Search Filters

Search Results

Found 3 results

510(k) Data Aggregation

    K Number
    K013612

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2002-03-19

    (134 days)

    Product Code
    Regulation Number
    884.1425
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The electrical stimulation component of the InCare Pelvic Floor Therapy System provides stimulation capabilities that are intended to provide electrical stimulation for the purpose of rehabilitation of weak pelvic floor muscles for the treatment of urinary incontinence.

    The biofeedback components of the InCare Pelvic Floor Therapy System are intended to provide electromyography or pressure biofeedback from the pelvic musculature for the purpose of rehabilitation of weak pelvic floor muscles and restoration of neuromuscular control for the treatment of urinary incontinence.

    Device Description

    The InCare Pelvic Floor Therapy System is an office based instrument that is intended to be used by physicians, nurses, nurse clinicians, and physiotherapists in a physicians office, clinic, or hospital for the purpose of providing electromyography or pressure biofeedback from pelvic musculature for the purpose of rehabilitation of weak pelvic floor muscles and the restoration of neuromuscular control for the treatment of urinary incontinence.

    The InCare Pelvic Floor Therapy System also provides electrical stimulation capabilities for the purpose of rehabilitation of weak pelvic floor muscles for the treatment of urinary incontinence.

    The proposed InCare Pelvic Floor Therapy System is comprised of a combination of an instrumentation unit and a personal computer. Other peripheral devices such as monitors and printers can be added for convenience or ease of use. The personal computer and instrumentation unit are physically separate devices "linked" to each other by a communication pathway. Features or functions that are "data manipulation and presentation" activities are properly associated with the personal computer. Features or functions that are" patient therapy" actions are controlled by the instrumentation unit.

    AI/ML Overview

    The provided text is a 510(k) premarket notification for the Hollister Incorporated InCare Pelvic Floor Therapy System. This type of document is for demonstrating substantial equivalence to a legally marketed predicate device, rather than proving a device meets specific acceptance criteria through a standalone study with detailed performance metrics.

    Therefore, the document does not contain the information requested in the prompt regarding:

    • A table of acceptance criteria and reported device performance: This is not a study assessing performance against pre-defined criteria.
    • Sample size used for the test set and data provenance: No new clinical study data is presented.
    • Number of experts used to establish ground truth and their qualifications: Not applicable, as no new ground truth was established for a test set.
    • Adjudication method: Not applicable.
    • Multi-reader multi-case (MRMC) comparative effectiveness study: Not applicable.
    • Standalone (algorithm only) performance: Not applicable, as this is a physical medical device, not an AI algorithm.
    • Type of ground truth used: Not applicable.
    • Sample size for the training set: Not applicable.
    • How the ground truth for the training set was established: Not applicable.

    What the document does provide:

    The core of this document is a statement of "substantial equivalence" to predicate devices, which means that the device has the same intended use and similar technological characteristics to devices already legally marketed.

    Here's a summary of the relevant details from the provided text:

    1. Device Name: InCare Pelvic Floor Therapy System
    2. Intended Use:
      • Electrical stimulation component: To provide electrical stimulation for the rehabilitation of weak pelvic floor muscles for the treatment of urinary incontinence.
      • Biofeedback components: To provide electromyography or pressure biofeedback from pelvic musculature for the rehabilitation of weak pelvic floor muscles and restoration of neuromuscular control for the treatment of urinary incontinence.
    3. Predicate Devices:
    4. Technological Characteristics (briefly described as justification for equivalence): The proposed system is an office-based instrument combining an instrumentation unit and a personal computer, with separate functions for "data manipulation and presentation" (PC) and "patient therapy" (instrumentation unit). The key update in this 510(k) is an "expanded stimulation parameter of 10 Hz and updated clinician application software."
    5. Conclusion of the 510(k): "Based upon the information presented within this pre-market notification it is concluded that the proposed InCare Pelvic Floor Therapy System with expanded stimulation parameter of 10 Hz and updated clinician application software is substantially equivalent to the predicate devices."

    In essence, this document is a regulatory submission for market clearance based on similarity to existing devices, not a scientific study proving performance against specific acceptance criteria.

    Ask a Question

    Ask a specific question about this device

    K Number
    K974048

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    1997-12-23

    (60 days)

    Product Code
    Regulation Number
    876.5320
    Age Range
    All
    Reference & Predicate Devices
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The biofeedback components of the InCare Pelvic Floor Therapy System with Desktop Computer are intended to provide electromyographic or pressure biofeedback from pelvic musculature for the purpose of rehabilitation of weak pelvic floor muscles and restoration of neuromuscular control for the treatment of urinary incontinence.

