Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K202915

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2021-04-23

    (206 days)

    Product Code
    Regulation Number
    884.4120
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The IRIS Thermocoagulator and Digital Colposcope is intended for the destruction of human tissue with high temperatures by tissue contact with an electrically heated probe, and to provide magnified visualization of the tissues of the vagina, cervix and external genitalia in order to aid in selecting areas for biopsy and diagnosing as needed for a colposcopy exam.

    Device Description

    The Liger Medical IRIS Thermocoagulator and Digital Colposcope is a portable thermal coagulator with integral colposcope that provides assistance for clinician examination of the cervix as well as utilizes a heated probe to ablate human tissue. It is specifically designed for use in resource-limited settings. The colposcope and thermal ablation modes can be used separately from each other. The device is a handheld, portable, battery-operated, ablator with an LCD display and optical camera to assist diagnosis and treatment, in a safe, effective, easy to use package with sufficient battery life to sustain work for 4 hours. Thermal coagulation can be used in hospital and non-hospital professional healthcare locations.

    The IRIS colposcope/thermal ablation device is not intended on into the vaginal canal during colposcopic examination. The camera remains outside the vaginal cavity and functions comparably to a standard non-invasive colposcope in terms of providing magnified visual assistance to the clinician. There is no patient contact during the colposcopic examination. Contact between the clinician and the device is mitigated through good clinical practice of wearing protective gloves, limiting contact to intact skin protected by operating gloves.

    The thermal ablation probe, which is reusable and provided non-sterile, is designed to perform low-power destruction of human tissue with high temperatures by tissue contact with an electrically heated probe tip. The biocompatibility of the probe materials conforms to all recommended endpoints per FDA's guidance document "Use of International Standard ISO 10993-1, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process" for limited duration mucosal membrane contact.

    AI/ML Overview

    The provided text is a 510(k) Summary for the Liger Medical IRIS Thermocoagulator and Digital Colposcope. It details the device's intended use, comparison to predicate devices, and non-clinical performance testing. However, it explicitly states "Not Applicable" under "9. Clinical Performance Data."

    Therefore, based solely on the provided text, a detailed response to points 1-7, 8, and 9 for a study proving the device meets acceptance criteria cannot be fully constructed, as the document states no clinical performance data was used for this submission. The device's substantial equivalence was demonstrated through non-clinical performance testing and comparison to predicate devices.

    Here's a breakdown of what can be extracted and what cannot:

    1. A table of acceptance criteria and the reported device performance:

    The document describes various non-clinical tests performed, implying these tests have acceptance criteria, but the specific numerical acceptance criteria and the precise quantified performance results are not detailed in a comparative table format. We only get a list of tests performed.

    Table of Acceptance Criteria (Implied) and Reported Device Performance (General):

    Acceptance Criteria (Implied from tests)Reported Device Performance (General Statement)
    Electrical/Thermal SafetyMet ANSI AAMI ES60601-1:2005/(R)2012 and A1:2012, C1:2009/(R)2012 and A2:2010/(R)2012
    Electromagnetic Compatibility (EMC)Met IEC 60601-1-2:2014 (4th Ed)
    Probe Fault DetectionTested; Implied to have met design input requirements
    Battery Fault DetectionTested; Implied to have met design input requirements
    Probe Heat-up time and Temperature Stability at Set-PointTested; Implied to have met design input requirements
    Heater Circuit Hardware ChecksTested; Implied to have met design input requirements
    Heater Performance on TissueTested; Implied to have met design input requirements
    Software Application TestingTested; Implied to have met design input requirements
    Use-Life TestingTested; Implied to have met design input requirements
    Interoperability and Connectivity TestingTested; Implied to have met design input requirements
    Endoscope General RequirementsMet ISO 8600-1:2015
    Field of View and Direction of ViewMet ISO 8600-3:2019
    Limiting Spatial ResolutionTested; Implied to have met design input requirements
    BiocompatibilityConforms to all recommended endpoints per FDA's guidance document "Use of International Standard ISO 10993-1" for limited duration mucosal membrane contact.

    2. Sample size used for the test set and the data provenance:

    • Sample size: Not specified as no clinical performance data was provided. The non-clinical tests would have involved specific test units, but the "sample size" in terms of patient data is not applicable here.
    • Data provenance: Not applicable.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • Not applicable as no clinical performance data/test set with human expert ground truth was mentioned.

    4. Adjudication method for the test set:

    • Not applicable as no clinical performance data/test set requiring adjudication was mentioned.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, if so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • No, a MRMC comparative effectiveness study was not done according to this document. The device includes a digital colposcope for magnified visualization to aid in selecting areas for biopsy and diagnosis, but the submission explicitly states "Not Applicable" for clinical performance data, meaning no study assessing human reader improvement with or without AI assistance was presented here.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

    • Not applicable. The device is a physical instrument (thermocoagulator and digital colposcope), not an AI algorithm in isolation. While it has a "Digital Colposcope" component capable of camera-assisted visualization, the document does not describe any standalone algorithm performance.

    7. The type of ground truth used:

    • For the non-clinical tests, the "ground truth" would be the established engineering and safety standards (e.g., ANSI AAMI ES60601-1, IEC 60601-1-2, ISO 8600-1, ISO 8600-3, ISO 10993-1) against which the device's technical performance was measured. No clinical ground truth (e.g., pathology, outcomes data) was used or presented in this 510(k) summary.

    8. The sample size for the training set:

    • Not applicable. This document is for a physical medical device, not an AI/ML algorithm that requires a training set of data.

    9. How the ground truth for the training set was established:

    • Not applicable.
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1