510(k) Data Aggregation

    K Number
    K231596

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2023-06-30

    (29 days)

    Product Code
    Regulation Number
    870.5900
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The IOB Temperature Management System Model IOB-507 is indicated for hypothermic patients or normothermic patients for whom induced hyperthermia or localized increase in temperature is clinically indicated.

    Device Description

    The IOB Temperature Management System Model IOB-507 consists of the IOB Warming Unit and the IOB Warming Blankets. The IOB Warming Unit, model IOB-507, draws ambient- temperature air through a 0.2 micron air filter. The filtered air is warmed to a selected temperature. The warmed air enters the IOB Warming Blanket through the hose and is distributed through delivery channels. Perforations on the patient side of the air delivery channels in the warming blanket gently disperse the warmed air over and around the patient. The warming unit has three temperature settings of 32℃, 38℃, and 43℃. These temperature settings are servo-controlled by a thermistor placed at the end of the hose where the hose connects to the blanket. The unit can also deliver ambient-temperature air. The temperature indicated on the control panel is the temperature of air being delivered to the average contact surface temperature of the blanket. A control thermistor in the warming unit adjusts the power applied to the heater to maintain the selected temperature. This enables the unit to maintain the selected temperature under variations in ambient temperature. Besides, the warming unit has high and low air flow options. A safety thermistor provides a signal to a separate high-temperature analog circuit. The safety thermistor activates and produces an alarm if the temperature exceeds the set point. The analog safety circuit provides an independent means of shutoff, which discontinues power to the heater and motor. This prevents patient exposure to excessive temperatures. IOB warming blankets are single-use and disposable. Each blanket consists of two layers of non-woven polypropylene fabric coated with a layer of polyethylene. The layers are bonded together to form a distribution network of air delivery channels. The warm air is distributed around the blanket through the delivery channels and exits the blanket through a specially designed series of perforations in the patient side of the blanket.

    AI/ML Overview

    The document describes the IOB Temperature Management System Model IOB-507, which includes the IOB Warming Unit and IOB Warming Blankets. The system is indicated for hypothermic patients or normothermic patients for whom induced hyperthermia or localized increase in temperature is clinically indicated. The device is a Class II device with product code DWJ (Thermal regulating system).

    No clinical studies were performed. The device's substantial equivalence determination is based on non-clinical tests. Therefore, there is no information about acceptance criteria or a study that proves the device meets specific acceptance criteria in the context of human performance or clinical outcomes.

    However, based on the non-clinical tests mentioned, we can infer some performance characteristics and the tests conducted to demonstrate them.

    1. A table of acceptance criteria and the reported device performance

    Since no specific acceptance criteria for clinical performance are provided, this table focuses on the performance characteristics assessed through non-clinical testing.

    Acceptance Criteria (Implied from tests)Reported Device Performance
    IOB Warming Unit Model IOB-507
    Temperature Settings Accuracy (from Table 1)43°C +/- 2°C, 38°C +/- 2°C, 32°C +/- 2°C, Ambient
    Forced Air Over Temperature Shut-off (from Table 1)Auto-shuts heater off at 47°C +/- 2°C
    Electrical Safety and EMC (from Section VII.1.a)Meets relevant applicable IEC standards (IEC 60601-1, IEC 60601-1-2)
    Usability and Alarm System (from Section VII.1.a)Meets relevant applicable IEC standards, meets design specifications, performs as intended.
    Software Validation (from Section VII.1.a)Software validation report shows device performs as intended.
    Air Velocity (from Table 1 and Section VII.1.a)30-45cfm (device performance reported from Table 1, air velocity testing performed per Section VII.1.a)
    IOB Warming Blankets
    Temperature Uniformity (from Section VII.1.b)All test results show temperature uniformity equivalence between the IOB Warming Blankets and the predicate. (Tested by measuring five testing points on blanket surface at different IOB Warmer settings)
    Shelf-life (from Section VII.1.c)3 years (stability tests performed)
    Transport Simulation (Container / Package Integrity) (from Section VII.1.d)No package damage was observed. All product hold integrity after the transport testing. (Tested according to ASTM D4169)
    Leakage (Blanket Integrity) (from Section VII.1.e)No leakage was found. (Bubble testing carried out according to ASTM F2096)
    Biocompatibility (Cytotoxicity, Irritation, Sensitivity) (from Section VII.1.f)Demonstrated to be suitable for the intended use of the product. (Tested according to ISO 10993 for a limited contact device)

