510(k) Data Aggregation

    K Number
    K130140

    Validate with FDA (Live)

    Date Cleared
    2013-08-01

    (191 days)

    Product Code
    Regulation Number
    872.3640
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    For 3.0mm Intra-Lock® OP Dental Implants:
    The 3.0mm Intra-Lock® OP Dental Implants are indicated for long-term maxillary and mandibular tissue-supported denture stabilization. They are also indicated for the rehabilitation of single or maxillary lateral incisors and mandibular lateral incisors. Multiple implants may be restored after a period of delayed loading or placed in immediate function when good primary stability is achieved with appropriate occlusal loading in order to restore normal teeth function.

    For 3.75mm, 4.0mm & 4.75mm Intra-Lock® OP Dental Implants:
    Intra-Lock® Implants have been designed to restore partially or fully edentulous patients. The implants have been designed to be used in either the maxilla and to support removable or fixed prostheses, from single tooth replacement to full arch reconstruction. Intra-Lock® Implants are intended for immediate function on single tooth and/or multiple tooth applications when good primary stability is achieved, with appropriate occlusal loading, in order to restore normal teeth functions.

    Device Description

    The Intra-Lock® OP Dental Implants are threaded dental implants diameter 3.0mm, 3.75mm, 4.0mm and 4.75mm, with an integrated straight abutment (implant and abutment arc one piece).

    AI/ML Overview

    The provided text is a 510(k) Summary for the Intra-Lock® OP Dental Implants. This type of document is for medical device regulation and describes the device, its intended use, and argues for its substantial equivalence to previously cleared devices. It outlines performance testing performed, but it does not detail specific acceptance criteria or report performance data against those criteria in a quantitative manner that would be typical for an AI/ML device study.

    Therefore, I cannot fulfill all parts of your request based on the provided text, particularly those related to a study proving the device meets acceptance criteria, sample sizes, expert ground truth, MRMC studies, or standalone performance for an AI/ML device. The document describes mechanical and material testing for an implant, not an AI software.

    However, I can extract the information that is present, especially regarding the performance testing performed and the general approach.

    Here's the closest I can get to your request based on the provided text:


    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not provide a table of quantitative acceptance criteria or specific numerical performance metrics. Instead, it lists the types of performance testing that were conducted to support the device's intended use and demonstrates substantial equivalence to predicate devices which have implicitly met regulatory acceptance.

    Acceptance Criteria Category (Implied)Reported Device Performance / Assessment
    Functional / Mechanical IntegrityStatic / Dynamic Fatigue: Testing was reviewed to support the performance. (No specific numerical results or criteria are provided, but the implication is that the implants withstood relevant forces to establish substantial equivalence.)
    Material BiocompatibilityMaterial: TI6AL4V, CP Titanium (These are well-established biocompatible materials for dental implants, aligning with predicate devices.) Differences in material (Titanium Alloy vs. TI6AL4V, CP Titanium) are presented but deemed not to raise new issues of safety or effectiveness.
    Surface PropertiesSurface Analysis: Testing was reviewed (No specific numerical results are provided, but implied to be acceptable and comparable to predicate 'Blasted' or 'Zimmer MTX Surface').
    SterilitySterilization Validation: Validation performed (Method: Gamma Irradiation, consistent with predicate devices, implying sterile product).
    Longevity / StabilityShelf Life: 5 years (Consistent with predicate devices, implying long-term stability of the packaged, sterile product).
    Design EquivalenceDesign Features: Comparisons with predicate devices (Zimmer® One-Piece Implant, MILO® Implant System, Intra-Lock® Implant System with Blossom™) indicate similar materials, intended use, indications, sterilization, and shelf life. "The minor differences between these designs and principals of operation raise no new issues of safety or effectiveness." This implies that the design performs acceptably for its intended use, analogous to the predicate devices. This includes thread patterns, diameter/length sizes, and abutment design.
    Risk MitigationRisk analysis: Performed. (No specific details, but standard regulatory practice to identify and mitigate risks).
    Intended Use FulfillmentThe device's indications for use are described, and the performance testing is stated to "support performance... for their intended use." This is an overarching criterion.

    2. Sample size used for the test set and the data provenance

    The document does not specify a "test set" in the context of an AI/ML study, nor does it provide sample sizes for the performance testing mentioned (risk analysis, shelf life, sterilization validation, surface analysis, static/dynamic fatigue).

    The data provenance is not mentioned. These are likely lab-based engineering and biological tests rather than clinical data from human subjects or retrospective/prospective medical data.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    Not applicable. This is a medical implant, not an AI/ML diagnostic or predictive device that relies on expert interpretation of data to establish ground truth for a test set. The validation is based on engineering and material science principles, and comparison to established predicate devices.

    4. Adjudication method for the test set

    Not applicable for the reasons stated above.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This is not an AI-assisted diagnostic or interpretation device.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. This is a physical medical device (dental implant).

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    For this type of device, "ground truth" is established through:

    • Engineering Standards: Compliance with recognized international and national standards for static and fatigue strength (e.g., ISO standards for dental implants, though not explicitly cited, they are implicit in such submissions).
    • Material Specifications: Verification of material composition and properties against established biocompatible requirements.
    • Sterilization Validation: Confirmation that the sterilization process achieves a predefined sterility assurance level (SAL).
    • Bench Testing: Physical testing results (e.g., fatigue cycles, surface roughness measurements) are compared against safety and performance thresholds derived from predicate devices and regulatory guidance.
    • Biocompatibility Testing: (Though not explicitly listed in "Performance Testing" for this specific summary, it's a standard requirement for implants and would involve testing for toxicity, sensitization, etc., against biological ground truth).

    8. The sample size for the training set

    Not applicable. There is no AI/ML component or "training set" in the context of this device.

    9. How the ground truth for the training set was established

    Not applicable for the reasons stated above.