Search Filters

Search Results

Found 2 results

510(k) Data Aggregation

    K Number
    K161158

    Validate with FDA (Live)

    Date Cleared
    2017-01-31

    (281 days)

    Product Code
    Regulation Number
    862.1545
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The IDS-iSYS Intact PTH+ assay is an in vitro diagnostic device intended for the quantitative determination of intact PTH in human serum or plasma on the IDS-iSYS Multi-Discipline Automated System. Results are to be used in the differential diagnosis of hypercalcemia and hypocalcemia resulting from disorders of calcium metabolism.

    Device Description

    The IDS-iSYS Intact PTH" assay is based on chemiluminescence technology. 100 uL of patient sample is incubated with a biotinylated polyclonal anti-PTH (39-84) antibody and an acridinium labelled PTH (13-34) antibody. Streptavidin labelled magnetic particles are added prior to a second incubation step. The magnetic particles are captured using a magnet and a wash step performed to remove any unbound analyte. Trigger reagents are added; the resulting light emitted by the acridinium label is directly proportional to the concentration of Intact PTH in the original sample.

    IDS-iSYS Intact PTH reagent kit consists of one (1) Immunoassay Cartridge, two (2) vials each of 2 concentration Calibrator levels and a mini-CD containing the Instructions For Use (IFU), CRY files and Certificate of Analysis.

    IDS-iSYS Intact PTHN Cartridge, sufficient for 100 tests, consists of reagents provided in individual compartment within a plastic container called the Cartridge. The IDS-iSYS Intact PTHN Cartridge contains the following ready to use reagents:

    • · Magnetic particles coated with streptavidin in a phosphate buffer containing sodium azide as preservative (<0.1%), 1 bottle.
    • Goat polyclonal antibody against 13-34 PTH labelled with an acridinium ester derivative, in buffer containing goat serum with sodium azide as preservative (<0.1%), 1 bottle.
    • Goat polyclonal antibody against 39-84 PTH 1abelled with biotin, in buffer containing bovine and goat proteins with sodium azide as preservative (<0.1%), 1 bottle.

    IDS-iSYS Intact PTH Calibrators are included in the reagent kit.

    • · Calibrator A: Two vials of lyophilized porcine serum matrix buffer containing low level PTH and sodium azide as preservative >1% (w/w%).
    • · Calibrator B: Two vials of lyophilized porcine serum matrix buffer containing high level PTH and sodium azide as preservative >1% (w/w%).

    The submission is due to a new source of antibody for the assay.

    AI/ML Overview

    This document describes the performance characteristics of the IDS-iSYS Intact PTHN assay, an in vitro diagnostic device for quantitative determination of intact PTH.

    Here's an analysis of the acceptance criteria and study information provided:

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria are not explicitly stated as distinct numerical targets but are implicitly defined by the reported performance and the comparison to the predicate device. The information below summarizes the performance characteristics provided, which in a regulatory context, demonstrate the device meets its intended use and is substantially equivalent to the predicate.

