Found 2 results

510(k) Data Aggregation

    K Number
    K160044

    Validate with FDA (Live)

    Date Cleared
    2016-02-01

    (24 days)

    Product Code
    Regulation Number
    876.1500
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The IMAGE1 SPIES is a camera control unit (CCU) for use with camera heads or video endoscopes for the visualization. image recording and documentation during general endoscopic and microscopic procedures.

    Device Description

    The modified Image1 SPIES (Storz Professional Image Enhancement System) camera control unit includes an additional camera head that can be connected to the X-link module and software updates for improvement of existing features and bug fixes. The SPIES Camera System is intended and designed for use during endoscopic and microscopic procedures.

    AI/ML Overview

    The document provided is a 510(k) summary for the Karl Storz Endoscopy America, Inc. Image1 SPIES System. This type of submission is for demonstrating substantial equivalence to a predicate device, not typically for proving a device meets
    clinical acceptance criteria through a specific study in the way a novel AI algorithm might.

    Based on the provided text, here's a breakdown of the acceptance criteria and the study that "proves" the device meets them, keeping in mind the context of a 510(k) for an endoscopic camera system:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not provide specific numerical acceptance criteria (e.g., minimum resolution in lp/mm, acceptable latency in ms) in a table format. Instead, it states that the device "met all its specifications." The performance validation involved a series of bench tests.

    Acceptance Criterion (Implied)Reported Device Performance
    ResolutionMet specifications
    BrightnessMet specifications
    White BalanceMet specifications
    Color PerformanceMet specifications
    LatencyMet specifications
    Temporal NoiseMet specifications
    Residual Spatial NoiseMet specifications
    Step Response of Automated Exposure SystemMet specifications
    Software Verification TestsMet specifications

    The "bench test data for the Image1 SPIES System demonstrates that the design characteristics used as the basis for the comparison have been met. The results show that the subject device has met all its specifications. The performance validation test report can be provided upon request."

    2. Sample Size Used for the Test Set and Data Provenance

    • Test Set Sample Size: The document does not specify a "sample size" in terms of cases or patients for the bench tests. The testing performed was on the device itself and its components.
    • Data Provenance: Not applicable in the context of clinical data. The tests were bench tests conducted on the device.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

    • Number of Experts: Not applicable. The ground truth for bench tests is based on engineering specifications and direct physical measurements, not expert consensus on medical images.

    4. Adjudication Method for the Test Set

    • Adjudication Method: Not applicable. Bench testing relies on objective measurements against predefined technical specifications.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    • MRMC Study: No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not performed. This type of study is typically done for diagnostic aids or AI algorithms that assist human readers in interpreting medical images, assessing the impact on diagnostic accuracy, sensitivity, or specificity. The Image1 SPIES System is a camera control unit for visualization, not a diagnostic AI tool.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    • Standalone Study: Yes, in a sense, the bench testing performed is analogous to a standalone performance evaluation for the device's technical specifications. The performance of the camera control unit itself (resolution, brightness, etc.) was evaluated independently of human interpretation of clinical outcomes. However, this is not an "algorithm only" study as would be described for AI.

    7. The Type of Ground Truth Used

    • Type of Ground Truth: The ground truth for the bench tests was based on engineering specifications and objective physical measurements. For example, resolution would be measured against a standard test pattern, and latency would be measured with specialized equipment.

    8. The Sample Size for the Training Set

    • Training Set Sample Size: Not applicable. This device is not an AI/ML algorithm that requires a training set.

    9. How the Ground Truth for the Training Set Was Established

    • Ground Truth for Training Set: Not applicable, as there is no training set for this device.
    Ask a Question

    Ask a specific question about this device

    K Number
    K131953

    Validate with FDA (Live)

    Device Name
    IMAGE1 SPIES
    Date Cleared
    2013-12-20

    (176 days)

    Product Code
    Regulation Number
    876.1500
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Image1 SPIES is a camera control unit (CCU) for use with camera heads or video endoscopes for visualization, image recording and documentation during general endoscopic and microscopic procedures.

    Device Description

    The Image1 SPIES (Storz Professional Image Enhancement System) camera control unit is a medical device which consists of an Image1 Connect display module (TC200), and a combination of a minimum of one to a maximum of three camera head input modules that is intended and designed for use during endoscopic procedures. Image1 H3-Link (TC300) and Image1 X-Link (TC301) are available as head input modules. The device's modularity enables customers to customize their Image1 SPIES system to their specific current and future video needs. Descriptions of these three modular devices are provided below.

