510(k) Data Aggregation

    K Number
    K140263

    Validate with FDA (Live)

    Date Cleared
    2014-03-05

    (30 days)

    Product Code
    Regulation Number
    884.1730
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Insuflow Synergy™ Port (5 mm) and Insylow® Synergy™ XL Port (8 mm, 10mm and 12mm) have applications in thoracic, abdominal and gynecologic minimally invasive endoscopic surgical procedures to establish a path of entry for endoscopic instruments and to heat, humidify, filter and introduce a CO2 gas stream for insufflation of the surgical cavity.

    Device Description

    The Insuflow® Synergy™ and Insuflow® Synergy™ XL Port devices are gas conditioning devices that attach to the outlet port of an insufflator or other regulated CO2 source and are design and humidify the CO2 gas stream prior to insufflation via an integral path of entry device during minimally invasive surgery. The Insuflow® Synergy™ XL Port consist of a sterile, disposable single use device with a filter, heater/humidifier, tubing set, and a path of entry access port device. A reusable control module houses the control and safety circuits for the system.

    AI/ML Overview

    The Insuflow® Synergy™ and Synergy™ XL Port devices and their modifications involve design changes to existing gas conditioning and path of entry devices used in minimally invasive surgery.

    Here's an analysis of the acceptance criteria and study information provided:

    Acceptance Criteria and Reported Device Performance

    The document does not explicitly state formal acceptance criteria with specific performance metrics. Instead, the "acceptance criteria" are implied by demonstrating that the modified devices perform in accordance with specifications and applicable standards, and are "substantially equivalent" to predicate devices. The performance testing focuses on ensuring that the changes do not negatively impact the established functionalities of the original devices.

    Acceptance Criteria (Implied)Reported Device Performance
    Gas Conditioning Functionality (Temperature and Humidity)The document states: "All of the modified devices utilize the same exact design/parts as the predicates for gas conditioning and sealing. As a result, the existing testing for gas temperature and humidity characterization and seal leak integrity testing previously submitted applies directly to these devices." This implies the modified devices meet the same performance as the predicate devices for gas temperature and humidity. No specific quantifiable metrics are provided in this summary, but it's asserted that prior testing for the predicates covers these aspects.
    Seal Leak IntegritySimilar to gas conditioning, the document states: "All of the modified devices utilize the same exact design/parts as the predicates for gas conditioning and sealing. As a result, the existing testing for gas temperature and humidity characterization and seal leak integrity testing previously submitted applies directly to these devices." This indicates the modified devices maintain the same seal leak integrity as the predicate devices. No specific quantifiable metrics are provided here.
    Flow/Pressure PerformanceThe document states: "Since the lumen designs were changing, flow/pressure performance and insertion/removal testing were repeated and successfully completed for the modified Insuflow® Synergy™ (5 mm) and Synergy™ XL (8 mm,10 mm and 12 mm) Port dual lumen models." While no specific pass/fail criteria or measurements are provided, the claim is that they were "successfully completed," implying they met the required performance for the new lumen designs.
    Insertion/Removal PerformanceThe document states: "Since the lumen designs were changing, flow/pressure performance and insertion/removal testing were repeated and successfully completed for the modified Insuflow® Synergy™ (5 mm) and Synergy™ XL (8 mm,10 mm and 12 mm) Port dual lumen models." Similar to flow/pressure, no specific metrics are given, but the successful completion implies the devices perform adequately during insertion and removal, despite the thinner inner lumen wall and smaller overall outer lumen diameter in the dual lumen configuration.
    Overall Safety and EffectivenessThe ultimate conclusion: "Based on a comparison to the predicate devices and information provided, it is the conclusion of LEXION Medical that the modified Insuflow Synergy™ Port (5 mm) and the Insuflow Synergy™ XL Port (8 mm, 10 mm and 12 mm) devices are substantially equivalent to devices already on the market being used for these applications... and present no new concerns about safety and effectiveness." This broad criterion is met by demonstrating equivalence through the aforementioned performance tests and comparison to predicates. No specific metrics are provided for "safety and effectiveness" beyond the functional performance.