    Performance CharacteristicPredicate Device (K103325) PerformanceCandidate Device (IDS-iSYS Intact PTHN) PerformanceImplied Acceptance Criteria (Demonstrated)
    Method Comparison (against a commercially available method)N/An=312 samples; Slope: 1.02 (95% CI: 0.99 to 1.04); Intercept: -0.7 pg/mL (95% CI: -5.4 to 4.1); Correlation coefficient (r): 1.00Demonstrate strong correlation and agreement with a commercially available quantitative chemiluminescence Intact PTH assay (e.g., r close to 1, slope close to 1, intercept close to 0) as per CLSI EP-9A2 guidelines.
    Sample Matrix Comparison (vs. K2 EDTA plasma)N/A
    - Serum (Red Top)N/An=52; Slope: 0.94 (95% CI: 0.92 to 0.97); Intercept: 2.55 (95% CI: 0.86 to 3.16); r: 1.00Demonstrate acceptable agreement across different sample matrices (serum, lithium heparin plasma) compared to the control (K2 EDTA plasma) as per CLSI EP9-A3 guidelines. Slopes close to 1 and intercepts close to 0 with good correlation coefficients.
    - Serum (SST)N/An=52; Slope: 0.93 (95% CI: 0.91 to 0.96); Intercept: 2.38 (95% CI: 1.25 to 3.15); r: 1.00
    - Lithium HeparinN/An=52; Slope: 0.98 (95% CI: 0.95 to 0.99); Intercept: 0.42 (95% CI: -0.43 to 1.63); r: 1.00
    Reference Interval11.5 to 78.4 pg/mL10.3 to 80.5 pg/mLEstablish a 95% reference interval for a healthy population, consistent with clinical expectations and CLSI C28-A3 guidelines. The new range should be comparable to the predicate.
    Sensitivity
    - Limit of Blank (LoB)1.2 pg/mL0.9 pg/mLDemonstrate low limits of blank, detection, and quantitation, indicating the ability to detect very low concentrations of PTH, as per CLSI EP17-A guidelines. Should be comparable or better than the predicate.
    - Limit of Detection (LoD)2.5 pg/mL2.3 pg/mL
    - Limit of Quantitation (LoQ)4.5 pg/mL4.5 pg/mL
    Precision
    - Within-run CV%1.1% to 6.3%1.5% to 9.9%Demonstrate consistent and reproducible results across different concentration levels, within acceptable ranges for clinical diagnostic assays, as per CLSI EP-5A2 guidelines. Total CV should be within acceptable limits (e.g., <10-15% depending on concentration).
    - Total CV%4.1% to 8.2%1.8% to 9.9%
    Linearity
    - K2 EDTA Plasma$y = 1.002 x – 4.748; R^2 = 1.00$$y = 0.96 x − 0.1; R^2 = 1.00$Demonstrate linearity across the claimed measurement range (5 to 3500 pg/mL) with a regression coefficient ($R^2$) close to 1 and minimal deviation from expected values, as per CLSI EP-6A guidelines.
    - SerumN/A (Only K2 EDTA provided for predicate)$y = 1.02x – 0.2; R^2 = 1.00$
    High Dose Hook EffectNot observed up to 95,000 pg/mLNot observed up to 100,000 pg/mLNo significant hook effect should be observed within a clinically relevant and extended high concentration range to prevent falsely low results in patients with very high PTH levels. The demonstrated range should be comparable or wider than the predicate.
    Interferences (≤±10% bias)
    - Bilirubin, conjugatedN/A22 mg/dLNo significant interference (≤±10% bias) from common endogenous and exogenous interfering substances at clinically relevant concentrations, as per CLSI EP7-A2 guidelines. The tested concentrations should be comparable or improved compared to the predicate device.
    - Bilirubin, unconjugated20 mg/dL40 mg/dL
    - Biotin300 nmol/L300 nmol/L
    - CholesterolN/A395 mg/dL
    - HAMA1000 ng/mL1000 ng/mL
    - Hemoglobin250 mg/dL500 mg/dL
    - Rheumatoid Factor1500 IU/mL1836 IU/mL
    - Total ProteinN/A10 g/dL
    - Triglycerides3000 mg/dL3000 mg/dL
    - AcetaminophenN/A200 µg/mL
    - CarbamazepineN/A200 µg/mL
    - IbuprofenN/A500 µg/mL
    Cross-Reactivity (vs. PTH (1-84))
    - PTH (1-84)100%100%Demonstrate high specificity to intact PTH (1-84) and minimal cross-reactivity with PTH fragments or structurally similar proteins, improving upon the predicate where possible, as per CLSI EP7-A2 guidelines.
    - PTH (7-84)60%83.6%
    - PTH (1-34)0.5%<0.01%
    - PTH (39-84)Not detectable<0.01%
    - PTH (53-84)2%<0.01%
    - PTH (39-68)N/A<0.01%
    - PTH (44-68)2%<0.01%
    - Human Calcitonin4%<0.01%
    - CTX-1 (β CrossLaps)2%<0.01%
    - Osteocalcin2%<0.01%

