Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K131791

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2014-02-26

    (253 days)

    Product Code
    Regulation Number
    866.5100
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The EUROIMMUN IFA 40: HEp-20-10 is an indirect immunofluorescence antibody test for the qualitative or semi-quantitative detection of antibodies against cell nuclei (ANA) in human serum. This test system is used as an aid in the diagnosis of systemic rheumatic diseases in conjunction with other laboratory and clinical findings.

    Device Description

    The test system consists of BIOCHIPs coated with HEp-20-10 cells. It includes a fluorescein-labeled goat anti-human IgG, a positive and negative control, salt for PBS, Tween 20, embedding medium, cover glasses and instruction booklet. Reagent trays for the TITERPLANE technique are required but ordered separately.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and study information for the EUROIMMUN IFA 40: HEp-20-10 device, based on the provided text:

    Acceptance Criteria and Device Performance

    CriteriaAcceptance Criteria (Implicit/Explicit)Reported Device Performance and Confidence Intervals
    Qualitative Agreement with Predicate DeviceNot explicitly stated as a numerical target in the provided text. However, the study aims to demonstrate substantial equivalence, implying high agreement. The predicate device is ImmunoConcepts® HEp-2000 ANA-Ro IFA.Overall Agreement: 94.5% (95% C.I.: 90.4% - 97.2%)Positive Agreement: 92.4% (95% C.I.: 83.2% - 97.5%)Negative Agreement: 95.5% (95% C.I.: 90.5% - 98.3%)
    Semi-Quantitative Agreement with Predicate Device (Positive/Negative)Not explicitly stated as a numerical target. The study aims to demonstrate substantial equivalence.Overall Agreement: 100.0% (95% C.I.: 97.7% - 100.0%)Positive Agreement: 100.0% (95% C.I.: 96.7% - 100.0%)Negative Agreement: 100.0% (95% C.I.: 92.3% - 100.0%)
    Semi-Quantitative Agreement with Predicate Device (Pattern Agreement)Not explicitly stated as a numerical target. Implicitly, high agreement is desired.Overall Pattern Agreement: 91.4% (Individual pattern agreements range from 66.7% (Nuclear Membrane) to 100.0% (Homogenous, Centromere, Nuclear Dot))
    Precision/ReproducibilityFluorescence Intensity: The results should not exceed an acceptable deviation of +/- 1 intensity level. Positive samples should not be found negative and vice versa. Observed patterns should not change. Endpoint Titer: Endpoint titer should not deviate more than +/- 1 titer level.Intra-Assay, Inter-Assay, Inter-Lot, Inter-Observer, Semi-quantitative Reproducibility: All studies met the criteria; results did not exceed +/- 1 intensity level deviation, positive/negative status remained consistent, and patterns did not change. Semi-quantitative reproducibility also showed endpoint titer did not deviate more than +/- 1 titer level.
    Linearity/Assay Reportable RangeMixed patterns should be distinguishable in every dilution. Samples should show a decrease in fluorescence intensity with increasing dilutions. The pattern of the samples should not change with dilution. Acceptable deviation of fluorescence intensity: ± 1 intensity level.Mixed patterns were distinguishable in every dilution. Samples showed a decrease in fluorescence intensity with dilution. The pattern did not change with dilution. Acceptable deviation of fluorescence intensity (± 1 intensity level) was met.
    Analytical Specificity (Cross-Reactivity)No significant cross-reactivity with ANCA-associated vasculitis, Crohn's disease, ulcerative colitis, celiac disease, Chlamydia pneumoniae, and Epstein-Barr virus samples. CDC reference panel results should be in line with CDC characterization.No significant cross-reactivity observed with the tested clinical samples. Results for the CDC reference panel were in line with CDC characterization (with one exception, CDC sample No. 12, which was negative across all three test systems).
    Analytical Specificity (Interfering Substances)Deviation in fluorescence intensity level should not exceed +/- 1. No significant interference.Hemoglobin (up to 1000 mg/dL), bilirubin (up to 40 mg/dL), triglyceride (up to 2000 mg/dL), HAMA, and RF at indicated concentrations had no effect on assay results. Deviation in fluorescence intensity level did not exceed +/- 1. No significant interference observed.
    Assay Cut-off VerificationPrevalence of ANAs in healthy individuals should be within the range of 3.0% - 15% as per the American College of Rheumatology.A prospective study with 138 samples (routine health screening) found a prevalence of 11.6% (95% C.I.: 6.8% - 18.6%) for ANA positivity at the 1:40 dilution, which falls within the 3.0% - 15% range.
    Expected Values/Reference RangePrevalence of ANAs in healthy individuals should be about 3.0% - 15% as per the American College of Rheumatology. Reference range determined as titer <1:40.A study with 200 healthy adult blood donors found a prevalence of ANA about 3.5%, which is within the 3.0% - 15% range. The reference range was determined as titer <1:40.