    Study Details

    The provided text describes a submission for a Special 510(k), which is typically used for modifications to a legally marketed device that do not significantly affect its safety or effectiveness. As such, the "study" is primarily a performance testing regimen to validate the design changes rather than a large-scale clinical trial or comparative effectiveness study.

    1. Sample size used for the test set and the data provenance:

      • Sample Size: Not explicitly stated as a number of units. The testing was conducted for "the modified Insuflow® Synergy™ (5 mm) and Synergy™ XL (8 mm, 10 mm and 12 mm) Port dual lumen models." This implies that representative samples of each modified size/configuration were tested.
      • Data Provenance: The testing was conducted by the manufacturer, LEXION Medical, LLC, as part of their risk analysis and product development. This would be retrospective in the sense that it's laboratory/bench testing performed after the design changes were implemented, but before market introduction of the modified device. The country of origin for the data is not specified, but the submitter is based in St. Paul, MN, USA.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

      • Not Applicable. This type of device modification and performance testing (bench testing for flow, pressure, and mechanical integrity) does not typically involve human expert "ground truth" establishment in the way an imaging AI algorithm would. The "ground truth" is defined by engineering specifications and physical measurements.
    3. Adjudication method for the test set:

      • None. Adjudication methods like 2+1 or 3+1 are used for establishing ground truth in subjective assessments, such as interpreting medical images. For physical device performance testing, objective measurements against engineering specifications are used, removing the need for expert adjudication in this context.
    4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

      • No. An MRMC study is not applicable as this device is a surgical instrument (gas conditioner and port) and not an AI-powered diagnostic or assistive tool for image interpretation.
    5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

      • Partially Yes (in a different context). The device itself is a standalone medical device that performs its function without a human "in the loop" for its primary operation (gas conditioning, port access). It's not an AI algorithm. The performance tests described (flow/pressure, insertion/removal) are standalone tests of the device's physical and functional properties.
    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

      • The ground truth for the performance testing is based on engineering specifications and measurable physical properties. For example, the "ground truth" for flow/pressure performance would be the required flow rates and pressure tolerances as defined by the device design and relevant industry standards. For insertion/removal, it would relate to acceptable forces, ease of use, and lack of damage.
    7. The sample size for the training set:

      • Not Applicable. This is not an AI/machine learning device, so there is no "training set."
    8. How the ground truth for the training set was established:

      • Not Applicable. As there is no training set, there is no ground truth established for it in this context.
    Ask a Question

    Ask a specific question about this device

    K Number
    K132204

    Validate with FDA (Live)

    Date Cleared
    2013-08-22

    (37 days)

    Product Code
    Regulation Number
    884.1730
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Insuflow® Synergy™ XL Port (10mm and 12mm) and Insuflow® Synergy™ XLR Port (10mm and 12mm with endoscopic cannula seal) have applications in thoracic, abdominal and gynecologic minimally invasive endoscopic surgical procedures to establish a path of entry for endoscopic instruments and to heat, humidify, filter and introduce a CO2 gas stream for insufflation of the surgical cavity.

    Device Description

    The Insuflow® Synergy™ XL and XLR Port are a gas conditioning device that attaches to the outlet port of an insufflator or other regulated CO2 source and is designed to warm and humidify the CO2 gas stream prior to insufflation via an integral path of entry device during minimally invasive surgery. The Insuflow Synergy™ XL and XLR Port consist of a sterile, disposable single use device with a filter, heater/humidifier, tubing set, and a path of entry access port device. A reusable control module houses the control and safety circuits for the system. The integral path of entry access device in the Insuflow Synergy™ XL and XLR Port is designed and constructed similarly to the Insuflow® Synergy™ Port but comes in larger sizes just as the predicate EndoPath trocar device. The Insuflow® Synergy™ XL Port and the Insuflow® Synergy™ XLR Port have access devices in 10 mm and 12 mm configurations with a single-lumen working channel configuration for conditioned gas delivery. The access device has a working channel with duckbill and tool seals for instrument entry into the surgical cavity and delivers conditioned insufflation gas through the working channel. The Insuflow® Synergy™ XLR Port has an endoscope compatible configuration that has an additional seal at the bottom of the main cannula, which serves as a means to seal off around the endoscope, forcing all gas flow out of the peripheral gas exhaust ports. This minor change is intended to reduce fogging and improve visualization. The obturator for the Insuflow Synergy™ XL and XLR Port has a transparent optical window at the distal end, which when used with an endoscope, provides visibility of individual tissue layers during insertion.

