510(k) Data Aggregation

    K Number
    K201916

    Validate with FDA (Live)

    Date Cleared
    2020-08-13

    (34 days)

    Product Code
    Regulation Number
    870.4350
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The INSPIRE 6M is intended for use in adult surgical procedures requiring cardiopulmonary bypass (CPB). It provides gas exchange support and blood temperature control. The INSPIRE 6M is intended to be used for 6 hours or less.

    The INSPIRE 7M is intended for use in adult surgical procedures requiring cardiopulmonary bypass (CPB). It provides gas exchange support and blood temperature control. The INSPIRE 7M is intended to be used for 6 hours or less.

    The INSPIRE 8M is intended for use in adult surgical procedures requiring cardiopulmonary bypass (CPB). It provides gas exchange support and blood temperature control. The INSPIRE 8M is intended to be used for 6 hours or less.

    Device Description

    The INSPIRE 6M, INSPIRE 7M and INSPIRE 8M oxygenators (hereinafter identified as INSPIRE) consist of an oxygenator with an integrated heat exchanger.

    The INSPIRE devices are intended for use in adult and small adult surgical procedures requiring cardiopulmonary bypass (CPB). It provides gas exchange support and blood temperature control.

    The INSPIRE consist of the following main components

    • a heat exchanger consisting of a bundle of polyurethane hollow fibers that are wound on a cylindrical core. it controls blood temperature and allows the use of hypothermia or aids in the maintenance of normothermia during surgery.
    • an oxygenating module element made of a coiled bundle of polypropylene microporous hollow fibers rolled on the heat exchanger sub assembly. The hollow fiber membrane provides oxygenation and carbon dioxide removal from venous blood or suction blood.

    The modified devices are modified versions of the currently marketed INSPIRE 6M, INSPIRE 7M and INSPIRE 8M products.

    AI/ML Overview

    The provided documents describe the Inspire 6M, 7M, and 8M Hollow Fiber Oxygenators, which are modifications of previously cleared devices. The submission focuses on demonstrating substantial equivalence to the predicate devices, rather than proving that the device meets acceptance criteria as would typically be done for a novel device. The primary change is the replacement of a purge line unidirectional valve.

    Therefore, the study focuses on showing that the modified device performs equivalently to the unmodified predicate device, rather than establishing absolute performance metrics against a defined set of acceptance criteria for a new device.

    Here's the information extracted from the document, tailored to your request, with an emphasis on the equivalence study:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not explicitly present a table of acceptance criteria for this specific submission. Instead, it relies on demonstrating that the modified device performs equivalently to the predicate devices, which are already presumed to meet established safety and effectiveness criteria. The acceptance is based on the modified device successfully meeting performance expectations in comparison to the unmodified (predicate) device.

    Performance Metric DemonstratedAcceptance Criteria (Implied by equivalence to predicate)Reported Device Performance (Modified device vs. Unmodified predicate)
    Recirculation Line Handling (specifically related to the purge line unidirectional valve)System successfully handles recirculation line operation, including gas exchange support and blood temperature control, as per predicate performance.The oxygenating module with the modified purge line unidirectional valve successfully met all acceptance criteria for the recirculation line handling. (This implies it performed equivalently to the unmodified device).
    All other relevant functional parameters (e.g., gas exchange, blood temperature control)Equivalent to predicate device.The results of in vitro studies demonstrate that the subject INSPIRE performs in a manner substantially equivalent to the Unmodified INSPIRE predicate device with respect to the relevant functional parameters.

    2. Sample Size Used for the Test Set and the Data Provenance

    • Sample Size: The document implies that "devices" were tested, but it does not specify the exact number of modified and unmodified devices used in the in vitro testing.
    • Data Provenance: The study was retrospective in the sense that it compared a modified device to an existing predicate. The testing itself was in-vitro. The country of origin of the data is not specified, but the applicant is Sorin Group Italia S.r.l. in Italy, suggesting the testing likely occurred there or at a contract lab.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts

    Not applicable. This was an in-vitro performance study comparing a modified medical device component to its predecessor, not a diagnostic or AI-driven study requiring expert interpretation or ground truth establishment in the traditional sense.

    4. Adjudication Method for the Test Set

    Not applicable. This was an in-vitro performance study. "Adjudication" typically refers to resolving disagreements among experts in establishing ground truth for diagnostic studies, which is not relevant here.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This is not an AI/diagnostic device.

    6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. This is not an AI/diagnostic device.

    7. The Type of Ground Truth Used

    The "ground truth" for this study was the established performance of the predicate (unmodified) devices as assessed through their own previous regulatory clearances and real-world usage. The study aimed to demonstrate that the modified device's performance was equivalent to this established "ground truth." The specific tests (e.g., for recirculation line handling, gas exchange, temperature control) are based on performance standards for cardiopulmonary bypass oxygenators (ISO 7199, FDA guidances).

    8. The Sample Size for the Training Set

    Not applicable. This is not an AI device that requires a training set.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable.

