Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K040036

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2004-03-18

    (69 days)

    Product Code
    Regulation Number
    880.5860
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Inviro Snap Safety Insulin Syringes are intended for the subcutaneous injection of insulins. In addition, the Inviro Snap Safety Syringe is designed to aid in the prevention of needle stick injuries.

    Device Description

    The Inviro Snap Safety Insulin Syringe is a 1 cc retractable type antineedlestick syringe. The syringe has a 100 unit scale in 1 unit increments. The Inviro Snap Safety Insulin Syringe is sterilized by gamma irradiation and supplied sterile in blister pack. One hundred blister packs are packaged in a chipboard box. Each Blister pack and chipboard box is labeled with the contents and the appropriate information per the FDA's Quality System Regulation and Labeling requirements. After use, with the plunger fully compressed, a 180 degree rotation of the plunger couples the Adapter/Needle to the end of the plunger. After this coupling occurs withdrawal of the plunger causes the Adapter and the attached needle to be withdrawn into the safety of the barrel. In this position against the flange, lateral pressure on the plunger results in a controlled fracture of the plunger. Both the syringe and plunger are discarded in a Sharps container.

    AI/ML Overview

    The Inviro Snap Safety Insulin Syringe is a medical device intended for the subcutaneous injection of insulins and designed to prevent needle stick injuries. The provided document is a 510(k) summary, which focuses on demonstrating substantial equivalence to a predicate device rather than conducting extensive clinical studies with human readers or large-scale outcome data. Therefore, many of the requested elements (like MRMC studies, effect sizes of AI assistance, sample sizes for training sets with detailed ground truth establishment, expert qualifications for ground truth) are not applicable or detailed in this type of submission for this device.

    Here's an analysis based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance:

    The document lists performance tests conducted, but it does not explicitly state numerical acceptance criteria for each test. It simply states that the "Inviro Snap Safety Insulin Syringe met all requirements." Without specific thresholds for "force to assemble/disassemble," "pressure leakage," "plunger action force," or "accuracy," a quantitative comparison in a table is not possible. However, the tests performed and the general conclusion are listed below:

    Acceptance Criteria (Implicitly Met)Reported Device Performance
    Biocompatibility Requirements:
    CytotoxicityMet all requirements
    Intracutaneous ReactivityMet all requirements
    Maximization Sensitization StudyMet all requirements
    In Vitro Hemolysis StudyMet all requirements
    USP and ISO Systemic Toxicity StudiesMet all requirements
    USP Prorogen StudyMet all requirements
    Component Interface Tests:
    Force to assemble/disassemble capMet all requirements
    Force to separate Adapter from CannulaMet all requirements
    Force to Snap off PlungerMet all requirements
    Cannula Bending TestMet all requirements
    Force required to press out Adapter from BarrelMet all requirements
    Syringe Assembly Tests:
    Freedom from air and liquid leakage past sealsMet all requirements
    Pressure Leakage TestMet all requirements
    Plunger Action ForceMet all requirements
    AccuracyMet all requirements

    2. Sample Size Used for the Test Set and Data Provenance:

    The document does not explicitly state the specific sample sizes for each performance test performed. It generally refers to "Prior to testing, First Article Inspections were conducted on all components" and "The Inviro Snap Safety Syringe assembly was tested per the requirements of the ISO-FDA Modified Matrix, FDA/ODE General Program Memorandum - # G95-1 and ANSI/AAMI/ISO 10993-1:1997."

    The data provenance is from Inviro Medical, based in Vancouver, BC, Canada. The testing appears to be retrospective in the sense that it was conducted on manufactured devices before the 510(k) submission.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications:

    This information is not provided as the device tests are engineering and biocompatibility tests, not diagnostic or interpretive tasks requiring human expert consensus for a "ground truth" in the traditional sense of medical image analysis or clinical data. The "ground truth" for these tests would be objective measurements against established engineering and biocompatibility standards.

    4. Adjudication Method for the Test Set:

    This is not applicable as the tests are objective engineering and biocompatibility measurements, not subjective human interpretations requiring adjudication.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done:

    No, an MRMC comparative effectiveness study was not done. This type of study is typically performed for diagnostic devices where human readers interpret medical data (e.g., images) and the AI's impact on their performance is evaluated. The Inviro Snap Safety Insulin Syringe is a physical medical device.

    6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Was Done:

    The "standalone" performance in this context refers to the device's own performance against engineering and biocompatibility standards, which was done. The listed tests (biocompatibility, component interface, and syringe assembly tests) are all evaluations of the device itself, independent of human interaction beyond normal use.

    7. The Type of Ground Truth Used:

    The ground truth used for the performance tests was based on established engineering standards, biocompatibility protocols (e.g., ISO, USP), and regulatory guidelines (FDA G95-1). This is a form of objective, quantitative measurement against predefined criteria.

    8. The Sample Size for the Training Set:

    This information is not applicable as this device is a physical medical product, not a software algorithm that undergoes machine learning training. There is no "training set" in the context of AI.

    9. How the Ground Truth for the Training Set Was Established:

    This information is not applicable for the reasons stated in point 8.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1