    AI/ML Overview

    The provided text describes a 510(k) summary for the Insuflow Synergy™ XL and XLR Port devices. This document focuses on demonstrating substantial equivalence to predicate devices rather than providing detailed performance studies with acceptance criteria, sample sizes, expert involvement, or statistical analysis typically associated with AI/ML device evaluations.

    Therefore, the requested information, particularly regarding specific acceptance criteria with numerical targets, the study proving the device meets them (including sample size, data provenance, expert qualifications, adjudication, MRMC studies, standalone performance, and ground truth establishment for test and training sets), is not present in the provided text.

    The text indicates that "Extensive performance testing has been conducted to assure that the Insuflow® Synergy™ XL and XLR Port perform in accordance with its specifications and applicable standards." and lists the types of tests performed. However, it does not provide the results of these tests in a quantifiable way against pre-defined acceptance criteria, nor does it detail the methodology (sample sizes, expert involvement, ground truth) as would be expected for a diagnostic AI/ML device.

    What is provided regarding "performance testing" are general categories of tests:

    • Flow/pressure performance
    • Gas temperature and humidity characterization
    • Insertion/removal testing
    • Seal leak integrity testing
    • A test to confirm that the cannula seal configuration (for the XLR Port) reduces scope fogging and improves visualization.

    Conclusion:

    Based on the provided text, it's not possible to generate the requested table of acceptance criteria and reported device performance, nor can we answer the subsequent questions about study specifics (sample size, data provenance, expert details, adjudication, MRMC, standalone performance, ground truth establishment) as these details are not included in this 510(k) summary. This document primarily focuses on establishing substantial equivalence to predicate devices through design description and a high-level summary of testing.

    Ask a Question

    Ask a specific question about this device

    K Number
    K120640

    Validate with FDA (Live)

    Date Cleared
    2012-07-18

    (138 days)

    Product Code
    Regulation Number
    884.1730
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Insuflow® Synergy™ Port has applications in thoracic, abdominal and gynecologic minimally invasive endoscopic surgical procedures to establish a path of entry for endoscopic instruments and to heat, humidify, filter and introduce a CO2 gas stream for insufflation of the surgical cavity.

    Device Description

    The Insuflow Synergy™ Port (Dual and Single Lumen) is a gas conditioning device that attaches to the outlet port of an insufflator or other regulated CO2 source and is designed to warm and humidify the CO2 gas stream prior to insufflation via an integral path of entry device during minimally invasive surgery. The Insuflow Synergy™ Port consists of a disposable single use device with a filter, heater/humidifier, tubing set, and a path of entry access port device. A reusable control module houses the control and safety circuits for the svstem. Regulated CO2 gas flows into the Insuflow® Synergy™ Port, through the in-line filter, continues along the tubing to enter the path of entry access device that contains the heating element and humidification media, and through the path of entry access device lumen for delivery into the patient's surgical cavity. The integral path of entry access device is designed and constructed similarly to the predicate trocar device with a sealed instrument access lumen. The Insuflow® Synergy™ Port (Dual and Single Lumen) has a dual-lumen or single-lumen path of entry access device configuration for conditioned gas delivery. Both configurations have a working channel with duckbill and tool seals for instrument entry into the surgical cavity. The single lumen device delivers conditioned insufflation gas through the working channel while the dual lumen configuration has a second outer lumen channel for conditioned gas delivery.

    AI/ML Overview

    The provided text is a 510(k) summary for the Insuflow® Synergy™ Port, a medical device. This type of document is a premarket notification to the FDA to demonstrate that a device is substantially equivalent to a legally marketed predicate device.

    This document focuses on regulatory approval based on demonstrating substantial equivalence to predicate devices, rather than presenting a detailed clinical study demonstrating the device's meeting specific, quantitative acceptance criteria via a performance study with a test set, ground truth and human readers.

    Therefore, I cannot fully complete all sections of your request as the information is not present in the provided text. However, I can extract what is available.