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    The document does not provide details on the sample size for non-clinical tests (e.g., how many units of the warming unit or blankets were tested for each criterion). The data provenance is not mentioned, but as these are non-clinical (laboratory) tests, the concept of "country of origin of the data," "retrospective," or "prospective" as typically applied to clinical studies is not directly relevant. These tests would have been performed in a controlled laboratory environment by the manufacturer or a third-party testing facility.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    This information is not applicable. The non-clinical tests described (electrical safety, temperature accuracy, material stability, etc.) do not involve human interpretation or the establishment of ground truth by clinical experts. These are objective engineering and material science tests.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    This information is not applicable. Adjudication methods are relevant for clinical studies or image interpretation tasks performed by multiple human readers, which is not the case here.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    No MRMC study was done, as this device does not involve AI or human readers for diagnostic interpretation. It is a temperature management system.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

    This information is not applicable. The device is a physical temperature management system, not an algorithm.

    7. The type of ground truth used (expert concensus, pathology, outcomes data, etc)

    This information is not applicable. The non-clinical tests described use objective physical measurements and established engineering and material science standards (e.g., IEC, ASTM, ISO) as their "ground truth" or reference for evaluating performance.

    8. The sample size for the training set

    This information is not applicable, as this is a non-AI/ML device.

    9. How the ground truth for the training set was established

    This information is not applicable, as this is a non-AI/ML device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K162679

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2017-04-28

    (214 days)

    Product Code
    Regulation Number
    870.5900
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The IOB Temperature Management system is indicated for hypothermic patients or normothermic patients for whom induced hyperthermia or localized increase in temperature is clinically indicated.

    Device Description

    The IOB Temperature Management System consists of the IOB warming unit and the IOB warming blankets. The IOB warming unit draws ambient- temperature air through a 0.2 micron air filter. The filtered air is warmed to a selected temperature. The warmed air enters the IOB Warming Blanket through the hose and is distributed through delivery channels. Perforations on the patient side of the air delivery channels in the warming blanket gently disperse the warmed air over and around the patient. The warming unit has three temperature settings of 32°C, 38°C, and 43°C. These temperature settings are servo-controlled by a thermistor placed at the end of hose where the hose connects to the blanket. The unit can also delivers ambient-temperature air. The temperature indicated on the control panel is the temperature of air being delivered to the blanket at the end of the hose. A control thermistor in the warming unit adjusts the power applied to the heater to maintain the selected temperature. This enables the system to maintain the selected temperature under variations in ambient temperature. Besides, the warming unit has high and low air flow for choose.

    The IOB Warming Blankets in this submission are the following:
    Torso IOB-001 Lower Body IOB-002 Upper Body IOB-003 Full Body IOB-004 Pediatric Underbody IOB-005 Adult Underbody IOB-006 Pediatric Full Body IOB-007 Full Body Surgical IOB-008 Large Pediatric Underbody IOB-009 Spinal Underbody IOB-010 Lithotomy Underbody IOB-011 Pediatric Lower Body IOB-012 Pediatric Long IOB-014 Cath Lab IOB-015 Surgical Access IOB-016 Chest Access IOB-017 Multi-Access IOB-018 Dual Port Torso IOB-019 Cardiac Access IOB-020 XL Upper Body IOB-021 Outpatient Care IOB-022 Cardiac IOB-023
    These blankets are single-use and disposable. Each blanket consists of two lavers of non-woven polypropylene fabric coated with a layer of polyethylene. The lavers are bonded together to form a distribution network of air delivery channels. The warm air is distributed around the blanket through the delivery channels and exits the blanket through a specially designed series of perforations in the patient side of the blanket.

