510(k) Data Aggregation

    K Number
    K040691

    Validate with FDA (Live)

    Date Cleared
    2004-03-23

    (6 days)

    Product Code
    Regulation Number
    872.6640
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Win-100D intra oral camera system and accessories is indicated for use to provide the dentist and the patient with a view of the mouth before and after the dental procedure, which assists the dentist in describing the dental procedure being performed as well as showing the results.

    Device Description

    Our device comprises of a light (0.6Kg), small (230mmX12mm), compact design and ergonomic handpiece along with a multi-docking system. The handpiece consists of a focusing mechanism and a capture button to assist the capture intraoral or full face images of the patient. The handpiece connects to a multi-docking system via a cable. The multi-docking system in turn can connect directly to a monitor and PC via a standard composite connection (RCA). It uses high definition imaging (460 TV-lines) to capture images at 80 degrees field of view. It can make the screen in 4 frames or full screen mode via a multi-docking system. It can change the side of the captured image.

    AI/ML Overview

    The provided text is a 510(k) summary for the Win-100D Intra-Oral Camera System and Accessories. This type of submission is for demonstrating substantial equivalence to a predicate device, not for proving a new device meets specific acceptance criteria through novel studies. Therefore, much of the requested information regarding acceptance criteria and performance studies in a traditional sense is not directly applicable or available in this document.

    The core of a 510(k) is to show that the new device has the same intended use and technological characteristics as a legally marketed predicate device, or, if there are differences, that those differences do not raise new questions of safety and effectiveness.

    Here's an analysis based on the available information:

    Acceptance Criteria and Reported Device Performance

    Instead of specific quantitative acceptance criteria, the "performance" is evaluated relative to the predicate device's characteristics. The acceptance criteria essentially boil down to demonstrating "substantial equivalence" in performance characteristics.

    Acceptance Criteria Category (implied by "Similar")Reported Device Performance (vs. Digital Doc, K981663)
    Indications for use is IdenticalIdentical
    Target population is IdenticalIdentical
    Design is SimilarSimilar (Light (0.6Kg), small (230mmX12mm), compact design and ergonomic handpiece; multi-docking system; focusing mechanism and capture button; composite connection (RCA); high definition imaging (460 TV-lines); 80 degrees field of view; 4 frames or full screen mode; change side of captured image)
    Materials is SimilarSimilar
    Performance is SimilarSimilar (Specific performance metrics are not given, but visual functions like 460 TV-lines and 80 degrees field of view establish similarity to accepted standards for intraoral cameras in 2004)
    Sterility is Not ApplicableNot Applicable
    Biocompatibility is SimilarSimilar
    Mechanical safety is SimilarSimilar
    Chemical safety is Not ApplicableNot Applicable
    Anatomical sites is IdenticalIdentical
    Human factors is SimilarSimilar
    Energy used and/or delivered is SimilarSimilar
    Compatibility with environment and other devices is SimilarSimilar
    Where used is IdenticalIdentical
    Standards met is SimilarSimilar
    Electrical safety is SimilarSimilar
    Thermal safety is SimilarSimilar
    Radiation safety is Not ApplicableNot Applicable

    The study that proves the device meets the acceptance criteria is the comparison presented in the 510(k) submission, which aims to demonstrate substantial equivalence to the predicate device K981663.


    Additional Requested Information:

    1. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective):

      • Not applicable / Not provided. A 510(k) submission for an intra-oral camera of this nature typically does not involve a "test set" in the sense of clinical trial data with patient samples. The performance evaluation is based on technical specifications and comparison to the predicate, rather than a clinical study with a patient sample. No specific clinical data or patient samples are mentioned.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

      • Not applicable / Not provided. As no clinical "test set" or ground truth based on expert review is mentioned, this information is not relevant to this 510(k) summary.
    3. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

      • Not applicable / None. No adjudication method is described because there is no clinical test set requiring adjudication.
    4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

      • No. This device is an intra-oral camera, not an AI-powered diagnostic tool. MRMC studies or AI assistance are not relevant to this submission.
    5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

      • No. This is a hardware device (camera) and does not involve an algorithm with standalone performance evaluation in the context of diagnostic interpretation.
    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

      • Not applicable / Not provided. The "ground truth" for this submission revolves around the technical performance specifications being comparable to the predicate device, not clinical outcomes or expert diagnoses.
    7. The sample size for the training set:

      • Not applicable / Not provided. This is not an AI/ML device requiring a training set.
    8. How the ground truth for the training set was established:

      • Not applicable / Not provided. This is not an AI/ML device requiring a training set with ground truth.