    Description of the Acceptance Criteria and the Study that Proves the Device Meets the Acceptance Criteria:

    The Insuflow® Synergy™ Port is a gas conditioning insufflator accessory device designed to warm, humidify, and filter CO2 gas for insufflation during minimally invasive surgeries, and to establish a path of entry for endoscopic instruments. The regulatory pathway for this device was a 510(k) premarket notification, which demonstrates substantial equivalence to predicate devices rather than a direct clinical performance study with specific acceptance criteria in the format you've requested.

    The "acceptance criteria" in this context are implicitly related to demonstrating that the Insuflow® Synergy™ Port performs safely and effectively in a manner substantially equivalent to its predicate devices. The study proving this involved extensive performance testing and biocompatibility testing.

    1. Table of Acceptance Criteria and Reported Device Performance

    As specific quantitative "acceptance criteria" for a clinical performance study (e.g., sensitivity, specificity thresholds) are not explicitly stated in this 510(k) summary, I can only infer what was tested and that the device "successfully completed" these tests, implying they met internal or standard-based criteria.

    Acceptance Criteria (Inferred from testing)Reported Device Performance (Summary)
    Flow/Pressure PerformanceSuccessfully completed
    Gas Temperature CharacterizationSuccessfully completed
    Humidity CharacterizationSuccessfully completed
    Insertion/Removal TestingSuccessfully completed
    Seal Leak Integrity TestingSuccessfully completed
    Biocompatibility (ISO 10993-1)Successfully conducted

    2. Sample Size Used for the Test Set and Data Provenance

    This information is not provided in the summary. The document mentions "extensive performance testing" but does not detail the sample sizes for these tests or the provenance of any data (e.g., country of origin, retrospective/prospective). It is likely these were bench/laboratory tests rather than tests on patient data.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    This information is not provided. Given the nature of the device (a gas conditioning insufflator and access port) and the type of testing described (flow/pressure, temperature, humidity, insertion/removal, seal leak, biocompatibility), it's highly improbable that "experts to establish ground truth" in the context of diagnostic performance (e.g., radiologists interpreting images) would be relevant or used. The "ground truth" for the performance tests would be instrument measurements and standardized test procedures.

    4. Adjudication Method for the Test Set

    This information is not provided and is not applicable for the type of testing described in this 510(k) summary. Adjudication methods like "2+1" or "3+1" are typically used in studies involving human interpretation of data where consensus is needed for ground truth (e.g., image interpretation).

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, and the effect size of how much human readers improve with AI vs without AI assistance

    No, a MRMC comparative effectiveness study was not performed. This device is a surgical instrument accessory, not an AI-powered diagnostic tool, so such a study would not be applicable or relevant to its regulatory approval.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

    This question is not applicable. The Insuflow® Synergy™ Port is a physical medical device, not an algorithm, so a "standalone" algorithmic performance study without human-in-the-loop would not be relevant.

    7. The Type of Ground Truth Used

    For the performance tests mentioned (Flow/Pressure, Gas Temperature, Humidity, Insertion/Removal, Seal Leak Integrity), the "ground truth" would be established by objective physical measurements using calibrated instruments and adherence to established engineering and medical device standards. For biocompatibility, it would be established by laboratory testing results against ISO 10993-1. Pathology or outcomes data would not directly serve as ground truth for these types of tests.

    8. The Sample Size for the Training Set

    This information is not provided and is not applicable. The device is a physical hardware product, not a software algorithm that requires a "training set."

    9. How the Ground Truth for the Training Set Was Established

    This information is not provided and is not applicable for the reasons stated in point 8.

    Ask a Question

    Ask a specific question about this device

    K Number
    K091366

    Validate with FDA (Live)

    Date Cleared
    2009-09-03

    (118 days)

    Product Code
    Regulation Number
    884.1730
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Insuflow® DewHeart™ is a Blower/Mister Gas conditioner device for use in cardiovascular surgical procedures intended to heat, humidify and filter a CO2 gas stream for introduction into the surgical cavity to improve visibility and reduce the risk of air embolism.