    Ask a Question

    Ask a specific question about this device

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    INSPIRE 6F M is intended for use in adult surgical procedures requiring cardiopulmonary bypass. It provides gas exchange support and blood temperature control. INSPIRE 6F M integrated arterial filter provides additional protection against air and solid emboli. INSPIRE 6F M is intended to be used for 6 hours or less.

    INSPIRE 6M is intended for use in adult and small adult surgical procedures requiring cardiopulmonary bypass. It provides gas exchange support and blood temperature control. INSPIRE 6M is intended to be used for 6 hours or less.

    INSPIRE 8F M is intended for use in adult surgical procedures requiring cardiopulmonary bypass. It provides gas exchange support and blood temperature control. INSPIRE 8F M integrated arterial filter provides additional protection against air and solid emboli. INSPIRE 8F M is intended to be used for 6 hours or less.

    The INSPIRE 8M is intended for use in adult and small adult surgical procedures requiring cardiopulmonary bypass. It provides gas exchange support and blood temperature control. INSPIRE 8M is intended to be used for 6 hours or less.

    Device Description

    The device consists of an oxygenator with an integrated heat exchanger. The INSPIRE model 6F M also has an integrated arterial filter. The INSPIRE 6F M device can be operated at flow rates up to 6 liters per minute (I/min). The hollow fiber membrane oxygenator provides oxygenation and carbon dioxide removal from venous blood or suction blood. The integrated heat exchanger controls blood temperature and allows the use of hypothermia or aids in the maintenance of normothermia during surgery. The integrated arterial filter provides additional protection against air and solid emboli. The device is a modified version of the currently marketed INSPIRE 6F M products.

    The device consists of an oxygenator with an integrated heat exchanger. The INSPIRE 6M device can be operated at flow rates up to 6 liters per minute (I/min). The hollow fiber membrane oxygenator provides oxygenation and carbon dioxide removal from venous blood or suction blood. The integrated heat exchanger controls blood temperature and allows the use of hypothermia or aids in the maintenance of normothermia during surgery. The device is a modified version of the currently marketed INSPIRE 6M products.

    The device consists of an oxygenator with an integrated heat exchanger. The INSPIRE model 8F M also has an integrated arterial filter. The INSPIRE 8F M device can be operated at flow rates up to 8 liters per minute (I/min). The hollow fiber membrane oxygenator provides oxygenation and carbon dioxide removal from venous blood or suction blood. The integrated heat exchanger controls blood temperature and allows the use of hypothermia or aids in the maintenance of normothermia during surgery. The integrated arterial filter provides additional protection against air and solid emboli. The device is a modified version of the currently marketed INSPIRE 8F M products.

    The device consists of an oxygenator with an integrated heat exchanger. The INSPIRE 8M device can be operated at flow rates up to 8 liters per minute (I/min). The hollow fiber membrane oxygenator provides oxygenation and carbon dioxide removal from venous blood or suction blood. The integrated heat exchanger controls blood temperature and allows the use of hypothermia or aids in the maintenance of normothermia during surgery. The device is a modified version of the currently marketed INSPIRE 8M products.

    AI/ML Overview

    The provided document is a 510(k) premarket notification for several models of the INSPIRE Hollow Fiber Oxygenator. This type of submission is for demonstrating substantial equivalence to a predicate device, not necessarily for proving novel clinical effectiveness or "AI" performance.

    Based on the information provided in the document:

    1. A table of acceptance criteria and the reported device performance

    Acceptance CriteriaReported Device Performance
    Moisture handlingSuccessfully met all acceptance criteria for moisture handling.

    2. Sample size used for the test set and the data provenance

    The document mentions "Testing was performed to compare devices with the modified and predicate gas outlet port." It states this was "in vitro test results" and conducted on "sterile aged devices; accelerated aging for a period of time equivalent to at least 3 years as per device labeling."

    • Sample size for the test set: Not explicitly stated. The document refers to "devices" in the plural, but the exact number is not provided.
    • Data provenance: In vitro testing, which means it's laboratory-based data, not human patient data. It is prospective testing designed to evaluate the modified device. The country of origin for the data is not specified, but the submitter is Sorin Group Italia S.r.l. in Mirandola (MO) ITALY.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    Not applicable. This was in vitro performance testing of a medical device, not a human reader study requiring expert opinion for ground truth.

    4. Adjudication method for the test set

    Not applicable. This was in vitro performance testing of a medical device against predefined engineering/performance criteria, not a study requiring adjudication of expert interpretations.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done

    No. This document describes in vitro testing of a device's physical and functional properties, not a reader study. There is no mention of human readers or AI assistance.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done

    No. This is not an AI/algorithm-based device. It's a cardiopulmonary bypass oxygenator. The testing described is for the device itself.

    7. The type of ground truth used

    The "ground truth" for this in vitro performance testing would be the predefined engineering specifications and performance standards for moisture handling for oxygenator devices. The document states "successfully met all acceptance criteria for moisture handling," implying these criteria were established prior to the testing.

    8. The sample size for the training set

    Not applicable. This is not an AI/machine learning device that requires a training set.

    9. How the ground truth for the training set was established

    Not applicable. This is not an AI/machine learning device.