    The electrical stimulation component of the InCare Pelvic Floor Therapy System with Desktop Computer provides stimulation capabilities for the purpose of rehabilitation of weak pelvic floor muscles for the treatment of urinary incontinence.

    Device Description

    The InCare Pelvic Floor Therapy System with Desktop Computer is an office based instrument that is intended to be used by physicians, nurses, nurse clinicians, and physiotherapists in a physicians office, clinic, or hospital for the purpose of providing electromyographic or pressure biofeedback from pelvic musculature for the purpose of rehabilitation of weak pelvic floor muscles and restoration of neuromuscular control for the treatment of urinary incontinence. The InCare Pelvic Floor Therapy System also provides electrical stimulation capabilities for the purpose of rehabilitation of weak pelvic floor muscles for the treatment of urinary incontinence.

    The premarket notification is being submitted to the Agency to include the electrical stimulation parameter of 200 Hz to the existing InCare Pelvic Floor Therapy System with Desktop Computer, K961872.

    AI/ML Overview

    The provided text is a 510(k) premarket notification for the InCare Pelvic Floor Therapy System with Desktop Computer. It focuses on demonstrating substantial equivalence to a predicate device (Hollister PRS9300 Pelvic Floor Therapy System, K961872), primarily due to the addition of a 200 Hz electrical stimulation parameter.

    This type of submission does not typically include detailed acceptance criteria in the sense of performance metrics (like sensitivity, specificity, accuracy) derived from a clinical study, nor does it present the results of such a study with sample sizes, ground truth establishment, or expert adjudication. Instead, the acceptance criteria are met by demonstrating that the new device's technological characteristics are substantially equivalent to a legally marketed predicate device, and that this new parameter does not raise new questions of safety or effectiveness.

    Therefore, many of the requested fields cannot be directly extracted from this document as the safety and effectiveness summary relies on substantial equivalence to a predicate device rather than a new clinical performance study with specific acceptance criteria that would be typical for a novel device or a device with new indications.

    However, I can extract information related to the device characteristics that are presented to support the claim of substantial equivalence.

    1. A table of acceptance criteria and the reported device performance

    Since this is a substantial equivalence claim, the "acceptance criteria" are implicitly that the proposed device's characteristics must be comparable to the predicate device, or any differences must not raise new questions of safety or effectiveness. The "reported device performance" is essentially the detailed specifications of the proposed device, shown in comparison to the predicate device.

    CharacteristicAcceptance Criteria (Predicate Device K961872)Reported Device Performance (Proposed Device)
    Stimulation Characteristics
    Intended UseTreatment of Urinary IncontinenceTreatment of Urinary Incontinence
    Computer TypeDesktop ComputerDesktop Computer
    Output (nominal)0-30 VDC0-30 VDC
    WaveformSquare, Symmetrical, Balanced, BiphasicSquare, Symmetrical, Balanced, Biphasic
    Charge/pulse at 500Ω60 µC/phase; net charge/pulse = 060 µC/phase; net charge/pulse = 0
    Frequency12.5, 20, 50, 100 Hz12.5, 20, 50, 100, 200 Hz (This is the difference)
    Peak Pulse Intensity30 VDC30 VDC
    Pulse width0.3, 1 ms0.3, 1 ms
    Ramps20%, 40%, 60%, 80%, 100% of “ON” time (no down ramping)20%, 40%, 60%, 80%, 100% of “ON” time (no down ramping)
    Duty CycleOn (sec): 1-80 in 1 sec increments; Off (sec): 0 - 80 in 1 sec incrementsOn (sec): 1-80 in 1 sec increments; Off (sec): 0 - 80 in 1 sec increments
    Session Duration (min)0-30, 1 minute increments0-30, 1 minute increments
    Programmable FeaturesBy physician: Pulse width, Frequency, Duty cycle, Session length (none by patient)By physician: Pulse width, Frequency, Duty cycle, Session length (none by patient)
    Current Density (Probe 9595)0.003 amperes/cm²0.003 amperes/cm²
    Current Density (Probe 9596)0.018 amperes/cm²0.018 amperes/cm²
    Power Density (Probe 9595)0.047 watts/cm²0.047 watts/cm²
    Power Density (Probe 9596)0.239 watts/cm²0.239 watts/cm²
    EMIMeets IEC 601-1, IEC 601-1-1, IEC 601-2-10 (except Para 36), EN 60601-1-2/IEC 601-1-2, EN 55011/CISPR 11, EN60801-2/IEC 801-2, IEC 801-3, IEC 801-4, IEC 801-5Meets IEC 601-1, IEC 601-1-1, IEC 601-2-10 (except Para 36), EN 60601-1-2/IEC 601-1-2, EN 55011/CISPR 11, EN60801-2/IEC 801-2, IEC 801-3, IEC 801-4, IEC 801-5
    Biofeedback Characteristics
    Intended UseTreatment of Urinary IncontinenceTreatment of Urinary Incontinence
    Measurement ChannelsAdjustable, 2 channels EMG, 2 channels pressure, or combinationAdjustable, 2 channels EMG, 2 channels pressure, or combination
    EMG Sensitivity (microvolts)0-5, 0-10, 0-25, 0-50, 0-100, 0-250, 0-5000-5, 0-10, 0-25, 0-50, 0-100, 0-250, 0-500
    EMG Bandwidth100-500 Hz100-500 Hz
    EMG Signal ProcessingRoot Mean Squared (RMS)Root Mean Squared (RMS)
    EMG DetectionBipolarBipolar
    Pressure Sensitivity (cm-H2O)0-10, 0-25, 0-50, 0-1000-10, 0-25, 0-50, 0-100
    Work Period (sec)1-80 in 1 second increments1-80 in 1 second increments
    Rest Period (sec)0-80 in 1 second increments0-80 in 1 second increments
    Session Duration (min)1-60 in 1 minute increments1-60 in 1 minute increments