    Device Description

    The Insuflow DewHeart™ device is a single use blower/mister device that attaches to the outlet port of the Insuflow controller with regulated CO2 source and is designed to warm and humidify the CO2 gas stream prior to introduction into the cardiovascular cavity. The Insuflow DewHeart™ device consists of a disposable filter heater/humidifier tubing set and a control module that houses the control and safety circuits for the system. Regulated CO2 gas flows into the Insuflow® DewHeart™ device, through the in-line filter, continues along the tube to enter the Insuflow® DewHeart™ device cassette that contains the heating element and humidification media, through a nozzle/wand for directional. localization of a CO2 gas stream entry into the patient's surgical cavity

    AI/ML Overview

    The provided text is a 510(k) summary for the Insuflow® DewHeart™ device, which is a blower/mister gas conditioner. It states that the device is technically the same as a previously cleared device (Insuflow®, K063546) and substantially equivalent to other predicate devices.

    Here's an analysis of the requested information based only on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not explicitly present a table of acceptance criteria with corresponding performance metrics. It generally states that "Extensive performance testing has been conducted to assure that the Insuflow DewHeart™ performs in accordance with its specifications and applicable standards." However, specific numerical acceptance criteria or performance results are not detailed in this summary.

    Acceptance CriteriaReported Device Performance
    Not specified in the document"performs in accordance with its specifications and applicable standards."

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not specify any sample size for a test set. It refers to "Extensive performance testing" but does not provide details on the experimental setup or the data used. Therefore, the data provenance (country of origin, retrospective/prospective) cannot be determined from this text.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

    The document does not describe any study involving human experts to establish ground truth or evaluate the device's performance. The "performance testing" mentioned appears to refer to engineering or laboratory testing against specifications, not clinical or expert-reviewed data.

    4. Adjudication Method for the Test Set

    As no human expert review or ground truth establishment is described, there is no mention of an adjudication method.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done

    No MRMC study is mentioned. The document primarily focuses on demonstrating substantial equivalence to predicate devices based on technical characteristics and intended use, rather than comparing performance with human readers with and without AI assistance.

    6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) was Done

    The device described is a physical medical device (blower/mister gas conditioner), not an AI algorithm. Therefore, the concept of "standalone (algorithm only)" performance is not applicable in this context. The performance testing refers to the device's physical functions like heating, humidifying, and filtering.

    7. The Type of Ground Truth Used

    The document discusses "performance testing" against "specifications and applicable standards." This implies the ground truth for evaluating the device's functionality would be engineering specifications and established standards for heating, humidifying, and filtering CO2 gas, rather than expert consensus, pathology, or outcomes data.

    8. The Sample Size for the Training Set

    As this is a physical medical device and not an AI/machine learning algorithm, the concept of a "training set" is not relevant.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable, as there is no training set for an AI algorithm.

    Ask a Question

    Ask a specific question about this device

    K Number
    K090456

    Validate with FDA (Live)

    Device Name
    INSUFLOW
    Date Cleared
    2009-08-26

    (184 days)

    Product Code
    Regulation Number
    884.1730
    Age Range
    All
    Reference & Predicate Devices
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Insuflow is a gas conditioner accessory device for use in minimally invasive surgery, including thoracoscopic, laparoscopic and endoscopic vein harvesting surgical procedures, intended to heat, humidify and filter a CO2 gas stream for insufflation of the surgical cavity.

    Device Description

    The Insuflow device is a single use device that attaches to the outlet port of an insufflator or other regulated CO2 source and is designed to warm and humidify the CO2 gas stream prior to insufflation into the cardiothoracic, vascular or abdominal surgical cavity. The Insuflow device consists of a disposable filter heater/humidifier tubing set and a control module that houses the control and safety circuits for the system. Regulated CO2 gas flows into the Insuflow® device, through the in-line filter, continues along the tube to enter the Insuflow® device cassette that contains the heating element and humidification media, through a tube that connects via a Luer lock connector to anno a insufflation needle/trocar and finally flows into the patient's surgical cayity over and finally

    AI/ML Overview

    The provided text is a 510(k) summary for the Insuflow device. It describes the device, its intended use, and argues for its substantial equivalence to previously cleared devices. However, this document does not contain any information about specific acceptance criteria, device performance metrics, or the details of a study (test set, training set, ground truth, expert involvement, or MRMC studies) that proves the device meets acceptance criteria.

