Search Filters

Search Results

Found 6 results

510(k) Data Aggregation

    K Number
    K121021

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2012-09-24

    (173 days)

    Product Code
    Regulation Number
    868.5165
    Age Range
    All
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The INOmax® DS delivery system delivers INOmax® (nitric oxide for inhalation) therapy gas into the inspiratory limb of the patient breathing circuit in a way that provides a constant concentration of nitric oxide (NO), as set by the user, to the patient throughout the inspired breath. It uses a specially designed injector module, which enables tracking of the ventilator waveforms and the delivery of a synchronized and proportional dose of NO. It may be used with most ventilators.

    The INOmax® DS provides continuous integrated monitoring of inspired O2, NO2, and NO, and a comprehensive alarm system.

    The INOmax® DS incorporates a battery that provides up to 6 hours of uninterrupted NO delivery in the absence of an external power source.

    The INOmax® DS includes a backup NO delivery capability that provides a fixed flow of 250 mL/min of NO which along with user supplied 10 L/min of oxygen provides 20 ppm in the gas flow to a patients breathing circuit. It may also use the INOblender® for backup.

    The target patient population is controlled by the drug labeling for INOmax® and is currently neonates. The primary targeted clinical setting is the Neonatal Intensive Care Unit (NICU) and secondary targeted clinical setting is the transport of neonates.

    Device Description

    The INOmax DSIR® uses a "dual-channel" design to ensure the safe delivery of INOmax®. The first channel has the delivery CPU, the flow controller and the injector module to ensure the accurate delivery of NO. The second channel is the monitoring system, which includes a separate monitor CPU, the gas cells (NO, NO2, and O2 cells) and the user interface including the display and alarms. The dual-channel approach to delivery and monitoring permits INOmax® delivery independent of monitoring but also allows the monitoring system to shutdown INOmax delivery if it detects a fault in the delivery system such that the NO concentration could become greater than 100 ppm. The delivery system can also shut down delivery if it detects certain serious problems with the monitoring system.

    AI/ML Overview

    1. Table of Acceptance Criteria and Reported Device Performance:

    The document describes software and labeling modifications to the INOmax DSIR, focusing on alarm thresholds and compatibility with new respiratory care devices. The acceptance criteria are implicitly met by successful verification testing. The performance is described by how the modified alarms function and successful compatibility testing.

    Feature/SpecificationAcceptance Criteria (Implied)Reported Device Performance
    Alarms/Alerts
    Low Battery Alarm ThresholdLow Battery alarm to activate at 60 minutes remaining battery life (internally)The Low Battery alarm threshold limit was changed from 30 minutes to 60 minutes in the software to provide additional margin to the system electronics. From the user perspective the alarm limit is still 30 minutes. Shutdown of the device when operating on battery is not dependent on the Low Battery alarm, therefore the length of runtime when operating on battery remains unchanged.
    Delivery Failure (NO > 100 ppm)Alarm and delivery shutdown when monitored NO > 100 ppm for 12 continuous seconds.The NO > 100 ppm condition must be present for 12 continuous seconds at monitored NO > 100 ppm before alarm and delivery shutdown. The 12 second timer will be reset when a sampling blackout occurs. The change in time to 12 seconds, from zero seconds was made to prevent loss of delivery conditions during transient monitored NO values briefly exceeding 100 ppm.
    Delivery Failure (NO > 2x setpoint)Alarm and delivery shutdown when monitored NO > 2x setpoint for 12 continuous seconds.To prevent loss of delivery during transient conditions, the overdelivery condition of NO > 2x setpoint must be present for 12 continuous seconds (versus 0 seconds in the prior software version) at monitored NO > 2x setpoint before alarm and delivery shutdown.
    Delivery Failure (Under Delivery)Under Delivery Alarm to trigger when under-delivery condition is present for 12 consecutive seconds.Under Delivery Alarm added. To prevent loss of delivery during transient conditions, the under delivery condition must be present for 12 consecutive seconds to trigger the under delivery alarm. In the previous version, the system would immediately shut-down as soon as the under delivery condition was detected.
    Delivery Failure (System Voltage)Elimination of three system voltage conditions that previously triggered this alarm.Three system voltage conditions during which this alarm was triggered have been eliminated.
    Compatibility with Respiratory DevicesMaintain essential performance (O2 dilution, effect on device, delivery accuracy, NO2 generation)Testing concluded four requirements necessary for operation of INOmax DSIR and the three respiratory care devices to be compatible: O2 dilution, Effect on respiratory care device, INOmax DSIR® delivery accuracy, NO2 generation.
    Backup Delivery Function (Fisher & Paykal)Proper function during backup delivery with Fisher & Paykal circuits.Testing with the Fisher & Paykal circuits additionally concluded proper function during backup delivery.
    INOmax DSIR® delivery accuracyAccurate delivery of NO at various settings (0, 5, 20, 40, 80 ppm).Performance testing (Verification) was conducted, using settings 0, 5, 20, 40, and 80 ppm for each setting and mode of ventilation. The measured values on the INOmax DSIR were recorded, and no anomalies were reported for delivery accuracy in the summary, implying satisfactory performance.
    Software ComplianceCompliant with system-level requirements, new/modified alarms function as specified, data communication.Software verification confirmed the INOmax DSIR® is compliant with its system level requirements, that the new/modified alarms function as specified and that data can be correctly communicated to third part data collection devices via the existing RS-232 port.
    Control System PerformanceStability/sensitivity analysis demonstrates response handling disturbances and component failures.Analysis incorporated stability/sensitivity analysis demonstrating the response of the control system using worst-case flow inputs and the handling of disturbances and component failures. Graphical presentation showing time-domain responses of overshoot and undershoot across various ventilator flow rates and flow profiles was also provided.

    2. Sample Size Used for the Test Set and Data Provenance:

    • Test Set Sample Size: The document does not specify a numerical sample size for the "test set" in terms of cases or patients. Instead, it refers to testing conducted with:
      • Five INOmax DSIR® settings: 0 (baseline), 5, 20, 40, and 80 ppm, "for each setting and mode of ventilation." This implies a systematic study across different operational parameters.
      • Three new respiratory care devices: Fisher & Paykal Infant Circuit Nasal Cannula (K020332), Fisher and Paykal Optiflow Breathing Circuit (K983112), and A-Plus Medical Babi Plus Bubble CPAP (K110471).
    • Data Provenance: The testing described is non-clinical and conducted by the manufacturer, INO Therapeutics/Ikaria. The data is thus prospective internal testing data. The country of origin is not explicitly stated for the testing, but the company is based in Madison, Wisconsin, USA.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of those Experts:

    This information is not provided in the document. The testing described is non-clinical performance and compatibility testing of a medical device, not a diagnostic algorithm that relies on expert interpretation for ground truth establishment.