    AI/ML Overview

    The provided text is a 510(k) summary for the IOB Temperature Management System, which is a medical device for warming patients. It addresses substantial equivalence to a predicate device (Bair Hugger Temperature Management System) rather than outlining acceptance criteria and a study proving device performance in the context of an AI/ML powered device.

    Therefore, the specific information requested about acceptance criteria and studies demonstrating performance as would be expected for an AI/ML driven diagnostic or assistive technology (e.g., sample sizes, ground truth establishment, expert adjudication, MRMC studies, standalone performance) is not present in the provided document.

    The document focuses on the physical and functional characteristics of a thermal regulating system and its blankets, demonstrating its equivalence to a legally marketed predicate device.

    However, I can extract the safety and performance characteristics mentioned in the document, which serve a similar purpose of demonstrating the device's suitability for its intended use, even if they are for a physical device rather than an AI/ML algorithm.

    Here's the information that can be extracted, framed to best fit your request:


    Acceptance Criteria and Study for the IOB Temperature Management System (as applicable to a physical thermal regulating device)

    Since the IOB Temperature Management System is a physical device (a thermal regulating system) and not an AI/ML powered diagnostic or assistive technology, the typical "acceptance criteria" and "study" framework for AI/ML performance metrics (like sensitivity, specificity, AUC, etc.) and related study design details (sample size for test/training, ground truth methods, expert adjudication, MRMC) does not apply.

    Instead, the provided document outlines the general "Safety and Performance Characteristics" and their verification methods to demonstrate substantial equivalence to a predicate device.

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria (Performance Characteristic)Reported Device Performance (Verification Method)
    Temperature Uniformity"Temperature uniformity tests show equivalence between the IOB warming blankets and the predicate."
    Shelf-Life"Accelerated stability tests show three years shelf-life of the IOB warming blankets."
    Electrical Safety / EMC"Electrical safety/EMC testing... show that the device meets its design specifications, performs as intended."
    Software Validation"Software validation show that the device meets its design specifications, performs as intended."
    Biocompatibility"Biocompatibility testing (cytotoxicity, irritation and sensitivity) according to ISO 10993 for a limited contact device was demonstrated to be suitable for the intended use of the product."

    2. Sample size used for the test set and the data provenance

    • Not applicable / Not specified in the provided document for the types of tests described (temperature uniformity, accelerated stability, electrical safety, software validation, biocompatibility). These tests typically involve a defined number of units or samples under specific conditions rather than patient "test sets" as understood for AI/ML evaluation.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    • Not applicable / Information not provided. The ground truth for these device performance tests is based on objective measurements and established standards (e.g., precise temperature readings, material degradation, electrical standards, ISO biocompatibility standards) rather than expert consensus on diagnostic interpretations.

    4. Adjudication method for the test set

    • Not applicable / Information not provided. Adjudication methods are not relevant for these types of engineering and material performance tests.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • No. This is a physical device, not an AI/ML diagnostic. A MRMC study is not relevant here.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • No. This is a physical device, not an AI/ML algorithm.

    7. The type of ground truth used

    • For Temperature Uniformity: Objective physical measurements against a standard for thermal performance (implied).
    • For Shelf-Life: Objective laboratory testing for material degradation and functional integrity over time (accelerated aging).
    • For Electrical Safety/EMC: Compliance with international safety standards (IEC60601-1, IEC 60601-1-2) through objective measurements.
    • For Software Validation: Verification against design specifications and functional requirements.
    • For Biocompatibility: Adherence to ISO 10993 standards (cytotoxicity, irritation, sensitivity) through laboratory testing.

    8. The sample size for the training set

    • Not applicable. There is no "training set" in the context of an AI/ML algorithm for this physical device.

    9. How the ground truth for the training set was established

    • Not applicable. There is no "training set" or corresponding ground truth establishment mentioned for this device.