    Image1 Connect (TC200):
    The Image1 Connect displav module is connected to the head module(s) via an inter-module link cable. The Image1 Connect accepts video from the head module(s) and offers additional functionality such as image capture, image printing, picture-in-picture representation and generation of a digital video output for monitor viewing purposes.

    Image1 H3-Link (TC300):
    The Image1 H3-Link head module processes raw video signals from the H3 family of camera heads.

    Image1 X-Link (TC301)
    The Image1 X-Link head module also processes raw video signal from a certain set of imagers housed in a camera head. Visually, the card edge connector receptacle of the X-Link is wider than the receptacle of the H3-Link.

    AI/ML Overview

    The provided text describes a 510(k) summary for the KARL STORZ Image1 SPIES, an endoscopic video imaging system. It details the device, its intended use, and substantial equivalence to predicate devices, but does not contain explicit acceptance criteria or a detailed study proving the device meets specific performance criteria through a clinical trial or a statistically designed non-clinical study that would typically generate the requested performance metrics.

    The document focuses on demonstrating substantial equivalence to previously cleared devices (Image1 Video Imaging System K070716 and Fujinon EPX-4440HD Video Processor and Light Source K102466) through non-clinical bench testing.

    Here's an analysis based on the provided text, addressing each point as much as possible:

    1. Table of acceptance criteria and the reported device performance:

      The document mentions the types of non-clinical tests performed but does not provide specific numerical acceptance criteria or the reported performance values. It states:

      Test CategoryAcceptance Criteria (Not explicitly stated with values)Reported Device Performance (Not explicitly stated with values)
      Electrical SafetyCompliance with IEC 60601-1 and IEC 60601-1-2 standards."demonstrates the continued electrical safety of the device"
      Optical Performance"includes white balance, brightness, image enhancement, video output format, zoom, image quality, image capture, latency and interface control"Implied to meet or exceed predicate devices' performance.
      Software Reliability(Not explicitly stated)"reliability of the software" is tested.
      Overall PerformanceAs safe, as effective, and performs as well as or better than legally marketed predicate devices."demonstrates that the device is as safe, as effective and performs as well as or better than the legally marketed devices."

      Note: The document only lists the categories of tests and a general conclusion of meeting requirements, rather than providing specific quantifiable acceptance criteria (e.g., "latency must be <X ms") and the measured results.

    2. Sample size used for the test set and the data provenance:

      • Sample Size: Not specified. The studies mentioned are described as "non-clinical bench testing," which implies physical device testing rather than testing on a dataset of images or clinical cases.
      • Data Provenance: Not applicable in the context of clinical/image data. The testing involves the device itself.
    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

      • Not applicable. The tests are non-clinical bench tests focusing on technical specifications and safety, not on human interpretation or diagnosis.
    4. Adjudication method for the test set:

      • Not applicable as no human interpretation or ground truth establishment by multiple experts is described for this type of non-clinical bench testing.
    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

      • No. The document explicitly states "non-clinical performance tests." There is no mention of an MRMC study, human readers, or AI assistance. The device is a camera control unit, not an AI-driven diagnostic tool.
    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

      • Not explicitly applicable in the context of an "algorithm only" performance because the device is hardware (a camera control unit) with integrated software for image processing and control. The performance tests are for the device's functional capabilities, such as image quality and safety, rather than a standalone algorithm's diagnostic accuracy. The non-clinical tests assess the device's technical performance.
    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

      • Not applicable. For non-clinical bench testing, "ground truth" would relate to engineering specifications and objective measurements (e.g., a calibrated light source for brightness, a specific pattern for image resolution, electrical safety standards compliance). No clinical ground truth (like pathology) is used as it's not a diagnostic device with outputs requiring such validation.
    8. The sample size for the training set:

      • Not applicable. The device is not described as using machine learning or AI that would require a "training set."
    9. How the ground truth for the training set was established:

      • Not applicable, as there is no training set described.

    In summary:

    The provided document describes the KARL STORZ Image1 SPIES as a camera control unit that underwent non-clinical bench testing to demonstrate electrical safety, optical performance (white balance, brightness, image enhancement, video output, zoom, image quality, image capture, latency, interface control), and software reliability. The conclusion is that these tests demonstrate the device is as safe and effective as, and performs as well as or better than, its predicate devices. However, the document lacks specific, quantifiable acceptance criteria and detailed numerical results for any of these tests, and it does not involve clinical studies with human readers or AI components that would typically require the detailed information requested in points 2-9.