    4. Adjudication Method for the Test Set:

    This information is not provided and is not applicable to the type of non-clinical device testing described. Adjudication methods like 2+1 or 3+1 are typically used in clinical studies where multiple experts evaluate cases and a consensus or tie-breaking mechanism is needed for ground truth.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    No, an MRMC comparative effectiveness study was not done. This document describes the premarket notification for a medical device (nitric oxide delivery system), not an AI-based diagnostic tool or system designed to assist human readers.

    6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    The device itself is a standalone system for delivering nitric oxide and monitoring gases. The document describes "software verification" and "performance testing" of the device, which inherently represents its standalone (without human intervention once settings are applied) functionality for its intended purpose. The modifications discussed pertain to the internal logic and alarm triggers of this standalone system.

    7. The Type of Ground Truth Used:

    The "ground truth" for this device's performance is established by:

    • Technical Specifications/Requirements: The device is tested against its established system-level requirements and specifications, as well as recognized consensus standards (e.g., IEC 60601 series).
    • Physical Measurements: For performance aspects like NO delivery accuracy, measured values from the INOmax DSIR® are recorded and compared against known input settings.
    • Functional Verification: For alarms and compatibility, the "ground truth" is whether the alarm triggers as expected under specified conditions or if the integration with other devices functions without issues and meets predefined operational criteria (e.g., no detrimental O2 dilution, no adverse effect on respiratory care device).

    8. The Sample Size for the Training Set:

    The concept of a "training set" is not applicable to this document. The INOmax DSIR® is a hardware device with embedded software; it does not utilize machine learning or AI that requires a training set in the conventional sense. The "training" for the software's development would implicitly come from the extensive design, development, and iterative testing processes, but not from a distinct, labeled "training set" of data.

    9. How the Ground Truth for the Training Set Was Established:

    As there is no "training set" in the context of machine learning, the question of how its ground truth was established is not applicable. The "ground truth" for the device's design and functionality is derived from engineering specifications, medical device standards, risk analysis, and clinical requirements for patient safety and efficacy.

    Ask a Question

    Ask a specific question about this device

    K Number
    K092545

    Validate with FDA (Live)

    Date Cleared
    2010-03-10

    (203 days)

    Product Code
    Regulation Number
    868.5165
    Age Range
    All
    Reference & Predicate Devices
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The INOmax DS delivery system delivers INOmax® (nitric oxide of inhalation) therapy gas into the inspiratory limb of the patient breathing circuit in a way that provides a constant concentration of nitric oxide (NO), as set by the user, to the patient throughout the inspired breath. It uses a specially designed injector module, which enables tracking of the ventilator waveforms and the delivery of a synchronized and proportional dose of NO. It may be used with most ventilators.

    The INOmax DS provides continuous integrated monitoring of inspired O2, NO2, and NO, and a comprehensive alarm system.

    The INOmax DS incorporates a battery that provides up to 6 hours of uninterrupted NO delivery in the absence of an external power source.

    The INOmax DS includes a backup NO delivery capability that provides a fixed flow of 250 mL/min of NO which along with user supplied 10 L/min of oxygen provides 20 ppm in the gas flow to a patients breathing circuit. It may also use the INOblender for backup.

    The target patient population is controlled by the drug labeling for INOmax and is currently neonates. The primary targeted clinical setting is the Neonatal Intensive Care Unit (NICU) and secondary targeted clinical setting is the transport of neonates.

    Device Description

    The INO Therapeutics INOmax DS represents the continuing evolution of the Datex-Ohmeda (now GE Healthcare) INOvent Delivery System design. Its published performance specifications are not significantly different from the INOvent. The INOmax DS design incorporates a reconfiguration of hardware, electronic circuitry and software resulting in a more compact, stylish, modern, user-friendly and easier manufactured product.

    The INOmax DS delivery system delivers INOmax® (nitric oxide of inhalation) therapy gas into the inspiratory limb of the patient breathing circuit in a way that provides a constant concentration of nitric oxide (NO), as set by the user, to the patient throughout the inspired breath. It uses a specially designed iniector module, which enables tracking of the ventilator waveforms and the delivery of a synchronized and proportional dose of NO. It may be used with most ventilators.

    The INOmax DS provides continuous integrated monitoring of inspired O2, NO2, and NO, and a comprehensive alarm system. The INOmax DS incorporates a battery that provides up to 6 hours of uninterrupted NO delivery in the absence of an external power source.

    The INOmax DS includes a backup NO delivery capability that provides a fixed flow of 250 mL/min of NO which along with user supplied 10 L/min of oxygen provides 20 ppm in the gas flow to a patients breathing circuit. It may also use the INOblender for backup.

    AI/ML Overview

    The provided text describes the INOmax DS (Delivery System), a nitric oxide administration apparatus. However, it explicitly states that clinical studies were not required to support substantial equivalence for this device. Therefore, information regarding acceptance criteria derived from clinical studies, comparative effectiveness studies, or clinical ground truth establishment is not available within the provided document.

    The document primarily focuses on non-clinical tests and the device's technological equivalence to a predicate device.

    Here's an analysis based only on the provided text, addressing the requested points:


    Acceptance Criteria and Device Performance

    Since clinical studies were not required for this 510(k) submission, there are no specific performance acceptance criteria or reported device performance metrics from clinical trials mentioned. The manufacturer's conclusion is that the INOmax DS is "as safe, as effective, and performance is substantially equivalent to the predicate device."

    Acceptance Criteria CategoryReported Device Performance
    Clinical PerformanceNot applicable; clinical studies were not required to support substantial equivalence. Device is considered "as safe, as effective, and performance is substantially equivalent to the predicate device."
    Non-Clinical PerformanceThe device complies with voluntary standards and underwent internal quality assurance measures: Risk Analysis, Requirements Reviews, Design Reviews, Unit-level Testing (Module verification), Integration Testing (System verification), Final Acceptance Testing (Validation), Performance Testing (Verification), Safety Testing (Verification).