    2. Sample Sizes and Data Provenance for Test Sets

    • Method Comparison:

      • Sample Size: 312 serum samples (291 native, 21 spiked).
      • Data Provenance: Not explicitly stated but clinical samples are generally considered prospective or retrospective from a clinical laboratory setting. The use of "commercially available quantitative chemiluminescence Intact PTH assay" as the comparator suggests real-world clinical samples.
    • Sample Matrix Comparison:

      • Sample Size: 52 matched samples (45 native, 7 spiked).
      • Data Provenance: Not explicitly stated, but clinical samples are typically used for this type of evaluation.
    • Reference Interval:

      • Sample Size: 129 individuals (67 males, 62 females; 21 to 89 years of age).
      • Data Provenance: K2 EDTA plasma samples collected from individuals with normal calcium, phosphate, and TSH values from the north, central and south regions of the United States. This is prospective or a well-characterized retrospective collection.
    • Sensitivity (LoB, LoD, LoQ):

      • Sample Size: 60 blanks and 10 low-level serum samples.
      • Data Provenance: Lab-prepared samples and low-level serum samples, likely from internal or commercially acquired biobanks.
    • Precision:

      • Sample Size: 7 samples (tested multiple times, N=80 for each sample).
      • Data Provenance: Lab-prepared controls and potentially pooled clinical samples ("Serum Sample," "EDTA Plasma Sample").
    • Linearity:

      • Sample Size: Not explicitly stated as a single number, but involved diluting high patient samples with low patient samples. Each sample was measured in 4 replicates.
      • Data Provenance: High and low intact PTH human K2 EDTA plasma and serum samples.
    • High Dose Hook Effect:

      • Sample Size: 3 samples (1 EDTA Plasma, 2 Serum) per spiked concentration.
      • Data Provenance: Spiked samples, likely from a laboratory setting using clinical matrices.
    • Interferences:

      • Sample Size: Two K2-EDTA samples (low and high Intact PTH levels) and two serum samples (different Intact PTH concentrations).
      • Data Provenance: Spiked samples, using clinical matrices (K2-EDTA and serum).
    • Cross-Reactivity:

      • Sample Size: A serum sample and a zero calibrator matrix.
      • Data Provenance: Spiked samples, using clinical serum matrix.

    3. Number of Experts and Qualifications for Ground Truth

    This device is an in vitro diagnostic assay, where the "ground truth" for performance characteristics such as sensitivity, precision, linearity, and interference is established by:

    • Reference Methods: Comparison against established, commercially available, and often FDA-cleared or gold-standard assays (e.g., the predicate device or a "commercially available quantitative chemiluminescence Intact PTH assay" for method comparison).
    • Analytical Standards/Spikes: Precisely known concentrations of analyte or interfering substances are used to create "spiked" samples, where the "true" concentration is known by design.
    • Clinical Laboratory Standard Institute (CLSI) Guidelines: Adherence to these guidelines (EP-9A2, EP9-A3, C28-A3, EP17-A, EP-5A2, EP-6A, EP7-A2) dictates the experimental design and statistical analysis which are considered "best practice" for establishing performance characteristics in IVDs.

    There are no "experts" in the sense of physicians or radiologists directly assessing images or patient data to establish ground truth for this type of analytical device performance. The "ground truth" is analytical and derived from established laboratory practices and reference methods.

    4. Adjudication Method for the Test Set

    Not applicable for this type of in vitro diagnostic device study. Adjudication methods like "2+1" or "3+1" are typically used in clinical studies, particularly for diagnostic imaging or subjective clinical assessments, where consensus among multiple expert readers is needed to establish a ground truth for a disease state. For IVD analytical performance, the ground truth is either the value from a reference method or a gravimetrically/volumetrically determined "true" concentration.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    No, an MRMC comparative effectiveness study was not done. This type of study is primarily relevant for diagnostic imaging AI algorithms where multiple human readers interpret cases with and without AI assistance. This document describes an in vitro diagnostic assay for a biochemical marker, not an imaging device or an AI application that assists human interpretation.

