Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K161286

    Validate with FDA (Live)

    Date Cleared
    2016-08-04

    (87 days)

    Product Code
    Regulation Number
    878.4810
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The VE2000 device is indicated for use in surgical, aesthetic applications in permanent hair reduction of benign pigmented lesions and benign vascular lesions.

    Permanent hair reduction is defined as the long-term, stable reduction in the number of hairs regrowing when measured at 6, 9, and 12 months after the completion of a treatment regimen

    Device Description

    The VE2000 device are intense pulsed light system which delivers intense pulsed light at a wavelength ranging from 480nm-1200nm. Intense Pulsed Light (IPL) systems work on the principles of selective thermolysis. That is, causing thermal damage to target chromophores by using light of appropriate wavelength in pulses that exceeds the chromophores' thermal relaxation time but sparing normal skin by limiting the pulse width below the thermal relaxation time for skin.

    IPL Systems are different from lasers in that they deliver many wavelengths in each pulse of light instead of just one wavelength. Generally, IPL enhances penetration without using excessive energy levels and enables targeting of specific chromophores.

    Based on this,The VE2000 device (inclusive of the handpiece used to deliver pulsed-light energy) are indicated for use in surgical, aesthetic and cosmetic applications in permanent hair removal, reduction of pigmented lesions and vascular lesions.

    AI/ML Overview

    The provided document outlines the substantial equivalence (SE) determination for the VE2000 IPL Therapy Machine (K161286) to a predicate device (Intense Pulsed Light (IPL) Systems, K122995).

    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not explicitly present "acceptance criteria" for clinical performance. Instead, it demonstrates substantial equivalence by comparing the proposed VE2000 device's performance parameters and intended uses to those of a legally marketed predicate device. The underlying "acceptance criteria" are therefore implied to be that the proposed device performs comparably to the predicate device for its stated indications.