    Other Information

    Due to the lack of clinical studies, much of the requested information is not present in the provided document.

    1. Sample sizes used for the test set and the data provenance: Not applicable, as no clinical test set for performance evaluation is mentioned. Non-clinical testing details (e.g., number of units tested, specific data provenance) are not provided.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable, as no clinical test set requiring expert ground truth is mentioned.
    3. Adjudication method for the test set: Not applicable, as no clinical test set is mentioned.
    4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable. This device is a nitric oxide delivery system, not an AI-assisted diagnostic tool. No MRMC study was conducted or mentioned.
    5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable. This is a medical device for nitric oxide delivery, not an algorithm.
    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.): Not applicable, as no clinical ground truth establishment is mentioned. The primary basis for equivalence is technological similarity and compliance with voluntary standards and internal quality assurance measures, rather than a clinical ground truth.
    7. The sample size for the training set: Not applicable, as this is a medical device, not an AI algorithm requiring a training set.
    8. How the ground truth for the training set was established: Not applicable.
    Ask a Question

    Ask a specific question about this device

    K Number
    K090958

    Validate with FDA (Live)

    Date Cleared
    2009-06-18

    (73 days)

    Product Code
    Regulation Number
    868.2375
    Age Range
    All
    Reference & Predicate Devices
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The INOmax DS delivery system delivers INOmax® (nitric oxide of inhalation) therapy gas into the inspiratory limb of the patient breathing circuit in a way that provides a constant concentration of nitric oxide (NO), as set by the user, to the patient throughout the inspired breath. It uses a specially designed injector module, which enables tracking of the ventilator waveforms and the delivery of a synchronized and proportional dose of NO. It may be used with most ventilators.

    The INOmax DS provides continuous integrated monitoring of inspired O2, NO2, and NO, and a comprehensive alarm system.

    The INOmax DS incorporates a battery that provides up to 6 hours of uninterrupted NO delivery in the absence of an external power source.

    The INOmax DS includes a backup NO delivery capability that provides a fixed flow of 250 mL/min of NO which along with user supplied 10 L/min of oxygen provides 20 ppm in the gas flow to a patients breathing circuit. It may also use the INOblender for backup.

    The target patient population is controlled by the drug labeling for INOmax and is currently neonates. The primary targeted clinical setting is the Neonatal Intensive Care Unit (NICU) and secondary targeted clinical setting is the transport of neonates.

    Device Description

    The INOmax DS Delivery System provides a constant dose of INOmax (nitric oxide) therapy gas into the inspiratory limb of the ventilator circuit. The INOmax DS Delivery System uses a "dual-channel" design to ensure the safe delivery of INOmax.

    The first channel has the delivery CPU, the flow controller and the injector module to ensure the accurate delivery of NO. The second channel is the monitoring system, which includes a separate monitor CPU, the gas cells (NO, NO2 and O2 cells) and the user interface including the display and alarms. The dual-channel approach to delivery and monitoring permits INOmax delivery independent of monitoring but also allows the monitoring system to shutdown INOmax delivery if it detects a fault in the delivery system such that the NO concentration could become greater than 100 ppm. INOmax drug is stored as a gas mixture of NO/N2 in an aluminum cylinder at a nominal pressure of 2200 psig.

    The cylinder is attached to a high pressure requlator which incorporates a pressure gauge that indicates cylinder pressure when the cylinder valve is open. The cylinder regulator is attached via tubing to the INOmax DS Delivery System using one of the two NO/NO2 quick connect inlets on the back of the machine.

    The INOmax enters the back of the INOmax DS Delivery System, passes through a filter, then a safety shutoff valve, which is open under normal operations.

    An injector module is placed in the ventilator gas flow between the ventilator inspiratory outlet and the humidifier. Based on the ventilator flow, the INOmax cylinder concentration and set INOmax dose, the proportional solenoid delivers INOmax into the ventilator circuit via the injector module. This allows the INOmax DS Delivery System to deliver a constant dose of INOmax regardless of the ventilator flow pattern or breath rate.

    A flow sensor inside the INOmax DS Delivery Systém also monitors the NO flow out of the machine. A check valve is included prior to the INOmax DS Delivery System drug outlet to prevent pressure effects from the ventilator breathing circuit interfering with the NO flow sensor reading.

    The INOmax DS Delivery System gas monitoring system provides monitored values for inspired NO, NO2, and O2 The sample gas is withdrawn from the breathing circuit and goes through a water trap to remove excess water, a zero valve, a sample pump and finally a sample flow sensor to the gas monitoring cells. The zero valve allows the gas cells to be zeroed (during low calibration) without having to disconnect the sample line from the breathing circuit. The pump and sample flow sensor ensure a constant sample gas flow rate is maintained to the monitoring cells.

    The gas monitoring cells are electrochemical; they are specific to each gas and provide an electronic signal, which is proportional to the concentration of gas present.

    If the delivery system does go into shutdown, the INOmax DS Delivery System has an integrated backup function which provides a fixed flow of INOmax (0.25L/min) into the injector module using a pneumatic on/off switch and a restrictor built into the delivery side of the system. This fixed flow of INOmax will provide 20 ppm of NO when the continuous ventilator gas flow is 10 L/min. The backup is only for short term use until a replacement delivery system can be obtained. An alarm will warn the user if the backup system is turned on while the main delivery system is in use for INOmax delivery.

    AI/ML Overview

    The provided text describes a medical device, the INOmax DS (Delivery System), and its regulatory submission (K0909258). However, it does not contain the specific details required to complete your request for acceptance criteria and the comprehensive study information.

    The document is a 510(k) summary for a medical device (nitric oxide administration apparatus). A 510(k) submission typically focuses on demonstrating substantial equivalence to a predicate device rather than presenting extensive clinical study data with detailed acceptance criteria for diagnostic performance metrics like accuracy, sensitivity, or specificity, as would be common for AI/ML devices.