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    Not applicable. This is a substantial equivalence submission based on device specifications comparison, not a clinical study with a test set of data.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    Not applicable. No ground truth was established by experts for a test set in this submission.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Not applicable. No test set adjudication method described as no clinical performance study was conducted.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This is not an AI-assisted device, and no MRMC study was conducted or described.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. This is a medical device for treatment/biofeedback, not an algorithm being evaluated for standalone performance.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    Not applicable. As noted, this submission is driven by substantial equivalence to a predicate device based on technical specifications rather than a clinical study requiring ground truth derived from expert consensus, pathology, or outcomes data. The "ground truth" for this application is the established safety and effectiveness of the predicate device and the demonstration that the new device remains within those bounds.

    8. The sample size for the training set

    Not applicable. This submission doesn't involve machine learning or a training set.

    9. How the ground truth for the training set was established

    Not applicable. This submission doesn't involve machine learning or a training set.

    Ask a Question

    Ask a specific question about this device

    K Number
    K961872

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    1996-08-09

    (86 days)

    Product Code
    Regulation Number
    876.5320
    Age Range
    All
    Reference & Predicate Devices
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The biofeedback components of the InCare Pelvic Floor Therapy System with Desktop Computer are intended to provide electromyographic or pressure biofeedback from pelvic musculature for the purpose of rehabilitation of weak pelvic floor muscles and restoration of neuromuscular control for the treatment of female urinary incontinence.

    The electrical stimulation component of the InCare Pelvic Floor Therapy System with Desktop Computer provides stimulation capabilities for the purpose of rehabilitation of weak pelvic floor muscles for the treatment of female urinary incontinence.

    Device Description

    The InCare Pelvic Floor Therapy System with Desktop Computer is an office based instrument that is intended to be used by physicians, nurse clinicians, and physiotherapists in a physicians office, clinic, or hospital for the purpose of providing electromyographic or pressure biofeedback from pelvic musculature for the purpose of rehabilitation of weak pelvic floor muscles and restoration of neuromuscular control for the treatment of female urinary incontinence. The InCare Pelvic Floor Therapy System also provides electrical stimulation capabilities for the purpose of rehabilitation of weak pelvic floor muscles for the treatment of female urinary incontinence.

    The "nucleus" of the system is the combination of an instrumentation unit and a desktop computer. Other peripheral devices such as monitors and printers can be added for convenience or ease of use. There is a clear separation of functional control between the two elements of the system. Features or functions that are "data manipulation and presentation" activities are properly associated with the personal computer. Features and functions that are "patient treatment" actions are controlled by the Instrumentation Unit.

    The desktop computer and Instrumentation Unit are physically separate devices "linked" to each other by a communication pathway.

    AI/ML Overview

    The provided text describes a 510(k) premarket notification for a medical device and, as such, does not contain a study demonstrating novel clinical effectiveness or "acceptance criteria" in the way understood for new efficacy studies.

