Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K013858

    Validate with FDA (Live)

    Date Cleared
    2002-01-22

    (62 days)

    Product Code
    Regulation Number
    892.1000
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Infinion 0.6T MR Imaging System is indicated for use as a NMR device that produces images that: (1) correspond to the distribution of protons exhibiting NMR, (2) depend upon the NMR parameters (proton density, flow velocity, spin-lattice relaxation time (T1), and spin-spin relaxation time (T2)) and (3) display the soft tissue structure of the head and whole body. When interpreted by a trained physician, these images yield information that can be useful in the determination of a diagnosis.

    Device Description

    This submission describes the hardware and software changes for the Philips' 0.6T Open Magnetic Resonance Imaging system, the Infinion 0.6T.The package also includes additional imagin sequences and acquisition/construction techniques. However, the basic features are unchanged.

    AI/ML Overview

    The provided document is a 510(k) premarket notification for the Philips Infinion 0.6T MR Imaging System. It describes the device and claims substantial equivalence to a predicate device (Philips Proview 0.23T System, K974844).

    It's important to understand that a 510(k) submission for a Magnetic Resonance Imaging (MRI) system is primarily a demonstration of "substantial equivalence" to a predicate device, focusing on technological characteristics and intended use, rather than a clinical study proving an AI algorithm's performance against specific acceptance criteria like an AI/CADe device would.

    Therefore, the typical acceptance criteria related to sensitivity, specificity, or reader performance improvements common for AI-powered diagnostic aids are not present in this document. This submission is for an MRI hardware and software system, not an AI algorithm performing a diagnostic interpretation.

    However, based on the information provided, we can infer "acceptance criteria" through the lens of substantial equivalence and report on the "device performance" by comparing it to its predicate.

    Here's the breakdown of what can be extracted from the document:


    1. Table of Acceptance Criteria and Reported Device Performance

    Since this is a 510(k) for an MRI system asserting substantial equivalence, the "acceptance criteria" are implied by demonstrating that the new device's parameters are either the same as or generally improved compared to the predicate device, or meet established safety norms (like IEC 60601-2-33). The "reported device performance" is essentially the specifications of the Infinion 0.6T MR Imaging System compared to the Proview 0.23T System.

    Parameter (Acceptance Criteria, implied as "substantially equivalent to predicate" or "meets standard")Infinion 0.6T MR Imaging System (Reported Device Performance)Predicate Device - Philips Proview 0.23T System (K974844)
    Computer Subsystem
    Display/database systemSame as predicate.Single display for system operation and image reviewing, workstation tower, moveable keyboard and mouse, CD-ROM. Allows for simultaneous scanning and image reconstruction/manipulation. Image storage via magneo-optical disk.
    Network communicationsDICOM Query/Retrieve, DICOM Modality WL Management.Display/Database computer and scan reconstruction hardware connected together by fast Ethernet communications system. DICOM compliant for transfer of images to other systems.
    Gradient Subsystem
    Gradient CoilsSame as predicate.Water-cooled self-shielded gradient system.
    Max. Gradient Strength20 mT/m (Improved from predicate)16 mT/m (XY), 18 mT/m (Z)
    Max. Slew Rate40 T/m/s (Improved from predicate)25 mT/m/msec (XY), 40 mT/m/msec (Z).
    RF Subsystem
    RF AmplifiersMax power 9 kW. (Improved from predicate)5 kW solid state.
    RF transmitter coilsSame as predicate, but with tuning adjusted to proton resonance at 0.6T field strength instead of 0.23 T, and with water cooling.Upper and lower, planar coils, integrated in the magnet.
    Receive only RF coilsHead (phased array two-channel) Head and Neck (phased array four-channel). Neck (phased array two-channel) Extremity (phased array two-channel) Body & Spine, different sizes (phased array two-channel) Multipurpose, different sizes (linear single-channel)Head (phased array two-channel). Brain (phased array two-channel). Head and Neck (phased array four-channel, formerly Vascular Head and Neck). Neck (phased array two-channel). Extremity (phased array two-channel). Body and Spine, different sizes (phased array two-channel) Flexible Spine (phased array two-channel). Flexible Multipurpose, different sizes (linear single-channel). Small Extremity (phased array two-channel, earlier name Small Joints). Large Neck (phased array two-channel).
    Receive only coil connectionAll receive only coils plug into the RF connectors in both ends of the patient couch.All receive only coils plug into the RF connector on the front of the magnet façade.
    Magnet Subsystem
    Magnet Type0.6 T superconducting (Significantly different, but compared for equivalence)0.23 T iron-core electromagnetic.
    Patient Handling
    Patient couchMotorized movement in three orthogonal dimensions. Optional trolley. Couch w/capacity 220 kg. Remote control horizontal movement. (Improved capacity)Computer and couch controlled patient transport system with 200 kg weight capacity. Optional couch with vertical movement.
    Patient positioningSame laser positioning marker. Additional automated stepping capabilities.Laser positioning marker for accurate placement of patient at isocenter.
    Patient communicationSame as predicate.Two-way intercom system and hand-held audio alarm.
    Magnet Enclosure
    Magnet FaçadeSame as predicate.Fiberglass enclosure.
    ControlsLCD display above bore, two keypads on either side of bore. (Improved/different from predicate)LED display, control panel in upper front pole.
    Power Distribution Subsystem
    Subsystem componentsSame as predicate.Isolation transformer, transient suppression circuitry, and power distribution center all contained in a single cabinet.
    Operating Software
    Base SoftwareSame as predicate.Windows 2000 based Graphical User Interface and scan / reconstruction software with multi-tasking capability.
    Imaging sequencesField Echo, Spin Echo, Dual Echo, Dual Spin Echo, Inversion Recovery, Dual Inversion Recovery, Fast Spin Echo, EXPRESS, CBASS, MRCP, MRA, STIR, FLAIR, TSHIRT, RF-FAST, CE-FAST, DWISE. (Expanded from predicate)Field Echo, Spin Echo, Dual Echo, Inversion Recovery, Fast Spin Echo, CBASS, MRCP, MRA, STIR.
    Acquisition and reconstruction techniquesSame as predicate, and additionally MAO, MAST, presaturation, chemical fat saturation, phase conjugate symmetry, read conjugate symmetry, true res, true slice, elliptic encoding, RAM, CODA, MTC, dynamic examination curve, contrast dynamic imaging, reconstruction filters, respiratory gating, multiplanar reconstruction, maximum intensity projection, phase correction, fat/water suppression, ADC map. (Expanded from predicate)Presaturation, MAST, ODA, PCS, no phase wrap-around, no slice wrap around, dynamic imaging, cardiac gating, optimized bandwidth, turbo multislice.
    Time Varying Magnetic Field Normal Operating ModeAccording to 60601-2-33 FDISscript 2001-11-02. (References current safety standard)dB/dt ≤ 20 T/s
    First Level Controlled Operating ModeAccording to 60601-2-33 FDISscript 2001-11-02. (References current safety standard)Not applicable.
    Radiofrequency Absorption Normal Operating ModeSame as predicate.Limited to a maximum level of 1.2 W/kg.
    First Level Controlled Operating ModeSame as predicate.Limited to a maximum value of 3.2 W/kg.
    Acoustic Noise Typical87 dBA (average), 102 dB (peak)<85 dBA (average, normal clinical sequences)
    Acoustic Noise Worst Case91 dBA (average), 104 dB (peak)84 dBA (average), 98 (peak)
    Intended Use and Indications for UseSame as predicate.The Proview System is indicated for use as a NMR device that produces images that: (1) correspond to the distribution of protons exhibiting NMR, (2) depend upon the NMR parameters (proton density, flow velocity, spin-lattice relaxation time (T1), and spin-spin relaxation time (T2)) and (3) display the soft tissue structure of the head and whole body. When interpreted by a trained physician, these images yield information that can be useful in the determination of a diagnosis.