    Here's a breakdown of what can be extracted and what is missing, based on your requested points:


    1. Table of Acceptance Criteria and Reported Device Performance

    • Acceptance Criteria: The document mentions that "Testing indicated the INOmax DS Delivery System met its design input specifications, design output specifications and risk analysis requirements." (page 4) and that "The INOmax DS Delivery System was designed to comply with the applicable portions of the following product standards" (page 5, listing IEC, ASTM, and FDA guidance documents). There are also specific performance specifications for gas monitoring and NO delivery (pages 1-2).
    • Reported Device Performance: The document provides specifications for the INOmax DS in comparison to a predicate device (INOvent Delivery System). These are more akin to design specifications rather than performance results from a clinical study for diagnostic accuracy.

    What's Missing for a comprehensive table:
    The document does not explicitly state "acceptance criteria" in a pass/fail quantifiable sense for a study evaluating diagnostic performance. It talks about meeting design specifications and complying with standards, which are different. It also does not provide specific numerical results from a study that can be directly compared against acceptance criteria for metrics like sensitivity, specificity, or accuracy, as would be relevant for an AI/ML device.


    2. Sample size used for the test set and the data provenance

    • Sample Size for Test Set: This information is not provided. The document mentions "Validation of ventilators" as part of testing (page 4), but no details on the number of cases or data points, or how the test set was constituted.
    • Data Provenance: This information is not provided. There's no mention of country of origin, or if the data was retrospective or prospective.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    • This information is not provided. The device is an administration and monitoring system, not one that requires human expert interpretation for ground truth in the way an AI diagnostic tool would.

    4. Adjudication method for the test set

    • This information is not provided. Not applicable in the context of this device and the type of testing described.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • Such a study was not done and is not applicable. This device is not an AI diagnostic tool that assists human readers.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • This concept is not applicable in the context of this device. The device is a system for delivering and monitoring nitric oxide; its "performance" is about accurate gas delivery, measurement, and alarms, not AI algorithm output.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    • "Ground truth" as understood for diagnostic AI/ML devices (e.g., pathology for disease presence) is not applicable here. The "ground truth" for this device's performance would be the actual concentration of gases, flow rates, and system functionality as measured by calibrated instruments and against established engineering specifications and physical principles.

    8. The sample size for the training set

    • This information is not provided. The document describes a medical device, not an AI/ML algorithm that is "trained" on a dataset in the typical sense.

    9. How the ground truth for the training set was established

    • This information is not provided. Not applicable as per point 8.

    Summary of what the document does provide regarding "testing":

    The document states:

    • "Testing indicated the INOmax DS Delivery System met its design input specifications, design output specifications and risk analysis requirements." (page 4)
    • "Testing completed included: Validation of ventilators." (page 4)
    • The device was designed to comply with several product standards (IEC, ASTM, FDA guidance) (page 5).
    • It provides a comparison of specifications between the INOmax DS and a predicate device, for features like battery backup duration, physical dimensions, ventilator compatibility, NO delivery resolution/accuracy, and gas monitoring resolution/accuracy (pages 1-2). For example, it lists INOmax DS NO accuracy as "+/- 20% indicated or 2 ppm whichever greater" and NO accuracy for gas monitoring as "+/- 20% of reading +0.5 ppm" (page 1-2). These could be interpreted as performance metrics, but they are stated specifications rather than results from a specific study.

    Conclusion:

    The provided K0909258 document is a 510(k) summary for a medical device that delivers and monitors Nitric Oxide. It focuses on demonstrating substantial equivalence to predicate devices and adherence to design specifications and relevant standards. It is not a submission for an AI/ML diagnostic or assistive device and therefore does not contain the detailed clinical study data, acceptance criteria for diagnostic metrics, or information about ground truth establishment, expert adjudication, or reader studies that your request asks for. The "testing" mentioned is likely engineering verification and validation testing to ensure the device performs according to its technical specifications and safety requirements, rather than a clinical study establishing diagnostic performance against a ground truth.

    Ask a Question

    Ask a specific question about this device

    K Number
    K081691

    Validate with FDA (Live)

    Date Cleared
    2008-08-28

    (72 days)

    Product Code
    Regulation Number
    868.5165
    Age Range
    All
    Reference & Predicate Devices
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The INOmax DS delivery system delivers INOmax® (nitric oxide for inhalation) therapy gas into the inspiratory limb of the patient breathing circuit in a way that provides a constant concentration of nitric oxide (NO), as set by the user, to the patient throughout the inspired breath. It uses a specially designed injector module, which enables tracking of the ventilator waveforms and the delivery of a synchronized and proportional dose of NO. It may be used with most ventilators.

    The INOmax DS provides continuous integrated monitoring of inspired O2, NO2, and NO, and a comprehensive alarm system.

    The INOmax DS incorporates a battery that provides up to 6 hours of uninterrupted NO delivery in the absence of an external power source.

    The INOmax DS includes a backup NO delivery capability that provides a fixed flow of 250 mL/min of NO which along with user supplied 10 L/min of oxygen provides 20 ppm in the gas flow to a patients breathing circuit. It may also use the INOblender for backup.

    The target patient population is controlled by the drug labeling for INOmax and is currently neonates. The primary targeted clinical setting is the Neonatal Intensive Care Unit (NICU) and secondary targeted clinical setting is the transport of neonates.

    Device Description

    The INOmax DS (Delivery System) provides a constant dose of INOmax (nitric oxide) therapy gas into the inspiratory limb of the ventilator circuit. The INOmax DS (Delivery System) uses a "dual-channel" design to ensure the safe delivery of INOmax.

    The first channel has the delivery CPU, the flow controller and the injector module to ensure the accurate delivery of NO. The second channel is the monitoring system, which includes a separate monitor CPU, the gas cells (NO, NO2 and O2 cells) and the user interface including the display and alarms. The dual-channel approach to delivery and monitoring permits INOmax delivery independent of monitoring but also allows the monitoring system to shutdown INOmax delivery if it detects a fault in the delivery system such that the NO concentration could become greater than 100 ppm. INOmax drug is stored as a gas mixture of NO/N2 in an aluminum cylinder at a nominal pressure of 2200 psig.

    The cylinder is attached to a high pressure regulator which incorporates a pressure gauge that indicates cylinder pressure when the cylinder valve is open. The cylinder requlator is attached via tubing to the INOmax DS (Delivery System) using one of the two NO/NO2 quick connect inlets on the back of the machine.

