection System is attached to the Control Unit, using the Laser Leveler for defining the height of the Drainage Collection System relative to the catheter's tip position in the patient's head. This positioning is used for controlling the speed of drainage. The tubing and catheter can be disconnected and connected by standard Luer-Lock connectors. Settings can be changed via the user interface on the Control Unit. The default mode provides drainage and measuring ICP, allowing bolus injections when indicated. The bolus injections allow the catheter to be flushed when it becomes clogged. CSF or intracranial fluid samples can be taken from the Drainage Collection System.

    AI/ML Overview

    The provided text is a 510(k) premarket notification decision letter from the FDA regarding the IRRAflow Active Fluid Exchange System (AFES). The purpose of this submission is to demonstrate substantial equivalence to a previously cleared predicate device, specifically regarding changes to the Tube Set and Drainage Collection System. Therefore, the "acceptance criteria" and "study that proves the device meets the acceptance criteria" are focused on demonstrating that the modified device's performance is equivalent to, or better than, the predicate device.

    Here's an analysis based on the provided document:

    1. A table of acceptance criteria and the reported device performance

    The document does not explicitly present a table of quantitative acceptance criteria alongside numerical performance results for the device. Instead, it relies on a "PASS" or "FAIL" outcome for various verification and validation tests. The acceptance criteria for these tests are implied to be established by the test protocols and industry standards (e.g., "Pressure accuracy per protocol," "Durability, flow and freedom from leakage per protocol").

    Here's a table summarizing the tests performed and their reported outcomes:

    Test CategoryTest PerformedReported Device Performance (Result)
    Biocompatibility TestingCytotoxicity (MEM Elution)PASS
    Sensitization TestPASS
    Irritation/Intracutaneous Reactivity TestPASS
    Acute Systemic Toxicity, Injection TestPASS
    Systemic Toxicity, Mediated PyrogenPASS
    Genotoxicity (on various strains of bacteria)PASS
    Genotoxicity (on mouse lymphoma cells)PASS
    Bench and Electrical TestingElectrical Requirements Verification (Pressure accuracy)PASS
    Mechanical Performance Verification (Durability, flow, freedom from leakage)PASS
    Life Cycle Verification Test (Simulated use for reliability)PASS
    Shelf Life / Package Integrity TestingSimulated Distribution Test (ASTM D4332-14, ASTM D4169-22 Cycle 13)PASS
    Package Integrity Test (ASTM F1886-16, ASTM F2096-11)PASS
    Package Seal Strength Test (EN 868-5:2009)PASS
    Aging Test (Accelerated and real-time aging)PASS
    Sterilization TestingSterilization Process Validation (Ethylene Oxide for Cassette and Drainage System)PASS

    2. Sample size used for the test set and the data provenance

    The document does not specify the sample sizes used for any of the tests listed in Table 3.
    The data provenance is not explicitly stated beyond general descriptions of the tests (e.g., "The Minimal Essential Media (MEM) Elution test," "This test was designed to evaluate the allergenic potential"). There is no mention of country of origin or whether the tests were retrospective or prospective. Given the nature of these tests (bench, electrical, biocompatibility, sterilization validation), they are typically conducted as prospective laboratory studies rather than clinical studies using patient data.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    This information is not provided in the document. The tests performed are primarily engineering and laboratory-based, often following standardized protocols. Therefore, the concept of "ground truth" established by clinical experts (like radiologists for imaging devices) would not directly apply to these types of performance tests. The "ground truth" here is implied by adherence to established test methods and acceptable performance limits defined by those methods.

    4. Adjudication method for the test set

    This information is not provided and is generally not applicable to the types of performance tests described (biocompatibility, electrical, mechanical, shelf life, sterilization). Adjudication methods like 2+1 or 3+1 are typically used in clinical studies where multiple human readers interpret data (e.g., medical images) and a consensus is needed to establish a definitive ground truth.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    There is no indication that a multi-reader multi-case (MRMC) comparative effectiveness study was done. The device is a physical system for fluid exchange and pressure monitoring, not an AI-powered diagnostic tool for image interpretation or similar tasks that would typically involve human readers and AI assistance.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This question is not applicable as the IRRAflow AFES is a medical device for intracranial pressure monitoring and fluid drainage, not an algorithm or AI system. Its performance is evaluated through engineering and biological safety tests.