    Instead, it states that:

    • "Extensive performance testing has been conducted to assure that the Insuflow performs in accordance with its specifications and applicable standards."
    • "Details of that testing were provided in 510(k) K063546 and are summarized in Section 5."

    To answer your request, the full 510(k) K063546 document, or specifically its Section 5, would be needed. The current document only makes a general declaration of testing.

    Therefore, I cannot populate the table or provide detailed answers to points 2 through 9 based solely on the provided text.

    Summary of what can be extracted from the provided text:

    1. A table of acceptance criteria and the reported device performance:

      • Cannot be provided. The document states "Extensive performance testing has been conducted to assure that the Insuflow performs in accordance with its specifications and applicable standards," but it does not specify what those specifications or standards are, nor does it report any performance metrics against them. It refers to a previous 510(k) (K063546) for these details.
    2. Sample size used for the test set and the data provenance:

      • Cannot be provided. No information on a "test set" or specific study details is available.
    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

      • Cannot be provided. No information on ground truth establishment or expert involvement in a test set is available.
    4. Adjudication method for the test set:

      • Cannot be provided. No information on a test set or adjudication method is available.
    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

      • Cannot be provided. This device is a medical device (gas conditioner), not an AI/software device. Therefore, an MRMC study with AI assistance is not applicable.
    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

      • Cannot be provided. This is not an algorithm/AI device.
    7. The type of ground truth used:

      • Cannot be provided. No information on ground truth for any testing is available.
    8. The sample size for the training set:

      • Cannot be provided. No information on a "training set" is available. This is not an AI/software device.
    9. How the ground truth for the training set was established:

      • Cannot be provided. No information on a "training set" or its ground truth establishment is available.

    In conclusion, the provided document is a regulatory summary focusing on substantial equivalence, and it explicitly defers details of performance testing to a prior submission (K063546). Without that prior submission, the requested information cannot be extracted.

    Ask a Question

    Ask a specific question about this device

    K Number
    K090879

    Validate with FDA (Live)

    Date Cleared
    2009-07-23

    (114 days)

    Product Code
    Regulation Number
    876.1500
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Insuflow® CO2lonShield™ is a gas conditioner accessory device for use during colonoscopy procedures, including CT Colonography (CTC or Virtual Colonoscopy) and conventional Colonoscopy procedures, and intraoperative endoscopy procedures, intended to heat, humidify and filter a CO2 gas stream for administration as a distention media.

    Device Description

    The Insuflow® CO2lonShield™ device is a single use device that attaches to the outlet port of a regulated CO2 source and is designed to warm and humidify the CO2 gas stream prior to gas delivery during colonoscopy and intraoperative endoscopy procedures. The Insuflow® CO2lonShield™ device consists of a disposable filter heater/humidifier tubing set and a control module that houses the control and safety circuits for the system. Regulated CO2 gas flows into the Insuflow® CO2lonShield™ device, through the inline filter, continues along the tube to enter the Insuflow® CO2lonShield™ device cassette that contains the heating element and humidification media, through a tube that connects to an endoscope/colonoscope or rectal probe for gas delivery into the patient's gastroenterological cavity.

    AI/ML Overview

    The provided text is for a medical device called Insuflow® CO2lonShield™, a gas conditioner accessory device. The 510(k) summary states that "Extensive performance testing has been conducted to assure that the Insuflow CO2lonShield™ performs in accordance with its specifications and applicable standards. Details of that testing were provided in 510(k) K063546 and are summarized in Section 5."

    However, the current document (K090879) does not contain the detailed acceptance criteria or the specific study results to prove the device meets these criteria. It explicitly refers to the previous 510(k) K063546 for those details. Therefore, I cannot extract the table of acceptance criteria and reported device performance from this document alone.