    The INOmax enters the back of the INOmax DS (Delivery System), passes through a filter, then a safety shutoff valve, which is open under normal operations.

    An injector module is placed in the ventilator gas flow between the ventilator inspiratory outlet and the humidifier. Based on the ventilator flow, the INOmax cylinder concentration and set INOmax dose, the proportional solenoid delivers INOmax into the ventilator circuit via the injector module. This allows the INOmax DS (Delivery Svstem) to deliver a constant dose of INOmax regardless of the ventilator flow pattern or breath rate.

    A flow sensor inside the INOmax DS (Delivery System) also monitors the NO flow out of the machine. A check valve is included prior to the INOmax DS (Delivery System) drug outlet to prevent pressure effects from the ventilator breathing circuit interfering with the NO flow sensor reading.

    The INOmax DS (Delivery System) gas monitoring system provides monitored values for inspired NO, NO2, and O2 The sample gas is withdrawn from the breathing circuit and goes through a water trap to remove excess water, a zero valve, a sample pump and finally a sample flow sensor to the gas monitoring cells. The zero valve allows the gas cells to be zeroed (during low calibration) without having to disconnect the sample line from the breathing circuit. The pump and sample flow sensor ensure a constant sample gas flow rate is maintained to the monitoring cells.

    The gas monitoring cells are electrochemical; they are specific to each gas and provide an electronic signal, which is proportional to the concentration of gas present.

    If the delivery system does go into shutdown, the INOmax DS (Delivery System) has an integrated backup function which provides a fixed flow of INOmax (0.25L/min) into the injector module using a pneumatic on/off switch and a restrictor built into the delivery side of the system. This fixed flow of INOmax will provide 20 ppm of NO when the continuous ventilator gas flow is 10 L/min. The backup is only for short term use until a replacement delivery system can be obtained. An alarm will warn the user if the backup system is turned on while the main delivery system is in use for INOmax delivery.

    AI/ML Overview

    This document describes the INOmax DS (Delivery System), a nitric oxide administration apparatus. However, it does not contain information about acceptance criteria or a study that specifically proves the device meets acceptance criteria in the context of diagnostic accuracy, which is typically what those terms imply for AI/ML devices.

    Instead, this submission is a 510(k) for a medical device (not an AI/ML algorithm) that relies on demonstrating substantial equivalence to a predicate device. The "acceptance criteria" here are more akin to engineering specifications and performance characteristics compared against the predicate device, rather than diagnostic performance metrics. The "study" refers to engineering testing and validation to ensure the device meets its design specifications and complies with relevant standards.

    Therefore, I cannot populate the requested table and sections as they would apply to an AI/ML diagnostic device. I will extract the relevant information regarding performance specifications and testing as presented for this hardware device.

    Here's the breakdown of what is available in the provided text:

    1. A table of acceptance criteria and the reported device performance

    The document provides a comparison table between the INOvent Delivery System (predicate) and the INOmax DS (Delivery System). These are more akin to performance specifications than acceptance criteria in a diagnostic sense.

    SpecificationINOmax DS (Delivery System) Performance (Reported)
    Indications for UseExcept battery backup is 6 hours; provides an integrated backup system for delivering a fixed flow of NO into a constant flow of breathing circuit gas to produce a constant concentration of NO.
    Physical DimensionsHeight 220mm, Width 350mm, Depth 160mm, Weight 5.5Kg
    Ventilator Compatibility2-60 L/min for inspiratory flow for neonatal and 4-120 L/min for adult.
    NO DeliveryNO set resolution is 0.1/1/5 ppm depending on range; accuracy is +/- 20% indicated or 2 ppm whichever greater; NO inlet pressure is 1.7 to 2.3 Bar (25 to 33 psig).
    NO Cylinder INOmaxSame as predicate.
    Gas MonitoringNO range resolution is 0 to 10 ppm +/- (20% of reading +0.5 ppm); calibration span is daily zero with span at pre-use test if needed; NO accuracy is +/- 20% of reading +0.5 ppm.
    Injector ModuleSame as predicate.
    NO Delivery Shut OffSame as predicate.
    Calibration Gas CylindersINOcal calibration gas, INO Therapeutics LLC; NO Cal gas 45 ppm +/- 4%; NO2 Cal gas 10 ppm +/- 10%.
    Electrical SpecificationsExcept no nurse call and battery back-up is 6 hours.
    Environmental SpecificationsAmbient operation and storage pressure are 57 to 110 kPa (430 to 825 mmHg).
    NO Back-up Delivery0.25 L/min when delivered into 10 L/min provides 20 ppm; INOblender may also be used.

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    This information is not provided in the document. The document describes engineering and software validation, not a clinical study with a "test set" in the context of diagnostic performance.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    This information is not provided and is not relevant for this type of device submission, which focuses on hardware and software function, not diagnostic interpretation.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    This information is not provided and is not relevant for this type of device submission.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    An MRMC study was not done. This device is a hardware delivery system, not an AI/ML diagnostic tool that assists human readers.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This is not applicable as the device is a hardware delivery system, not an algorithm.

    7. The type of ground truth used (expert concensus, pathology, outcomes data, etc)

    The "ground truth" for this device's performance validation is its design input specifications and risk analysis requirements. The testing confirmed that the device performed according to these engineering specifications and safety requirements, and in compliance with applicable standards.

    8. The sample size for the training set

    This information is not provided and not applicable. This is not an AI/ML device that requires a "training set."

    9. How the ground truth for the training set was established

    This is not applicable as the device is a hardware delivery system, not an AI/ML algorithm.

    Ask a Question

    Ask a specific question about this device

    K Number
    K080484

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2008-08-06

    (166 days)

    Product Code
    Regulation Number
    868.5165
    Age Range
    All
    Reference & Predicate Devices
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The INOmax DS delivery system delivers INOmax® (nitric oxide for inhalation) therapy gas into the inspiratory limb of the patient breathing circuit in a way that provides a constant concentration of nitric oxide (NO), as set by the user, to the patient throughout the inspired breath. It uses a specially designed injector module, which enables tracking of the ventilator waveforms and the delivery of a synchronized and proportional dose of NO. It may be used with most ventilators.

    The INOmax DS provides continuous integrated monitoring of inspired O2, NO2, and NO, and a comprehensive alarm system.