    2. Sample Size Used for the Test Set and Data Provenance

    This document describes a 510(k) submission for an MRI system, not a clinical trial for an algorithm. There is no mention of a specific "test set," "sample size," or "data provenance" (country of origin, retrospective/prospective) in the context of evaluating diagnostic performance against ground truth. The submission relies on engineering specifications and comparison to a predicate device to establish substantial equivalence.


    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of those Experts

    As there is no "test set" in the context of evaluating diagnostic performance of an AI algorithm, there is no mention of experts establishing ground truth for such a purpose. The device produces images, and the "ground truth" for its clinical utility is implied by its ability to produce diagnostic quality images, which is assessed through technical specifications and comparison to a legally marketed predicate MRI system. The document states that the images "yield information that can be useful in the determination of a diagnosis" when interpreted by a "trained physician."


    4. Adjudication Method for the Test Set

    Since there is no "test set" for diagnostic performance evaluation of an AI algorithm, there is no adjudication method described.


    5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve with AI vs Without AI Assistance

    This document is for an MRI system, not an AI or CADe device. Therefore, no MRMC study was conducted, and no effect size for human reader improvement with AI assistance is reported.


    6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) Was Done

    This document describes an MRI imaging system, not an AI algorithm intended for standalone performance. Therefore, no standalone performance study was conducted. The device produces images for a trained physician to interpret.


    7. The Type of Ground Truth Used (expert consensus, pathology, outcomes data, etc.)

    For the purpose of this 510(k) submission for an MRI system, the "ground truth" for its functionality and safety is based on engineering specifications, compliance with relevant safety standards (e.g., IEC 60601-2-33), and demonstrated equivalence to a legally marketed predicate MRI device. There isn't a "ground truth" in the diagnostic sense (like pathology or expert consensus on disease presence) because the device's role is to generate images, not to provide an automated diagnosis.


    8. The Sample Size for the Training Set

    This document is for an MRI system's hardware and embedded software. It is not an AI algorithm that requires a "training set" in the machine learning sense.


    9. How the Ground Truth for the Training Set Was Established

    As this is not an AI algorithm, there is no "training set" and no method for establishing its ground truth in the context of machine learning. The "ground truth" for the development of the MRI system would involve engineering principles, physics, and medical imaging standards to ensure functional and safe image acquisition.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1