510(k) Data Aggregation

    K Number
    K190268

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2019-11-06

    (271 days)

    Product Code
    Regulation Number
    N/A
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Hydrofera Blue READY - Border is intended as an external dressing for use in local management of moderate to heavily exuding wounds such as pressure ulcers, donor sites, venous stasis ulcers, diabetic ulcers, abrasions, lacerations, superficial burns, post-surgical incisions, and other external wounds inflicted by trauma.

    Device Description

    Hydrofera Blue READY - Border is a sterile absorptive foam dressings made of polyurethane (PU) foam, with methylene blue and gentian violet concentrations of at least 0.00005 g/g each with a combined total concentration of less than or equal to 0.0007 g/g. The dressings include an outer polyurethane film backing with silicone adhesive border.

    AI/ML Overview

    This document describes the acceptance criteria and performance of the Hydrofera Blue READY - Border device, a wound dressing, as part of its 510(k) submission (K190268) to the FDA.

    1. Table of Acceptance Criteria and Reported Device Performance:

    Parameter EvaluatedAcceptance CriteriaReported Device Performance
    Antibacterial Efficacy (ASTM E2149-10, modified)≥ 4 log reduction over 7 days- 8 gram-positive Organisms tested: 8 experienced ≥ 6 log reduction - 10 gram-negative Organisms tested: 10 experienced ≥ 6 log reduction
    Zone of Inhibition (AATCC 147-2011)No zone of inhibitionNo evidence of growth surrounding the Dressing
    Dye Leachable< 0.5 ppm for each dyeLeachable amounts were measured at far less than 0.5 ppm for each dye (Methylene Blue and Gentian Violet)
    LAL Endotoxin (ANSI/AAMI ST72:2011)< 20 EU/deviceDevice contains less than 0.5 EU/device and meets the requirements of the test for Blood/Lymph Contact and Cerebrospinal Fluid Contact.
    Device Performance - Wicking SpeedAbsorb fluid in < 30 secondsAbsorbs fluid in < 3 seconds
    Device Performance - Fluid Retention≥ 30% Fluid Retained37.47% (σ 1.35)
    Device Performance - Border AdhesionComparable to market leadersThe adhesion performance of the perforated silicone adhesive was evaluated for peel strength in comparison to Hollister Triact Foam Dressing with silicone border and Coloplast Biatain silicone (with silicone border). The peel strength met the acceptance criteria.
    Antibacterial Efficacy - Low Specification Dye Limit (Simulated Wound Fluid)≥ 4 log reduction over 7 daysAll low specification dye limit test articles evaluated exceeded the antibacterial efficacy acceptance criteria by demonstrating complete kill of the bacteria Staphylococcus aureus, Streptococcus pneumoniae, Bacillus subtilis, Pseudomonas aeruginosa, Escherichia coli, and Enterobacter cloacae after 7 days of incubation, following a 7-day pre-conditioning period.
    Antibacterial Efficacy - Lot Consistency (Simulated Wound Fluid)≥ 4 log reduction over 7 daysAll lot consistency test articles evaluated exceeded the antibacterial efficacy acceptance criteria by demonstrating complete kill of the bacteria Staphylococcus aureus, Streptococcus pneumoniae, Bacillus subtilis, Pseudomonas aeruginosa, Escherichia coli, and Enterobacter cloacae after 7 days of incubation, following a 7-day pre-conditioning period.
    Biocompatibility (Cytotoxicity per ISO 10993-5)Non-cytotoxicDevice is considered non-cytotoxic.
    Biocompatibility (Sensitization per ISO 10993-10)Non-sensitizingDevice is considered non-sensitizing.
    Biocompatibility (Irritation per ISO 10993-11)Non-irritantDevice is considered non-irritant.
    Biocompatibility (Pyrogenicity per ISO 10993-11 and USP <151>)Non-pyrogenicDevice is considered non-pyrogenic.
    Biocompatibility (Local effects after implantation per ISO 10993-6)Considered passingIn the rabbit muscle study, the device was considered passing.
    Biocompatibility (Acute systemic toxicity per ISO 10993-11)No signs of acute systemic toxicityDevice showed no signs of acute systemic toxicity.
    Biocompatibility (Hemolysis/Interaction with blood per ASTM Hemolysis Test)PassedPassed
    Biocompatibility (Toxicology Risk Assessment per ISO 10993-1, 10993-11, 10993-3, 10993-17, 10993-18, 21726, and 14971)Toxicity not expected from intended useToxicity is not expected from the intended use of the device. (Note: product has not been tested or evaluated on pediatric or neonatal patient populations).

    2. Sample Size Used for the Test Set and Data Provenance:

    The document does not explicitly state the sample sizes for each test in the acceptance criteria table. However, it indicates:

    • Antibacterial Efficacy (ASTM E2149-10 modified): 8 gram-positive organisms and 10 gram-negative organisms were tested. The specific number of replicates or samples per organism is not provided.
    • Other In Vitro Tests: Standard laboratory testing was performed (e.g., LAL Endotoxin, Wicking Speed, Fluid Retention, Dye Leacheable). The number of samples for these tests is generally small and typical for in vitro device characterization, but not specifically quantified in this summary.
    • Biocompatibility: Testing was conducted according to ISO 10993 standards and included animal studies (rabbit for pyrogenicity, implantation, and acute systemic toxicity), but the specific number of animals or samples per test is not detailed in this summary.

    Data Provenance: All data appears to be from non-clinical, in vitro, and animal studies conducted to demonstrate device performance and safety. There is no information regarding country of origin, or whether the data is retrospective or prospective, beyond the tests being performed "in support of the substantial equivalence determination." This is typically laboratory data derived from testing performed by or for the manufacturer.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts:

    • This submission relies entirely on non-clinical (bench and animal) testing. It does not involve a clinical study with a "test set" requiring expert ground truth establishment in the context of diagnostic or interpretive performance.
    • Therefore, there were no experts used to establish ground truth for a test set in the way one would for an AI-powered diagnostic device (e.g., radiologists interpreting images). The "ground truth" for these tests is defined by the objective metrics and outcomes of the established laboratory and animal test methods (e.g., bacterial log reduction, ppm of leachable dye, absence of cytotoxic effects).

    4. Adjudication Method for the Test Set:

    • Since there is no "test set" requiring expert interpretation or consensus for ground truth (as explained above), there was no adjudication method used. The results are direct measurements or observations from standardized tests.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done:

    • No, an MRMC comparative effectiveness study was not done. This device is a wound dressing, not a diagnostic imaging device or an AI-assisted interpretation tool. The submission states, "No clinical data was required to support substantial equivalence."

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done:

    • Not applicable. This device is a physical wound dressing and does not involve any algorithm or AI component. Therefore, no standalone performance evaluation in that context was performed.

    7. The Type of Ground Truth Used:

    The ground truth for the performance and safety evaluations are based on objective, standardized laboratory and animal testing results (bench data and in vivo animal study outcomes), verified against predefined acceptance criteria for each test. This includes:

    • Microbiological growth/reduction data.
    • Chemical concentration measurements.
    • Physical property measurements (wicking speed, fluid retention, peel strength).
    • Biological responses in cytotoxicity, sensitization, irritation, pyrogenicity, implantation, acute systemic toxicity, and hemolysis tests.

    8. The Sample Size for the Training Set:

    • Not applicable. This device does not use machine learning or AI models, so there is no "training set."

    9. How the Ground Truth for the Training Set Was Established:

    • Not applicable. As there is no training set, no ground truth was established for it.