    The INOmax DS incorporates a battery that provides up to 6 hours of uninterrupted NO delivery in the absence of an external power source.

    The INOmax DS includes a backup NO delivery capability that provides a fixed flow of 250 mL/min of NO which along with user supplied 10 L/min of oxygen provides 20 ppm in the gas flow to a patients breathing circuit. It may also use the INOblender for backup.

    The target patient population is controlled by the drug labeling for INOmax and is currently neonates. The primary targeted clinical setting is the Neonatal Intensive Care Unit (NICU) and secondary targeted clinical setting is the transport of neonates.

    Device Description

    The INOmax DS provides a constant dose of INOmax (nitric oxide) therapy gas into the inspiratory limb of the ventilator circuit. The INOmax DS uses a "dual-channel" design to ensure the safe delivery of INOmax.

    The first channel has the delivery CPU, the flow controller and the injector module to ensure the accurate delivery of NO. The second channel is the monitoring system, which includes a separate monitor CPU, the gas cells (NO, NO2 and O2 cells) and the user interface including the display and alarms. The dual-channel approach to delivery and monitoring permits INOmax delivery independent of monitoring but also allows the monitoring system to shutdown INOmax delivery if it detects a fault in the delivery system such that the NO concentration could become greater than 100 ppm. INOmax drug is stored as a gas mixture of NO/N2 in an aluminum cylinder at a nominal pressure of 2200 psig.

    The cylinder is attached to a high pressure regulator which incorporates a pressure gauge that indicates cylinder pressure when the cylinder valve is open. The cylinder regulator is attached via tubing to the INOmax DS using one of the two NO/NO2 quick connect inlets on the back of the machine.

    The INOmax enters the back of the INOmax DS, passes through a filter, then a safety shutoff valve, which is open under normal operations.

    An injector module is placed in the ventilator gas flow between the ventilator inspiratory outlet and the humidifier. Based on the ventilator flow, the INOmax cylinder concentration and set INOmax dose, the proportional solenoid delivers INOmax into the ventilator circuit via the injector module. This allows the INOmax DS to deliver a constant dose of INOmax regardless of the ventilator flow pattern or breath rate.

    A flow sensor inside the INOmax DS also monitors the NO flow out of the machine. A check valve is included prior to the INOmax DS drug outlet to prevent pressure effects from the ventilator breathing circuit interfering with the NO flow sensor reading.

    The INOmax DS gas monitoring system provides monitored values for inspired NO, NO2, and O2. The sample gas is withdrawn from the breathing circuit and goes through a water trap to remove excess water, a zero valve, a sample pump and finally a sample flow sensor to the gas monitoring cells. The zero valve allows the gas cells to be zeroed (during low calibration) without having to disconnect the sample line from the breathing circuit. The pump and sample flow sensor ensure a constant sample gas flow rate is maintained to the monitoring cells.

    The gas monitoring cells are electrochemical; they are specific to each gas and provide an electronic signal, which is proportional to the concentration of gas present.

    If the delivery system does go into shut down, the INOmax DS has an integrated backup function which provides a fixed flow of INOmax (0.25L/min) into the injector module using a pneumatic on/off switch and a restrictor built into the delivery side of the system. This fixed flow of INOmax will provide 20 ppm of NO when the continuous ventilator gas flow is 10 Limin. The backup is only for short term use until a replacement delivery system can be obtained. An alarm will warn the user if the backup system is turned on while the main delivery system is in use for INOmax delivery.

    AI/ML Overview

    The provided document is a 510(k) summary for the INOmax DS (Delivery System), indicating its substantial equivalence to previously cleared predicate devices. It describes the device's technical specifications and intended use, but it does not contain information about a study that establishes acceptance criteria through performance metrics like sensitivity, specificity, or accuracy, nor does it detail a study involving human readers or expert consensus for ground truth.

    Instead, the document focuses on demonstrating that the INOmax DS meets various engineering and safety standards, and that its performance is comparable to predicate devices. The "acceptance criteria" here are primarily framed as compliance with design specifications and relevant medical device standards.

    Here's an attempt to extract the requested information based on the provided text, recognizing that much of it is not explicitly present for a typical AI/diagnostic device study.


    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria are generally implied by the design input and output specifications, and compliance with various standards. The "performance" is described in terms of the device's functional capabilities and its similarity to predicate devices.

    Acceptance Criterion (Implied)Reported Device Performance
    Nitric Oxide (NO) Delivery AccuracyNO set resolution: 0.1/1/5 ppm depending on range. Accuracy: +/- 20% indicated or 2 ppm, whichever is greater. NO inlet pressure: 1.7 to 2.3 Bar (25 to 33 psig).
    Backup NO Delivery FunctionalityProvides a fixed flow of 0.25 L/min of NO. When delivered into 10 L/min of continuous ventilator gas flow (user-supplied 10 L/min O2), provides 20 ppm of NO. Integrated backup function with pneumatic on/off switch and restrictor.
    Gas Monitoring (NO)Range resolution: 0 to 10 ppm +/- (20% of reading + 0.5 ppm). Accuracy: +/- 20% of reading + 0.5 ppm. Calibration: Daily zero with span at pre-use test if needed.
    Gas Monitoring (NO2)Monitored values for inspired NO2 provided. (Specific accuracy/range not detailed beyond general "gas monitoring" similar to predicate).
    Gas Monitoring (O2)Monitored values for inspired O2 provided. (Specific accuracy/range not detailed beyond general "gas monitoring" similar to predicate).
    Ventilator CompatibilityCompatible with most ventilators. Inspiratory flow range: 2-60 L/min for neonatal, 4-120 L/min for adult (differs from predicate). New ventilators "qualified" and "validated."
    Battery Backup DurationProvides up to 6 hours of uninterrupted NO delivery in the absence of an external power source.
    Safety Shutoff (High NO Concentration)Monitoring system can shut down INOmax delivery if it detects a fault where NO concentration could become greater than 100 ppm.
    Compliance with Medical Electrical Equipment StandardsDesigned to comply with IEC 60601-1, IEC 60601-1-2 (EMC/EMI), IEC 60601-1-4 (Programmable Systems), IEC 60601-1-8 (Alarms).
    Compliance with Gas-Related StandardsDesigned to comply with CGA V-1 (Medical Cylinder Connections), ASTM F-1462-93 (Oxygen Analyzers), ASTM F-1054-87 (Conical Fittings).
    Compliance with FDA Guidance for Nitric Oxide Delivery DevicesDesigned to comply with FDA Guidance Document for Premarket Notification Submissions for Nitric Oxide Delivery Apparatus, Nitric Oxide Analyzer and Nitrogen Dioxide Analyzer (Special Controls).
    Physical DimensionsHeight 220mm, Width 350mm, Depth 160mm, Weight 5.5Kg (significantly smaller/lighter than predicate)
    Materials SelectionMaterials based on INOmax DS Material(s) and Datex-Ohmeda INOvent Delivery System.
    Environmental SpecificationsAmbient operation and storage pressure: 57 to 110 kPa (430 to 825 mmHg) (differs from predicate). Similar to INOmax DS for other environmental aspects.
    Calibration Gas CylindersUses INOcal calibration gas (INO Therapeutics LLC). NO Cal gas 45 ppm +/- 4%. NO₂ Cal gas 10 ppm +/- 10%.