    Study Details

    2. Sample Size and Data Provenance

    • Qualitative Comparison Study (Method comparison with predicate device):
      • Test Set Sample Size: 200 blinded prospective samples.
      • Data Provenance: Department of Clinical Pathology & Laboratory Medicine at the University of Pennsylvania (USA). Samples obtained from patients sent for routine ANA screening.
    • Semi-Quantitative/Quantitative Comparison Study (Method comparison with predicate device):
      • Test Set Sample Size: 156 blinded prospective samples.
      • Data Provenance: Samples obtained from patients sent for routine ANA screening (location not explicitly stated but implied to be the same clinical setting as the qualitative study given the continuous description and similar sample acquisition method).
    • Assay Cut-off Verification:
      • Test Set Sample Size: 138 samples.
      • Data Provenance: EUROIMMUN US Inc. laboratory (USA). Serum samples sent in for antibody testing, collected in sequential order for two days.
    • Expected Values/Reference Range Study:
      • Test Set Sample Size: 200 sera from normal healthy adult blood donors.
      • Data Provenance: University hospital Lübeck, Germany.
    • Analytical Performance Studies (Precision/Reproducibility, Linearity, Analytical Specificity, Interfering Substances):
      • Data Provenance: All performed at EUROIMMUN Medizinische Labordiagnostika AG, Lübeck, Germany, unless stated otherwise. The Inter-Observer Reproducibility Study was performed in a US Laboratory setting.
      • Sample Sizes: Vary per study (e.g., 10-fold repeated measurements for Intra-Assay, 20 repeated for Inter-Assay, 6 for Inter-Lot, 14 positive and 1 negative for Semi-quantitative Reproducibility, 6 for Linearity, 20 for Interfering substances). Specific numbers for each analytical study are provided in the "Performance Characteristics" section.

    3. Number of Experts used to establish the ground truth for the test set and the qualifications of those experts

    • Qualitative and Semi-Quantitative/Quantitative Comparison Studies:
      • Number of Experts: Two primary technicians, with a third decisive technician in case of discrepancies.
      • Qualifications: "Technicians" are mentioned. No specific background (e.g., years of experience, specific certifications) is provided.
    • Analytical Performance Studies (e.g., Reproducibility, Linearity, Analytical Specificity, Interfering Substances):
      • Generally, the evaluation was performed by "the same technician" or "two different technicians," with a third technician for resolution in Inter-Observer Reproducibility. No specific qualifications are detailed.
    • Assay Cut-off Verification & Expected Values/Reference Range Studies:
      • "As per the American College of Rheumatology" is cited for an expected prevalence range, indicating reliance on established medical guidelines/expert consensus for context, but not for direct ground truth labeling of individual samples in these specific studies. Samples were tested as per instructions for use by technicians.

    4. Adjudication method for the test set

    • Qualitative and Semi-Quantitative/Quantitative Comparison Studies:
      • Adjudication Method: "Visual fluorescence microscopy evaluation was performed by two different technicians independently. In case of a discrepant result a 3rd technician was decisive." This is a 2+1 adjudication method.
    • Inter-Observer Reproducibility:
      • Adjudication Method: "separate independent reading of the Inter-Assay slides by a 2nd technician. In case of discrepancies a 3th technician would have been decisive." This is also a 2+1 adjudication method.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance.

    • No, a multi-reader multi-case (MRMC) comparative effectiveness study focusing on human readers improving with AI vs. without AI assistance was not performed. This device is an in-vitro diagnostic test kit that relies on human microscopy reading, not an AI-assisted diagnostic tool for image interpretation. The comparison studies were between the new device and a predicate device, both relying on human visual interpretation.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done

    • No, a standalone algorithm-only performance study was not done. The device is an IFA test system that requires human visual interpretation via a fluorescent microscope.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    • Qualitative and Semi-Quantitative/Quantitative Comparison Studies: The ground truth was established by the predicate device (ImmunoConcepts® HEp-2000 ANA-Ro IFA), interpreted by technicians with a 2+1 adjudication. Discrepancies were sometimes confirmed using "FDA cleared assays" (e.g., blot tests), which serve as a higher-level reference standard.
    • Analytical Specificity (Cross-reactivity): The CDC Centers for Disease Control and Prevention reference panel (characterized by the CDC) was used as ground truth.
    • Assay Cut-off Verification & Expected Values/Reference Range Studies: The ground truth for individual samples was their status as positive or negative based on the device's interpretation at a specific dilution. The reference range and prevalence context were established by comparing against published ranges from the American College of Rheumatology.
    • Precision/Reproducibility, Linearity, Interfering Substances: Ground truth implicitly involved the expected behavior of the samples (e.g., 0-4+ intensity, specific patterns, decreasing intensity with dilution) as evaluated by trained technicians.

    8. The sample size for the training set

    • The document does not mention a training set, as this is an in-vitro diagnostic test kit that is read visually by technicians, not a machine learning or AI-based diagnostic tool that would typically require a training set.

    9. How the ground truth for the training set was established

    • Not applicable, as no training set for an algorithm was mentioned or used.
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1