    Summary of the "Study" (510(k) Comparison):

    The "study" in this context is the side-by-side comparison of technological characteristics between the WIN-100D and the predicate device (Digital Doc, K981663). This comparison is presented in a table format and detailed explanations within the main submission (which are not fully provided here). The core of the evidence is showing that the new device's features (design, materials, stated performance, safety aspects) are "identical" or "similar" to a legally marketed device. Performance is largely demonstrated through technical specifications (e.g., 460 TV-lines, 80 degrees field of view) which are deemed acceptable for the intended use of providing visual information to dentists and patients.

    Ask a Question

    Ask a specific question about this device

    K Number
    K032341

    Validate with FDA (Live)

    Date Cleared
    2003-08-25

    (27 days)

    Product Code
    Regulation Number
    872.6640
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Claris i310 intra oral camera system and accessories is indicated for use to provide the dentist and the patient with a view of the mouth before and after the dental procedure, which assists the dentist in describing the dental procedure being performed as well as showing the results.

    Device Description

    Claris i310 comprises of a light (0.1 lbs), small (8"x0.86"), and ergonomic handpiece along with a docking station.

    The handpiece consists of a focusing mechanism and a capture button to assist the doctor in taking intraoral or full face images of the patient.

    The handpiece connects to a docking station via a cable. The docking station in turn can connect directly to a monitor. PC, or a printer via a S-Video or a standard composite connection.

    Claris i310 is also available in a wireless option.

    It uses high definition imaging (494 lines/inch) to capture images at 60 degrees field of view.

    Images may be stored using third party software vendors.

    Claris i310 includes also an optional footswitch.

    AI/ML Overview

    The provided text describes a 510(k) summary for the Claris i310 Intra Oral Camera System and Accessories. This type of submission focuses on demonstrating substantial equivalence to a predicate device, rather than providing detailed acceptance criteria and a rigorous study on the device's performance.

    Therefore, many of the requested details about acceptance criteria, specific studies, sample sizes, ground truth establishment, and MRMC studies are not available in the provided document.

    Here's a breakdown of what can be extracted and what information is missing:

    1. Table of Acceptance Criteria and Reported Device Performance:

    Acceptance CriteriaReported Device Performance
    Specific functional or performance metrics (e.g., image resolution, accuracy in detecting conditions, error rates)Not Available. The document states "Performance: Similar" to the predicate device. It mentions "high definition imaging (494 lines/inch)" but doesn't provide specific performance criteria or tests against them.

    2. Sample size used for the test set and data provenance:

    • Sample Size: Not available. No specific test set or study validating performance is described.
    • Data Provenance: Not available.

    3. Number of experts used to establish the ground truth for the test set and qualifications of those experts:

    • Number of Experts: Not applicable, as no specific performance test set with ground truth is described.
    • Qualifications of Experts: Not applicable.

    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

    • Adjudication Method: Not applicable, as no specific performance test set is described.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • MRMC Study: No, an MRMC study was not done or described. This device is an intra-oral camera, not an AI-powered diagnostic tool, so such a study would not be relevant in this context.

    6. If a standalone (i.e. algorithm only, without human-in-the-loop performance) was done:

    • Standalone Performance: No, a standalone algorithm performance study was not done or described. The device is purely an imaging tool, not a diagnostic algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    • Type of Ground Truth: Not applicable, as no specific performance test with ground truth is described.

    8. The sample size for the training set:

    • Sample Size: Not applicable. This device is a hardware imaging system, not an AI model requiring a training set.

    9. How the ground truth for the training set was established:

    • Ground Truth Establishment: Not applicable.

    Summary of Device and 510(k) Claims:

    The Claris i310 Intra Oral Camera System and Accessories is an imaging device intended to provide dentists and patients with a view of the mouth before and after dental procedures to assist in describing procedures and showing results.

    The 510(k) submission (K032341) claims substantial equivalence to the predicate device "DIGITAL DOC from Digital Doc. Inc." (510k # K981663). The basis for this claim is a comparison of technological characteristics, finding them "Identical" or "Similar" across various categories, including:

    • Indications for use: Identical
    • Target population: Identical
    • Design: Similar
    • Materials: Similar
    • Performance: Similar (stated as "high definition imaging (494 lines/inch)")
    • Biocompatibility: Similar
    • Mechanical safety: Similar
    • Chemical safety: Similar
    • Anatomical sites: Identical
    • Human factors: Similar
    • Energy used and/or delivered: Similar
    • Compatibility with environment and other devices: Similar
    • Where used: Identical
    • Standards met: Similar
    • Electrical safety: Similar
    • Thermal safety: Similar
    • Radiation safety: Similar

    The document does not provide a detailed "study that proves the device meets the acceptance criteria" in terms of specific performance metrics because the 510(k) pathway for this type of device relies on demonstrating equivalence to a legally marketed predicate, rather than novel performance claims requiring extensive new clinical or performance studies. The "study" here is essentially the comparison of technological characteristics to the predicate device, showing that it performs similarly and is thus safe and effective for its stated intended use.