    7. The type of ground truth used

    For the performance tests described (biocompatibility, electrical, mechanical, shelf life, sterilization), the "ground truth" is based on:

    • Established Test Standards and Protocols: Such as ASTM D4332-14, ASTM D4169-22 Cycle 13, ASTM F1886-16, ASTM F2096-11, EN 868-5:2009.
    • Defined Acceptance Criteria: These criteria are inherent to the test methods and are designed to ensure safety and effectiveness (e.g., pressure accuracy per protocol, freedom from leakage per protocol).
    • Laboratory Measurements and Observations: The results are direct measurements or observations within controlled laboratory environments.

    It's not "expert consensus," "pathology," or "outcomes data" in the clinical sense, but rather adherence to predefined engineering and biological safety specifications.

    8. The sample size for the training set

    This question is not applicable because the IRRAflow AFES is a hardware medical device with specific Tube Set and Drainage Collection System modifications, not a machine learning or AI model that requires a "training set."

    9. How the ground truth for the training set was established

    This question is not applicable as there is no training set for this device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K222471

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2022-12-08

    (114 days)

    Product Code
    Regulation Number
    882.5550
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The use of the IRRAflow Active Fluid Exchange System is indicated when intracranial pressure monitoring is required and for externally draining intracranial fluid as a means of reducing intracranial pressure in patients where an external drainage and monitoring system is needed.

    Device Description

    The IRRAflow® Active Fluid Exchange System (formerly branded as the CNS System) is an intracranial pressure (ICP) monitoring and drainage system, substantially equivalent to the predicate 510(k) K200807. Changes described herein were evaluated using design controls. The drainage flow of cerebrospinal fluid (CSF) into the IRRAflow Catheter is uni-directional and gravity-driven; there is no recirculation of the CSF. A parallel line from the saline infusion bag is used in case clearance at the tip of the catheter is required. The IRRAflow Tube Set has a cassette that clicks on to the IRRAflow Control Unit and aligns the tubing against a peristaltic pump and pinch valve. A drainage bag is attached to the Control Unit, using the Laser Leveler for defining the height of the bag relative to the catheter's tip position in the patient's head. This positioning is used for controlling the speed of drainage. The tubing and catheter can be disconnected and connected by standard Luer-Lock connectors. Settings can be changed via the user interface on the Control Unit. The default mode provides drainage and measuring ICP, allowing bolus injections when indicated. The bolus injections allow the catheter to be flushed when it becomes clogged. CSF or intracranial fluid samples can be taken from the aspiration port.

    AI/ML Overview

    The provided text is a 510(k) summary for the IRRAflow® Active Fluid Exchange System (AFES). It details the device's technical specifications and compares it to predicate devices to demonstrate substantial equivalence. However, it does not contain the information requested regarding acceptance criteria and performance data for an AI/ML powered device.

    The document explicitly states that the device is an "intracranial pressure (ICP) monitoring and drainage system" and that "Changes made only to the Control Unit software required additional testing, including verification and validation of new and enhanced functionality." It refers to "Control Unit 4.0," which includes software enhancements.

    Despite mentioning "User software enhancements" and "pressure accuracy improvement," the document focuses on engineering verification and validation tests for a hardware/software medical device, not an AI/ML algorithm. The tests listed in "Table 4" (Verification by Inspection, BSM Requirements Verification, Electrical Requirements Verification, Mechanical Performance Verification, Life Cycle Verification Test, Packaging Verification) are standard for medical device hardware and software, but do not align with the typical criteria and studies for proving the performance of an AI/ML-powered algorithm, especially none that would require ground truth from experts, multi-reader studies, or specific AI acceptance metrics like sensitivity/specificity or AUC.

    Therefore, I cannot fulfill the request for a table of AI acceptance criteria and performance, sample sizes for test/training sets, expert qualifications, adjudication methods, MRMC studies, or standalone algorithm performance, as these details are not present in the provided document. The document describes a conventional medical device clearance, not an AI/ML clearance.

    Without further information that explicitly details a clinical study demonstrating the performance of an AI/ML component with human-in-the-loop or standalone, I cannot generate the requested output.