Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K110518

    Validate with FDA (Live)

    Device Name
    IBAG
    Date Cleared
    2011-06-24

    (121 days)

    Product Code
    Regulation Number
    876.5250
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    This device is intended to measure/ display collected urine volume. It detects fill level and temperature.

    Device Description

    iBag™ is a urine management system that includes bladder and temperature measurement for catheterized patients who desire to be fully mobile. · iBag™ provides accurate, real-time assistive data. It does this through use of a calibrated drainage bag integrated with a Future Path Medical sensor, a base transmitter unit called Wi-Mitter™ for wireless transmission of urine bag fill and bladder temperature information. Once the Wi-Mitter™ is snapped in place, the container is polled and the measurement of fluid level begins and this data is then transmitted to the patient, the care provider (an aid) a family member, or even to the hospital's base station. Data is transmitted via traditional data networking and may be interpreted and displayed on a user's laptop computer, and other types monitors. Since the data from the Wi-Mitter™ is time and date stamped, time of fill level may be automatically calculated, noted and threshold values used for fill level alarming.

    AI/ML Overview

    Here's an analysis of the provided text to extract information about the acceptance criteria and the study conducted for the iBag™ Urine Management System:

    1. Table of Acceptance Criteria and Reported Device Performance

    CharacteristicAcceptance CriteriaReported Device Performance
    Volume AccuracyNot explicitly stated as a separate acceptance criteria, but for predicate device "Not stated"±5% of the actual fill for 70 – 100% fill level readings.
    Temperature Accuracy± 0.2 °F (for predicate device)± 0.2 °F
    Temperature Range50°F to 110°F (10°C to 43.3°C) (for predicate device)50 to 110 °F in increments of 0.1 °F
    Electrical SafetyElectrical Safety per IEC-60601 (for predicate device)SAME (Implies compliance with IEC 60601-1)
    Electromagnetic Comp.N/A (not explicitly stated for predicate)IEC 60601-1-2
    FCC ComplianceN/A (not explicitly stated for predicate)FCC Compliance Testing
    RF Dosimetric Asses.N/A (not explicitly stated for predicate)RF Dosimetric Assessment Testing
    Transmitter RangeN/ATransmitter effective range testing was done
    User FeedbackN/AUser Feedback Study was done
    Battery LifeN/ABattery life computations were done
    Simulated UseN/ASimulated Use Testing was done
    Leak TestingN/ALeak Testing was done
    Sterility TestingN/ASterility Testing was done

    Note: The predicate device's specifications for volume accuracy and temperature range are used as a proxy for acceptance criteria where the iBag™ device states "SAME" or explicitly meets/exceeds them. For other tests, the text indicates that the testing was "successfully conducted" or that the device meets the respective standard.

    2. Sample Size and Data Provenance for the Test Set

    • Sample Size for Test Set: The document does not specify the sample sizes for the "Fill level accuracy" or "Temperature measurement accuracy" tests, nor for the "User Feedback Study" or "Simulated Use Testing."
    • Data Provenance: The document does not provide details on the country of origin of the data or whether the studies were retrospective or prospective. It only states that "Bench and test laboratory results indicate that the new device is as safe and effective as the predicate device" and lists the tests conducted.

    3. Number and Qualifications of Experts for Ground Truth

    The document does not provide any information about the number or qualifications of experts used to establish ground truth for any of the tests.

    4. Adjudication Method for the Test Set

    The document does not specify any adjudication methods for the test set.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    No mention of a Multi-Reader Multi-Case (MRMC) comparative effectiveness study being done, or any effect size related to human readers improving with or without AI assistance. This device is not an AI-based system.

    6. Standalone Performance (Algorithm Only)

    The document primarily focuses on the device's mechanical and electronic performance, not on an algorithm's standalone performance. It states that "iBag™ provides accurate, real-time assistive data" and "Software is designed to be used with little or no training," but does not detail an algorithm's standalone performance separate from the hardware.

    7. Type of Ground Truth Used

    Based on the provided information, the ground truth for volume and temperature accuracy would likely be established through:

    • Reference measurements: Using highly accurate, calibrated instruments to measure actual urine volume and temperature.
    • Bench testing: Controlled laboratory conditions for precise measurements.

    For other tests like electrical safety, EMC, leak testing, and sterility, the ground truth would be defined by the relevant industry standards (e.g., IEC 60601-1, IEC 60601-1-2) and their specific test protocols.

    8. Sample Size for the Training Set

    No information is provided about a "training set" as this device is not an AI/ML-based system requiring a training phase in the typical sense. The testing described is for performance validation, not algorithm training.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable, as there is no mention of a training set for an AI/ML algorithm.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1