    2. Sample Size Used for the Test Set and Data Provenance

    The document states "Testing indicated the INOmax DS met its design input specifications, design output specifications and risk analysis requirements. Testing completed included: Validation of ventilators." However, it does not specify:

    • The exact sample size for the ventilator validation or other electrical/mechanical tests.
    • The data provenance (e.g., country of origin, retrospective/prospective). This type of device's "testing" is typically bench testing and laboratory validation, not clinical data collection on patients in the way a diagnostic AI would be.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    This information is not provided in the document. The device is an electro-mechanical gas delivery system, not a diagnostic tool requiring expert interpretation of images or other clinical data for "ground truth" establishment in the typical sense of AI/medical imaging studies. Ground truth here would be established by reference instruments and engineering standards.

    4. Adjudication Method (e.g., 2+1, 3+1, none) for the Test Set

    This information is not provided and is not applicable for this type of device validation. Adjudication methods are typically used in clinical studies comparing human interpretations.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    No, an MRMC study was not done. This document describes a device for delivering therapeutic gas, not an AI-powered diagnostic tool that assists human readers.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

    Yes, implicitly. The device is a "standalone" system in that it performs its functions (gas delivery, monitoring) automatically based on user settings and its internal algorithms. Its performance is evaluated against engineering specifications and predicate devices, not typically in a human-in-the-loop context for a diagnostic output. The "testing" mentioned ("Validation of ventilators") would inherently be a standalone evaluation of the device's ability to interface and deliver gas with various ventilators.

    7. The Type of Ground Truth Used (expert consensus, pathology, outcomes data, etc.)

    For this type of device, the "ground truth" for performance evaluations would be:

    • Reference measurements: From calibrated scientific instruments for gas concentrations (NO, NO2, O2), flow rates, pressure, etc.
    • Engineering specifications: The device is designed to meet specific numerical parameters, and testing validates these parameters using metrology.
    • Standard compliance: Adherence to national and international medical device standards (e.g., IEC 60601 series, ASTM, CGA).
    • Predicate device performance: Functional equivalence to prior cleared devices serves as a benchmark.

    8. The Sample Size for the Training Set

    This information is not applicable and not provided. This is not an AI/machine learning device that requires a "training set." Its functionality is based on established physical principles and control algorithms, not learned patterns from data.

    9. How the Ground Truth for the Training Set Was Established

    This information is not applicable as there is no training set for this device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K070867

    Validate with FDA (Live)

    Date Cleared
    2007-04-27

    (29 days)

    Product Code
    Regulation Number
    868.5165
    Age Range
    All
    Reference & Predicate Devices
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The INOmax DS delivery system delivers INOmax® (nitric oxide of inhalation) therapy gas into the inspiratory limb of the patient breathing circuit in a way that provides a constant concentration of nitric oxide (NO), as set by the user, to the patient throughout the inspired breath. It uses a specially designed injector module, which enables tracking of the ventilator waveforms and the delivery of a synchronized and proportional dose of NO. It may be used with most ventilators.

    The INOmax DS provides continuous integrated monitoring of inspired O2, NO2, and NO, and a comprehensive alarm system.

    The INOmax DS incorporates a battery that provides up to 6 hours of uninterrupted NO delivery in the absence of an external power source.

    The INOmax DS includes a backup NO delivery capability that provides a fixed flow of 250 mL/min of NO which along with user supplied 10 L/min of oxygen provides 20 ppm in the gas flow to a patients breathing circuit. It may also use the INOblender for backup.

    The target patient population is controlled by the drug labeling for INOmax and is currently neonates. The primary targeted clinical setting is the Neonatal Intensive Care Unit (NICU) and secondary targeted clinical setting is the transport of neonates.

    Device Description

    The INOmax DS provides a constant dose of INOmax (nitric oxide) therapy gas into the inspiratory limb of the ventilator circuit. The INOmax DS uses a "dual-channel" design to ensure the safe delivery of INOmax. The first channel has the delivery CPU, the flow controller and the injector module to ensure the accurate delivery of NO. The second channel is the monitoring system, which includes a separate monitor CPU, the gas cells (NO, NO2 and O2 cells) and the user interface including the display and alarms. The dual-channel approach to delivery and monitoring permits INOmax delivery independent of monitoring but also allows the monitoring system to shutdown INOmax delivery if it detects a fault in the delivery system such that the NO concentration could become greater than 100 ppm. INOmax drug is stored as a gas mixture of NO/N2 in an aluminum cylinder at a nominal pressure of 2200 psig. The cylinder is attached to a high pressure regulator which incorporates a pressure gauge that indicates cylinder pressure when the cylinder valve is open. The cylinder regulator is attached via tubing to the INOmax DS using one of the two NO/NO2 quick connect inlets on the back of the machine. The INOmax enters the back of the INOmax DS, passes through a filter, then a safety shutoff valve, which is open under normal operations. An injector module is placed in the ventilator gas flow between the ventilator inspiratory outlet and the humidifier. Based on the ventilator flow, the INOmax cylinder concentration and set INOmax dose, the proportional solenoid delivers INOmax into the ventilator circuit via the injector module. This allows the INOmax DS to deliver a constant dose of INOmax regardless of the ventilator flow pattern or breath rate. A flow sensor inside the INOmax DS also monitors the NO flow out of the machine. A check valve is included prior to the INOmax DS drug outlet to prevent pressure effects from the ventilator breathing circuit interfering with the NO flow sensor reading. The INOmax DS gas monitoring system provides monitored values for inspired NO, NO2, and O2 The sample gas is withdrawn from the breathing circuit and goes through a water trap to remove excess water, a zero valve, a sample pump and finally a sample flow sensor to the gas monitoring cells. The zero valve allows the gas cells to be zeroed (during low calibration) without having to disconnect the sample line from the breathing circuit. The pump and sample flow sensor ensure a constant sample gas flow rate is maintained to the monitoring cells. The gas monitoring cells are electrochemical; they are specific to each gas and provide an electronic signal, which is proportional to the concentration of the gas present. If the delivery system does go into shut down, the INOmax DS has an integrated backup function which provides a fixed flow of INOmax (0.25L/min) into the injector module using a pneumatic on/off switch and a restrictor built into the delivery side of the system. This fixed flow of INOmax will provide 20 ppm of NO when the continuous ventilator gas flow is 10 L/min. The backup is only for short term use until a replacement delivery system can be obtained. An alarm will warn the user if the backup system is turned on while the main delivery system is in use for INOmax delivery.