    Parameter/CharacteristicAcceptance Criteria (Implied from Predicate)Reported Device Performance (VE2000)Meets CriteriaNotes on Comparison
    Product CodeONFONFYesSubstantially Equivalent (SE)
    Regulation No.21 CFR 878.481021 CFR 878.4810YesSE
    ClassIIIIYesSE
    Intended Use (Overall)Surgical, aesthetic, and cosmetic applications in permanent hair removal, reduction of pigmented lesions, acne therapy, freckles, vascular lesions and facial blemish.Surgical, aesthetic applications in permanent hair reduction, reduction of benign pigmented lesions and benign vascular lesions.YesSE (Proposed device has a narrower IFU scope, making it SE)
    Light SourceIntense pulsed lightIntense pulsed lightYesSE
    Wavelength430-1200nm, 530-1200nm, 640-1200nm, Optional: 480-1200nm, 560-1200nm, 590-1200nm, 690-1200nm, 750 -1200nm480nm-1200nm, 590nm-1200nm, 640nm-1200nmYesSE (Ranges overlap substantially, and proposed device ranges fall within or are comparable to predicate's capabilities)
    Delivery SystemSapphireSapphireYesSE
    Energy Density10-60 J/cm²1-50 J/cm²Yes, with analysisAnalysis 1: "the difference is very slight, and only in the range, but because the proposed device has narrow IFU scope and same IFU setting with the predicate device, the current parameter range is available for the indication for use of proposed device." Considered SE.
    Pulse Delay5-50 ms1-50 msYes, with analysisAnalysis 1: "the difference is very slight, and only in the range... Considered SE.
    Pulse Width1-20 ms1-25 msYes, with analysisAnalysis 1: "the difference is very slight, and only in the range... Considered SE.
    Max. Power2000 W2200 WYesSE (Slight increase, likely not detrimental for a Class II device)
    Spot SizeMED-210: 15mmX50mm (optional: 12mmX33mm, 15mmX35mm); MED-230: A: 12mm X33mm; B: 15mmX50mm (optional: 15mmX35mm)12x30mmYes, with analysisAnalysis 1: "the difference is very slight, and only in the range... Considered SE.
    Permanent Hair Reduction Settings:
    - Wavelength Range (nm)640-1200/690-1200/750-1200640-1200YesComparison shows overlap and acceptable ranges.
    - Energy Range (J/cm²)10-4410-44YesSE
    - Pulse Width (ms)3-143-14YesSE
    - Pulse Delay (ms)16-3216-32YesComparison shows overlap and acceptable ranges.
    - Spot Size (mm)12mm X33mm; 15mmX50mm; 15mmX35mm12x30mmYesComparison shows overlap and acceptable ranges.
    Pigmented Lesions Settings:
    - Wavelength Range (nm)480-1200/530-1200/560-1200480-1200YesComparison shows overlap and acceptable ranges.
    - Energy Range (J/cm²)12-4412-44YesSE
    - Pulse Width (ms)3-93-9YesComparison shows overlap and acceptable ranges.
    - Pulse Delay (ms)16-3216-32YesComparison shows overlap and acceptable ranges.
    - Spot Size (mm)12mm X33mm; 15mmX50mm; 15mmX35mm12x30mmYesComparison shows overlap and acceptable ranges.
    Vascular Lesions Settings:
    - Wavelength Range (nm)530-1200/560-1200/590-1200590-1200YesComparison shows overlap and acceptable ranges.
    - Energy Range (J/cm²)10-4210-42YesSE
    - Pulse Width (ms)3-83-8YesComparison shows overlap and acceptable ranges.
    - Pulse Delay (ms)16-3216-32YesComparison shows overlap and acceptable ranges.
    - Spot Size (mm)12mm X33mm; 15mmX50mm; 15mmX35mm12x30mmYesComparison shows overlap and acceptable ranges.
    Electrical SafetyCompliance with IEC 60601-1Compliance with IEC 60601-1YesSE
    EMCCompliance with IEC 60601-1-2Compliance with IEC 60601-1-2YesSE
    Patient Contact MaterialHandpieceHandpieceYesSE
    CytotoxicityNo toxicity (ISO 10993-5)No toxicity (ISO 10993-5)YesSE
    IrritationNo irritation to skin (ISO 10993-10)No irritation to skin (ISO 10993-10)YesSE
    SensitizationNo significant evidence of sensitization (ISO 10993-10)No significant evidence of sensitization (ISO 10993-10)YesSE

    2. Sample Size Used for the Test Set and Data Provenance

    The document explicitly states: "No clinical study is included in this submission."

    Therefore, there is no test set sample size or data provenance related to clinical performance described. The safety and effectiveness determination is based on non-clinical testing and comparison to a predicate device.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

    Since no clinical study was conducted and no clinical test set was used for performance evaluation, this information is not applicable and is not provided in the document.

    4. Adjudication Method for the Test Set

    Not applicable, as no clinical test set was used.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done

    No, an MRMC comparative effectiveness study was not done. The submission relies on non-clinical testing and substantial equivalence to a predicate device, without involving human readers or AI assistance in a comparative effectiveness study.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

    Not applicable. The device is an IPL Therapy Machine, not an algorithm, and its performance evaluation for substantial equivalence was based on non-clinical testing and comparison to a physical predicate device, not standalone algorithmic performance.

    7. The Type of Ground Truth Used

    The "ground truth" for the substantial equivalence determination for this device is primarily established by:

    • Compliance with recognized electrical safety and EMC standards: IEC 60601-1, IEC 60601-1-2, IEC 60601-2-57.
    • Biocompatibility testing: ISO 10993-5 (Cytotoxicity) and ISO 10993-10 (Irritation and Sensitization).
    • Direct functional and parametric comparison to a legally marketed predicate device (K122995) for intended use, light source, wavelength, energy density, pulse characteristics, and spot size. The predicate device itself has presumably been accepted by the FDA based on its own evidence of safety and effectiveness.

    8. The Sample Size for the Training Set

    Not applicable. This device is an IPL therapy machine, not an AI/ML algorithm that requires a training set.

    9. How the Ground Truth for the Training Set was Established

    Not applicable, as there is no training set for this type of device.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1