    6. Standalone Performance

    Yes, the studies described are standalone performance assessments of the IDS-iSYS Intact PTHN assay. The device is an automated system for measuring PTH levels, and the performance characteristics (method comparison, precision, linearity, etc.) directly evaluate the algorithm/assay's ability to accurately quantify PTH in patient samples. There is no human-in-the-loop component being evaluated in these tests; the device's output is directly compared to established analytical standards or reference methods.

    7. Type of Ground Truth Used

    • Reference Method Results: For method comparison, the results from a "commercially available quantitative chemiluminescence Intact PTH assay" served as the ground truth comparator.
    • Assigned Values of Controls/Standards: For precision, linearity, sensitivity, interference, and cross-reactivity studies, the ground truth was based on:
      • Known concentrations: Spiked samples with precisely known concentrations of analyte or interfering substances.
      • Expected values: Derived from dilution series or preparation of controls.
      • Blank matrices: For limit of blank determination.
    • Clinical Criteria: For the reference interval study, the ground truth for "normal" individuals was based on clinical criteria (normal calcium, phosphate, and TSH values).

    8. Sample Size for the Training Set

    The document does not specify a "training set" in the context of machine learning. This is an analytical medical device, not an AI/ML algorithm that is "trained" on data. The characteristics and parameters of the assay (e.g., antibodies, reagent concentrations, calibration curve algorithm) are developed and optimized during the product development phase, which is analogous to "training" in AI, but it's not described in terms of a distinct dataset.

    9. How Ground Truth for the Training Set Was Established

    As explained above, there isn't a "training set" in the common AI/ML sense. The "ground truth" for the development and optimization of the assay's chemical and instrumental parameters typically comes from:

    • Biochemical principles: Understanding of antigen-antibody interactions, chemiluminescence, and PTH biology.
    • Calibration materials: Use of international standards or highly characterized primary calibrators with assigned values.
    • Internal validation studies: Extensive testing during R&D to optimize reagent formulations, incubation times, and instrumental settings to achieve desired performance characteristics against known samples and reference methods.
    • Iterative development and testing: Performance data from pilot studies or earlier assay versions would inform adjustments to achieve acceptable analytical performance compared to the established medical consensus for PTH measurement.
    Ask a Question

    Ask a specific question about this device

    K Number
    K103325

    Validate with FDA (Live)

    Date Cleared
    2012-02-03

    (448 days)

    Product Code
    Regulation Number
    862.1545
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The IDS-iSYS Intact PTH assay is intended for the quantitative determination of PTH in human serum or plasma on the IDSiSYS Multi-Discipline Automated Analyzer. Results are to be use in the differential diagnosis of hypercalcemia and hypocalcemia resulting from disorders of calcium metabolism.

    The kit is for in vitro diagnostic use only.

    The IDS-iSYS Intact PTH assay Control Set is intended for use as assayed quality control samples to monitor the precision of the IDS-iSYS Intact PTH assay.

    The IDS-iSYS Intact PTH Calibrator is intended for the calibration of the IDS-iSYS Intact PTH assay.

    The IDS-iSYS iPTH Calibration Verifier is intended for verification of calibration of the IDS-iSYS iPTH Assay when performed on the IDS-iSYS Multi-Discipline Automated Analyzer.

    Device Description

    Not Found

    AI/ML Overview

    This document is a 510(k) clearance letter from the FDA for the IDS-iSYS Intact PTH assay. It states that the device is substantially equivalent to legally marketed predicate devices. However, this letter does not contain the acceptance criteria, device performance data, study details, or ground truth information as requested in your prompt.

    The FDA 510(k) clearance letter primarily confirms that the device meets the regulatory requirements for marketing based on substantial equivalence. It refers to a "premarket notification" which would typically include the study data, but that specific document is not provided here.

    Therefore, I cannot provide the requested table and study details based solely on the provided text. To answer your prompt, you would need the actual premarket notification submission (510(k) documentation) for K103325, which would contain the study results and acceptance criteria.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1