    Instead, the document asserts substantial equivalence to a predicate device (Hollister PRS9300 Pelvic Floor Therapy System, K930530/C). This means the device is considered safe and effective because it is essentially the same as a legally marketed device.

    Therefore, the "acceptance criteria" here are essentially the technological characteristics and intended use being substantially equivalent to the predicate device. The "study" is a comparative analysis demonstrating this equivalence.

    Here's the information broken down based on the provided text, while acknowledging that a traditional clinical study as you might expect for a new drug or novel device with performance metrics against a clinical endpoint is not present:

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria (based on Predicate Device Characteristics)Reported Device Performance (InCare Pelvic Floor Therapy System with Desktop Computer)Evidence/Basis for Performance
    Intended Use: Treatment of Urinary IncontinenceIntended Use: Treatment of Urinary IncontinenceStated as identical.
    Computer Type: Laptop ComputerComputer Type: Desktop ComputerThis is the one difference. The document argues this difference does not raise new questions of safety or effectiveness because patient treatment actions are controlled by the Instrumentation Unit (which is identical to the predicate).
    Stimulation Characteristics (e.g., Output, Waveform, Frequency, Pulse Width, Duty Cycle, etc.)Identical to Predicate Device (e.g., Output: 0-30 VDC, Waveform: Square, Symmetrical, Balanced, Biphasic, Frequency: 12.5, 20, 50, 100 Hz, Pulse Width: 0.3, 1 ms, Duty Cycle: On (sec): 1-80, Off (sec): 0-80)Explicitly listed as identical in the comparison table.
    Biofeedback Characteristics (e.g., Measurement Channels, EMG Sensitivity, Pressure Sensitivity)Identical to Predicate Device (e.g., Measurement Channels: Adjustable, 2 channels EMG, 2 channels pressure, EMG Sensitivity: 0-5 to 0-500 microvolts, Pressure Sensitivity: 0-10 to 0-100 cm-H2O)Explicitly listed as identical in the comparison table.
    EMI/EMC Compliance: Meets EN 55011 and IEC 801-3 (for predicate)Meets EN 60601-1-2/IEC 601-1-2, EN 55011/CISPR 11, EN60801-2/IEC 801-2, IEC 801-3, IEC 801-4, IEC 801-5The proposed device provides a list of specific standards it aims to comply with, which are considered equivalent or updated versions of the predicate's compliance.
    Safety and Effectiveness: Shown to be safe and effective for intended use (by predicate clearance)Safety and Effectiveness: Concluded to be safe and effective for its intended useBased on the assertion of substantial equivalence due to identical instrumentation unit, software for data manipulation/patient therapy, and the desktop computer difference not impacting critical functions.

    Regarding a "Study" to prove the device meets acceptance criteria:

    The document describes a comparison analysis rather than a traditional clinical or performance study in the sense of collecting new patient data.

    2. Sample size used for the test set and the data provenance:

    • Test Set Sample Size: Not applicable. This document is a predicate comparison for a 510(k) submission, not a new clinical study. There is no "test set" of patients or data in the typical sense.
    • Data Provenance: The "data" used is a comparison of the device's technical specifications and intended use against a previously cleared device (the predicate). The provenance is the design specifications of the new device and the known characteristics of the predicate device. This is a retrospective analysis by comparing specifications.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • Not applicable. Ground truth, in the context of clinical studies, refers to definitive diagnoses or outcomes. For a 510(k) submission, the "ground truth" for regulatory clearance is that the device is substantially equivalent to a legally marketed predicate. This "truth" is established by the manufacturer's regulatory team and reviewed by the FDA, based on documented technical specifications.

    4. Adjudication method for the test set:

    • Not applicable. There is no multi-person adjudication of a "test set" in this type of submission. The comparison is made against the clear specifications and intended use of the predicate device.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • Not applicable. This device is a pelvic floor therapy system that provides biofeedback and electrical stimulation. It is not an AI-powered diagnostic imaging device or an AI assistant for human readers. Therefore, an MRMC study is irrelevant to this device.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

    • Not applicable. This is a physical therapy device used by clinicians with patients, not a standalone algorithm.

    7. The type of ground truth used:

    • The "ground truth" in this context is the design specifications and functional equivalence to the legally marketed predicate device (Hollister PRS9300 Pelvic Floor Therapy System, K930530/C), as defined by FDA 510(k) regulations.

    8. The sample size for the training set:

    • Not applicable. As this is not an AI/ML device or a de novo clinical study, there is no "training set."

    9. How the ground truth for the training set was established:

    • Not applicable for the same reasons as #8.
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1