    Based on the information available in K090879:

    • 1. A table of acceptance criteria and the reported device performance: This information is not present in K090879. It refers to K063546.
    • 2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective): Not present in K090879. It refers to K063546.
    • 3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience): Not applicable, as this is a gas conditioning device, not an interpretative imaging device requiring expert ground truth for classification. The performance testing would likely involve technical specifications of gas temperature, humidity, and filtration.
    • 4. Adjudication method (e.g. 2+1, 3+1, none) for the test set: Not applicable for this type of device.
    • 5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable. This is a medical device for warming and humidifying CO2, not an AI-powered diagnostic tool.
    • 6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable. This is a physical device, not an algorithm. Performance would be measured against its technical specifications.
    • 7. The type of ground truth used (expert consensus, pathology, outcomes data, etc): Not applicable in the traditional sense for this device. The "ground truth" would be the target temperature, humidity, and filtration efficiency specifications that the device is designed to meet.
    • 8. The sample size for the training set: Not applicable for this type of physical device. "Training set" typically refers to machine learning algorithms.
    • 9. How the ground truth for the training set was established: Not applicable for this type of physical device.

    In summary, the provided document K090879 serves as a re-submission or update that references a previous 510(k) (K063546) for the detailed performance testing information. Without access to K063546, the specific performance data and acceptance criteria cannot be extracted.

    Ask a Question

    Ask a specific question about this device

    K Number
    K063546

    Validate with FDA (Live)

    Date Cleared
    2007-01-10

    (47 days)

    Product Code
    Regulation Number
    884.1730
    Age Range
    All
    Reference & Predicate Devices
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Lexion Medical Insuflow® device is an accessory to an insufflator intended to heat, humidify and filter a gas stream used for inflation during laparoscopic surgery.

    Device Description

    The Insuflow® device is a single use device that attaches to the outlet port of an insufflator and is designed to warm and humidify the gas stream prior to insufflation into the abdominal cavity. The Insuflow® device consists of a disposable filter heater/humidifier tubing set and a control module that houses the control and safety circuits for the system.

    Gas from the insufflator flows into the Insuflow® device, through the in-line filter, continues along the tube to enter the Insuflow® device cassette that contains the heating element and humidification media, through a tube that connects via a Luer lock connector to a trocar or insufflation needle, and finally flows into the patient's abdomen.

    AI/ML Overview

    The provided text is a 510(k) summary for the Lexion Medical Insuflow® Device. This document is a premarket notification for a medical device and describes its equivalence to a previously marketed device (predicate device), rather than presenting a study that establishes acceptance criteria and then proves the device meets those criteria from an AI/ML perspective.

    Therefore, the requested information, which pertains specifically to performance metrics, sample sizes, expert ground truth establishment, adjudication methods, MRMC studies, standalone performance, and training data for AI/ML models, cannot be extracted from this document, as it is not relevant to the type of device and the submission presented.

    The device described is a physical accessory to an insufflator that heats, humidifies, and filters gas for laparoscopic surgery. The studies mentioned are primarily focused on system performance (CO2 gas temperature and humidity characteristics), electrical safety, EMC testing, and biocompatibility. These are engineering and biological performance characteristics, not AI/ML model performance metrics.

    Specifically, for each point requested:

    1. A table of acceptance criteria and the reported device performance: Not applicable. The document mentions "The Insuflow® device met all specified design and performance requirements" for electrical safety and EMC, and "All specified biocompatibility requirements were met." However, it does not provide a table of these specific criteria or quantitative performance results in a way that would be comparable to AI/ML model output (e.g., sensitivity, specificity, AUC).
    2. Sample sizes used for the test set and the data provenance: Not applicable. The studies mentioned (system performance, electrical safety, biocompatibility) are laboratory or bench tests, not clinical studies with "test sets" in the AI/ML sense. Data provenance is not specified beyond being "verification testing."
    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable. Ground truth for a physical device's engineering and biocompatibility performance is typically established by engineering standards, chemical analysis, and certified test protocols, not expert consensus in the diagnostic sense.
    4. Adjudication method for the test set: Not applicable. There is no "test set" and thus no adjudication method as understood in AI/ML performance evaluation.
    5. If a multi-reader, multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable. This device does not involve human "readers" or AI assistance.
    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done: Not applicable. This is not an algorithm.
    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.): The "ground truth" for the tests performed would be established by engineering specifications, validated measurement equipment, and standardized biological assessments.
    8. The sample size for the training set: Not applicable. There is no AI/ML model to train.
    9. How the ground truth for the training set was established: Not applicable.

    In summary, the provided K063546 document for the Insuflow® device does not describe an AI/ML device and therefore the requested information fields related to AI/ML model performance and validation are not present.