    AI/ML Overview

    The provided document describes the INOmax DS (Delivery System), a device for delivering nitric oxide (NO) therapy, primarily to neonates. The document is a 510(k) summary for a premarket notification (K070867). It details the device's characteristics, its similarities to a predicate device, and the testing conducted to ensure its compliance with design specifications and applicable standards.

    Here's an analysis of the acceptance criteria and study information:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document doesn't explicitly present a table of acceptance criteria with corresponding performance metrics in a pass/fail format typical of many medical device studies for software algorithms. Instead, it describes compliance with various standards and design specifications. The performance is primarily characterized by comparing the INOmax DS to its predicate device (INOvent Delivery System) and outlining its features.

    However, based on the narrative, we can infer some key performance aspects and their intended or reported capabilities:

    Criterion (Inferred from Design/Standards)Acceptance Criteria (Implied)Reported Device Performance (INOmax DS)
    NO Delivery AccuracyConstant concentration of NO as set by user, synchronized and proportional to ventilator waveforms.Achieves constant dose regardless of ventilator flow or breath rate. NO set resolution is 0.1/1/5 ppm depending on range, accuracy is +/-20% indicated or 2 ppm whichever greater.
    Integrated Gas MonitoringContinuous monitoring of inspired O2, NO2, and NO.Provides continuous integrated monitoring of inspired O2, NO2, and NO. NO range resolution 0-10ppm +/- (20% of reading +0.5ppm).
    NO Monitoring AccuracySufficient accuracy for clinical use.NO accuracy is +/-20% of reading +0.5ppm.
    Battery Backup DurationSufficient backup power in absence of external source.Provides up to 6 hours of uninterrupted NO delivery.
    Backup NO Delivery CapabilityFixed flow of NO for emergency situations.Fixed flow of 250 mL/min of NO, providing 20 ppm with 10 L/min oxygen. Also can use INOblender for backup.
    Safety - Fault Detection/ShutdownSystem can detect faults and shut down NO delivery if concentration exceeds safe limits.Monitoring system can shut down INOmax delivery if NO concentration could become greater than 100 ppm.
    Ventilator CompatibilityCompatibility with most ventilators.May be used with most ventilators. New ventilators validated.
    Compliance with StandardsDevice complies with applicable medical device standards.Designed to comply with FDA Guidance, IEC 60601-1 (and amendments), IEC 60601-1-2, IEC 60601-1-4, IEC 60601-1-8, CGA V-1, ASTM F-1462-93, ASTM F-1054-87.
    Risk Analysis RequirementsDevice meets risk analysis requirements.Testing indicated the INOmax DS met its design input specifications, design output specifications and risk analysis requirements.

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    The document refers to "Testing" that indicated the INOmax DS met its design input, output, and risk analysis requirements. It also states "Validation of ventilators" was performed, and "newly validated ventilators have been qualified with the INOmax DS."

    • Sample Size for Test Set: Not explicitly stated for any specific test. The phrase "Validation of ventilators" implies a test set of ventilators, but the number is not given.
    • Data Provenance: Not specified. Given the nature of a 510(k) submission in the US, the testing would generally be expected to be conducted under a US-based quality system, but the geographic origin of the specific test data is not detailed.
    • Retrospective or Prospective: Not specified. Given the nature of device validation, these would typically be prospective tests conducted on the manufactured device.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    This information is not applicable to this type of device submission. The INOmax DS is a physical medical device (gas delivery system), not an AI/software algorithm that requires expert interpretation for establishing ground truth in a diagnostic or image-based context. The "ground truth" for this device would be established by physical measurements against engineering specifications and validated standards.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    This information is not applicable. Adjudication methods like '2+1' or '3+1' are typically used in clinical studies or for establishing ground truth from multiple expert readings in diagnostic imaging or similar fields. For a device like the INOmax DS, performance is verified through objective measurements and engineering standards, not through adjudication of expert opinions.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    This information is not applicable. An MRMC study is designed to evaluate the performance of diagnostic systems (often AI-assisted) by comparing multiple human readers' interpretations of multiple cases. The INOmax DS is a therapeutic gas delivery system, not a diagnostic imaging or AI-based diagnostic tool.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This information is not applicable. The INOmax DS is a physical device with integrated hardware and software for gas delivery and monitoring. It is not an "algorithm only" device, nor is its performance typically evaluated in a standalone "algorithm only" context, as it functions as a complete system with human interaction (setting dose, monitoring, etc.).

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    For this device, the "ground truth" is derived from:

    • Engineering Specifications: The device's performance against its own design input and output specifications.
    • Physical Measurements: Quantifiable measurements of gas concentration, flow rates, battery life, and alarm functions.
    • Compliance with Standards: Verification against established industry and regulatory standards (e.g., IEC, ASTM, FDA guidance documents).

    8. The sample size for the training set

    This information is not applicable. The INOmax DS is not an AI/machine learning device that requires a "training set" in the conventional sense. The device's operation is based on predefined algorithms and control logic, not learned patterns from data.

    9. How the ground truth for the training set was established

    This information is not applicable for the same reasons as in point 8.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1